HP 4262A Service manual

OPERATING
AND
SERVICE
MANUAL
SERIAL
NUMBERS
This
manual
applies
directly
to
instruments
with
serial
numbers
prefixed
2514J
With
changes
described
in
Section
VS,
this
manual
also
applies
to
instruments
with
serial
numbers
prefixed
171QJ,
1739J,
and
2022
J
For
additional
important
information
about
serial
numbers,
see
INSTRUMENTS
COVERED
BY
MAN¬
UAL
in
Section
L
©COPYRIGHT:
YOKOGAWA-HEWLETT-PACKARD,
LTD.,
1977
9-1.
TAKAKURA-CHO,
HACHIOJI-SHI,
TOKYO,
JAPAN
Manual
Part
No.
04262-90007
Microfiche
Part
No.
04262-90057
Printed
:
JUN.

The
following
general
safety
precautions
must
be
observed
during
all
phases
of
op¬
eration,
service,
and
repair
of
this
instrument.
Failure
to
comply
with
these
pre¬
cautions
or
with^
specific
warnings
given
elsewhere
in
this
manual
violates
safety
standards
of
design,
manufacture,
and
intended
use
of
the
instrument.
Hewlett-
Packard
Company
assumes
no
liability
for
the
customer's
failure
to
comply
with
these
requirements.
GROUND
THE
INSTRUMENT
To
minimize
shock
hazard,
the
instrument
chassis
and
cabinet
must
be
connected
to
an
electrical
ground.
The
instrument
is
equipped
with
a
three-conductor
ac
power
cable.
The
power
cable
must
either
be
plugged
into
an
approved
three-contact
electrical
outlet
or
used
with
a
three-contact
to
two-contact
adapter
with
the
grounding
wire
(green)
firmly
connected
to
an
electrical
ground
(safety
ground)
at
the
power
outlet.
The
power
jack
and
the
mating
plug
of
the
power
cable
meet
International
Electrotechnical
Commission
(IEC)
safety
standards.
DO
NOT
OPERATE
IN
AN
EXPLOSIVE
ATMOSPHERE
Do
not
operate
the
instrument
in
the
presence
of
flammable
gases
or
fumes.
Op¬
eration
of
any
electrical
instrument
in
such
an
environment
constitutes
a
definite
safety
hazard.
KEEP
AWAY
FROM
LIVE
CIRCUITS
Operating
personnel
must
not
remove
instrument
covers.
Component
replacement
and
internal
adjustments
must
be
made
by
qualified
maintenance
personnel.
Do
not
replace
components
with
power
cable
connected.
Under
certain
conditions,
dan¬
gerous
voltages
may
exist
even
with
the
power
cable
removed.
To
avoid
injuries,
always
disconnect
power
and
discharge
circuits
before
touching
them.
DO
NOT
SERVICE
OR
ADJUST
ALONE
Do
not
attempt
internal
service
or
adjustment
unless
another
person,
capable
of
rendering
first
aid
and
resuscitation,
is
present.
Because
of
the
danger
of
introducing
additional
hazards,
do
not
install
substitute
parts
or
perform
any
unauthorized
modification
to
the
instrument.
Return
the
in¬
strument
to
a
Hewlett-Packard
Sales
and
Service
Office
for
service
and
repair
to
ensure
that
safety
features
are
maintained.
DANGEROUS
PROCEDURE
WARNINGS
Warnings,
such
as
the
example
below,
precede
potentially
dangerous
procedures
throughout
this
manual.
Instructions
contained
in
the
warnings
must
be
followed.
WARNING
Dangerous
voltages,
capable
of
causing
death,
are
present
in
this
instru¬
ment,
Use
extreme
caution
when
handling,
testing,
and
adjusting.

Herstellerbescheinigung
Hiermit
wird
bescheinigt,
dafi
das
Gerat
HP
4262A
digital
LCR
meter
in
Ubereinstimmung
mit
den
Bestimmungen
von
Postverfiigung
1046/84
funkentstort
ist.
Der
Deutschen
Bundespost
wurde
das
Inverkehrbringen
dieses
Cerates
angezeigt
und
die
Berechtigung
zur
Uberpriifung
der
Serie
auf
Einhaltung
der
Bestimmungen
eingeraumt.
Anm:
Werden
Mefi-
und
Testgerate
mit
ungeschirmten
Kabeln
und/oder
in
offenen
Mefiaufbauten
verwendet,
so
ist
vom
Betreiber
sicherzustellen,
dafi
die
Funk-
Entstorbestimmungen
unter
Betriebsbedingungen
an
seiner
Grundstiicksgrenze
eingehalten
werden.
GERAUSCHEMISSION
Lpa
<
70
dB
am
Arbeitsplatz
normaler
Betrieb
nach
DIN
45635
T.
19
Manufacturer’s
Declaration
This
is
to
certify
that
this
product,
the
HP
4262A
digital
LCR
meter,
meets
the
radio
frequency
interference
requirements
of
directive
1046/84.
The
German
Bundespost
has
b&f
notified
that
this
equipment
was
put
into
circulation
and
was
granted
the
right
to
checj^.
fch^>
product
type
for
compliance
with
these
requirements.
r
Note:
If
test
and
measurement
equipment
is
operated
with
unshielded
cables
and/or
used
fot^
measurements
on
open
set-ups,
the
user
must
insure
that
under
these
operating
conditions,
the
radio
frequency
interference
limits
are
met
at
the
border
of
his
premises.
,.Ja
ACOUSTIC
NOISE
EMISSION
Lpa
<
70
dB
operator
position
normal
operation
per
ISO
7779

OPERATING
AND
SERVICE
MANUAL
MODEL
4262A
LCR
METER
(including
Options
001,004,
010,
and
101)
SERIAL
NUMBERS
This
manual
applies
directly
to
instruments
with
serial
numbers
prefixed
2514J
With
changes
described
in
Section
VH,
this
manual
also
applies
to
instruments
with
serial
numbers
prefixed
1710J,
1739J,
and
2022J
For
additional
important
information
about
serial
numbers,
see
INSTRUMENTS
COVERED
BY
MAN¬
UAL
in
Section
I.
©COPYRIGHT:
YOKOGAWA-HEWLETT-PACKARD,
LTD.,
1977
9-1,
TAKAKURA-CHO,
HACHIOJI-SHI,
TOKYO,
JAPAN
Manual
Part
No.
04262-90007
Microfiche
Part
No.
04262-90057
Printed
:
JUN.
1987

