HP 41802A Quick start guide

HP 41802A 1 M Input Adapter
Operation Note
SERIAL NUMBERS
This manual applies directly to HP 41802As with serial number prex 3103J.For additional
important information about serial numbers, read \Serial Number" in Chapter 1.
ABCDE
HP Part No. 41802-90000
Printed in Japan February, 1999
2nd Edition
c

Copyright 1991,1999 Hewlett-Packard Japan, LTD.
Certication
Hewlett-Packard Company certies that this product met its published specications at the
time of shipment from the factory. Hewlett-Packard further certies that its calibration
measurements are traceable to the United States National Institute for Standards and
Technology, to the extent allowed by the Institution's calibration facility, or to the calibration
facilities of other International Standards Organization members.
Warranty and Assistance
All Hewlett-Packard products are warranted against defects in materials and workmanship.
This warranty applies for one year from the date of delivery,or, in the case of certain
major components listed in the operating manual, for the specied period. We will repair or
replace products which prove to be defective during the warranty period provided they are
returned to Hewlett-Packard. No other warranty is expressed or implied. We are not liable for
consequential damages.
For any assistance, contact your nearest Hewlett-Packard Sales and Service Oce.Addresses
are provided at the back of this manual.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper or adequate
maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modication or
misuse, operation outside of the environment specications for the product, or improper site
preparation or maintenance.
No other warranty is expressed or implied. HP specically disclaims the implied warranties
of merchantability and tness for a particular purpose.
iii
Caution
The HP 41802A is sensitive to electrostatic discharge (ESD). The followings
must be adhered to when using the HP 41802A.
Do
NOT
touch the center conductor of the BNC connector of the HP
41802A.
Do
NOT
touch the pin of the probe which is connected to the HP 41802A.
Eliminate ESD on your body.
Eliminate ESD on the work surface.
Do
NOT
introduce ESD into the DUT, while the HP 41802A and the probe is
in use.
iv
Contents
1. General Information
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Compatible Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Compatible Instruments .......................... 1-2
Compatible Probes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Serial Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Specications ................................ 1-4
Supplemental Performance Characteristics .... .. .. .. .. .. .. .. 1-5
2. Installation
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Mating Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Environmental Requirements . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
3. Operation
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Operating Precautions ............................ 3-2
Anti-static Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Maximum Allowable Level . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Discharging the Probe .... .. .. .. .. .. .. .. .. .. .. .. . 3-3
Probe Power Plug .... .. .. .. .. .. .. .. .. .. .. .. .. . 3-3
Preparation for Use ............................. 3-3
Operating Check Using a Network Analyzer . . . . . . . . . . . . . . . . . 3-4
Using an HP 8751A or HP 3577A/B Network Analyzer . . . . . . . . . . . 3-4
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Using HP 8753A/B/C Network Analyzer . . . . . . . . . . . . . . . . . . 3-5
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Operating Check Using a Spectrum Analyzer .... .. .. .. .. .. .. 3-6
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Operating Check Using a Network/Spectrum Analyzer ............ 3-7
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Adjusting the Probe ............................ 3-8
Using a Network Analyzer .. .. .. .. .. .. .. .. .. .. .. . . 3-8
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Using a Spectrum Analyzer . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Typical Measurement Setups .... .. .. .. .. .. .. .. .. .. .. . 3-11
Network Measurements .......................... 3-12
Contents-1
Using One HP 41802A .......................... 3-12
Using One HP 41802A with a Transmission/Reection Test Set .... .. . 3-13
Using Two HP 41802As ......................... 3-14
Spectrum Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Using HP 41802A with instruments which have BNC connectors .. .. .. . 3-16
4. Performance Test
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Calibration Cycle .............................. 4-2
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Gain Accuracy/Flatness Tests . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Description ................................ 4-3
Specications ............................... 4-3
Test Equipment .............................. 4-3
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Performance Test Record . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
Gain Accuracy/Flatness Tests . . . . . . . . . . . . . . . . . . . . . . . . 4-6
5. Adjustment
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Equipment Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Gain Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Description ................................ 5-2
Equipment ................................ 5-2
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
6. Replaceable Parts
Replaceable Parts Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
A. Manual Changes
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Manual Changes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1
Contents-2
Figures
1-1. Serial Number Plate .. .. .. .. .. .. .. .. .. .. .. .. .. .. 1-3
1-2. Specications .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . 1-4
1-3. Supplemental Characteristics . . . . . . . . . . . . . . . . . . . . . . . . 1-5
2-1. Product Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
2-2. Probe Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
3-1. Operating Check Setup Using HP 8751A or HP 3577A/B Network Analyzer .. 3-4
