HP 8590 E-Series Use and care manual

About this Manual
We’ve added this manual to the Agilent website in an effort to help you support
your product. This manual is the best copy we could find; it may be incomplete
or contain dated information. If we find a more recent copy in the future, we will
add it to the Agilent website.
Support for Your Product
Agilent no longer sells or supports this product. Our service centers may be able
to perform calibration if no repair parts are needed, but no other support from
Agilent is available. You will find any other available product information on the
Agilent Test & Measurement website, www.tm.agilent.com.
HP References in this Manual
This manual may contain references to HP or Hewlett-Packard. Please note that
Hewlett-Packard's former test and measurement, semiconductor products and
chemical analysis businesses are now part of Agilent Technologies. We have
made no changes to this manual copy. In other documentation, to reduce
potential confusion, the only change to product numbers and names has been in
the company name prefix: where a product number/name was HP XXXX the
current name/number is now Agilent XXXX. For example, model number
HP8648A is now model number Agilent 8648A.

HP 8590 EM-Series EMC Analyzer
Calibration Guide
ABCDE
HP P
art No
.5962-0467
Printed in US
A December 1995

Notice
The information contained in this document is subject to change without notice.
Hewlett-Packard makes no warranty of any kind with regard to this material,
including but not limited to, the implied warranties of merchantability and
tness for a particular purpose. Hewlett-Packard shall not be liable for errors
contained herein or for incidental or consequential damages in connection with
the furnishing, performance
,or use of this material.
Restricted Rights Legend
Use
,duplication, or disclosure by the U.S. Government is subject to restrictions
as set forth in subparagraph (c) (1) (ii) of the Rights of T
echnical Data and
Computer Software clause at DF
ARS 252.227-7013 for DOD agencies
,and
subparagraphs (c) (1) and (c) (2) of the Commercial Computer Software
Restricted Rights clause at F
AR 52.227-19 for other agencies
.
c
Copyright Hewlett-P
ackard Company 1995
All Rights Reserved. Reproduction, adaptation, or translation without prior
written permission is prohibited, except as allowed under the copyright laws
.
1400 Fountaingrove Parkway, Santa Rosa CA, 95403-1799, USA

Certication
Hewlett-Packard Company certies that this product met its published
specications at the time of shipment from the factory. Hewlett-Packard
further certies that its calibration measurements are traceable to the United
States National Institute of Standards and Technology, to the extent allowed
by the Institute's calibration facility, and to the calibration facilities of other
International Standards Organization members.
W
arranty
This Hewlett-P
ackard instrument product is warranted against defects in
material and workmanship for aperiod of one year from date of shipment.
During the warranty period, Hewlett-P
ackard Company will, at its option, either
repair or replace products which prove to be defective
.
F
or warranty service or repair
,this product must be returned to aservice
facility designated by Hewlett-Packard. Buyer shall prepay shipping charges to
Hewlett-P
ackard and Hewlett-Packard shall pay shipping charges to return the
product to Buyer
. However
,Buyer shall pay all shipping charges
,duties
,and
taxes for products returned to Hewlett-P
ackard from another country
.
Hewlett-P
ackard warrants that its software and rmware designated by
Hewlett-P
ackard for use with an instrument will execute its programming
instructions when properly installed on that instrument. Hewlett-P
ackard does
not warrant that the operation of the instrument, or software, or rmware will
be uninterrupted or error-free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from improper or
inadequate maintenance by Buyer, Buyer-supplied software or interfacing,
unauthorized modication or misuse, operation outside of the environmental
specications for the product, or improper site preparation or maintenance
.
NO OTHER W
ARRANTY IS EXPRESSED OR IMPLIED
. HEWLETT-
PACKARD SPECIFICALL
Y DISCLAIMS THE IMPLIED W
ARRANTIES OF
MERCHANTABILITY AND FITNESS FOR A P
ARTICULAR PURPOSE.
Exclusive Remedies
iii

THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE AND EXCLUSIVE
REMEDIES. HEWLETT-PACKARD SHALL NOT BE LIABLE FOR ANY DIRECT,
INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES,
WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER LEGAL THEORY.
Assistance
Product maintenance agreements and other customer assistance agreements
are available for Hewlett-Packard products.For any assistance, contact your
nearest Hewlett-P
ackard Sales and Service Oce
.
Compliance
This instrument has been designed and tested in accordance with IEC
Publication 348, Safety Requirements for Electronic Measuring Apparatus
,and
has been supplied in a safe condition. The instruction documentation contains
information and warnings which must be followed by the user to ensure safe
operation and to maintain the instrument in asafe condition.
Safety Notes
The following safety notes are used throughout this manual. Familiarize
yourself with each of the notes and its meaning before operating this
instrument.
W
ARNING
W
arning denotes ahazard. It calls attention to aprocedure
which, if not correctly performed or adhered to
,could
result in injury or loss of life. Do
not
proceed beyond
a warning note until the indicated conditions are fully
understood and met.
iv

CAUTION
Caution denotes a hazard. It calls attention to a procedure
that, if not correctly performed or adhered to, would result in
damage to or destruction of the instrument. Do
not
proceed
beyond a caution sign until the indicated conditions are fully
understood and met.
v

General Safety Considerations
WARNING
No operator serviceable parts inside. Refer servicing to
qualied personnel. To prevent electrical shock, do not
remove covers.
If this instrument is not used as specied, the protection
provided by the equipment may be impaired. This
instrument must be used in a normal condition (in which
all means for protection are intact) only.
For continued protection against re hazard, replace line
fuse only with same type and rating ([F 5A/250V]). The
use of other fuses or material is prohibited.
CA
UTION
Before switching on this instrument, make sure that the line
voltage selector switch is set to the voltage of the power
supply and the correct fuse is installed.
Always use the three-prong ac power cord supplied with this
instrument. F
ailure to ensure adequate earth grounding by
not using this cord may cause instrument damage
.
Only clean the instrument cabinet using a damp cloth.
L
The instruction documentation symbol. The product is marked
with this symbol when it is necessary for the user to refer to the
instructions in the documentation.
CE The CE mark is a registered trademark of the European
Community. (If accompanied by a year, it is when the design was
proven.)
ISM1-A This is asymbol of an Industrial Scientic and Medical Group 1
Class Aproduct.
CSA The CSA mark is a registered trademark of the Canadian
Standards Association.
vi

Manual Conventions
4
Front-Panel Key
5
This represents a key physically located on the instrument.
NNNNNNNNNNNNNNNNNNNNNNN
Softkey
This indicates a \softkey," a key whose label is determined by
the rmware of the instrument.
Screen Text
This indicates text displayed on the instrument's screen.
vii

HP 8590 EM-Series EMC Analyzer Documentation
Description
The following documents are provided with the HP 8590 EM-Series EMC
Analyzer.
Verication Guide
provides information for verifying the
instrument's performance, specications and
characteristics, and customer support.
User's Guide
describes instrument features and how to make
measurements with your EMC analyzer.
Quick R
eference Guide
provides menu maps
,errormessages, and key
descriptions
,remote programming codes
,and
other helpful charts and tables
.
In This Book
This book helps you test the performance of the EMC analyzer
.The
rst three chapters of this guide provide information for performing the
performance verication tests. The next six chapters provide specications and
characteristics for the six EMC analyzer products
.The last chapter gives helpful
information when your require support.
Chapter 1
contains information on which performance verication tests
that you will perform and lists the equipment required to
perform these tests.
Chapter 2
contains the performance verication tests to ensure your EMC
analyzer is operating properly.
Chapter 3
contains the test records for your use when performing the
performance verication tests
.
Chapter 4
provides the specications and characteristics for the
HP 8591EM EMC analyzer .
Chapter 5
provides the specications and characteristics for the
HP 8593EM EMC analyzer .
Chapter 6
provides the specications and characteristics for the
HP 8594EM EMC analyzer .
viii

Chapter 7
provides the specications and characteristics for the
HP 8595EM EMC analyzer .
Chapter 8
provides the specications and characteristics for the
HP 8596EM EMC analyzer .
Chapter 9
provides the characteristics for the EMC analyzer with the RF
lter section.
Chapter 10
contains information for providing customer support to you if
you have a problem with your EMC analyzer.
1
ix