Table
of
Contents
Model
4262A
TABLE
OF
CONTENTS
Section
Title
Page
I
GENERAL
INFORMATION...
1-1
1-1.
Introduction...
1-1
1-4.
Description.
1-1
1-8.
Specifications.
1-7
1-10.
Safety
Considerations.
1-7
1-13.
Instruments
Covered
by
Manual
1-7
1-18.
Options.
1-8
1-20.
Option
001.
1-8
1-22,
Option
004
.
1-8
1-24.
Option
101....
1-8
1-26.
Option
010.
1-8
1-28.
Other
Options.
1-8
1-31.
Accessories
Supplied...
1-8
1-
33.
Equipment
Available
...
1-8
n
INSTALLATION...
2-1
2-
1.
Introduction
.
2-1
2-3.
Initial
Inspection..
2-1
2-5.
Preparation
for
Use
..
2-1
2-6.
Power
Requirements
..........
2-1
2-8.
Line
Voltage
and
Fuse
Selection
2-1
2-10.
Power
Cable
...
2-1
2-14.
Interconnections..............
2-3
2-16.
Operating'Environment........
2-3
2-19.
Installation
Instructions
.......
2-3
2-21.
Installation
of
Options
907,
908
and
909
......
2-3
2-23.
Storage
and
Shipment
.............
2-3
2-24.
Environment...
2-3
2-26.
Packaging
..
2-3
2-29.
Option
Installation
................
2-6
2-31.
Option
001
BCD
Data
Output
Installation..
2-6
2-32.
Option
004
Comparator
Installation...
2-6
2-33.
Coupling
Option
004
Comparator
with
Option
001
BCD
Data
Output
Installation.
2-6
2-34.
Option
101
HP-IB
Remote
Control
and
Data
Output
Installation...
2-6
HI
OPERATION........
3-1
3-1.
Introduction...
3-1
3-3.
Panel
Features
...
3-1
3-5.
Self
Test
(Basic
Operating
Check)
..
3-1
3-7.
Test
Signals.
3-6
3-9.
Measurement
Range
...
3-6
3-11.
Initial
Display
Test
...
3-7
3-13.
Initial
Control
Settings...
3-7
3-15.
D/Q
Measurement.
3-7
3-17.
LCR
Measurement..
3-7
3-19.
D/Q
Blanking
Function
(Switch
selectable
function
inside
cabinet).
3-9
3-21.
General
Component
Measurement
..
3-9
3-23.
Semiconductor
Device
Measurement
3-9
3~25.
External
DC
Bias
.................
3-9
3-28.
External
Triggering
..............
3-9
Section
Title
Page
3-30.
Terminal
Configuration.3-10
3-32.
Offset
Adjustment
.3-10
3-40.
Option
Operation
.
3-27
3-42.
Option
001:
BCD
Parallel
Data
Output
..
3-27
3-44.
Output
Data
and
Pin
Assignment.
3-27
3-46.
Alternate
Output
of
LCR
and
D/Q
Data.
3-27
3-47.
Output
Timing...
3-27
3-49.
Option
004-COMPARATOR
.
3-30
3-51.
Front
Panel
Features
(Figure
3-42)
..
3-30
3-52.
LIMIT
Setting
Warning..
3-31
3-53.
OUTPUT
Connector
Decision
Output
.
3-31
3-60.
Option
101:
HP-EB...
3-32
3-62.
Connection
to
HP-IB
Controller
3-32
3-63.
HP-IB
Status
Indicator..
3-32
3-64.
LOCAL
Switch...
3
-32
3-65.
HP-IB
Interface
Capabilities
.
„
3-32
3-66.
Source
and
Acceptor
Handshake:
SHI,
AH1.
„
3-32
3-67.
Talker
Capability:
T5
...
3-32
3-68.
Functions
Related
to
Talker
Capability
.....
3-33
3-69.
Listener
Capability:
L4
.
3-33
3-70,
Service
Request
Capability:
SRI
.......
3-33
3-71.
Remote/Local
Capability:
RLl.......
3-36
3-72.
Device
Clear
Capability:
DC1.......
3-36
3-73.
Device
Trigger
Capability:
DTI..
3-36
3-74.
ADDRESS
Switch.
3-36
3-
75.
Remote
Message
Coding.
3-36
IV
PERFORMANCE
TEST
..................
4-1
4-
1.
Introduction...
4-1
4-3.
Equipment
Required.
4-1
4-5.
Test
Record
........
.
.
.
.
.
4-1
4-7.
Calibration
Cycle.
4-1
•-
Preliminary
Operations
..
4-2
-—-
Calibration
of
DUT's
..4-4
4-9.
Measurement
Frequency
Test......
4-7
4-10.
Capacitance
Accuracy
Test...
4-8
4-11.
Resistance/ESR
Accuracy
Test....
4-12
4-12.
Dissipation
Factor
Confirmation
Check
.
4-14
4-13.
Inductance
Accuracy
Test.
4-17
4-14.
Internal
DC
Bias
Source
Test.
4-19
4-15.
Offset
Adjustment
Test...
4-20
4-16,
Comparator
Test
(Option
004
Only)
4-22
4-17.
HP-IB
Interface
Test
(Option
101
Only)...............
4-25