3-2. Operating Check Setup Using HP 8753A/B/C Network Analyzer . . . . . . . . 3-5
3-3. Operating Check Setup Using Spectrum Analyzer .... .. .. .. .. .. 3-6
3-4. Operating Check Setup Using Network/Spectrum Analyzer . . . . . . . . . . 3-7
3-5. Probe Adjustment Setup Using a Network Analyzer|1 . . . . . . . . . . . . 3-8
3-6. Probe Adjustment Setup Using a Network Analyzer|2 . . . . . . . . . . . . 3-9
3-7. Probe Adjustment Setup Using Spectrum Analyzer . . . . . . . . . . . . . . 3-10
3-8. Network Measurement Setup Example (Using One HP 41802A) . . . . . . . . 3-12
3-9. Network Measurement Setup Example (Using with a Transmission/Reection Set) 3-13
3-10. Network Measurement Setup Example (Using Two HP 41802As) .... .. . 3-14
3-11. Spectrum Measurement Setup Example . . . . . . . . . . . . . . . . . . . 3-15
3-12. Measurement Setup Example (BNC input connector) . . . . . . . . . . . . . 3-16
4-1. Test Setup 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4-2. Test Setup 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
5-1. Adjustment Setup 1 ............................ 5-2
5-2. Adjustment Setup 2 ............................ 5-3
5-3. Adjustment Setup 3 ............................ 5-4
6-1. Replaceable Parts (1 of 2) ......................... 6-2
6-2. Replaceable Parts (2 of 2) ......................... 6-4
Tables
2-1. Probe Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
3-1. Anti-static Products Available ....................... 3-2
4-1. Recommended Test Equipment . . . . . . . . . . . . . . . . . . . . . . . 4-1
6-1. Component Manufactures .... .. .. .. .. .. .. .. .. .. .. . 6-1
6-2. Replaceable Parts (2 of 2) ......................... 6-3
6-3. Replaceable Parts (2 of 2) ......................... 6-4
A-1. Manual Changes by Serial Number ..................... A-1
Contents-3
1
General Information
Introduction
The HP 41802A 1 M Input Adapter is used with high impedance passive probes for circuit
signal analysis using network and spectrum analyzers.
This is an operation note for the HP 41802A, containing information on installation, operation,
and service in the following chapters.
Chapter 1, General Information
Provides the specications and the information necessary for preparing the HP 41802A for
use.
Chapter 2, Installation
Provides the installation information, including initial inspection, power requirements
,
mating connectors, and environmental requirements.
Chapter 3, Operation
Provides the information for preparation and typical measurement setups
, including cautions
on use.
Chapter 4, Performance Test
Provides performance test procedures to ensure that the HP 41802A is within specications
.
Chapter 5, Adjustment
Provides the adjustment procedure to bring it within specications
, if the HP 41802A failed
the performance test or it has been repaired.
Chapter 6, Replaceable Parts
Provides the replaceable parts information to service the HP 41802A.
General Information 1-1
Compatible Equipment
Compatible Instruments
Instruments compatible with the HP 41802A are network and spectrum analyzers which meet
the following conditions.
Measurement frequency range includes 5 Hz to 100 MHz.
Input resistance is 50 .
Probe power supplies are furnished.
The following HP network and spectrum analyzers are compatible with the HP 41802A.
Network Analyzers: HP 8751A, HP 8753A/B/C, HP 3577A/B
Spectrum Analyzers: HP 3585A/B, HP 8568B
Network/Spectrum Analyzer: HP 4195A
If the instrument does not have a probe power supply, use a separate power supply which
meets the requirements described in \Power Requirements" in Chapter 2.
Compatible Probes
Compatible probes with the HP 41802A are high impedance passive probes which meet the
following conditions.
Input resistance is 1 M.
Compensates oscilloscope input capacitance includes 12 pF.
The following HP Oscilloscope Probes are compatible with the HP 41802A.
HP 10432A, HP 10433A, HP 10435A, HP 10440A
HP 10017A, HP 10018A, HP 10080A, HP 10081A
Refer to \Adjusting the Probe" in Chapter 3 how to adjust the probe which is connected to a
network analyzer or a spectrum analyzer.
1-2 General Information
Serial Number
Hewlett-Packard uses a two-part, nine character serial number which is stamped on the serial
number plate (see Figure 1-1) attached to the instrument's rear panel. The rst four digits
and a letter are the prex, and the last ve digits are the sux of the serial number. The
letter in the serial number identies the country where the instrument was manufactured.
The prex is same for all identical instruments; it changes only when a change is made to the
instrument. The sux, however, is assigned sequentially and is dierent for each instrument.
This operation note applies to instruments with the serial number prexes listed under SERIAL
NUMBERS on the title page.
Figure 1-1. Serial Number Plate
Units manufactured after this operation note was printed may have a serial number prex that
is not listed on the title page. An unlisted serial number prex indicates that the instrument
may be dierent from those described in this operation note
. Operation notes for new
instruments may be accompanied by a yellow Manual Supplement page or have a dierent
manual part number. This supplement contains \Change Information" explaining how to adapt
this operation note to the newer instruments.
In addition to change information, the supplement may contain information for correcting
errors (Errata) in the previous operation notes.To keep this manual as current and accurate as
possible, Hewlett-Packard recommends that you periodically request the latest Manual Changes
supplements. The supplement for this operation note is identied by the Print Date and HP
Part No., both of which appear on the operation note's title page.
For information concerning the serial number prex not listed on the title page or in the
Manual Change supplements, contact your nearest Hewlett-Packard oce.
General Information 1-3
Specications
Figure 1-2 lists the specications for the HP 41802A. The specications are performance
standards or limits. The HP 41802A meets all of the specications listed in Figure 1-2, when it
is shipped from the factory.
Specications describe the instrument's warranted performance over the temperature
range of 23
6
5

