OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDEOPERATIONAL GUIDE
OPERATIONAL GUIDE
TAKING MEASUREMENTS
• Turn the meter on by connecting the 12 VDC power
adapter to the meter and to the mains.
• Immerse the tip (4 cm/1½”)
of the EC or TDS probe in the
sample.
Note: In order not to affect the
accuracy of measurements,
it is important that the
probe body does not touch
nor stand close to the side
walls of the vessel. The tip
may lay on the bottom of
the beaker.
• The LCD will show the EC or
TDS value. Any initial varia-
tionmaybeduetotemperature
compensationand thefactthat
the probe is adjusting itself to
thenewsample.Allowtheread-
ing to stabilize and the meter
will start continuous monitor-
ing.
ADJUSTINGTHESETPOINT
Withthe ECandTDS , you can selectyourown
setpoint and be alerted with a visual LED alarm when an
abnormal situation arises.
• Unscrew and remove the rear panel and gasket seal to
access the MEASURE/SET switch.
• Move the switch to the left (SET Mode).
• With a small screwdriver adjust the setpoint trimmer to
display the desired value in the 0.70 to 3.50 mS/cm or
in the 500 to 1600 ppm range for the EC and TDS
,respectively.
• Make sure the switch is moved back to
the right (MEASURE Mode).
• Replace the rear panel and
the gasket, ensuring the unit
is properly closed.
• Whenever the EC or TDS read-
ing varies by more than
±0.20mS/cm or ±100 ppm
from the setpoint, respectively,
the red alarm LED will blink.
PROBE MAINTENANCE
To minimize clogging and provide longer life for the probe,
itis recommended to clean it often or atleast once a month.
• Immerse the tip of the electrode in HI 7061 Cleaning
Solution for one hour or clean it with detergents used to
wash irrigation pipelines and fertilizer vessels.
• If a more thorough cleaning is required, brush the
metal pins with very fine sandpaper.
• After cleaning or when not in use, rinse the probe with
tap water.
For the greatest accuracy, frequent calibration of the instru-
ment is recommended.
• Turn the meter on and make sure that
the MEASURE/SET switch is on the MEA-
SURE mode.
• Pour a small quantity of
1413µS/cm calibration so-
lution or 1500 ppm TDS
calibrationsolutionina bea-
ker. If possible, use plastic
beakers to minimize any
EMCinterference.
CALIBRATIONCALIBRATION
CALIBRATIONCALIBRATION
CALIBRATION
• Immerse the EC or TDS probe in the
solution, making sure that metal pins
are completely submerged.
Note: the probe should be submerged ap-
proximately 4 cm (1½”) in the
solution. Moreover, in order not to
affect the accuracy of measurements,
it is important that probe body does not touch nor
stand close to the side walls of the beaker. The tip
can lay on the bottom of the beaker.
• Wait for a couple of minutes for thermal equilibrium to
be reached.
• Tap the probe gently on the bottom, then shake it
while rotating to make sure no air bubbles have
remained trapped.
• Adjust the calibration trimmer
with the supplied screwdriver
until the display shows
"1.41 mS" (HI 983302) or
“1500ppm” (HI 983301).
• The calibration is now complete and the instrument is
ready for use.
The instrument should be recalibrated at least once a
month and after performing probe cleaning procedure.
ACCESSORIESACCESSORIES
ACCESSORIESACCESSORIES
ACCESSORIES
HI 7632 * EC probe with Automatic Temperature Compen-
sation and 2 m (6.6’) cable
HI 7634 * TDS probe with Automatic Temperature Com-
pensation and 2 m (6.6’) cable
HI70031P 1413 µS/cm (EC) calibration solution, 20 mL
sachet(25 pcs)
HI70442P 1500 ppm (TDS) calibration solution, 20 mL
sachet(25 pcs)
HI 7052L 1500ppm (TDS) calibrationsol.,460 mL bottle
HI7031L 1413µS/cm(EC)calibrationsol.,460 mLbottle
HI 7061L Electrode cleaning solution, 460 mL bottle
HI710005 12 VDC power adapter, US plug
HI710006 12 VDC power adapter, European plug
HI710012 12 VDC power adapter, Australian plug
HI710013 12 VDC power adapter, Southern Africa plug
HI710014 12 VDC power adapter, UK plug
*Tobereplacedonly by authorized service technicians
CE DECLARATIONCE DECLARATION
CE DECLARATIONCE DECLARATION
CE DECLARATION
OF CONFORMITYOF CONFORMITY
OF CONFORMITYOF CONFORMITY
OF CONFORMITY
RecommendationsforUsers
Before using these products, make sure that they are entirely suitable for the environ-
mentinwhichtheyare used.
Operation of these instruments in residential areas could cause unacceptable interfer-
encestoradioandTVequipment.
Themetalbandattheendoftheprobeissensitivetoelectrostaticdischarges.Avoidtouching
thismetal band at alltimes.
Duringoperation,ESDwriststrapsshouldbeworntoavoidpossibledamagetotheprobeby
electrostaticdischarges.
Anyvariationintroducedbytheusertothesuppliedequipmentmaydegradetheinstrument’s
EMCperformance.
Toavoidelectricalshock,donotusetheseinstrumentswhenvoltagesatthemeasurement
surfaceexceed24VACor60VDC.
UseplasticbeakerstominimizeanyEMCinterferences.
Toavoiddamagesorburns,donotperformanymeasurementinmicrowaveovens.
ISTRECGRR1 09/98PRINTED IN PORTUGAL