List
of
Tables
list
of
Illustrations
Model
4262A
LIST
OF
TABLES
Number
Title
Page
Number
Title
Page
1-1.
Specifications
...
.
1-2
5-1.
Adjustable
Components...........
.
5-2
1-2.
General
Information...
.
1-6
5-2.
Factory
Selected
Components.
.
5-3
1-3.
Equipment
Available.
.
1-9
5-3.
DUT's
Recommended
for
1-4.
Recommended
Test
Equipment
....
1-10
Making
Adjustments
...
.
5-4
5-4.
Adjustment
Requirements.»„.
.
5-5
2-1.
Option
Components
..............
.
2-8
6-1.
List
of
Reference
Designators
and
Abbreviations.
.
6-1
3-1.
Test
Signal
Level..
3-6
6-2.
Manufacturers
Code
List
.........
.
6-2
3-2.
Measurement
Ranges..
3-6
6-3.
Replaceable
Parts......
.
6-3
3-3.
Annunciation
Display
Meanings....
3-12
6-4.
Option
010
Modification
..
.
6-24
3-4.
Unusual
Operating
Indications
.....
3-21
7-1.
Manual
Changes
by
Serial
Number
.
.
7-1
3-60.
Remote
Program
Codes..
3-32
7-2.
Summary
of
Changes
by
Assembly.
.
7-1
3-61.
Remote
Message
Coding...
3-33
8-1.
Currently
Available
Options
......
8-14
4-1.
Recommended
Components
for
8-2.
Symptoms
Likely
to
Mislead..
8-18
Accuracy
Checks
............
4-6
8-3.
Front
Panel
Symptoms
of
Internal
Control
Misadjustment..
8-19
8-4.
Front
Panel
Isolation
Procedure
..
8-20
LIST
OF
ILLUSTRATIONS
Number
Title
Page
Number
Title
Page
1-1.
Model
4262A
and
Accessories
.....
.
1-1
4-4.
Dissipation
Factor
Confirmation
Check
1-2.
Serial
Number
Plate
.............
.
1-7
Setups
.......
,4-14
4-5.
Inductance
Accuracy
Test
Setup
...
4-17
2-1.
Voltage
and
Fuse
Selection........
.
2-2
4-6.
Internal
DC
Bias
Source
2-2.
Power
Cable..
.
2-4
Test
Setup..
4-19
2-3.
Rack
Mount
Kit..
.
2-5
4-7.
Offset
Adjustment
Test
Setup......
4-20
2-4.
Option
Installation
Illustrations....
.
2-7
4-8.
Comparator
Test
Setups..........
4-22
4-9.
Comparator
Output
(J6)
data
format
4-24
3-1.
Front
Panel
Features
............
.
3-2
4-10.
HP-IB
Interface
Test
Setup
...
4-25
3-2.
Rear
Panel
Features
.............
.
3-4
4-11.
SRQ
Service
Routing
.............
4-29
3-3.
Test
Fixture
and
Leads....____
.
3-8
3-4.
Measurement
Error
due
to
5-1.
Power
Supply
Voltage
Adjustment...
,
5-6
Misadjusted
ZERO
ADJ
controls
.
.3-10
5-2.
Nanoprocessor
Operating
Power
3-5.
Conversion
between
Parallel
and
Voltage
Adjustment
Location
.....
,
5-7
Series
Equivalents..
.3-12
5-3.
A12
Board
Offset
Adjustment.
,
5-8
3-6.
Relationship
of
Dissipation
to
5-4.
Waveform
at
A.12Q11
Source
.......
,
5-9
Series
and
Parallel
Resistance...
.3-13
5-5.
A13
Board
Offset
Adjustment.......
,
5-9
3-7.
General
Component
Measurements.
.3-15
5-6.
Waveform
at
A13TP1
............
5-10
3-8.
Semiconductor
Device
Measurement
3-18
5-7.
Waveform
at
A13TP2
............
5-11
3-9.
External
DC
Bias
Circuits...
.3-20
5-8.
Waveform
at
A13TP3
..
5-12
3-40.
Pin
Assignments
of
Output
5-9.
A14
Phase
Detector
&
Integrator
Connector
and
Output
Format....
.3-28
Adjustment..
5-13
3-41.
Timing
Chart
of
BCD
Data
Output..
.3-29
5-10.
A
22SI
Switch
Setting
..
5-14
3-42.
Front
Panel
Features
............
.3-30
5-11.
Zero
Detector
&
APAO
Adjustments
5-15
3-43.
Option
004:
COMPARATOR_...
.3-30
5-12.
10kHz
Measurement
Accuracy
3-44.
Pin
Location
of
Comparator
Adjustment
....................
5-17
Data
Output....................
.3-31
5-13.
Offset
Adjustment
Setup
..........
5-22
4-1.
Measurement
Frequency
Test
Setup
.
4-7
6-1.
Major
Mechanical
Parts
4-2.
Capacitance
Accuracy
Test
Setup..
.
4-8
-
Exploded
View
...............
6-23
4-3.
Resistance
Accuracy
Test
Setup...
.4-12
6-2.
Mechanical
Parts
-
Exploded
View
...............
6-25
iv

Model
4262A
Section
I
Paragraphs
1-1
to
1-6
SICTION
I
GENERAL
INFORMATION
1-1.
INTRODUCTION.
1-2.
This
operating
and
service
manual
contains
the
information
required
to
install,
operate,
test,
adjust
and
service
the
Hewlett-Packard
Model
4262A
Digital
LCR
Meter.
Figure
1-1
shows
the
in¬
strument
and
supplied
accessories.
This
section
covers
specifications,
instrument
identification,
description,
options,
accessories,
and
other
basic
information.
1-3.
Listed
on
the
title
page
of
this
manual
is
a
microfiche
part
number.
This
number
can
be
used
to
order
4x6
inch
microfilm
transparencies
of
the
manual.
Each
microfiche
contains
up
to
60
photo¬
duplicates
of
the
manual
pages.
The
microfiche
package
also
includes
the
latest
manual
changes
supplement
as
well
as
all
pertinent
service
notes.
To
order
an
additional
manual,
use
the
part
num¬
ber
listed
on
the
title
page
of
this
manual.
1-4.
DESCRIPTION.
1-5.
The
HP
Model
4262A
LCR
Meter
is
a
general
purpose,
fully
automatic
test
instrument
designed
to
measure
the
parameters
of
an
impedance
ele¬
ment
with
high
accuracy
and
speed.
The
4262A
measures
capacitance,
inductance,
resistance
(equi¬
valent
series
resistance)
and
dissipation
factor
or
quality
factor
over
a
wide
range
at
test
frequencies
of
120Hz,
1kHz
and
10kHz
employing
a
five-term¬
inal
connection
configuration
between
the
com¬
ponent
and
the
instrument.
The
measuring
circuit
for
the
device
to
be
measured
is
capable
of
both
parallel
and
series
equivalent
circuit
measurements
and
the
measured
values
are
displayed
by
the
two
three-full
digits
LED
displays
on
the
front
panel.
A
convenient
diagnostic
function,
also
featured
in
the
4262A,
is
actuated
by
a
SELF
TEST
switch.
This
confirms
functional
operation
of
the
instrument.
1-6.
The
measuring
range
for
capacitance
is
from
O.OlpF
to
19.99mF,
inductance
from
O.OImH
to
1999H,
and
resistance
from
lm£2
to
19.99MJ2,
which
are
measured
with
a
basic
accuracy
of
0.2
to
0.3%
depending
on
test
signal
level,
frequency,
and
measuring
equivalent
circuit,
and
at
typical
measuring
speeds
of
220
to
260
milliseconds
at
Figure
1-1.
Model
4262A
and
Accessories.

