HP 8509A User manual

Installation and Quick Start
Guide
HP 8509A/B Lightwave
Polarization Analyzer
ABCDE
HP P
art No. 08509-90015
Printed in USA October 1994

Notice.
The information contained in this do cument is sub ject to
change without notice.
Hewlett-Packard makes no warrantyof any kind with
regard to this material, including but not limited to,
the implied warranties of merchantability and tness
for a particular purp ose. Hewlett-Packard shall not be
liable for errors contained herein or for incidental or
consequential damages in connection with the furnishing,
performance, or use of this material.
c
Copyright Hewlett-P
ackard Compan
y 1992, 1994
All Righ
ts Reserv
ed. Reproduction, adaptation,
or translation without prior written p ermission is
prohibited, except as allowed under the cop
yrightla
ws.
1400 F
ountaingroveP
arkway,San
ta Rosa, CA 95403,
USA

Certification
Hewlett-Packard Company certies that this pro duct
met its published specications at the time of shipment
from the factory. Hewlett-Packard further certies
that its calibration measurements are traceable to the
United States National Institute of Standards and
Technology, to the extent allowed by the Institute's
calibration facility, and to the calibration facilities of
other International Standards Organization members.
W
arranty
This Hewlett-P
ac
k
ard instrumen
tpro duct is w
arran
ted
against defects in material and w
orkmanship for aperio d
of one y
ear from date of shipment. During the w
arran
t
y
period, Hewlett-P
ac
k
ard Compan
ywill, at its option,
either repair or replace pro ducts whic
hpro
v
eto be
defectiv
e.
F
or w
arran
t
yservice or repair, this pro duct m
ust
bereturned to aservice facilitydesignated b
y
Hewlett-P
ack
ard. Buy
er shall prepa
yshipping c
harges to
Hewlett-P
ac
k
ard and Hewlett-P
ac
k
ard shall pa
yshipping
c
harges to return the pro duct to Buyer. How
ever, Buyer
shall pay all shipping charges, duties, and taxes for
pro ducts returned to Hewlett-Packard from another
country.
Hewlett-Packard warrants that its software and
rmware designated by Hewlett-Packard for use
with an instrumen
twill execute its programming
instructions when properly installed on that instrumen
t.
Hewlett-P
ac
k
ard does not w
arran
tthat the op eration
of the instrument, or softw
are, or rm
ware will b e
uninterrupted or error-free.
iii

Limitation of Warranty
The foregoing warranty shall not apply to defects
resulting from improp er or inadequate maintenance
by Buyer, Buyer-supplied software or interfacing,
unauthorized mo dication or misuse, op eration outside
of the environmental sp ecications for the pro duct, or
improp er site preparation or maintenance.
NO OTHER WARRANTY IS EXPRESSED OR
IMPLIED. HEWLETT-PACKARD SPECIFICALLY
DISCLAIMS THE IMPLIED WARRANTIES OF
MER
CHANT
ABILITY AND FITNESS F
OR A
P
ARTICULAR PURPOSE.
Ex
clusive Remedies
THE REMEDIES PR
O
VIDED HEREIN ARE
BUYER'S SOLE AND EX
CLUSIVE REMEDIES.
HEWLETT-P
A
CKARD SHALL NOT BE LIABLE
F
OR ANY DIRECT, INDIRECT, SPECIAL,
INCIDENT
AL, OR CONSEQUENTIAL D
AMA
GES,
WHETHER BASED ON CONTRA
CT, TOR
T, OR
ANY OTHER LEGAL THEOR
Y.
Assistance
Product maintenance agreements and other customer
assistance agreements are available for Hew lett-Packard
products.
For any assistance, contact your nearest Hew lett-Packard
Sales and Service Oc
e.
iv

Safety Notes
The following safety notes are used throughout this
manual. Familiarize yourself with each of the notes and
its meaning b efore op erating this instrument.
Caution
Caution denotes a hazard. It calls attention to a
pro cedure that, if not correctly p erformed or adhered
to, would result in damage to or destruction of the
instrument. Do not pro ceed beyond a caution sign until
the indicated conditions are fully understoo d and met.
W
arning
W
arning denotes ahazard. It calls attention to a
procedure which, if not correctly performed or adhered
to, could result in injury or loss of life. Do not proceed
bey
ond aw
arning note until the indicated conditions are
fully understood and met.
L
The instruction documen
tation sym
bol.
The pro duct is mark
ed with this sym
bol when it is
necessary for the user to refer to the instructions in the
documen
tation.
\CE" The CE mark is a registered trademark of the
European Comm
unit
y
.(If accompanied b
yay
ear, it is
when the design w
as pro
v
en.)
\ISM1-A" This is a symbol of an Industrial Scientic
and Medical Group 1 Class A pro duct.
\CSA" The CSA mark is a registered trademark of the
Canadian Standards Asso ciation.
v