C (except where noted).
Supplemental characteristics are intended to provide information useful in applying the
instrument by giving non-warranted performance parameters. These are denoted as
\typical," \nominal," or \approximate."
Frequency Range:
5 Hz to 100 MHz
Accuracy
1
: (Input Level =
0
3 dBm, at 1 MHz)
0dB
6
2dB
0dB
6
0.5 dB (typical)
Flatness
1
(relative to 1 MHz): (Input Level =
0
3 dBm)
6
2.8 dB
6
1 dB (typical, at 50 Hz to 100 MHz)
+1/
0
2 dB (typical, at 5 Hz to 50 Hz)
Damage Level:
>
2 Vrms,
>
50 Vdc
Power:
+15 V, 45 mA nominal
0
12.6 V, 40 mA nominal
Operating Temperature:
0to55

C
Operating Relative Humidity:

95% at 40

C (Non-condensing)
Dimension:
approximately 30(H) x 42(W) x 116(D) mm (excluding the cable)
Weight:
approximately 250 g
1
The performance is specied when the HP 41802A is terminated with an HP 10100C 50
Feedthrough Termination.
Figure 1-2. Specications
1-4 General Information
Supplemental Performance Characteristics
Figure 1-3 lists the supplemental performance characteristics for the HP 41802A. Supplemental
performance characteristics are not specications, but are typical characteristics included as
additional information for the operator. Supplemental performance characteristics are not
guaranteed.
Supplemental Characteristics describe the instrument's performance over the
temperature range of 23
6
5