Section
I
Model
4262A
Table
1-1
Table
1-1.
Specifications
(Sheet
1
of
4).
COMMON
SPECIFICATIONS
Parameters
Measured:
C
-
D
or
Q
(1/D)
Deviation
Measurement:
When
^LCR
key
is
de-
L
-
D
or
Q
(1/D)
pressed,
the
existing
measured
value
R
(ESR)
(Loss
measure-
is
stored
as
a
reference
value
and
ment
can
be
negated
by
displayed
value
is
offset
to
zero'.
The
switch
on
internal
board).
range
is
held
and
deviation
is
dis¬
played
as
the
difference
between
the
Display:
3-1/2
Digit,
Maximum
Display
1999
referenced
value
and
subsequent
re-
(When
D
value
is
more
than
10,
max-
suit.
(Deviation
spread
in
counts
mum
display
is
199).
from
-999
to
1999).
Measurement
Circuit
Modes:
Offset
Adjustment:
Stray
capacitance
and
re-
Auto,
Parallel,
and
Series
sidual
inductance
of
test
jig
can
be
compensated
for
as
follows:
Measurement
Terminals:
5-terminal
configura-
tion
(high
and
low
terminals
for
both
C:
up
to
lOpF
potential
and
current
leads
plus
guard).
L:
up
to
ljuH
Range
Modes:
LCR
-
Auto
and
Manual
Self
Test:
Annunciates
either
Pass,
or
Fail
for
(up-down)
performance
in
each
of
the
five
basic
DQ
-
Auto
and
Manual
(step)
ranges.
DC
Bias:
Measurement
Frequencies:
120(100)Hz,
1kHz
Internal:
1.5V,
2.2V,
6V
(Selectable
at
front
panel).
Accuracy
±5%
External:
External
DC
bias
connector
on
rear
panel.
Maximum
+40V,
and
10kHz
±3%.
Test
Signal
Level:
Normal
level:
iVrms.
Low
level
:
50m
Vrms
(parallel
Trieeer:
Internal.
External,
or
Manual
capacitance
mode
only)
Warm-up
Time:
15
minutes
GENERAL
Operating
Temperature
&
Humidity:
Power
Consumption:
55VA
with
any
option
0
#
C
to
55°C
at
95%
RH(to40°C)
Dimensions:
426(W)
x
147(H)
x
345(D)mm
Power
Requirements:
100/120/220V
±10%,
(16-3/4”
x
5-3/4”
x
13-3/4”)
240V
+5%
-10%
48
-
66Hz
Weight:
Approximately
8kg
(Std)

Model
4262A
Table
1-1.
Specifications
(Sheet
2
of
4).
C-D,
C-Q
MEASUREMENT
Ranges
120Hz
C
1kHz
10kHz
D
Qn
Test
Signal
Level
*2
db-
HPM-
AUTO
C
Accuracy
*3
AUTO
D(l/Q)
Accuracy
*3
-O-
AUTO
[iViiW
i
i
(iiiVinl
b
iiiiiinJ
*
uVi!3a
IO.OOmF
100.0/jF
lOOOnF
IO.OOmF
lOO.OnF
lOOOnF
.001'
■
'-'19.9
(2
Ranges)
0.05^1000
(4
Ranges)
IV
or
50mV
(LOW
LEVEL)
lOQ^A
1mA
e
0.2%
+
1
counts
0.3%
+
2
counts
(At
120Hz,
1kHz)
|
0.3%
+
2
counts
(At
10kHz)
|
0.3%
+
2
counts
10mA
40mA
(Test
signal
level;
IV)
(Test
signal
level;
50mV)
1%
+
2
5%+
2
■
Same
as
-tt-v*-
0.2%
+
(2
+
200/Cx)
counts
0.5%
+
(2
+
200/Cx)
counts
0.3%
+
(2
+
1000/Cx)
counts
2
+
10
(At
120Hz,
1kHz)
1
0.3%
+
(2
+
Cx/500)
c
ounts
(At
10kHz)
0.5%
+
(2
+
Cx/500)
counts
ode
Same
as
+*+*+
■
At
120Hz,
1kHz
(Test
signal
level;
IV)
At
10kHz
At
120Hz,
1kHz
(Test
signal
level;
50mV)
At
10kHz
*1
Calculated
from
D
value
as
a
reciprocal
number.
*2
Typical
data,
varies
with
value
of
D
and
number
of
counts.
*3
±{%
of
reading
+
counts).
Cx
is
capacitance
readout
in
counts.
This
accuracy
only
applies
for
D
values
to
1.999.
(For
higher
D
values,
refer
to
General
Information).
*4
(5%
+
2
counts)
at
1kHz.
Accuracy
applies
over
a
temperature
range
of
23°C
±
5°C
(At
0
c
C
to
55°C,
error
doubles).
Note:
C
accuracy
for
higher
D
values
are
unspecified.
Section
I
Table
1-1
1-3

Section
I
Table
1-1
Model
4262A
Ranges
Test
Signal
Level
*2
D(l/Q)
Accuracy
Table
1-1.
Specifications
(Sheet
3
of
4).
L-D,
L-Q
MEASUREMENT
lO.OOmH
lOO.OmH
lOOOmH
10.00H
100.OH
lOOOpH
lO.OOmH
lOO.OmH
lOOOmH
10.00H
100.OpH
lOOOpH
lO.OOmH
lOO.OmH
lOOOmH
1000H
100.
OH
.001^19.9
(2
Ranges)
0.05^1000
(4
Ranges)
L
Accuracy
*3
-rjnrMWr
AUTO
-O
AUTO
-O-
-'WHW-
AUTO
(At
120Hz,
1kHz)
(At
10kHz)
0
.2%
+
2
counts
0.3%
+
2
0.2%
+
2
counts
Same
as
Mode
0.3%
+
2
counts
0.3%
+
2
counts
1
%
+
2
counts
1%
+
2
5%+
2
(At
120Hz,
1kHz)
(AtlOkHz)
e
as
-GD-
M
(At
120Hz,
1kHz)
|0.3%
+
(At
10kHz)
[0.5%
+
(3
+
Lx/500)
0.2%
+
(3
+
200/Lx)
counts
0.5%
+
(3
+
200/Lx)
counts
Same
as
-wmwt
Mode
1
%
+
-tool
^'(S'-soo
(At
120Hz,
1kHz)
(At
1kHz)
*1
Calculated
from
D
value
as
a
reciprocal
number.
*2
Typical
data,
varies
with
value
of
D
and
number
of
counts.
*3
i(%
of
reading
+
counts).
Lx
is
inductance
readout
in
counts.
This
accuracy
only
applies
for
D
values
to
1.999.
Accuracy
applies
over
a
temperature
range
of
23°C
±
5*0.
(At
0°C
to
55°C,
error
doubles).
R/ESR
MEASUREMENT
120Hz
Ranges
R/ESR
1kHz
lOOOrofi
lkOOft
100.0S2
1000£2
lO.OOkft
10kHz
Test
Signal
Level
Accuracy
*2
AUTO
*1
Typical
data,
varies
with
number
of
counts.
*2
±(%
of
reading
+
counts).
*3
(0.5%
+
2
counts)
on
10.00MQ
range
at
10kHz.
**
Measurement
range
for
ESR
(equivalent
series
resistance)
is
from
Imfi
to
19.99kn
(typical),
which
varies
with
series
capacitance
and
inductance
value
....
refer
to
“REFERENCE
DATA”.
Accuracy
applies
over
a
temperature
range
of
23°C
±
5°C.
(At
0°C
to
55°C
S
error
doubles.)

