HyQuest Solutions iLevel-GW User manual

iLevel-GW
User Manual

© 2022 HyQuest Solutions 2
Table of Contents
Disclaimer 3I
II Safety Instructions 4
III Specific Safety Instructions 5
Part I Introduction 6
Part II Installation 7
2.1 Mechanics ........................................................................................................................................................... 7
2.2 Fitting into Measuring Tube ......................................................................................................................... 7
2.3 Fitting the Antenna ......................................................................................................................................... 7
2.4 Fitting the SIM Card ........................................................................................................................................ 7
2.5 Assembling and Changing Batteries .......................................................................................................... 8
2.6 Initial Commissioning ..................................................................................................................................... 8
Part III Configuration 9
3.1 Functions / Settings ........................................................................................................................................ 9
3.1.1 Setup ........................................................................................................................................................... 10
3.1.2 Service ........................................................................................................................................................ 10
3.1.3 Main Window ........................................................................................................................................... 11
3.1.4 Parameters ............................................................................................................................................... 12
3.1.5 Scale ............................................................................................................................................................ 13
3.1.6 More Buttons ............................................................................................................................................ 13
3.2 GPRS / GSM Setup ......................................................................................................................................... 14
3.3 Two Point Calibration .................................................................................................................................. 15
3.4 Tare .................................................................................................................................................................... 16
Part IV Maintenance 17
4.1 HyQuest Solutions GP-Shell – local Software ....................................................................................... 17
4.2 Loggers ............................................................................................................................................................. 17
Part V Repair 18
Part VI Technical Data 19
Part VII Obligations of the Operator and Disposal 20
7.1 Obligations of the Operator ....................................................................................................................... 20
7.2 Dismantling / Disposal ................................................................................................................................ 20
Part VIII Appendices 22
8.1 FAQ – Frequently asked questions .......................................................................................................... 22

Copyright Notice: No parts of this work may be
reproduced in any form or by any means without the
written permission of the publisher. HyQuest
Solutions waives copyright for users to print out
parts of the documentation in hard copy for their
own use only.
Trademark Notice: HyQuest Solutions (HS) and
KISTERS products and services referred to in this
document are trademarks or registered trademarks
of HyQuest Solutions or KISTERS AG. Other product
names used may or may not be the trademarks of
their respective owners.
© 2020 HyQuest Solutions, a KISTERS Group
Company. Any rights not expressly granted herein
are reserved.
The information provided in this manual was deemed
accurate as of the publication date. However, updates
to this information may have occurred.
This manual does not include all of the details of
design, production, or variation of the equipment
nor does it cover every possible situation which may
arise during installation, operation or maintenance.
HyQuest Solutions shall not be liable for any
incidental, indirect, special or consequential damages
whatsoever arising out of or related to this
documentation and the information contained in it,
even if HyQuest Solutions has been advised of the
possibility of such damages.
Any errors found in any HyQuest Solutions product
should be reported to HyQuest Solutions where
every effort will be made to quickly resolve the
problem.
This document is public.
3© 2022 HyQuest Solutions
I Disclaimer

