Hanna Instruments HI 991401 User manual

WARRANTY
HI 991401 is warranted for two years against defects in workmanship
and materials when used for their intended purpose and maintained accord-
ing to instructions. Electrodes and probes are warranted for six months.
This warranty is limited to repair or replacement free of charge.
Damages due to accident, misuse, tampering or lack of prescribed mainte-
nance are not covered.
If service is required, contact the dealer from whom you purchased the instru-
ment. If under warranty, report the model number, date of purchase, serial
number and the nature of the failure. If the repair is not covered by the
warranty, you will be notified of the charges incurred. If the instrument is to be
returned to Hanna Instruments, first obtain a Returned Goods Authorization
Number from the Customer Service department and then send it with ship-
ment costs prepaid. When shipping any instrument, make sure it is properly
packaged for complete protection.
All rights are reserved. Reproduction in whole or in part is prohibited with-
out the written consent of the copyright owner.
Remove the instrument from the packing material and exam-
ine it carefully. If any damage has occurred during shipment,
immediately notify your dealer or the nearest Hanna Customer
Service Center.
The meter is supplied with:
• HI 1293D pH-electrode
• temperature probe (fixed)
• pH 4.01 and pH 7.01 buffer solutions (20 mL each)
• 12 Vdc power adapter and instructions
Note: Conserve all packing material until the instrument has
been observed to function correctly. Any defective item
must be returned in its original packing.
PRELIMINARY EXAMINATION
HI 991401 indicator features a large, dual-level backlit LCD
to give istantaneous readings of both pH and temperature,
even from a distance.
This waterproof indicator has been specially designed to meet
the grower’s need for equipment suited to the aggressive
environments of agricultural and hydroponics applications.
Measurements are highly accurate and the meter can be
calibrated at one or two points. Low battery and battery
changes are no longer a problem thanks to a 12 Vdc power
supply included with the meter.
The HI 1293D pH electrode & the temperature probe have
been molded with pipe thread (1/2’’ NPT), which allows the
user to attach them to an in-line system.
Moreover, the pH electrode has been specially designed to
guarantee greater clogging resistance in fertilizer solutions
with high concentrations of phosphate and nitrate.
GENERAL DESCRIPTION
FUNCTIONAL DESCRIPTION SPECIFICATIONS
Recommendations for Users
Before using this product, make sure that it is entirely suitable for the
environment in which it is used. Operation of this instrument in residential
areas could cause unacceptable interferences to radio and TV equipment.
The glass bulb at the end of the electrode is sensitive to electrostatic dis-
charges. Avoid touching this glass bulb at all times. During operation, ESD wrist
straps should be worn to avoid possible damage to the electrode by electro-
static discharges.
Any variation introduced by the user to the supplied equipment may degrade
the instrument’s EMC performance.
To avoid electrical shock, do not use this instrument when voltages at the
measurement surface exceed 24 Vac or 60 Vdc.
To avoid damages or burns, do not perform any measurement in microwave
ovens.
Range 0.0 to 14.0 pH
0.0 to 60ºC / 32.0 to 140ºF
Resolution 0.1 pH / 0.1ºC / 0.1ºF
Accuracy (@20ºC/68ºF) ±0.1 pH / ±0.5 ºC / ±1ºF
Typical EMC Deviation
±0.1 pH / ±1ºC / ±2 ºF
pH Calibration Automatic, 1 or 2 point
with auto-buffer recognition
Temperature Compensation Automatic
Probes HI 1293D pH electrode (included)
HI 1294 temperature probe (fixed)
Power supply 12 Vdc power adapter (included)
Environment 0 to 50ºC (32 to 122ºF);
RH max 95% non-condensing
Dimensions 160 x 105 x 31 mm (6.2 x 4.1 x 1.2’’)
Weight 190 g (6.7 oz.)
1. Liquid Crystal Display
2. ON/OFF/MODE button
3. Power supply connector
4. 12 Vdc power adapter
5. HI 1293D pH electrode with differential input and pipe
thread 1/2’’ NPT
6. Temperature probe with pipe thread 1/2’’ NPT
(also working as matching pin for pH electrode)
7. DIN connector
8. SET/HOLD button
www.hannainst.com
InstructionManual
(HI991401)
Hanna Instruments reserves the right to modify the design, construction and
appearance of its products without advance notice.
Dear Customer,
Thank you for choosing a Hanna product.
This manual will provide you with the necessary information
for a correct operation. Please read it carefully before using
the meter. If you need additional technical information, do
not hesitate to e-mail us at [email protected].
This instrument is in compliance with the directives.
IST991401R2 07/05
Preliminary operations
Connect the pH electrode to the meter.
Connect the 12 Vdc adapter to the meter and to the mains;
the display will lit.
Press and hold the MODE button for 2-3 seconds. All the used
segments on the LCD will be visible for a few seconds.
Tochange the temperature unit
To change the temperature unit (from °C to °F), from
measurement mode, press and hold the MODE button until
“TEMP” and the current temperature unit are displayed on
the lower LCD (E.g. TEMP °C).
Use the SET/HOLD button to change the temperature unit,