Model
4262A
Table
1-1.
Specifications
(Sheet
4
of
4).
Section
I
Table
1-1
OPTIONS
Option
001:
Simultaneous
BCD
output
of
LCR
and
DQ
data
(positive
true).
Max.
sink
current
16mA.
Mating
connector
(P/N
1251-0086).
(Alternate
BCD
output
of
LCR
and
DQ
data
selectable
by
switch
on
internal
board).
Option
004:
Digital
comparator
(can
not
be
used
with
OPT
101).
Compares
meas¬
ured
value
with
high
and
low
limit
set¬
tings
for
LCR
or
DQ
and
provides
HIGH,
IN,
LOW
comparison
outputs.
Limit
setting
range:
0000
-
1999
for
each
limit
switch.
Comparison
output:
Visual,
relay
contact,
and
TTL
level.
Visual:
3
LED’s
indicate
HIGH(red),
IN
(green),
or
LOW
(red).
Relay
contacts:
SPST
contacts
to
circuit
common
for
each
HIGH,
IN
and
LOW
output.
TTL
level:
Open
collector
circuits
to
high
level
(open)
for
each
HIGH,
IN
and
LOW
outputs
(fanout
max.
30mA).
Option
101:
HP-IB
data
output
&
remote
con¬
trol.
Remotely
controllable
functions:
Function
(L,
C,
R/ESR,
a
LCR)
Loss
(D,
Q)
LCR
range
DQ
range
Circuit
mode
Test
frequency
&
level
Trigger
Self
test
Data
output:
C
-
D/Q,
L
-
D/Q,
R/ESR
Internal
function
allowable
subsets:
SHI,
AH1,
T5,
L4,
RL1,
DC1,
SRI
and
DTI.
Data
output
format:
Either
of
two
formats
may
be
selected.
Switchable
at
rear
panel
(no
+
sign
outputs).
Format
A.
SFFT±N.NNNE+NN,
SF±N.NN(6r)@
Format
B.
SFFT±N.NNNE±NN(CR)@
SF±N.NN@)@)
SFFT
L-Test
Frequency
—Measurement
Equivalent
Circuit
^-Measurement
Status
SF
L-Loss
measurement
D
or
Q
Measurement
status
Option
010:
100Hz
test
frequency
instead
of
120Hz.
ACCESSORIES
AVAILABLE
16061A:
Test
fixture,
direct
coupled,
5-terminal
Two
kinds
of
inserts
are
included
for
components
with
either
axial
or
radial
leads.
Usable
on
all
ranges
of
4262A.
16062A:
Test
cable
with
alligator
clips,
4-termi-
nal.
Useable
for
low
impedance
meas¬
urements.
Measurement
range
at
1kHz
is
L
2H,
C
>
200nF
and
R
<,
10ka
[For
L
and
C
measurements,
these
ranges
increase
by
xlO
at
120
(100)Hz
and
decrease
by
same
factor
at
10kHz].
16063A:
Test
cable
with
alligator
clips,
3-termi-
nal.
Useable
for
high
impedance
meas¬
urements.
Measurement
range
at
1kHz
is
L
^
3mH,
C
£
10
m
F
and
R
2:200«.
[For
L
and
C
measurement,
these
ranges
increase
by
xlO
at
120(100)Hz
and
decrease
by
same
factor
at
10kHz].
1
-

Section
I
Table
1-2
Model
4262A
Table
1-2.
General
Information.
Measurement
Times
(typical):
For
a
1000
count
measurement
on
a
low
loss
component
on
a
fixed
range:
1kHz,
10kHz
120(100)Hz
When
autorange
is
selected
the
following
times
per
range
step
must
be
added
to
the
above
times:
1kHz,
10kHz
120(100)Hz
45ms/180ms
150ms/670ms
When
U-CL
is
displayed,
the
faster
ranging
time
is
selected.
Reading
Rate:
Internal
-
Approx.
30ms
between
end
of
measurement
and
start
of
next
cycle.
External
-
Measurement
cycle
is
initiated
by
external
trigger
input.
High
D
Factor
Accuracies:
Typical
(
2,
on
10.00
range).
Circuit
Mode
Accuracy
5%
+
(2
+
1000/Cx)
HhW—
0
5
%
4-
(5
4-
Cx/500)
5%+(5
+
Lx/500)
o^innr^wo
5
%
4-(3
4-
200/Lx)
ESR
(Equivalent
Series
Resistance)
Measurement:
Following
tables
show
ESR
measure¬
ment
range
for
capacitors
and
inductors.
Function
Meas.
Time
c
&
C/L
220-260ms
fl
u
R
120-160ms
s
C/L
900ms
R
700mS
I?
£
5
8
100m
1
10
10m
100m
I
710004
10m
100m
a
«
>
100
e
1000*
10m
s
I
10^
100^
1000
j
k
§
In
-
lOn
■
100t/
lu-
In-
li'pf
G1p
lu
-
1-
In
.Olp
-
.Ip
2
or
1
Digit
Duplay
3
or
4
Digit
Display
_i__JJ
Measurement
Range
of
ESR
(SI)
:
1
10
100-
100m-
1
••
10-
X
7
lOtnr-
100m-
1-
o
■
a
**
lm»-
10®
-
100m-
<o
u
J
lOOp
-
lm
-
10m-
I
10p-
!00p-
ImU*^
Ip-
I0p-
lOOpii
2
or
I
3
or
4
Digit
Display
Digit
Display
.
...
sFor
|
Indf
fc
frra*
.
»
if
.
.1
pi-
Ip*-
10p*
Measurement
Range
of
ESR
(SI)
(
mS)
IOOO
—
aoo
—
0OO
—
TOO-
|
600
—
500
-.
£
400
—
MEASURING
TIME
—
When
LCR
RANGE
is
set
to
MANUAL
Cp,l_«
Ca,l-p
|
it
hi
R«
Rp
4000
C
Counts
3
Sum
of
Counts
on
LCR
and
DISSIPATION
FACTOR
Dlsployt