§
Read the user manual including all operating instructions prior to installing, connecting and powering up the HyQuest
Solutions iLevel-GW. The manual provides information on how to operate the product. The manual is intended to be
used by qualified personnel, i.e. personnel that have been adequately trained, are sufficiently familiar with
installation, mounting, wiring, powering up and operation of the product.
§
Keep the user manual on hand for later reference!
§
If you encounter problems understanding the information in the manual (or part thereof), please consult the
manufacturer or its appointed reseller for further support.
§
HyQuest Solutions iLevel-GW is intended to be used in hydrometeorological or environmental monitoring
applications.
§
Before starting to work, you have to check the functioning and integrity of the system.
§
Check for visible defects on the iLevel-GW, this may or may not include any or all of the following mounting
facilities, connectors and connections, mechanical parts, internal or external communication devices, power
supplies or power supply lines, etc.
§
If defects are found that jeopardize the operational safety, work must be stopped. This is true for defects found
before starting to work as well as for defects found while working.
§
Do not use the HyQuest Solutions iLevel-GW in areas where there is a danger of explosion.
§
The present user manual specifies environmental/climatic operating conditions as well as mechanical and electrical
conditions. Installation, wiring, powering up and operating the HyQuest Solutions iLevel-GW must strictly comply with
these specifications.
§
Perform maintenance only when tools or machinery are not in operation.
§
If guards are removed to perform maintenance, replace them immediately after servicing.
§
Never make any electrical or mechanical diagnostics, inspections or repairs under any circumstances. Return the
sensor to the manufacturer’s named repair centre. You can find information on how to return items for repair in the
relevant section of the HyQuest Solutions web site.
§
Disposal instructions: After taking the HyQuest Solutions iLevel-GW out of service, it must be disposed of in
compliance with local waste and environmental regulations. The HyQuest Solutions iLevel-GW is never to be disposed
in household waste!
§
Inputs and outputs of the device are protected against electric discharges and surges (so-called ESD). Do not
touch any part of the electronic components! If you need to touch any part, please discharge yourself, i.e. by touching
grounded metal parts.
II Safety Instructions
© 2022 HyQuest Solutions 4

III Specific Safety Instructions
© 2022 HyQuest Solutions 5
§
The manual provides information on how to operate the sensor system.
§
HyQuest iLevel-GW is intended to be used in hydrological surveys, and more specifically to determine groundwater
and surface water levels.
§
Never use the the HyQuest Solutions iLevel-GW to transport people!
§
Special caution is needed for all works and operations in the vicinity of water streams that may result in a risk of
drowning. Work places must be equipped by the operator of the site with the required safety and rescue equipment
(life vest, buoys, ropes, etc.).

© 2022 HyQuest Solutions 6
Introduction
1Introduction
Thank you for choosing our product. We hope you will enjoy using the device.
HyQuest Solutions manufactures, sells, installs and operates quality instrumentation, data loggers and communication
technology. Products are designed with passion for environmental monitoring and with a deep understanding of the
quality, accuracy and robustness needed to fulfil the requirements of measurement practitioners in the field.
The present User Manual will help you understand, install and deploy the device. If, however, you feel that a particular
information is missing, incomplete or confusing, please do not hesitate to contact us for further support!
iLevel-GW is an IP data logger designed for very low power data acquisition of level and temperature in groundwater and
surface waters. The system can be easily expanded into a multi-channel monitoring device by adding optional sensors for
water quality monitoring. By default, an iLevel-GW comes with a coupled water level and water temperature sensor
accurate to 0.1 % FS. The vented pressure sensors auto-compensate for atmospheric pressure and temperature variations
providing a corrected reading at all times.
iLevel-GW loggers are equipped with a 433 MHz near field radio interface for all local communications. The wireless
technology allows to connect to an iLevel-GW installed in a borehole or other 2” pole without any need to remove or
even touch the device.
At the time of order, users can choose from a series of distinct communication types:
1. 2G/GPRS (General Package Radio Service).
2. 3G/UMTS
3. 4G/LTE Cat M
All of these data communication methods use IP protocols to transmit data directly to the server via cellular network.
Using that technology, it becomes possible to set up even large monitoring networks which can still be handled
comfortably and require little maintenance. Furthermore, using the cellular networks is a cost-saving way to transfer the
data, as it doesn't require any further hardware to be installed in the field.
Network time information is used to sync the internal clock of the data logger, allowing for perfect timing at all times.

