Capacitive touchscreen
Users Settings Log Recall Help
Channel
Settings
Back key Home key Menu key
• return to a previous hierarchical menu level
• access to measurement screen
• access to:
`Users login & rights configuration
`Settings system configuration, connectivity & printing items
`Log Recall logged measurement data
`Help support
• from measurement screen, tap to access measurement menu
Probe usage guidelines & Calibration tips
Accessories: membrane cap and O-ring, electrolyte solution, zero
oxygen solution, calibration beakers
• Unscrew the shipping cap and save for calibration.
• Take one O-ring and one membrane cap and position the O-ring
(1) in the cap (2).
1
2
• Rinse the membrane cap with electrolyte and discard.
• Fill the cap, above the O-ring, with electrolyte and tap the side
walls to dislodge bubbles that may adhere to the threads.
• Over a sink, with the catode facing down, screw the cap counter
clockwise until the threads are fully engaged.
• Rinse the probe and inspect the membrane for trapped bubbles.
If any, discard the electrolyte, refill, and tap the sides. Reinstall.
Calibration tips
• Ensure the probe is correctly plugged in and turn on the meter.
• Allow 15 minutes for probe polarization conditioning.
• 100% saturation calibration: pour some water into supplied
cap and screw the cap onto the probe, kept in vertical position.
Allow 15 minutes for air to become water saturated.
• 0% saturation calibration: pour HI7040 zero oxygen solution
into a beaker and gently stir for 2-3 minutes. Gently tap the
probe to dislodge any air bubbles.
• Wait for stable reading to calibrate.
Measurement tips
• Work with recently calibrated probes.
• Place probe in sample and allow 1 minute for reading to stabilize.
• For accurate measurements the membrane needs constant
oxygen replenishment. Ensure adequate water movement either
manually or by use of a stirrer.
Logging in & Creating a new account
1. Power the meter and tap .
2. Tap followed by .
3. Tap Edit Users to enter Account Management.
4. Tap to enable Account Creation and Logins. Tap to return.
5. Tap Logout to enter Users screen. Tap the plus symbol avatar.
08:37:41 31/08/2022
Users
Admin
HI6421 +
6. Input user name and tap .
7. Enter password and tap . Re-enter password to confirm.
All Hanna instruments conform to the CE European Directives and UK Standards, and our production facilities are ISO 9001 certified. HI6421P is warranted for a period
of two years and the probe for six months against defects in workmanship and materials when used for their intended purpose and maintained according to instructions.
For
technical
information
e-mail
us
at
[email protected].
Visit
www.hannainst.com for
more
information
about
Hanna
Instruments
and
our
products.
Please retain for future use. QR6421P 12/22