Model
4262A
1kHz
and
10kHz
and
about
900
milliseconds
at
120Hz.
The
wide
range
capability
of
the
4262A
enables
a
measurement
range
from
small
capaci¬
tances
such
as
mica
capacitors
and
the
para¬
sitic
capacitance
of
a
semiconductor
device
through
high
capacitances
such
as
the
measurement
of
electrolytic
capacitors
to
be
covered.
A
wide
range
of
inductance
measurements
from
the
in¬
ductance
of
a
high
frequency
transformer
to
that
of
a
power
transformer
can
be
measured.
The
wide
resistance
range
permits
the
measurement
of
wire-
wound
resistors
through
the
measurement
of
solid
resistors.
In
parallel
capacitance
measurements,
either
a
test
signal
level
of
lVrms,
or
50mVrms
can
be
selected.
1-7.
The
4262A
has
the
capability
of
making
capacitance,
inductance,
and
resistance
deviation
measurements.
This
function
is
enabled
by
pushing
the
A
LCR
switch
to
display
the
deviation
of
a
re¬
ference
value.
When
the
A
LCR
switch
is
depressed
the
reference
value
is
obtained
and
memorized
from
the
preceding
measurement.
The
practi¬
cal
use
of
this
feature
is
evident
when
it
is
desired
to
make
a
measurement
on
a
variable
capacitor:
First,
the
minimum
value
is
measured,
then
the
A
LCR
button
is
pushed.
Minimum
to
maximum
capa¬
citance
is
now
displayed
as
the
capacitor
is
rotated
through
its
range.
For
parallel
capacitance
meas¬
urements,
test
signal
levels
of
either
lVrms
or
SOmVrms
may
be
selected.
Other
versatile
4262A
capabilities
and
features
are,
for
example,
the
use
of
internal
and
external
dc
bias
voltages,
LC
zero
adjustment,
and
options
providing
BCD
output,
HP-IB
interfacing
capability,
or
a
comparator
func¬
tion.
1-8.
SPECIFICATIONS.
1-9.
Complete
specifications
of
the
Model
4262A
LCR
Meter
are
given
in
Table
1-1.
These
specifi¬
cations
are
the
performance
standards
or
limits
against
which
the
instrument
is
tested.
The
test
procedures
for
the
specifications
are
covered
in
Section
IV
Performance
Tests.
Table
1-2
lists
gen-
Figure
1-2.
Serial
Number
Plate.
Section
I
Paragraphs
1-7
to
1-16
eral
information.
General
information
is
not
speci¬
fications
but
is
typical
characteristics
included
as
additional
information
for
the
operator.
When
the
4262A
LCR
Meter
is
shipped
from
the
factory,
it
meets
the
specifications
listed
in
Table
1-1.
1-10.
SAFETY
CONSIDERATIONS.
1-11.
The
Model
4262A
LCR
Meter
has
been
designed
to
conform
to
the
safety
requirements
of
an
IEC
(International
Electromechanical
Com¬
mittee)
Safety
Class
I
instrument
and
is
shipped
from
the
factory
in
a
safe
condition.
1-12.
This
operating
and
service
manual
contains
information,
cautions,
and
warnings
which
must
be
followed
by
the
user
to
ensure
safe
operation
and
to
maintain
the
instrument
in
a
safe
condition.
1-14.
Hewlett-Packard
uses
a
two-section
nine
character
serial
number
which
is
marked
on
the
se¬
rial
number
plate
(Figure
1-2)
attached
to
the
in¬
strument
rear
panel.
The
first
four
digits
and
the
letter
are
the
serial
prefix
and
the
last
five
digits
are
the
suffix.
The
letter
placed
between
the
two
sec¬
tions
identifies
country
where
instrument
was
manufactured.
The
prefix
is
the
same
for
all
iden¬
tical
instruments;
it
changes
only
when
a
change
is
made
to
the
instrument.
The
suffix,
however,
is
as¬
signed
sequentially
and
is
different
for
each
in¬
strument.
The
contents
of
this
manual
apply
to
in¬
struments
with
the
serial
number
prefix(es)
listed
under
SERIAL
NUMBERS
on
the
title
page.
1-15.
An
instrument
manufactured
after
the
printing
of
this
manual
may
have
a
serial
number
prefix
that
is
not
listed
on
the
title
page.
This
un¬
listed
serial
number
prefix
indicates
the
instrument
is
different
from
those
described
in
this
manual.
The
manual
for
this
new
instrument
may
be
ac¬
companied
by
a
yellow
Manual
Changes
supple¬
ment
or
have
a
different
manual
part
number.
This
supplement
contains
“change
information”
that
explains
how
to
adapt
the
manual
to
the
newer
instrument.
1-16.
In
addition
to
change
information,
the
supplement
may
contain
information
for
correct¬
ing
errors
(called
Errata)
in
the
manual.
To
keep
this
manual
as
current
and
accurate
as
possible,
Hewlett-Packard
recommends
that
you
periodically
request
the
latest
Manual
Changes
supplement.
The
supplement
for
this
manual
is
identified
with
this
manual’s
print
date
and
part
number,
both
of
which
appear
on
the
manual’s
title
page.
Com¬
plimentary
copies
of
the
supplement
are
available
from
Hewlett-Packard.
If
the
serial
prefix
or
number
of
an
instrument
is
lower
than
that
on
title
page
of
this
manual,
see
Section
VII
Manual
Changes.
1-7
M3.
INSTRUMENTS
COVERED
BY
MANUAL,

Section
I
Paragraphs
1-17
to
1-33
Model
4262A
147.
For
information
concerning
a
serial
number
prefix
that
is
not
listed
on
the
title
page
or
in
the
Manual
Changes
supplement,
contact
your
nearest
Hewlett-Packard
office.
1-18.
OPTIONS.
1-19.
Options
for
the
Model
4262A
LCR
Meter
are
available
for
adding
the
following
capabilities:
Option
001:
BCD
Parallel
Data
Output.
Option
004:
Comparator.
A
comparator
func¬
tion
providing
GO/NO-GO
judge¬
ment
with
HIGH
and
LOW
limits
for
LCR
and
D/Q.
Option
101:
HP-IB
Interface.
Option
010:
100Hz
Test
Frequency.
(instead
of
120Hz)
Options
907,
908
or
909
are
handle
or
rack
mount
kits.
See
paragraph
1-29
for
details.
Option
910:
Extra
Manual.
1-20.
OPTION
001.
1-21.
The
4262A
option
001
provides
separate
BCD
parallel
data
output
for
L,
C,
R/ESR
and
dissipation
factor
or
quality
factor
simultaneously
from
the
two
rear
panel
connectors.
With
this
op¬
tion,
external
data
processing
devices
such
as
a
digital
printer
can
be
used
with
the
4262A.
1-22.
OPTION
004.
1-23.
The
4262A
Option
004
provides
for
GO/NO-GO
judgement
by
comparing
L,
C,
R/ESR
and
D/Q
values
to
HIGH
and
LOW
limits.
Three
judgement
outputs
are
provided:
LED
lamp
dis¬
play,
relay
contacts,
or
TTL
level
voltages
(open
collectors):
HIGH
.
.measured,
value
is
not
less
than
HIGH
limit.
IN
.
..
.measured
value
is
less
than
HIGH
limit
and
not
less
than
LOW
limit.
LOW
...measured
value
is
less
than
LOW
limit.
1-24.
OPTION
101.
1-25.
The
4262A
Option
101
provides
interfac¬
ing
functions
to
both
transfer
L,
C,
R/ESR
and
D/Q
data
to
HP
Interface
Bus
line
and
to
receive
remote
control
signals
from
HP
Interface
Bus
line.
1-26.
OPTION
010.
1-27.
The
4262A
Option
010
provides
test
fre¬
quencies
of
100Hz,
1kHz,
and
10kHz
(100Hz
is
used
instead
of
standard
120Hz).
All
other
electri¬
cal
performance
is
the
same
as
that
of
standard
in¬
strument.
1-28.
OTHER
OPTIONS.
1-29.
The
following
options
provides
mechanical
parts
necessary
for
rack
mounting
and
hand
carry
¬
ing:
Option
907:
Front
Handle
Kit.
Option
908:
Rack
Flange
Kit.
Option
909:
Rack
Flange
and
Front
Handle
Kit.
The
installation
procedures
for
these
options
are
detailed
in
section
II.
1-30.
The
4262A
Option
910
provides
an
extra
copy
of
the
operating
and
service
manual.
1-31.
ACCESSORIES
SUPPLIED.
1-32.
Figure
14
shows
the
HP
Model
4262A
LCR
Meter,
power
cord
(HP
Part
No.
8120-1378),
and
fuses
(HP
Part
No.
2110-0007
and
2110-0202).
1-33.
ACCESSORIES
AVAILABLE.
1-34.
For
effective
and
easy
measurement*
three
styles
of
fixtures
and
leads
for
the
measurement
of
various
components
are
available.
These
are
listed
in
Table
1-1,
A
brief
description
of
each
of
these
fixtures
and
leads
is
given
in
Table
1-3.
Refer
to
Section
III
Figure
3-3
on
page
3-8
for
detailed
information
on
these
devices.