1
Calibrating
This chapter identies the performance test procedures which test the electrical
performance of the analyzer.
Allow the analyzer to warm up in accordance with the temperature stability
specications before performing the tests called out in this chapter
.
None of the test procedures involve removing the cover of the analyzer
.
Calibration
Calibration veries that the analyzer performance is within all specications
.It
is time consuming and requires extensive test equipment. Calibration consists of
all
the performance tests
.F
or acomplete listing of the performance tests
,see
the performance verication tests table for your specic analyzer
.
Operation Verication
Operation verication only tests the most critical specications. These tests
are recommended for incoming inspection, troubleshooting, or after repair.
Operation verication requires less time and equipment than the calibration.
See the performance verication tests table for your analyzer
.
Calibrating 1-1

Calibration Cycle
The performance tests in Chapter 2 should be used to check the analyzer against
its specications once every year. Specications are listed in this calibration
guide.
The 300 MHz frequency of the CAL OUT signal must be checked at the same
time and adjusted if necessary. Refer to the \10 MHz Frequency Reference
Adjustment" procedure in the assembly-level repair service guide.
1-2 Calibrating

Performance Verication Test Tables
The tables on the following pages list the performance tests in chapter 2. Select
the analyzer option being calibrated and perform the tests marked in the option
column.
A dot indicates that the test is required for calibration. Note that some of the
tests are used for both calibration and operation verication (marked with
4
5
).
Calibrating 1-3

Table 1-1. HP 8591EM Performance Verication Tests
Performance Test Name
Calibration for Instrument Option:
Std
1
004 010 101
1. 10 MHz Reference Output Accuracy
2. 10 MHz Precision Frequency Reference Output
Accuracy
4. Frequency Readout and Marker Count Accuracy
4
5 4
5 4
5 4
5
6. Noise Sidebands
4
5 4
5 4
5 4
5
7. System Related Sidebands
8. Frequency Span Readout Accuracy
4
5 4
5 4
5 4
5
10. Residual FM
12. Sweep Time A
ccuracy
13. Scale Fidelity
4
5 4
5 4
5 4
5
14. Reference Level A
ccuracy
4
5 4
5 4
5 4
5
16. Absolute Amplitude Calibration and Resolution (IF)
Bandwidth Switching Uncertainties
4
5 4
5 4
5 4
5
17. Resolution (IF) Bandwidth A
ccuracy
18. Calibrator Amplitude A
ccuracy
4
5 4
5 4
5 4
5
19. Frequency Response
4
5 4
5 4
5 4
5
24. Other Input Related Spurious Responses
29. Spurious Response
2
4
5 4
5 4
5 4
5
34. Gain Compression
39. Displayed A
verage Noise Level
4
5 4
5 4
5 4
5
44. Residual Responses
47. F
ast Time Domain Sweeps
49. Absolute Amplitude,Vernier,andPower Sweep
Accuracy
52. Tracking Generator Level Flatness
54. Harmonic Spurious Outputs
56. Non-Harmonic Spurious Outputs
58. Tracking Generator F
eedthrough
62. CISPR Pulse Response
1
Use this column for all other options
not
listed in this table
.
2
\Part 2: Third Order Intermodulation Distortion, 50 MHz" is not required for operation verication.
1-4 Calibrating