C (except where noted).
1 dB Gain Compression
1
: (Input Level = +3 dBm)
0.32 Vrms (at 1 MHz)
Input Resistance:
1M
Input Capacitance:
12 pF
1
The following performance is specied when the HP 41802A is terminated with an HP
10100C 50  Feedthrough Termination.
Figure 1-3. Supplemental Characteristics
General Information 1-5
2
Installation
Introduction
This chapter contains the following information.
Initial inspection
Power Requirements
Mating connector
Environmental requirements
Installation 2-1
Initial Inspection
Caution
Electrostatic Discharge (ESD) can damage the HP 41802A's highly sensitive
input amplier.Do
NOT
touch the center conductor of the BNC connector of
the HP 41802A.
The HP 41802A meets all of the specications listed in Figure 1-2. Upon receipt, inspect
the shipping container for damage. If the shipping container or the cushioning has been
damaged, keep the container and packing material until the contents have been checked
for completeness and the HP 41802A has been mechanically and electrically checked out.
Figure 2-1 shows the product overview of the HP 41802A. The procedures for checking the
general electrical operation are given in Chapter 4.
If anything is damaged (scratches, dents, broken connectors, etc.), or if performance does not
meet the specied performance test limits, notify the nearest HP sales oce (see the list at
the back of this operation note). The HP sales oce will immediately arrange for repair or
replacement without waiting for a claim settlement.
Figure 2-1. Product Overview
2-2 Installation
Power Requirements
Power for the HP 41802A is supplied by the compatible instruments described in \Compatible
Instruments" in Chapter 1 by connecting the HP 41802A's power plug to the probe power jack
on the instrument. If the instrument used with the HP 41802A does not have a probe power
supply, use a separate power supply which meets the requirements listed in Table 2-1 . The HP
1122A and HP 11899A Probe Power Supply meets these requirements and it accepts the HP
41802A's probe power plug directly.
Figure 2-2. Probe Power Requirements
Table 2-1. Probe Power Requirements
Pin Voltage Current
1
0
12.6 V
6
10% 40 mA
2 Ground
3 +15 V
6
6% 45 mA
Mating Connectors
The output connector of the HP 41802A is a 50  N-type male connector
.
Note
Keep the N-type output connector clean.
Caution
Trying to mate this 50  N-type connector to a 75  N-type connector will
result in damage to both connectors.
Installation 2-3
Environmental Requirements
The HP 41802A may be stored or shipped under the following environmental conditions.
Temperature:
0
40

C to +70

C
Humidity:

95% at 40

C (Non-condensing)
The unit must be protected from temperature extremes which can cause condensation.
2-4 Installation
3
Operation
Introduction
This chapter provides the following information.
Operating precautions
Preparation for use
Typical measurement setups
Operation 3-1
Operating Precautions
This paragraph describes precautions for using the HP 41802A to prevent from damage on the
HP 41802A and your device under test (DUT).
Anti-static Precautions
Caution
The HP 41802A is sensitive to electrostatic discharge (ESD). The following must
be adhered to when using the HP 41802A.
Do
NOT
touch the center conductor of the BNC connector of the HP
41802A.
Do
NOT
touch the pin of the probe which is connected to the HP 41802A.
The probe input circuit is highly susceptible to damage by ESD introduced
through the input connector.
Eliminate ESD on your body.
Wear a snug-tting ground strap that is connected to earth ground through a
high resistance.Table 3-1 lists the available anti-static products.
Eliminate ESD on the work surface.
A grounded anti-static bench mat is recommended. Optional oor mats
provide an extra measure of protection, especially in areas with oor carpet.
Do
NOT
use the HP 41802A on a carpeted work surface
.
Do
NOT
introduce ESD into the DUT while the probe is in use
.
If an unprotected person touches a part of the DUT
, a static surge could
damage the DUT as well as the probe.
Table 3-1. Anti-static Products Available
Description HP Part Number
Liquid Anti-static Agent 8550-3397
Static Control Table Mat with Ground Cord 9300-0797
Small Wrist Strap with Ground Cord 9300-1099
Medium Wrist Strap with Ground Cord 9300-1117
Large Wrist Strap with Ground Cord 9300-1242
3-2 Operation