© 2022 HyQuest Solutions 7
Installation
2Installation
This chapter contains the following subsections:
§
Mechanics
§
Fitting into Measuring Tube
§
Fitting the Antenna
§
Fitting the SIM Card
§
Assembling and Changing Batteries
§
Initial Commissioning
2.1 Mechanics
The housing consists of two V4A-steel casing pipes and a plastic fitting inside, in which the battery and the electronic
system are located.
The joints and gaskets must not get dirty or damaged!
Prior to assembling the device, joints and gaskets must be checked and lubricated (silicone grease or O ring grease).
Attention: Don't use force!
Replace the desiccant (i.e. Silica gel(R)) as necessary. The desiccant bag is located at the cover panel of the electronics.
Important: Check the O-ring seal every time the device is reassembled. Any damaged seal must be replaced! Depending on
which type of desiccant is used, the humidity should be not higher than 20%. Humidity above 20% might indicate broken
seals. At a humidity higher than 50% the electronic is at risk! Advice: most of desiccants can be re-generated by drying at
120 °C…150 °C for 20 minutes.
2.2 Fitting into Measuring Tube
For installation in a 2 " tube, the border of the cap of the data logger is placed on the edge of the tube.
To fit the logger into a 3” or 4” tube, an adapter ring is required. Adapter rings of 3” or larger have an opening for light
plummets. The antennas are fitted individually: For underfloor installations underfloor antennas are available. Those are
clamped between the frame and the lid.
Helical antennas are fitted onto the lid for above-floor measuring points. For fitting the antenna, the lid is pierced with an
appropriate drill bit in the centre. Using that hole the antenna can be screwed onto the lid. Inside buildings, a stub
antenna can be screwed onto the logger. Once the logger is fitted, a comparative measurement should be performed
using a light plummet.
2.3 Fitting the Antenna
Note: The device comes with a standard GSM-plug (FME) for the antenna. Various antennas can be connected using the
FME plug. Your cable may be brought out through a small hole or a small notch. It is recommended to mount the antenna
on a metallic surface. When connecting the antenna, plug in the plug as far as possible for optimal contact. Please do not
unnecessarily tighten the knurled screw. Note: In case of low signal strength a slight change of the antenna position can
cause a significant improvement in signal strength (just a few cm). If the signal is too low at all, antennas with higher
transmitting / receiving power or directional effect can be a good solution. We will be glad to help you finding the
optimum antenna for your area!
2.4 Fitting the SIM Card
Important: To open the card holder, press your finger on the drawer slightly and push it in the direction which is marked
"Push Open", so that the latch releases the drawer. To lock, simply press your finger on the drawer slightly again and push
it in the other direction which is marked "Push Open" until the latch closes.
7
7
7
7
8
8

© 2022 HyQuest Solutions 8
Installation
2.5 Assembling and Changing Batteries
Possible types of batteries are 2 mono-D-cells with 1.5 Volt or 1 lithium DD 35Ah with 3.6V.
Caution: To avoid damage, use appropriate batteries only!
To assemble the battery, remove the steel casing pipe which covers the battery compartment.
For this, twist off the ring nut and pull off the casing pipe.
Attention: Please mind the polarity of the batteries. Inserting the battery incorrectly may blow the fuse (which can only
be repaired at the factory) or even damage the electronics.
The electronic comes with a backup capacitor which can compensate a power outage for a short time (i.e. during a battery
change), so it doesn't lose data like date and time.
If the outage was too long, the device will reboot, which is indicated by the green LED blinking about 10 times. After that
the device is ready to use. The time has to be synchronized again.
2.6 Initial Commissioning
Due to European law, devices will be shipped deenergized only. See Assembling and Changing Batteries on how to fit
the batteries.
8

© 2022 HyQuest Solutions 9
Configuration
3Configuration
The software HyQuest Solutions GP shell is used for local communication between a PC and the iLevel-GW data logger. For
the communication, a 433 MHz radio link is used. Using HyQuest Solutions GP shell, the parameters of the logger can be
configured and measured values are shown directly.
Please install the HyQuest Solutions GP shell before you connect the dongle to the PC for the first time! The drivers
needed for the installation of the dongle are installed with the HyQuest Solutions GP shell.
Please use the supplied CD or download the latest version of the software from the customer homepage. Double-click the
*. Exe file to start the installation. When the installation was successful, click "Exit Installation".
This chapter contains the following subsections:
§
Functions / Settings
§
GPRS / GSM Setup
§
Two Point Calibration
§
Tare
3.1 Functions / Settings
The start window of the shell is divided into 3 areas:
§
Menu with buttons
§
Main window with current information
§
List of available loggers
§
Setup
§
Service
§
Main Window
§
Parameters
§
Scale
§
More Buttons
9
14
15
16
10
10
11
12
13
13

