Svantek SV 111 User manual

USER MANUAL
SV 111
VIBRATION
CALIBRATOR
Warsaw, 2022-04-21
Rev. 01.03
Copyright © 2022 SVANTEK.
All rights reserved.

2
SV 111 User Manual
Note: Because of continuous product improvement SVANTEK reserves the right to make
changes to product specifications without notice. To download the most up to date user's
manual please visit our web site at www.svantek.com. This user’s manual presents the
firmware revision named 1.16.x.
The succeeding software revisions (marked with the higher numbers) can change the view of
some displays presented in the text of the manual.
WEEE Notice: Do not throw thedevice away with theunsorted municipal waste at the end of its
life. Instead, hand it in at an official collection point for recycling. By doing this you will help to preserve
the environment.
The software described in this manual is furnished under a license agreement and may be used only in
accordance with the terms of that agreement.
Copyright Notice
Copyright © 2022 Svantek Sp. z o.o.
All rights reserved.
Reproduction without permission is prohibited.
Trademarks
Trademarks or registered marks in this manual belong to their respective manufacturers.
Microsoft and Windows are registered trademarks of Microsoft Corporation.
Disclaimer
Information in this document is subject to change without notice and does not represent a commitment
on the part of Svantek.
Svantek provides this document “as is”, without warranty of any kind, either expressed or implied,
including, but not limited to, its particular purpose. Svantek reserves the right to make improvements
and/or changes to this manual, or to the products and/or the programs described in this manual, at
any time.
Information provided in this manual is intended to be accurate and reliable. However, Svantek assumes
no responsibility for its use, or for any infringements on the rights of third parties that may result from
its use.
This product might include unintentional technical or typographical errors. Changes are periodically
made to the information herein to correct such errors, and these changes are incorporated into new
editions of the publication.
Technical Support Contact Information:
web: www.svantek.com
e-mail: [email protected]m.pl

3
SV 111 User Manual
CONTENTS
1 General safety summary .............................................................................................................4
2 Calibration ..................................................................................................................................5
3 Accuracy of calibration................................................................................................................5
4 SV 111 model information...........................................................................................................5
5 Unpacking and Inspecting the package contents .........................................................................6
6 Optional accessories ...................................................................................................................7
7 SV 111 vibration calibrator..........................................................................................................8
8 Control panel ..............................................................................................................................9
9 Operating diagram....................................................................................................................10
10 Submenu diagram.................................................................................................................11
11 Calibration of the whole-body sensor....................................................................................12
12 Calibration of the SV 105 Hand-Arm sensor...........................................................................14
13 Calibration of the SV 3023M2 accelerometer ........................................................................16
14 Calibration of the general-purpose accelerometer.................................................................17
15 Programming ........................................................................................................................18
16 General Care and Cleaning ....................................................................................................20
17 Technical data .......................................................................................................................21
18 Declaration of Conformity .....................................................................................................22
Appendix A Recalibration.............................................................................................................23
Appendix B Defining calibration coefficient..................................................................................26

4
SV 111 User Manual
1General safety summary
Review the following safety precautions to avoid injury and prevent from damaging
this product or other products connected with it. To avoid potential hazards, use this
product only as specified. Qualified personnel should only perform the service
procedures.
Warnings, precautions and maintenance:
•Before using the calibrator be sure that the shaker shipping lock is detached. Starting
the vibration calibrator with the shipping lock may destroy it.
•Use a proper AC/DC adapter, specified for this product and certified for the country of
its use.
•Keep the product’s surfaces clean and dry.
•Even when the device is not in use it is recommended to charge the battery once a
month to keep it in good condition.
•It is recommended to carry out a technical inspection and recalibration of the device
every 12 months to ensure accurate calibration level.
•Recalibration can be made by the calibration laboratory. If there is no possibility to
recalibrate the device by calibration laboratory the device should be send to the
manufacturer.
•All maintenance work and repairs can only be done by the personnel trained by the
manufacturer.
•Handle with care.
Safety terms and symbols:
Symbol
Symbol meaning
Do not throw into standard municipal waste containers. The user is obliged to
deliver used equipment to the manufacturer or to the recycling collection point.
This product can be recycled
This product has met EU consumer safety, health or environmental
requirements