Other HP Adapter manuals

HP NC6170 User manual

HP

HP NC6170 User manual

HP Ethernet 10Gb 2-port 530T Adapter User manual

HP

HP Ethernet 10Gb 2-port 530T Adapter User manual

HP E2492A User manual

HP

HP E2492A User manual

HP Compaq NC6136 User instructions

HP

HP Compaq NC6136 User instructions

HP 647590-B21 User manual

HP

HP 647590-B21 User manual

HP 381513-B21 - Smart Array P800 Controller... User manual

HP

HP 381513-B21 - Smart Array P800 Controller... User manual

HP 90W User manual

HP

HP 90W User manual

HP BT400 User manual

HP

HP BT400 User manual

HP 571SFP User manual

HP

HP 571SFP User manual

HP FlexFabric 650FLB User instructions

HP

HP FlexFabric 650FLB User instructions

HP NC1020 User manual

HP

HP NC1020 User manual

HP NC364m User instructions

HP

HP NC364m User instructions

HP NC364m User instructions

HP

HP NC364m User instructions

HP 366T User manual

HP

HP 366T User manual

HP HP A5513A Parts list manual

HP

HP HP A5513A Parts list manual

HP 530SFP+ User manual

HP

HP 530SFP+ User manual

HP PP2080 User manual

HP

HP PP2080 User manual

HP TouchSmart 300-1000 - Desktop PC User manual

HP

HP TouchSmart 300-1000 - Desktop PC User manual

HP ScanJet 100 User manual

HP

HP ScanJet 100 User manual

HP 561T User manual

HP

HP 561T User manual

HP Ethernet 1Gb 4-port 366M User instructions

HP

HP Ethernet 1Gb 4-port 366M User instructions

HP P1980A User manual

HP

HP P1980A User manual

HP NC380T User manual

HP

HP NC380T User manual

HP 562FLR-SFP+ User instructions

HP

HP 562FLR-SFP+ User instructions

Popular Adapter manuals by other brands

Brandywine NTA-100 user guide

Brandywine

Brandywine NTA-100 user guide

Gateway WBU-110 installation guide

Gateway

Gateway WBU-110 installation guide

Rockwell Automation Allen-Bradley PowerFlex 20-COMM-B user manual

Rockwell Automation

Rockwell Automation Allen-Bradley PowerFlex 20-COMM-B user manual

Linksys PAP2-T Setup and configuration guide

Linksys

Linksys PAP2-T Setup and configuration guide

TP-Link TL-WA890EA user guide

TP-Link

TP-Link TL-WA890EA user guide

One Stop Systems Superswitch OSS-PCIe-HIB5-x8 user manual

One Stop Systems

One Stop Systems Superswitch OSS-PCIe-HIB5-x8 user manual

Scientific Atlanta WebSTAR DPW941 user guide

Scientific Atlanta

Scientific Atlanta WebSTAR DPW941 user guide

Savio AK-38 user manual

Savio

Savio AK-38 user manual

MoCA KMC1000 Quick installation guide

MoCA

MoCA KMC1000 Quick installation guide

Adaptec Adaptec RAID Controller user guide

Adaptec

Adaptec Adaptec RAID Controller user guide

Sena Rhio232 user manual

Sena

Sena Rhio232 user manual

Sony CLIE PEGA-AC10 operating instructions

Sony

Sony CLIE PEGA-AC10 operating instructions

ANTAIRA USB-14010-SI installation guide

ANTAIRA

ANTAIRA USB-14010-SI installation guide

Gefen VGA-2-DVI2 user manual

Gefen

Gefen VGA-2-DVI2 user manual

Digitus eSATA CARDBUS ADAPTER user manual

Digitus

Digitus eSATA CARDBUS ADAPTER user manual

Humminbird 1810300 Information

Humminbird

Humminbird 1810300 Information

ZyXEL Communications NWD210N quick start guide

ZyXEL Communications

ZyXEL Communications NWD210N quick start guide

ZyXEL Communications ZyXEL ZyAIR G-260 user guide

ZyXEL Communications

ZyXEL Communications ZyXEL ZyAIR G-260 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.