© 2022 HyQuest Solutions 10
Configuration
3.1.1 Setup
Figure 1 – Setup Dialogue: General Settings
§
Data Path ('SERIAL'): At this path, one folder for each new logger will be created. Each Folder will be named after the
serial number of the logger. Parameters / Settings and all raw data will be saved in each folder.
§
Path for data export ('EXPORT'): All exported data (i.e. MS Excel) will be stored to the folder entered in that field.
§
Open Data with built-in Viewer (Text mode): The measured values will be shown as Text files.
§
Open Data with CSVies (Graphical Viewer): The measured values will be shown as hydrograph and table.
§
Open exported Data as CSV-File (Spreadsheet): The measured values will be shown as CSV File (Comma separated
values).
§
Decimal Comma: Use comma (,) as decimal separator when exporting data.
§
Extended Info in exported Data: Check this option, if you want additional Information on actions (i.e. GPRS-
transmissions, parameter adjustment) to be added to your export-file.
§
Optional Internet Upload: Enabling this option, data from third-party loggers can be transmitted to the server.
§
Language: Language for the user interface can be chosen from a drop-down menu.
§
Product activation: Additional features can be activated by entering serial numbers. Further information on these
features can be found in the feature’s separate manuals.
3.1.2 Service
§
GPRS Mail: By clicking this button, a test mail will be sent to the server in order to check the GPRS connection. The
Logger has to be identified by the software before this option becomes available. Once the button is clicked, a terminal
window shows all steps of the GPRS-transmission.
§
Terminal: Opens a terminal window. The terminal mode offers the option of showing text output of the logger. This
option can be useful for monitoring data transmissions to the server, as every step of the transmission will be listed.
Under normal circumstances, this option will not be needed. As the data is not coming from HyQuest Solutions GP-
Shell but directly from the logger, the language settings of GP-Shell will be ignored. The output will always be in
English.

© 2022 HyQuest Solutions 11
Configuration
§
Echotest: Clicking this button, the signal strength of the 433 MHz near field radio connection will be tested. Data
packages will be sent, until the test is aborted manually by the user.
§
Update Logger Firmware: Perform an update of the logger’s firmware. After a security prompt, new firmware files can
be selected a transferred to the logger. Please refer to the chapter “Updates” for further information.
§
Update USB-Dongle Firmware: Perform an update of the USB-dongles firmware. After a security prompt, new
firmware files can be selected and transferred to the USB-dongle. Please refer to the chapter “Updates” for further
information.
§
All Radio OFF: All loggers in range of the dongle will be set to 'flight mode', so the they will not send any more data.
Important: The loggers will also disappear from the list in GP-Shell.
§
All Radio ON: All loggers in range of the dongle will be set from 'flight mode' back to 'normal mode'.
3.1.3 Main Window
§
Identify Logger: Establishes a connection to a logger. To connect to a logger, select a logger from the list before clicking
“Identify Logger”.
§
Synchronize Clock: Synchronizes logger time with time of the current system.
§
Display Measure: Show current measurement values (status of logger and measured values of the activated measuring
probe). The values measured when clicking this button will not be stored and not be transmitted to the server, but can
be used for local calibration of data loggers.
§
Load to Disk: Download measured values from the data logger to your local computer. You can choose between
incremental download (only values that have not been transmitted to the server yet will be downloaded) and full
download (all values stored on the logger will be downloaded). The local path for saving the data can be set in the
setup menu (see Setup ).
§
Attention: If the modem of the logger is switched off, any data downloaded to your computer will be marked as sent
and will not be transmitted during the next GPRS-transmission. In doing so, the loggers can be operated even without
the use of the GPRS-connection.
§
Clear Logger: By clicking this button, the whole memory of the logger will be cleared. (All measured data will be
deleted!)
§
Parameters: Opens a menu for configuring a logger. The logger must be identified first.
§
Data from Disk: Open Files that have been downloaded to the local computer previously, i.e. measured values (i.e.
*.g2p files).
§
Exit: Closes the application.
10

