
Safety instructions
6A2781-3.1 HBM: public CMD
SAs specified in EN 61326-1, Section 3.6, the connect
ing cables of the device must be no longer than 30 m
(when laid within a building) and must not leave the
building.
SThe CMD housing must be grounded to achieve EMC.
SComply with the maximum permissible ambient tem
peratures and the data on maximum humidity as
stated in the specifications.
SThe device must not be modified from the design or
safety engineering point of view except with our
express agreement. In particular, any repair or solder
ing work on motherboards (exchanging components)
is prohibited. When exchanging complete modules,
use only original parts from HBM.
SThe device is delivered from the factory with a fixed
hardware and software configuration. Changes can
only be made within the possibilities documented in
the manuals.
SThe device is maintenance free.
SPlease note the following points when cleaning the
housing (also see Chapter 13, page 61):
SDisconnect the device from all current and voltage
supplies before cleaning it.
SClean the housing with a soft, slightly damp (not wet!)
cloth. Never use solvent as this could damage the
labeling or the housing.
SWhen cleaning, ensure that no liquid gets into the
device or connections.
SIn accordance with national and local environmental
protection and material recovery and recycling regula
tions, old equipment that can no longer be used must