5
SV 111 User Manual
2Calibration
One of the fundamental questions, that are most frequently asked while taking a
measurement, is whether its result is accurate. Proceeding, with a measurement without
having a positive answer to this question, may result in obtaining data of no practical use and
wasting our time. However, we may easily obtain the answer by performing a calibration of
the vibration level meter using a vibration calibrator. This device should be used before every
set of measurements.
The vibration calibrator is a device, which produces the vibration of the defined levels and
frequency. It allows to calibrate the vibration meter in comparative way.
Calibration procedure is also the best way for the complete measuring system (Meter, Cable
and Transducer connected together) check. This is an essential action for the reliable
measurements performed in the field!
3Accuracy of calibration
Each measurement performed by any measurement device is burdened with an error. Result
obtained from such measurement is only an estimate of the real value of the measured
quantity. Hence, the purpose of calibration is to limit this inevitable error to a certain
acceptable level. Maximum absolute value of the error of generated vibration signals is called
the tolerance and is strictly defined by the standard ISO8041:2005.
4SV 111 model information
The model SV 111 is a Portable Vibration Calibrator. It is designed to check/calibrate various
types of vibration level meters according to the ISO 8041-1:2017.
Three standard frequencies 15.92 Hz, 79.58 and 159.2 Hz (plus one more extra 636,6Hz)
combined with high loading mass brings opportunity to calibrate almost any existing
transducer.
Because of its dedicated adapters, SV 111 allows to take measurements in three directions.
So, to calibrate effectively tri-axial transducers!
Because of its own internal rechargeable battery, it is a truly mobile and flexible device
designed to use either in a laboratory or during fieldwork.
The unique feature of SV 111 is capability to calibrate complete “seat accelerometer”, without
necessity to extract accelerometer form the rubber cushion!
The robust case allows placing the calibrator directly on the ground, what is extremely useful
in the field applications. Built in electronic levelling system helps with correct positioning of
the device for correct vibration calibration. Levelling of the SV 111 is important for keeping
transversal vibrations levels within tolerances specified by the ISO 8041-1:2017.

6
SV 111 User Manual
5Unpacking and Inspecting the package contents
Despite careful packing, the risk of the device damage cannot be entirely eliminated. Upon
delivery, please make sure that the device is not damaged and verify that you received the
ordered equipment and optional accessories (if ordered). In case of any problems, please
contact an authorized Svantek representative, the service staff or the manufacturer directly.
The complete set includes the following items:
•SV 111 Vibration calibrator
•SA 33 AC/DC adapter
Before using the device:
•Be sure that the shaker shipping lock is detached. Starting the vibration calibrator with
the shipping lock may destroy it.
To detach the shaker shipping lock, unscrew four bolts with the hex wrench and two screwdriver
tips attached to the magnet. Screw the central bolt into the loose sleeve.
Don’t throw away the transport lock with three bolts because it is necessary to secure the shaker
during transportation.

7
SV 111 User Manual
•Charge SV 111 to the full. In order to charge the battery, connect SA33 AC/DC adapter
plug to SV 111 EXT_DC supply socket and then connect it to the mains.
The SA33 AC/DC adapter is placed in the charger slot.
•If the device has been kept or transported in low temperature (below 0 °C), it is
recommended to leave it for a few hours in the room temperature before connecting
it to the power supply. If any steam condensation appears, it is recommended not to
connect the device to the mains electricity for 4 to 8 hours until the exterior surface is
dry.
6Optional accessories
The SV 111 calibrator set doesn’t include the SA 105 adapter used for calibration of the
SV 105B hand-arm transducers and the SA 44 adapter for calibration of the SV 3023M2
transducers. These adapters are optional and should be ordered separately.

8
SV 111 User Manual
7SV 111 vibration calibrator
SV 111 vibration calibrator
1.Magnet with two screwdriver tips; 2.Adapter; 3.Hex wrench slot with support
sleeves; 4.Charger slot; 5.Shaker; 6.Transportation lock; 7.Support; 8.Fastening screw;
9.Display; 10.Keyboard; 11.EXT_DC supply socket; 12.USB port;
1
2
3
4
5
7
8
9
10
11
12
6

9
SV 111 User Manual
8Control panel
SV 111 control panel
1.Vibrations level button/Power on button A; 2.Start/Stop button; 3.Serial number
box; 4.Display box; 5.Vibration frequency button/Power button B; 6. Spirit level; 7.USB
port
1
2
3
4
5
6
7

10
SV 111 User Manual
9Operating diagram
SV 111
levelling
Vibration
parameters
setting
menu
Shaker
vibration
s
SV 111 switching on
Sensor
installation
Skipping the levelling
process
SV 111 levelled
properly
…
Sample vibration
parameters
Start
shaker
Stop
shaker
Vibrations
parameters not
reached properly
Shaker vibrates
with set
parameters

11
SV 111 User Manual
10 Submenu diagram
•Program submenu can be reached from “Vibrations parameters setting
menu” by pressing
•To return to “Vibrations parameters setting menu” press

12
SV 111 User Manual
11 Calibration of the whole-body sensor
The whole-body sensor should be tested/calibrated at 15.92 Hz only!
•dismantle the adapter from the case cover and screw it to the shaker with one or
more bolts (one central bolt should be sufficient for the system check)
•dismantle the support system from the case holder, put it on the adapter and tight the
fixing bolt:
•unscrew the central screw of the whole-body sensor and install the sensor on the
support system screwing them altogether:

13
SV 111 User Manual
•position the sensor according to the calibrated axis (channel) and tight the fixing bolt:
•put the support system with the tested sensor in the horizontal position on the adapter
and tight the fixing bolt:

14
SV 111 User Manual
12 Calibration of the SV 105 Hand-Arm sensor
The SV 105 Hand-Arm sensors should be calibrated at 79,58 Hz only.
The description of calibration of SV 105 Hand-Arm sensors is given based on its SV 105D
modification with the use of the SA 105D dedicated adapter. This adapter is compatible
with all previous modifications of SV 105. At the same time all previous modifications
of SV 105 can be calibrated with the use of their dedicated adapters and the calibration
description presented in the previous SV 111 User manuals.
To perform calibration of the SV 105D Hand-Arm sensor, use the SA 150C belt with the
greatest curvature.
•Screw the SA 150C belt to the SV 105D sensor with the special screw from the SV 105D
kit using for that a special 1.5mm Allen screwdriver (both included in the SV 105D kit).
If the belt was screwed to the sensor with the normal screw it is necessary to unscrew
it and screw again with the SA 105D screw.
•Apply the twisted sensor and belt to the SA 105D adapter (SV 105D must be positioned
so that the cable is facing the adapter side without thread hole). The adapter has a nut
that will tighten the SV 105D sensor to the SA 105D adapter when twisted. In order for
the belt not to dangle, it should be positioned as shown in the photo below.

15
SV 111 User Manual
•Install the calibration adapter with the vibration sensor to the calibrator’s shaker using
special stud (included in the SV 105 kit):
Positioning of the adapter with the sensor for calibration of the X-axis, Y-axis and Z-axis should
be as per photos below. For this use the appropriate hole in the adapter for the stud.
X
Y
Z

16
SV 111 User Manual
13 Calibration of the SV 3023M2 accelerometer
The SV 3023M2 accelerometer is mounted on the shaker with special adapter (SA 44). The SA 44
adapter set includes two studs M5/M5 and M5/M3. The M5/M3 stud is used for fixing the
accelerometer to the adapter. Another M5/M5 stud is used for fixing adapter with the accelerometer
to the shaker.
The SA 44 adapter allows to calibrate the SV 3023M2 accelerometer in all directions.

17
SV 111 User Manual
14 Calibration of the general-purpose accelerometer
The general-purpose accelerometer (for example, SV 80) is mounted directly on the shaker with
special stud, normally included in the accelerometer set.

18
SV 111 User Manual
15 Programming
•Switch on power by pressing two buttons at the same
time for a while
•The device type information will be displayed
•When the battery level is low, ‘LOW BATTERY’ is
displayed every 30 sec.
•Level the device according to the spirit level information.
oMove front side up
oMove front side down
oMove left side up
oMove left side down
•When the level is reached the information ‘levelling ok’
is projected on the display for 5 seconds, then
automatically switches to vibrations parameters setting
menu
It is possible to skip leveling process, but it is strongly recommended to proceed the
leveling process before every calibration.
Spirit level can be launched anytime using the program submenu.
•After leveling, the device displays the default vibrations
parameters values
•You can change values by pressing the ‘Freq.’ button in
the following order: 15,92; 79,58; 159,2; 636,6 Hz
•Depending on settled frequency value, vibration level
can be set according the table bellow
Frequency
15,92
79,58
159,2 Hz
636,6
Hz
Level
1,2
1; 2; 3; 4;
5; 6; 7; 8;
9; 10
1; 2; 3; 4;
5; 6; 7; 8;
9; 10
1
𝐦
𝐬𝟐
•When vibration parameters are set, run the shaker by
pressing ‘Start/Stop’ button.

19
SV 111 User Manual
During levelling of the calibrator, the diodes are
blinking yellow (orange)
•When the vibrations are steady, the diodes are lighting
with continuous green light and the information ‘Level
OK’ is projected on the display
+
•In special cases described below some diodes may have
red lights. The explanation of red diodes is displayed on
the screen.
THD (Total harmonic distortion) –means that the
harmonics on Z axis exceed threshold 5% (-26 dB)
of reference vibration level.
For example: When the vibrations are set with ≈16
Hz, 1 m/s2(120 dB), then the total vibrations
amount with frequencies n*16 HZ (32, 48, 64,
80,...) cannot be higher than 0,95 m/s2(94 dB)
SV 111 is compensating harmonic distortion
X Y Vibration means that vibrations’ level in X or Y
(corresponding diode is lighting red) direction are
higher than 10% (-20dB) of vibrations level
generated in Z direction
This means that both kinds of mistakes were made
at the same time.
The device cannot reach set parameters, and the
device stops generating the vibrations
•While the shaker is working, set parameters can be
checked by pressing either the ‘Level’ or ‘Freq.’ button:
Set parameters will be displayed by 5 seconds
or
•To stop the shaker, press the ‘Start/Stop’ button

20
SV 111 User Manual
•Switch off by pressing two power buttons at the same
time for a while
16 General Care and Cleaning
•Remove the sensor and switch the device off
•Disconnect the device from the power supply.
•Wipe the device’s surface with the cloth damped with the mixture of warm
water and detergent.
•After cleaning, wipe the device with dry cloth and wait until the surface is
completely dry.
•Do not immerse the device in any fluids as this may damage the device and
cause electrical shock. Only the external parts of the device should be cleaned
Other manuals for SV 111
1
Table of contents
Other Svantek Test Equipment manuals