General Safety Considerations
Warning
This is a Safety Class I product (provided with a
protective earthing ground incorporated in the power
cord). The mains plug shall only be inserted in a socket
outlet provided with a protective earth contact. Any
interruption of the protective conductor, inside or outside
the instrument, is likelytomake the instrument dangerous.
Intentional interruption is prohibited.
W
arning
No operator serviceable parts inside. Refer servicing to
qualified personnel. T
oprev
ent electrical shock, do not
remo
v
eco
vers.
Caution
Before switc
hing on this instrumen
t, mak
esure that the
line v
oltage selector switc
his set to the v
oltage of the
po
w
er supply and the correct fuse is installed.
vi

How to Use This Guide
This guide uses the following conventions:
4
Button
5
This represents a button or key that
may b e selected in the HP 8509 user
interface on the Vectra Display.
NNNNNNNNNNNNNNNNNNNNNNNNNNNNN
Menu Item
This indicates a menu item in the
HP 8509 user interface main menuor
pull-do
wn men
us.
Documentation Description
This man
ual pro
vides information on installing y
our
ligh
t
w
a
ve polarization analyzer and ho
wto perform an
operation v
erication to ensure that no damage w
as
incurred during shipmen
t, and that y
our analyzer is
operating prop erly
.
The operation v
erication is afully automated pass or
fail indicator. Users should familiarize themselves with
MS Windo
ws operation and con
v
en
tions before using this
pro duct.
Microsoft, MS, MS-DOS, and the Microsoft logo are
registered trademarks, and Windows is a trademark, of
Microsoft Corp oration.
vii


Contents
1. Introduction
2. Installation
T
ools needed ........... . 2-1
Installation o
v
erview .... . . . . 2-1
Unpac
kand Insp ect the System ....2-3
P
o
w
er Requiremen
ts .........2-4
Stac
kand Connect the System .....2-6
P
o
w
er Up the System .. . ......2-8
3. Operation V
erication
Operation V
erication for HP 8509A ..3-2
Required Equipmen
t. . ......3-2
Preliminary Steps ...........3-3
Adjust the Laser Fib er F
or Maximum
P
o
w
er .. . . . . . . . .....3-4
Setup Y
our Source . . ........3-5
Operation Verication . . . . . . . . . 3-6
Operation Verication for HP 8509B
(With Internal Optical Source) . . . 3-8
Required Equipment . . . . . . . . 3-8
Preliminary Steps . . . . . . . . . . . 3-9
Operation V
erication . . .......3-10
Operation V
erication for HP 8509B
(Without In
ternal Source) . . . . . 3-12
Required Equipmen
t . . . . . . . . 3-12
Preliminary Steps . . . . . . . . . . . 3-13
Setup Y
our Source . . . . . . . . . . 3-14
Adjust F
or Maximum P
ower . . . . . . 3-15
Operation Verication . . . . . . . . . 3-16
Contents-1

4. If Things Go Wrong
Problems requiring additional resources . 4-1
The system app ears to be dead . . . . . 4-2
If the computer do es not respond . . . . 4-3
If you get the \EEprom not formatted or
Box on?" error message . . . . . . 4-4
If op eration verication fails . . . . . . 4-5
DOP measurement is out of sp ecication 4-5
Power displayreads
underrange
. . . 4-6
To return for service . . . . . . . . . 4-7
Contents-2