© 2022 HyQuest Solutions 12
Configuration
3.1.4 Parameters
§
Global Parameters: The measuring time intervals are set here.
§
Period: Enter measuring interval. The minimum for this parameter is 1 minute for GPRS-transmission. If
Monitoring on the site, a 2 second interval can be used in combination with the live mode (i.e. pumping tests) If the
interval is set as min, sec or hour can be selected from the drop-down menu next to the field. If you set an interval
of less than 1 minute, the program will refuse to transfer the parameters to the logger if the transmission intervals
are too large to all data measured in the interval to be sent in a single transmission.
§
Period-Offset: From a period of 5 minutes or more, the logger automatically selects the first measurement, the
next "modular" time, i.e. at a period of 1 h, this would be the next full hour. If for example a period of 12 hours is
selected, the measurements would be made, at 0:00 and 12:00. With the "Period Offset" the time of measurement
can be postponed. If an offset of 2 hours is selected for example, the measurements would be made at 2:00 am
and 2:00 pm.
§
Alarm period: Interval of measurements in case of an alarm
§
Name: Custom name for each logger (for example, tag number, street name). IMPORTANT: The name cannot
contain special characters.
§
Live mode (Real-Time): The live mode is used for continuous local transmission of data via radio to a receiver.
When the real time mode is active, the device sends the currently measured data to the shell software which
displays this data in the main window. This allows to monitor the water level in real-time, for example when
changing water levels by pumping. The shortest possible measuring period is two seconds.
§
There are 3 modes for the live mode:

© 2022 HyQuest Solutions 13
Configuration
§
Intelligent: Logger detects when a receiver is in range and then sends data automatically in the measuring
interval set.
§
Always: The logger will always send data, not matter whether there is a receiver in range or not. This setting
requires continuous power!
§
Off: The logger does not send any data.
§
HK-Counter: Set transmission interval of internal self-monitoring values. 1 = with every normal transmission. 10 =
with every 10th transmission. In case of alarm the self-monitoring values will be collected at each measurement.
§
Log HK: Select, which self-monitoring values shall be collected. The following options can be selected: Supply
(battery), temperature (Housing temperature) and humidity (Housing inside humidity).
§
Parameter Set: Shows which set of parameters is currently selected and cannot be changed. The number contains
the logger ID, year, month, day, hour, minute, second (e.g. 100012_20090812080050: Logger Nr 100012 _ 12.
August 2009 at 08:00:50) The number is set every time parameters are sent to the logger.
§
GPRS / GSM Setup: Opens the window for transmission and modem settings.
§
Channel Parameters: Settings of the channel parameters can be changed here. Channel 1 is for the measurement of
water level / tapping, channel 2 is for measuring the water temperature.
§
# 1: means channel 1 is selected. Switch the channel to be edited by clicking the arrow up / down buttons.
§
Type: indicates which value is measured in this channel and cannot be changed manually.
§
Unit: Choose the Unit to be measured.
§
ID: This parameter is intended to mark the measuring point, for example, in a database. As long as "0" is set, this
value has no effect. Values from 1 to 65535 can be entered. The value entered here will be listed in the table
headings of the exported data. Whether and how this information will be taken into consideration when importing
the data is the responsibility of the importing software.
§
Action: Each action can be activated by checking.
§
Log channel: Measured data of this channel will be saved when this option is checked.
§
Check alarms: Check this option if you want to set special alarm events (e.g. shorter measuring and
transmissions intervals).
3.1.5 Scale
§
Offset: Offset (available if no pressure sensor is used for the water level): This value will be subtracted from the
measured value (e.g. if the value should be displayed as height above sea level). Note: positive numbers are
subtracted from the measured value, negative numbers are added!
§
Multi: Default setting is 1.000000. If a minus (-) is set before the first digit, the value entered becomes negative. Use
this to select whether water column or tapping should be measured. The adjustment value can also be calculated
under the tare button. Furthermore, the density of the water can be set here. The value 1.000000 is calibrated for
fresh water. Depending on the salinity, the value has to be reduced (i.e.: 3% salt content = 0.997000).
§
Tare: Opens the tare window (see Tare ).
§
Alarm: If the thresholds entered here are exceeded or fallen below, the device will switch to alarm mode. High: If the
measured value falls below the value entered here, the alarm is activated.
§
Low: If the value entered here is reached or exceeded, the alarm is activated.
§
Power Up Wait: warm-up time for the sensor (enter 1000 ms for PTM/N from STS)
§
Sensor min / max.: Not required if a pressure sensor is used. If slower sensors are used, a period of time can be set for
the measuring time. Settings depend on the sensor type.
§
Type: Drop Down Menu. Choose between SDI (for general SDI sensors i.e. temperature), SDI12-Pressure (for the
pressure channel of the iLevel-GW and wireless sensor (to connect more sensors via near field radio)
§
No measure, use cached values: must be enabled if a sensor sends multiple values (i.e. pressure and temperature)
§
Index: Numbers of sensor channels (i.e. pressure = 0 as first channel and temperature = 1 as second channel.
§
S.Nr.: SDI 12 address of the selected sensor (default is 0)
§
Cmd.: 0 means SDI command m!, which triggers measuring
3.1.6 More Buttons
§
Transfer: Transfers the new Parameter Set to the logger.
§
Cancel: Cancel without transferring the new parameter set to the logger.
§
Write Par...: Save the new parameter set to the local computer without transferring it to the logger.
§
Read Par...: Read a parameter set that has been saved to the local computer before.
16