Section
I
Table
1-3
Table
1-3.
Accessories
Available.
Extender
Board
used
for
4262A
troubleshooting.

Section
I
Table
1-4
Model
4262A
Table
1-4.
Recommended
Test
Equipment.
Instrument
Critical
Specifications
Recommended
Model
*Use
Frequency
Counter
Frequency
Range:
40Hz
to
10kHz
Sensitivity:
SOmVrms
min.
HP
5300A/
w
5306A
P
Capacitance
Standard
(See
para.
4-3)
Capacitance
Values:
lOOpF,
lOOOpF,
lOnF,
lOOnF,
lOOOnF
and
10pF
GR
Type
1413
GRType
1417
P*
A
Resistance
Standard
(See
para.
4-3)
Resistance
Values:
Ikfi,
lOkft,
lOOkfl
and
10M
n
GR
Type
1443-Y
P
s
A
Inductance
Standard
(See
Para.
4-3)
Inductance
Value:
lOOmH
GR
Type
1482-L
P
DC
Voltmeter
Voltage
Range:
IV
to
10V
Sensitivity:
lOmV
min.
HP
5300A/
w
5306A
P,
A
Oscilloscope
Bandwidth:
10MHz
min.
Vertical
Sensitivity:
5mV/div.
Horizontal
Sweep
Rate:
lps/div.
HP
1740A
A,
T
Signature
Analyzer
HP
5004A
T
Current
Tracer
HP
547A
T
Service
Kit
Signature
Analysis
Test
Board
HPP/N:
04262-87002
T
DUT
Box
Comprises
L
s
C
and
R
components
whose
values
are
calibrated
at
120Hz
and
1kHz.
HP
16361A
m
1
DUT
Box
Comprises
L,
C
and
R
components
whose
values
are
calibrated
at
10kHz.
HP
16362A
P»
A
*P=Performance
Test
A=Adjustments
Troubleshooting

Model
4262A
Section
II
Paragraphs
2-1
to
2-11
SECTION
II
INSTALLATION
2-1.
INTRODUCTION.
2-2.
This
section
provides
installation
instructions
for
the
Model
4262A
LCR
Meter.
The
section
also
includes
information
on
initial
inspection
and
damage
claims,
preparation
for
using
the
4262A,
packaging,
storage,
and
shipment.
2-3.
INITIAL
INSPECTION.
2-4.
The
4262A
LCR
Meter,
as
shipped
from
the
factory,
meets
all
the
specifications
listed
in
Table
1-
1.
On
receipt,
inspect
the
shipping
container
for
damage.
If
the
shipping
container
or
cushioning
material
is
damaged,
notify
the
carrier
as
well
as
the
Hewlett-Packard
office
and
be
sure
to
keep
the
shipping
materials
for
carrier’s
inspection
until
the
contents
of
the
shipment
have
been
checked
for
completeness
and
the
instrument
has
been
checked
mechanically
and
electrically.
The
contents
of
the
shipment
should
be
as
shown
in
Figure
1-1.
The
procedures
for
checking
the
general
electrical
op¬
eration
are
given
in
Section
III
(Paragraph
3-5
Ba¬
sic
Operating
Check)
and
the
procedures
for
checking
the
4262A
LCR
Meter
against
its
specifi¬
cations
are
given
in
Section
IV.
Firstly,
do
the
self
test.
If
the
4262A
LCR
Meter
is
electrically
ques¬
tionable,
then
do
the
Performance
Tests
to
deter¬
mine
whether
the
4262A
has
failed
or
not.
If
con¬
tents
are
incomplete,
if
there
is
mechanical
damage
or
defects
(scratches,
dents,
broken
switches,
etc.),
or
if
the
performance
does
not
meet
the
self
test
or
performance
tests,
notify
the
nearest
Hewlett-
Packard
office
(see
list
at
back
of
this
manual).
The
HP
office
will
arrange
for
repair
or
replacement
without
waiting
for
claim
settlement.
2-
5.
PREPARATION
FOR
USE.
2-6.
POWER
REQUIREMENTS.
2-7.
The
4262A
requires
a
power
source
of
100,
120,
220Volts
ac
±10%,
or
240Volts
ac
+5%,
-10%,
48
to
66Hz
single
phase.
Power
consumption
is
approximately
55
watts.
WARNING
IF
THIS
INSTRUMENT
IS
TO
BE
ENERGIZED
VIA
AN
EXTER¬
NAL
AUTOTRANSFORMER
FOR
VOLTAGE
REDUCTION,
BE
SURE
THAT
THE
COMMON
TERMINAL
IS
CONNECTED
TO
THE
NEUTRAL
POLE
OF
THE
POWER
SUPPLY.
2-8.
LINE
VOLTAGE
AND
FUSE
SELECTION.
CAUTION
BEFORE
TURNING
THE
4262A
LINE
SWITCH
TO
ON,
VERIFY
THAT
THE
INSTRUMENT
IS
SET
TO
THE
VOLTAGE
OF
THE
POWER
SUPPLIED.
2-9.
Figure
2-1
provides
instructions
for
line
volt¬
age
and
fuse
selection.
The
line
voltage
selection
card
and
the
proper
fuse
are
factory
installed
for
the
voltage
appropriate
to
instrument
destination.
CAUTION
USE
PROPER
FUSE
FOR
LINE
VOLTAGE
SELECTED.
CAUTION
MAKE
SURE'THAT
ONLY
FUSES
FOR
THE
REQUIRED
RATED
CURRENT
AND
OF
THE
SPECI¬
FIED
TYPE
ARE
USED
FOR
RE¬
PLACEMENT.
THE
USE
OF
MENDED
FUSES
AND
THE
SHORT-CIRCUITING
OF
FUSE-
HOLDERS
MUST
BE
AVOIDED.
2-10.
POWER
CABLE.
2-11.
To
protect
operating
personnel,
the
2-1

