H43.0.01.6C-11 Operating Manual GOX 20 Page 3 of 6
2.3 Safety guidelines
This device has been designed and tested in accordance with the safety regulations for electronic devices.
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and
special safety advises given in this manual will be adhered to when using the device.
1. Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any
other climatic conditions than those stated under "Specification".
If the device is transported from a cold to a warm environment condensation may cause in a failure of the func-
tion. In such a case make sure the device temperature has adjusted to the ambient temperature before trying a
new start-up.
2. If there is a risk whatsoever involved in running it, the device has to be switched off immediately
and to be marked accordingly to avoid re-starting.
Operator safety may be a risk if:
- there is visible damage to the device
- the device is not working as specified
- the device has been stored under unsuitable conditions for a longer time.
In case of doubt, please return device to manufacturer for repair or maintenance.
3. When connecting the device to other devices the connection has to be designed most thoroughly as internal
connections in third-party devices (e.g. connection GND with protective earth) may lead to undesired voltage po-
tentials that can lead to malfunctions or destroying of the device and the connected devices.
4. Do not use these products as safety or emergency stop devices or in any other application where
failure of the product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material dam-
age.
5. Caution, acid! The electrode contains KOH. KOH can cause severe chemical burns! If leaking,
avoid contact.
If there was contact:
- to skin: flush contacted area with large amounts of water for several minutes.
- to clothing: remove contaminated clothing.
- to eyes: flush with large amounts of water for several minutes, obtain medical treatment.
After swallowing:
- give large volumes of water. DO NOT induce vomiting!
- obtain medical treatment.
3 Product Specification
3.1 Scope of supply
The scope of supply includes:
Device, incl. 9V-battery block
Operating Manual
3.2 Operation and maintenance advice
•Battery operation:
If ‘bAt’ is shown in the lower display the battery has been used up and needs to be replaced. However, the device
will operate correctly for a certain time. If ‘bAt’ is shown in the upper display the voltage is too low to operate the
device; the battery has been completely used up.
The battery has to be taken out, when storing device above 50 °C.
We recommend taking out battery if device is not used for a longer period of time.
•For measuring remove the protective flask of the electrode.
•Please note: the electrode measurement is sensitive against shocks!
By stirring of the electrode in the measured liquid be careful that the electrode does not hit the container. A vibra-
tion of the electrode has a effect to the measured value.
•Store the oxygen electrode always in water, either in the protective flask, or directly in a vessel filled with dechlo-
rinated water. For storage in the protective flask first push the cap oh the electrode second the o-ring and thereaf-
ter put the electrode in the flask and screw the cap down the flask.
Note: for measuring remove the protective flask!
If the diaphragm has dried up the electrode must be soaked in water for approx. 2 h prior to measuring start-up.
Then a calibration can be carried out without any problems.