Table 1-2. HP 8593EM Performance Verication Tests
Performance Verication Test Name
Calibration for Instrument Option:
Std
1
004 010 026 027 101
1. 10 MHz Reference Output Accuracy
2. 10 MHz Precision Frequency Reference Output
Accuracy
3. Comb Generator Frequency Accuracy
5. Frequency Readout and Marker Count Accuracy
4
5 4
5 4
5 4
5 4
5 4
5
6. Noise Sidebands
4
5 4
5 4
5 4
5 4
5 4
5
7. System Related Sidebands
9. Frequency Span Readout Accuracy
4
5 4
5 4
5 4
5 4
5 4
5
11. Residual FM
12. Sweep Time A
ccuracy
13. Scale Fidelity
4
5 4
5 4
5 4
5 4
5 4
5
15. Reference Level A
ccuracy
4
5 4
5 4
5 4
5 4
5 4
5
16. Absolute Amplitude Calibration and Resolution (IF)
Bandwidth Switching Uncertainties
4
5 4
5 4
5 4
5 4
5 4
5
17. Resolution (IF) Bandwidth A
ccuracy
18. Calibrator Amplitude A
ccuracy
4
5 4
5 4
5 4
5 4
5 4
5
20. Frequency Response
4
5 4
5 4
5 4
5 4
5 4
5
25. Other Input Related Spurious Responses
30. Spurious Response
2
4
5 4
5 4
5 4
5 4
5 4
5
35. Gain Compression
40. Displayed A
verage Noise Level
4
5 4
5 4
5 4
5 4
5 4
5
46. Residual Responses
48. Fast Time Domain Sweeps
50. Absolute Amplitude Accuracy
51. Power Sweep Range
53. Tracking Generator Level Flatness
55. Harmonic Spurious Outputs
57. Non-Harmonic Spurious Outputs
60. Tracking Generator F
eedthrough
61. Tracking Generator LO Feedthrough Amplitude
62. CISPR Pulse Response
1
Use this column for all other options
not
listed in this table
.
2
\Part 2: Third Order Intermodulation Distortion, 50 MHz" is not required for operation verication.
Calibrating 1-5

Table 1-3. HP 8594EM Performance Verication Tests
Performance Verication Test Name
Calibration for Instrument Option:
Std
1
004 010 101
1. 10 MHz Reference Output Accuracy
2. 10 MHz Precision Frequency Reference Output
Accuracy
4. Frequency Readout and Marker Count Accuracy
4
5 4
5 4
5 4
5
6. Noise Sidebands
4
5 4
5 4
5 4
5
7. System Related Sidebands
9. Frequency Span Readout Accuracy
4
5 4
5 4
5 4
5
11. Residual FM
12. Sweep Time A
ccuracy
13. Scale Fidelity
4
5 4
5 4
5 4
5
15. Reference Level A
ccuracy
4
5 4
5 4
5 4
5
16. Absolute Amplitude Calibration and Resolution (IF)
Bandwidth Switching Uncertainties
4
5 4
5 4
5 4
5
17. Resolution (IF) Bandwidth A
ccuracy
18. Calibrator Amplitude A
ccuracy
4
5 4
5 4
5 4
5
21. Frequency Response
4
5 4
5 4
5 4
5
26. Other Input Related Spurious Responses
31. Spurious Response
2
4
5 4
5 4
5 4
5
36. Gain Compression
41. Displayed A
verage Noise Level
4
5 4
5 4
5 4
5
45. Residual Responses
48. F
ast Time Domain Sweeps
50. Absolute Amplitude Accuracy
51. Power Sweep Range
53. Tracking Generator Level Flatness
55. Harmonic Spurious Outputs
57. Non-Harmonic Spurious Outputs
59. Tracking Generator F
eedthrough
61. Tracking Generator LO F
eedthrough Amplitude
62. CISPR Pulse Response
1
Use this column for all other options
not
listed in this table
.
2
\Third Order Intermodulation Distortion" is not required for operation verication.
1-6 Calibrating

Table 1-4. HP 8595EM Performance Verication Tests
Performance Verication Test Name
Calibration for Instrument Option:
Std
1
004 010 101
1. 10 MHz Reference Output Accuracy
2. 10 MHz Precision Frequency Reference Output
Accuracy
5. Frequency Readout and Marker Count Accuracy
4
5 4
5 4
5 4
5
6. Noise Sidebands
4
5 4
5 4
5 4
5
7. System Related Sidebands
9. Frequency Span Readout Accuracy
4
5 4
5 4
5 4
5
11. Residual FM
12. Sweep Time A
ccuracy
13. Scale Fidelity
4
5 4
5 4
5 4
5
15. Reference Level A
ccuracy
4
5 4
5 4
5 4
5
16. Absolute Amplitude Calibration and Resolution (IF)
Bandwidth Switching Uncertainties
4
5 4
5 4
5 4
5
17. Resolution (IF) Bandwidth A
ccuracy
18. Calibrator Amplitude A
ccuracy
4
5 4
5 4
5 4
5
22. Frequency Response
4
5 4
5 4
5 4
5
27. Other Input Related Spurious Responses
32. Spurious Response
2
4
5 4
5 4
5 4
5
37. Gain Compression
42. Displayed A
verage Noise Level
4
5 4
5 4
5 4
5
46. Residual Responses
48. F
ast Time Domain Sweeps
50. Absolute Amplitude Accuracy
51. Power Sweep Range
53. Tracking Generator Level Flatness
55. Harmonic Spurious Outputs
57. Non-Harmonic Spurious Outputs
60. Tracking Generator F
eedthrough
61. Tracking Generator LO F
eedthrough Amplitude
62. CISPR Pulse Response
1
Use this column for all other options
not
listed in this table
.
2
\Third Order Intermodulation Distortion" is not required for operation verication.
Calibrating 1-7