Figures
1-1. HP 8509B System . . . . . . . . . 1-2
1-2. The HP 8509 User Interface Basic
Measurement Screen . . . . . . . 1-3
2-1. Items Needed for Installation . . . . . 2-2
2-2. V
oltage Selector . . ........2-5
2-3. System Connections .........2-7
3-1. HP 8509A Equipmen
tSetup .....3-3
3-2. HP 8509 Basic Measurement Screen . .3-4
3-3. Optical Source Setup Windo
w....3-5
3-4. Operation V
erication Windo
w....3-6
3-5. Equipmen
tSetup ....... . ..3-9
3-6. Operation Verication Windo
w....3-10
3-7. Equipmen
tSetup ..........3-13
3-8. Optical Source Setup Windo
w....3-14
3-9. HP 8509 Basic Measuremen
tScreen ..3-15
3-10. Operation V
erication Windo
w. . . .3-16
Tables
2-1. F
uses .......... . ....2-4
3-1. Recommended P
atc
hcords . . ....3-2
3-2. Recommended P
atchcords . . . . . . 3-8
3-3. Recommended P
atchcords . . . . . . 3-12
4-1. Hewlett-P
ackard Sales and Service
Oces . . . . . . . . . . . . . 4-8
Contents-3


1
1
Introduction
For the quickest installation of your HP 8509 system,
use the
HP 8509A/B Instal lation Quick ReferenceCard
that came attached to the top cover of your lightwave
polarization analyzer.
Chapter 1, Introduction
Figure 1-1 sho
ws at
ypical HP 8509B system. Figure 1-2
sho
ws the basic measuremen
tscreen of the HP 8509 user
in
terface.
Chapter 2, Installation
Pro
vides detailed installation instructions.
Chapter 3, Op eration V
erication
P
erform the operation v
erication after installing
y
our system to v
erify that your instrumen
tis w
orking
properly
.
Chapter 4, If Things Go Wrong
If you encounter obstacles or diculties, check this
section for p ossible solutions.
Note
The HP 8509 user in
terface is an MS Windo
ws
application. Y
ou should familiarize yourself with
Windo
ws con
v
en
tions and operation before using this
pro duct. Refer to the Microsoft Windo
ws
User's Guide
for information on using MS Windo
ws.
Introduction 1-1

1
Figure 1-1. HP 8509B System
1-2 Introduction

1
Figure 1-2. The HP 8509 User Interface Basic Measurement Screen
Introduction 1-3


2
2
Installation
The lightwave polarization analyzer system is
precongured at the factory and installation requires
minimal time to complete.
T
ools needed
No special to ols are required for the basic installation of
the HP 8509 system. A small at-blade screwdriv
er ma
y
beused to secure the monitor and analyzer in
terconnect
cables.
Installation o
v
erview
The installation consists of unpacking the pieces of the
system, correctly connecting them, and then turning
them on.
After installation is complete, operation v
erication
is performed to v
erify that the system is operating
properly
.
Installation 2-1

2
Figure 2-1. Items Needed for Installation
2-2 Installation

2
Unpack and Inspect the System
1. Unpack the lightwave p olarization analyzer system
from its shipping containers. Inspect it thoroughly to
ensure that it was not damaged during shipment. If
anycontainer or piece of the system is damaged or
incomplete, save the packing materials and notify
both the carrier and Hewlett-Packard.
2. Refer to Figure 2-1. Gather the items shown, and put
all other hardware, do cumentation, and software aside
for no
w.
Note
The ligh
t
w
a
v
epolarization analyzer in Figure 2-1 is an
HP 8509B.The HP 8509A diers in app earance. Its
fron
tpanel con
tains only aline po
wer switc
hand an
OPTICAL INPUT connector.
Installation 2-3

2
Power Requirements
The power requirements include line frequency in the
range of 47 to 66 Hz, and 200 VA maximum. Other
power requirements for the lightwave polarization
analyzer are listed in Table 2-1.
Table 2-1. Fuses
Available ac Voltage Voltage Range Fuse Type
100 V90 to 110 Vrms 2.00 AF(HP p.n. 2110-0002)
120 V108 to 132 Vrms 2.00 AF(HP p.n. 2110-0002)
220 V198 to 242 Vrms 1.00 AF(HP p.n. 2110-0001)
240 V216 to 250 Vrms 1.00 AF(HP p.n. 2110-0001)
On the ligh
t
w
a
v
epolarization analyzer, set the
line-v
oltage selector to the v
oltage corresp onding to the
po
w
er source used. The line v
oltage selector is lo cated
on the rear panel of the p olarization analyzer.
Note
If y
ou need to c
hange the switc
hsetting or the fuse, refer
to Figure 2-2 and use a small at-blade screwdriver to
pry op en the line mo dule cover door.
2-4 Installation