and then press MODE button twice to return to normal
measurement mode.
To freeze the display
Press and hold the SET/HOLD button for 2-3 seconds, until
“HOLD” appears on the secondary display.
Press either button to return to normal mode.
Taking measurements
Immerse the electrode and the tem-
perature probe in the solution to be
tested. For better accuracy, the elec-
trode should not touch or stand close to
the walls or bottom of the sample
vessel.
The pH value automatically compen-
sated for temperature is shown on the primary LCD, while the
secondary LCD displays the temperature of the sample.
Measurements should be taken when the stability symbol 
on the top left of the LCD disappears.
IMPORTANT NOTE
Measurements have to be taken with both pH elec-
trode and temperature probe (also working as
matching pin) immersed in the same vessel.
OPERATIONAL GUIDE
Calibration buffer set
• From measurement mode, press and hold the MODE
button until “TEMP” and the current temperature unit are
displayed on the lower LCD (E.g. TEMP °C).
• Press the MODE button again to show the current buffer set:
“pH 7.01 BUFF” for standard set (pH 4.01/7.01/10.01) or
“pH 6.86 BUFF” for NIST set (pH 4.01/6.86/9.18).
• Press the SET/HOLD button to change the buffer value.
• Press the MODE button to return to the normal mode.
Calibration procedure
From measurement mode, press and hold the MODE button
until “CAL” is displayed on the lower LCD. Release the button.
The LCD will display “pH 7.01 USE” (or “pH 6.86 USE” if you
have selected the NIST buffer set). The CAL tag blinks on the
display.
• For a single-point pH calibration, immerse the electrode
and the temperature probe in any buffer from the selected
buffer set (eg. pH 7.01 or pH 4.01 or pH 10.01). The
meter will recognize the buffer value automatically.
If using pH 4.01 or pH 10.01, the meter will display “OK”
for 1 second and then return to measurement mode.
If using pH 7.01, after recognition of the buffer the meter
will ask for pH 4.01 as second calibration point. Press the
MODE button to return to measurement mode or, if
desired, proceed with the 2 point calibration as explained
below.
Note: For better accuracy it is always recommended to carry
out a two-point calibration.
• For a two-point pH calibration, place the electrode and
the temperature probe in pH 7.01 buffer (or pH 6.86 if
you have selected the NIST buffer set). The meter will
recognize the buffer value and then display “pH 4.01
USE”.
Rinse the electrode thoroughly and immerse it in the
second buffer value (pH 4.01 or 10.01, or, if using NIST,
pH pH 4.01 or 9.18). When the second buffer is also
recognized, the LCD will display “OK” for 1 second and the
meter will return to the normal mode.
pH ELECTRODE MAINTENANCE
• When not in use, rinse the electrode with water and store
it with a few drops of HI 70300 storage solution in the
protective cap.
NEVER DISTILLED OR DEIONIZED WATER FOR STORAGE
PURPOSES.
• If the electrode has been left dry, soak in storage or pH 7
solution for at least one hour to reactivate it.
• To prolong the life of the pH electrode, it is recommended
to clean it monthly by immersing it in the HI 7061
cleaning solution for half an hour. Afterwards, rinse it
thoroughly with tap water and recalibrate the meter.
ACCESSORIES
HI 1293D Spare pH electrode with differential input,
pipe thread 1/2’’ NPT and DIN connector
HI 1294 (*) Temperature probe, pipe thread 1/2’’
NPT (also working as matching pin for the
pH electrode)
HI 70004P pH 4.01 solution, 20 mL sachet (25 pcs)
HI 70006P pH 6.86 solution, 20 mL sachet (25 pcs)
HI 70007P pH 7.01 solution, 20 mL sachet (25 pcs)
HI 70009P pH 9.18 solution, 20 mL sachet (25 pcs)
HI 70010P pH 10.01 solution, 20 mL sachet (25 pcs)
HI 7004M pH 4.01 solution, 230 mL bottle
HI 7006M pH 6.86 solution, 230 mL bottle
HI 7007M pH 7.01 solution, 230 mL bottle
HI 7009M pH 9.18 solution, 230 mL bottle
HI 7010M pH 10.01 solution, 230 mL bottle
HI 7061M Electrode cleaning solution, 230 mL bottle
HI 70300M Electrode storage solution, 230 mL bottle
HI 710005 12 Vdc power adapter, US plug
HI 710006 12 Vdc power adapter, European plug
HI 710012 12 Vdc power adapter, Australian plug
HI 710013 12 Vdc power adapter, South Africa plug
HI 710014 12 Vdc power adapter, UK plug
(*) To be replaced by authorized technical personnel only
To turn the meter off
Press the MODE button while in normal measurement mode.
“OFF” will appear on the lower part of the display. Release
the button. The display still lit, until the power supply is
connected.
Notes:
• Before taking any measurement make sure the meter has
been calibrated.
• To clear a previous calibration, press the MODE button
after entering the calibration mode. The lower LCD will
display “ESC” for 1 second and the meter will return to
normal measurement mode. The “CAL” symbol on the LCD
will disappear. The meter will be reset to the default
calibration.
• If measurements are taken in different samples successively,
rinse the electrode thoroughly with water, and then with
some of the sample to be measured.
pH CALIBRATION
Note: For storing calibration data in the non-volatile memory,
turn the meter OFF and then ON again through the
MODE button.
The “CAL” symbol on the LCD means that the meter is
calibrated.