Table 1-5. HP 8596EM Performance Verication Tests
Performance Verication Test Name
Calibration for Instrument Option:
Std
1
004 010 101
1. 10 MHz Reference Output Accuracy
2. 10 MHz Precision Frequency Reference Output
Accuracy
3. Comb Generator Frequency Accuracy
4
5 4
5 4
5 4
5
5. Frequency Readout and Marker Count Accuracy
4
5 4
5 4
5 4
5
6. Noise Sidebands
4
5 4
5 4
5 4
5
7. System Related Sidebands
9. Frequency Span Readout Accuracy
4
5 4
5 4
5 4
5
11. Residual FM
12. Sweep Time A
ccuracy
13. Scale Fidelity
4
5 4
5 4
5 4
5
15. Reference Level A
ccuracy
4
5 4
5 4
5 4
5
16. Absolute Amplitude Calibration and Resolution (IF)
Bandwidth Switching Uncertainties
4
5 4
5 4
5 4
5
17. Resolution (IF) Bandwidth A
ccuracy
18. Calibrator Amplitude A
ccuracy
4
5 4
5 4
5 4
5
23. Frequency Response
4
5 4
5 4
5 4
5
28. Other Input Related Spurious Responses
33. Spurious Response
2
4
5 4
5 4
5 4
5
38. Gain Compression
43. Displayed A
verage Noise Level
4
5 4
5 4
5 4
5
46. Residual Responses
48. Fast Time Domain Sweeps
50. Absolute Amplitude Accuracy
51. Power Sweep Range
53. Tracking Generator Level Flatness
55. Harmonic Spurious Outputs
57. Non-Harmonic Spurious Outputs
60. Tracking Generator F
eedthrough
61. Tracking Generator LO F
eedthrough Amplitude
62. CISPR Pulse Response
1
Use this column for all other options
not
listed in this table
.
2
\Third Order Intermodulation Distortion" is not required for operation verication.
1-8 Calibrating

Safety
Familiarize yourself with the safety symbols marked on the analyzer, and read
the general safety instructions and the symbol denitions given in the front of
this guide
before
you begin verifying performance of the EMC analyzer.
Before You Start
There are four things you should do before starting a performance verication
test:
Switch the analyzer on and let it warm up in accordance with the
temperature stability specication.
Read \Making aMeasurement" in your analyzer user's guide
.
After the analyzer has warmed up as specied, perform the self-calibration
procedure documented in \Improving A
ccuracy With Self-Calibration
Routines" in the
HP 8590 EM-Series EMC Analyzer User's Guide
.The
performance of the analyzer is only specied after the analyzer calibration
routines have been run and if the analyzer is autocoupled.
Read the rest of this section before you start any of the tests
,and make
acopy of the P
erformance V
erication T
est Record described below in
\Recording the test results
."
T
est equipment you will need
Tables 1-6 through 1-9 list the recommended test equipment for the performance
tests. Any equipment that meets the critical specications given in the table can
be substituted for the recommended model.
Recording the test results
P
erformance verication test records
,for each EMC analyzer
,are provided in
the chapter following the tests
.
Each test result is identied as a
TR Entry
in the performance tests and on the
performance verication test record. W
e recommend that you make a copy of
the performance verication test record, record the test results on the copy
,and
Calibrating 1-9
Other manuals for 8590 E-Series
1
Table of contents
Other HP Measuring Instrument manuals