V:\HMC-Europe\Beschreibungstexte\HV\Bedienungsanleitungen\Kurzbedienungsanleitungen\Kurzbedienungsanleitung engl..doc Stand Juli 2003
Modification of other programparameters (temperature, time, exhaust)
Select program
Press switch temperatur blinks
Press switch, successsive blinks min (time), warming temperature oPN (80,90,95) and take-off-
temperature
Press switch for modifiing (blinking) parameters
Press switch for saving modivication
Press switch to start program
5. START PROGRAM
Test the feeding water level (demineralised-water)
Choose the suitable program.
Liquid sterilisation: Look for correct take-of-temperature.
Place the flex. temperature sensor in suitable vessel, therefore turn the lock/unlock lever
to (left) „LOCK“ position
Press the switch
First you see the take-off-temperature and the steamsterilizer starts. „HEATG“ blinks.
6. STOP PROGRAM
Press the switch
The program can be stopped at every time .After reaching the selected lid-opening temperature and the ambient air
pressure (Check that the pressure is at “0Mpa” ) the lid can be opened. Turn the
lock/unlock lever
to (right) „UNLOCK“
position. Open the lid
7. PRINTER
The printer will be actevated with
Func-Switch and documents the program parameter
date, sterilising program, time and exhaust speed.
Description of paper leading edge
Activate the printer.
Press careful the bottom of printer door and open it.
Cut the paper roll straight.
Set the roll exactly as shown in the drawing. Then, slowly insert the leading edge, square and
flat, into the leading slot and press the feed switch.
8. EXHAUST AIR FILTER
The exhaust air filter system filters the steam-air mixture when sterilising microbiological contaminated material and waste .
The filter system and the filter cartridge is sterilised with each cycle. („In-line“-sterilisation).
The filter cartridge is integrated in a stainless steel case.
This filter case you find inside of the autoclave, it is outside not visible.
The filter system don’t need any adjustment. The average running time is 1.000 hours. The filter cartridges should be
changed regularly. Therefore replace the right cover of the steamsterilizer. You can order Filter cartridges (Part.No.50019)
on HMC Europe or your distributor.
9. ERROR
If you press switches to fast or wrong, you hear a periodic peep-sound and the programming can’t continue. You cancel this
error when you turn short time the
lock/unlock lever
to (right) „UNLOCK“-position.
Error descriptions in details you find in the operation-manual.
European Service Center
HMC Europe GmbH, Hafing 21, D-84549 Engelsberg, Germany
Tel.: 0049 08634/625 994, Fax.: 0049 08634/625 996, email: info@hmc-europe.com