Liquid Crystal
Display
CAL button
Operation
Thank You
Thank you for choosing a Hanna Instruments product. Please read this
instruction manual carefully before using the instrument.
For more information about Hanna Instruments and our products,
For technical support, contact your local Hanna Instruments Office or
e-mail us at tech@hannainst.com
Find your local Hanna Instruments Office on www.hannainst.com
Preliminary Examination
Remove the meter from the packing material and examine it carefully to make sure that
no damage has occurred during shipment. If noticeable damage is evident, contact your
local Hanna Instruments Office.
Each meter is supplied with:
• CR2032 battery (1 pc.)
• Storage / Protection sleeve
• Instruction manual
• Quality Certificate
Note: Save all packing material until you are sure that the instrument functions
correctly. All defective items must be returned in the original packaging with the
supplied accessories.
Preparation
The probe is shipped dry. Before using the electrode, remove the protective cap. Then
follow the calibration procedure.
• Turn the tester on by pressing ON/OFF button.
• Remove the protective cap and immerse the tip of the probe in the sample to
be tested.
• Stir gently and wait for a stability tag to disappear.
• The electrode automatically compensates for temperature variations.
• The reading on display is directly expressed in: mS/cm or ppm.
• For best results, recalibrate periodically.
• After use rinse the probe with water.
• Always replace the protective cap after each use.
Specifications
TDS Conversion factor
(CF)*
0.5 (500 ppm) or
0.7 (700 ppm)
Calibration
Solutions
HI50036 HI70031
(Quick Cal) (1.41 mS/cm)
Calibration automatic,single point (1.41 mS/cm)
Quick calibration one point calibration using Quick Cal pH and EC
one point calibration solution HI50036
Temperature
Compensation automatic, 0.0 to 50.0 ºC (32.0 to 122.0 ºF)
Battery type CR2032 Li-Ion 3V (1 pc.)
Battery life approximately 250 hours of continuous use
Environment 0 to 50 ºC (32 to 122 ºF);
RH 100% max
Weight
(without battery) 68 g (2.4 oz.)
ON/OFF button
Temperature
Sensor (inside)
EC Probe
Range EC/TDS
0.00 to 6.00 mS/cm
0 to 3000 ppm (0.5) CF
0 to 4000 ppm (0.7) CF
0.0 to 50.0 °C (32.0 to 122.0 °F)
Resolution EC/TDS
0.01 mS/cm
10 ppm (0.5) CF
10 ppm (0.7) CF
0.1 °C (0.1 °F)
Accuracy
(@25 °C/77 °F) EC/TDS ±2% F.S.
±0.5 °C (±1.0 °F)
All rights are reserved. Reproduction in whole or in part is prohibited without the
written consent of the copyright owner.
Hanna Instruments reserves the right to modify the design, construction, or
appearance of its products without advance notice.
*1000 μS/cm =500 ppm with 0.5 CF
• Quick Cal pH and EC one point calibration solution sachet (4 pcs.)
Operational Guide
To turn the meter ON and check the battery status
Press the ON/OFF button to turn the meter on. At start-up, all the LCD segments are
displayed for 1 second, then the percent indication of the remaining battery life is
displayed for another second. The meter then enters the normal measuring mode.
Note: Keeping the ON/OFF button pressed while turning the meter on will
display all LCD segments as long as the button is pressed.
To enter calibration mode
Press the CAL button. “CAL“ message is displayed.
To enter setup mode
Remove the battery cover and press the Setup button located on the side of the battery.
Meter Setup
While in measurement mode, remove the battery cover. Press the Setup button located
on the side of the battery in the battery compartment. The meter will enter in setup
mode. Press the ON/OFF button to move through setup parameters. Change option
by pressing CAL button. The default settings are: “Set t” measure unit - °C, “AOFF” -
8 min, “CAL” - standard and EC unit.
To select the temperature unit (°C/°F)
To select the temperature unit when “SET t” is displayed press the CAL button to
change between °C or °F.
To select the Auto-Off time
To select the AUTO OFF when “AOFF” is displayed press the CAL button to change
between 8 min, 60 min or --- (disabled).
To select calibration type
To select calibration type when “CAL“ is displayed press the CAL button to choose
from “CAL STD“ (standard calibration) or “CAL quic “(one-point quick calibration).
To select EC or TDS
To select TDS measurement with 0.5 TDS factor press CAL button until “ppm 500 UNIT“ is
displayed. To select TDS measurement with 0.7 TDS factor press CAL button until “ppm
700 UNIT“ is displayed. By pressing CAL button again “mS/cm EC UNIT“ is selected.
To return to measurement mode
Press the ON/OFF button.
Setup
Button
Before using Hanna Instruments products, make sure that they are entirely suitable for
your specific application and for the environment in which they are used. Any variation
introduced by the user to the supplied equipment may degrade the instrument’s
performance. For yours and the instrument safety do not use or store the instrument
in hazardous environments.
Recommendations for Users