G4100 NOx/O2 Analyzing System
9
The analyzing board shall be placed and installed where the ambient tempera-
ture is below 55 °C at any time.
The system shall be installed in a clean place away from dust, oil mist, and
moisture. The elements of the system shall be installed at viewing level so that
it is easy accessible in connection with operation and service.
3.3 Safety Aspects
Warning
Make sure to follow the operating instructions!
Read the manual before operating and servicing the system.
Hot sensor/probe!
The sensor/probe is hot and can cause severe burning of personnel if not handled
with care.
Installation and operation!
It should be pointed out that installation and operation of the G4100 NOX/O2Ana-
lyzing System and associated equipment must be carried out by skilled, trained,
and certified personnel, and that Green Instruments A/S does not take any respon-
sibility for the operation of the system and associated equipment whatsoever. The
successful and safe operation of this equipment is dependent upon proper han-
dling, installation, operation, and maintenance.
Recycling!
Please do not dispose the G4100 NOX/O2Analyzing System with regular refuse.
Disposal should be in accordance with the requirements of the current statutory
regulations.
3.4 Ejector Probe
The ejector probe is supplied with a mounting socket, gasket, sample tube, con-
nection for ejector air, connection for test gas and back-flushing air, sensor, and
housing with mounting plate.
The ejector probe has a small air operated ejector, which will propel the sample
gas into the sensor and back into the stack. The sample gas flow is drawn through
the sample filter element into the sensor area via a narrow passage securing quick
response time. The sample gas velocity and responding time is controlled by the
flow of the ejector air.