Section
II
Model
4262A
Paragraph
2-12
SELECTION
OF
OPERATING
VOLTAGE
1.
Disconnect
power
cable
and
slide
module
window
to
left.
2.
Pul!
FUSE
PULL
lever
and
rotate
to
left.
This
removes
line
fuse
3.
Select
operation
voltage
by
orienting
PC
board
to
position
desired
voltage
on
top-
left
side.
Push
board
firmly
into
module
slot.
4.
Rotate
FUSE
PULL
lever
back
to
its
normal
position
and
re-insert
fuse
in
holder
be
careful
to
select
correct
fuse
value.
Operating
voltage
is
shown
in
module
window
and
is
usually
set
to
120V
at
factory.
Operating
Voltage
Fuse
HP
Part
No.
Description
lOOVac
or
120Vac
220Vac
or
240Vac
2110-0007
2110-0202
1
A
250V
Slow
Blow
0.5A
250V
Slow
Blow
Figure
2-1.
Voltage
and
Fuse
Selection.
National
Electrical
Manufacturer’s
Association
(NEMA)
recommends
that
the
instrument
panel
and
cabinet
be
grounded.
The
Model
4262A
is
e-
quipped
with
a
three-conductor
power
cable
which,
when
plugged
into
an
appropriate
receptacle,
grounds
the
instrument.
The
offset
pin
on
the
power
cable
is
the
ground
wire,
2-12.
To
preserve
the
protection
feature
when
operating
the
instrument
from
a
two
contact
outlet,
use
a
three
prong
to
two
prong
adapter
(HP
Part
No.
5080-3149)
and
connect
the
green
grounding
tab
on
the
adapter
to
power
line
ground.
CAUTION
THE
MAINS
PLUG
MUST
ONLY
BE
INSERTED
IN
A
SOCKET
OUTLET
PROVIDED
WITH
A
PROTECTIVE
EARTH
CONTACT.
THE
PROTECTIVE
ACTION
MUST
NOT
BE
NEGATED
BY
THE
USE
OF
AN
EXTENSION
CORD
(POWER
CABLE)
WITH¬
OUT
PROTECTIVE
CONDUCTOR
(GROUNDING).

Model
4262A
Section
II
Paragraphs
2-13
to
2-28
2-13.
Figure
2-2
shows
the
available
power
cords,
which
may
be
used
in
various
countries
including
the
standard
power
cord
furnished
with
the
in¬
strument.
HP
Part
number,
applicable
standards
for
power
plug,
power
cord
color,
electrical
character¬
istics
and
countries
using
each
power
cord
are
listed
in
the
figure.
If
assistance
is
needed
for
se¬
lecting
the
correct
power
cable,
contact
nearest
Hewlett-Packard
office.
2-14.
Interconnections.
2-15.
When
an
external
bias
is
applied
to
the
sample
capacitor
through
DC
BIAS
input
con¬
nectors
on
the
4262A
rear
panel,
both
plus
and
minus
sides
of
the
external
power
supply
should
be
connected
to
the
plus
and
minus
sides
of
the
4262A
EXT
DC
BIAS
connector,
respectively,
CAUTION
THE
MAINS
PLUG
MUST
BE
IN¬
SERTED
BEFORE
EXTERNAL
CONNECTIONS
ARE
MADE
TO
MEASURING
AND/OR
CON¬
TROL
CIRCUITS.
2-16.
Operating
Environment.
2-17.
Temperature.
The
instrument
may
be
op¬
erated
in
temperatures
from
0°C
to
+55
9
C.
2-18.
Humidity.
The
instrument
may
be
operated
in
environments
with
relative
humidities
to
95%
to
40°C.
However,
the
instrument
should
be
protected
from
temperature
extremes
which
cause
condensation
within
the
instrument.
2-19.
Installation
Instructions.
2-20.
The
HP
Model
4262A
can
be
operated
on
the
bench
or
in
a
rack
mount.
The
4262A
is
ready
for
bench
operation
as
shipped
from
the
factory.
For
bench
operation
a
two-leg
instrument
stand
is
used.
For
use,
the
instrument
stands
are
de¬
signed
to
be
pulled
towards
the
front
of
instrument.
2-21.
Installation
of
Options
907,
908
and
909.
2-22.
The
4262A
can
be
installed
in
a
rack
and
be
operated
as
a
component
of
a
measurement
system.
Rack
mounting
information
for
the
4262A
is
presented
in
Figure
2-3.
2-23.
STORAGE
AND
SHIPMENT.
2-24.
Environment.
2-25.
The
instrument
may
be
stored
or
shipped
in
environments
within
the
following
limits:
Temperature.-40°C
to
+75°C
Humidity.
to
95%
Altitude...50,000ft
The
instrument
should
be
protected
from
temper¬
ature
extremes
which
cause
condensation
inside
the
instrument.
2-26.
Packaging.
2-27.
Original
Packaging.
Containers
and
mat¬
erials
identical
to
those
used
in
factory
packaging
are
available
through
Hewlett-Packard
offices.
If
the
instrument
is
being
returned
to
Hewlett-
Packard
for
servicing,
attach
a
tag
indicating
the
type
of
service
required,
return
address,
model
number,
and
full
serial
number.
Also
mark
the
container
FRAGILE
to
assure
careful
handling.
In
any
correspondence,
refer
to
the
instrument
by
model
number
and
full
serial
number.
2-28.
Other
Packaging.
The
following
general
in¬
structions
should
be
used
for
re-packing
with
com¬
mercially
available
materials:
a.
Wrap
instrument
in
heavy
paper
or
plastic.
If
shipping
to
Hewlett-Packard
office
or
service
center,
attach
tag
indicating
type
of
service
required,
return
address,
model
number,
and
full
serial
number.
b.
Use
strong
shipping
container.
A
double-wall
carton
made
of
350
pound
test
material
is
adequate.
c.
Use
enough
shock
absorbing
material
(3
to
4
inch
layer)
around
all
sides
of
instrument
to
provide
firm
cushion
and
prevent
movement
inside
container.
Protect
control
panel
with
cardboard.
d.
Seal
shipping
container
securely.
e.
Mark
shipping
container
FRAGILE
to
ensure
careful
handling.
f.
In
any
correspondence,
refer
to
instrument
by
model
number
and
full
serial
number.
2-3

Section
I!
Figure
2-2
Model
4262A
Figure
2-2.
Power
Cables
Supplied
Other manuals for 4262A
2
Table of contents
Other HP Measuring Instrument manuals