Other manuals for HI 991401

1

This manual suits for next models

1

Other Hanna Instruments Measuring Instrument manuals

Hanna Instruments HI 93738 User manual

Hanna Instruments

Hanna Instruments HI 93738 User manual

Hanna Instruments HI96720C User manual

Hanna Instruments

Hanna Instruments HI96720C User manual

Hanna Instruments HI 98186 User manual

Hanna Instruments

Hanna Instruments HI 98186 User manual

Hanna Instruments HI747 User manual

Hanna Instruments

Hanna Instruments HI747 User manual

Hanna Instruments HI 99163 User manual

Hanna Instruments

Hanna Instruments HI 99163 User manual

Hanna Instruments HI 93728 User manual

Hanna Instruments

Hanna Instruments HI 93728 User manual

Hanna Instruments HI 93640N User manual

Hanna Instruments

Hanna Instruments HI 93640N User manual

Hanna Instruments HI96718C User manual

Hanna Instruments

Hanna Instruments HI96718C User manual

Hanna Instruments HI 96769C User manual

Hanna Instruments

Hanna Instruments HI 96769C User manual

Hanna Instruments HI 96762C User manual

Hanna Instruments

Hanna Instruments HI 96762C User manual

Hanna Instruments HI97752 User manual

Hanna Instruments

Hanna Instruments HI97752 User manual

Hanna Instruments HI7640-18 Series User manual

Hanna Instruments

Hanna Instruments HI7640-18 Series User manual

Hanna Instruments HI 4321 User manual

Hanna Instruments

Hanna Instruments HI 4321 User manual

Hanna Instruments HI 84181 User manual

Hanna Instruments

Hanna Instruments HI 84181 User manual

Hanna Instruments HI 96104C User manual

Hanna Instruments

Hanna Instruments HI 96104C User manual

Hanna Instruments HI 99161 User manual

Hanna Instruments

Hanna Instruments HI 99161 User manual

Hanna Instruments HI96704C User manual

Hanna Instruments

Hanna Instruments HI96704C User manual

Hanna Instruments Pronto HI 982401 User manual

Hanna Instruments

Hanna Instruments Pronto HI 982401 User manual

Hanna Instruments 98192 User manual

Hanna Instruments

Hanna Instruments 98192 User manual

Hanna Instruments HI 98186 User manual

Hanna Instruments

Hanna Instruments HI 98186 User manual

Hanna Instruments HI 713 User manual

Hanna Instruments

Hanna Instruments HI 713 User manual

Hanna Instruments HI 4110 User manual

Hanna Instruments

Hanna Instruments HI 4110 User manual

Hanna Instruments HI 93711 User manual

Hanna Instruments

Hanna Instruments HI 93711 User manual

Hanna Instruments HI 9033 User manual

Hanna Instruments

Hanna Instruments HI 9033 User manual

Popular Measuring Instrument manuals by other brands

Vutlan HAT-100Q1 Quick installation manual

Vutlan

Vutlan HAT-100Q1 Quick installation manual

ITRON Intelis Setup guide

ITRON

ITRON Intelis Setup guide

BRUEL & KJAER 2203 Instructions and applications

BRUEL & KJAER

BRUEL & KJAER 2203 Instructions and applications

Sequoia LISST-VSF user manual

Sequoia

Sequoia LISST-VSF user manual

TTI 1604 Service manual

TTI

TTI 1604 Service manual

Goobay 67000 instruction manual

Goobay

Goobay 67000 instruction manual

Mitchell Instrument WENS 900C user manual

Mitchell Instrument

Mitchell Instrument WENS 900C user manual

Syco SWR quick start guide

Syco

Syco SWR quick start guide

Pacific Data Systems DPM-RT-2 operating manual

Pacific Data Systems

Pacific Data Systems DPM-RT-2 operating manual

YOKOGAWA Power Series Plus IM 2493WVP-1 user manual

YOKOGAWA

YOKOGAWA Power Series Plus IM 2493WVP-1 user manual

Lutron Electronics MHT-381SD Operation manual

Lutron Electronics

Lutron Electronics MHT-381SD Operation manual

Myzone Inbody 230 Setup instructions

Myzone

Myzone Inbody 230 Setup instructions

Endress+Hauser Prosonic T FMU 130 operating manual

Endress+Hauser

Endress+Hauser Prosonic T FMU 130 operating manual

XS Instruments Temp70 PT100 vio instruction manual

XS Instruments

XS Instruments Temp70 PT100 vio instruction manual

RadComm RC2PLUS user manual

RadComm

RadComm RC2PLUS user manual

ZBM MINI OTDR user manual

ZBM

ZBM MINI OTDR user manual

Dostmann Electronic P705 Operation manuals

Dostmann Electronic

Dostmann Electronic P705 Operation manuals

Simrad PI50 -  REV A installation manual

Simrad

Simrad PI50 - REV A installation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.