© 2022 HyQuest Solutions 14
Configuration
3.2 GPRS / GSM Setup
§
General Settings
§
GSM/GPRS Mode: Switch the modem on or off. By default, the modem is switched on.
§
GSM PIN: Enter the PIN of the SIM-card here, if required.
§
Min. Temp (°C): Enter the temperature until which the modem should remain active. The default is -10 °C for
alkaline manganese batteries and -25 °C for lithium batteries. It is not recommended to enter temperatures lower
than the default temperatures. If the temperature falls below the temperature entered here, the logger will keep
measuring, but send the data not until the temperature exceeds the limit entered here.
§
Allow Roaming: When enabled, the modem is allowed to use foreign networks. When disabled, only the home
network will be used. This option becomes important when the device is operated in regions close to the border.
§
Stay online: The modem will be operated continuously, so data can be sent every minute. Attention: This feature
increases the power consumption!
§
Verbose: Show detailed information about the internet connection. Useful for finding connection errors.
§
GSM period: Set the transmission intervals in which the data is transmitted. The minimum for this setting is five
minutes. Use the drop-down menu to switch between seconds, minutes and hours.
§
GSM Offset: Sets the time for the data transmission to the server. If, for example, the interval is set to 24 hours and
the offset is set to 6 hours, a transfer is performed daily at 6 o’clock. If the interval is set to 12 hours and the offset
to 6 hours, a transfer is performed at 6 o’clock and 18 o’clock.
§
GSM alarm period: Set the transmission intervals in which the data is transmitted in case of an alarm. The minimum
for this setting is five minutes. Use the drop-down menu to switch between seconds, minutes and hours.
§
Network parameter
§
GPRS APN: (Access Point Name) The GSM Access point provided by the GSM network provider.
§
PPP-User and PPP-Password: Also provided by the GSM provider. Can be left empty often.
§
GPRS repeat: By activating the check box the logger is allowed to repeat transmissions up to 3 times in case a
transmission has failed. For different reasons, such as high network load, transmissions can fail from time to time.
If this option is enabled, waiting for the repeated transmission and sending again, consumption battery power.

© 2022 HyQuest Solutions 15
Configuration
Therefore, it can make sense to disable the option. If disabled, the collected data will be transmitted during the
next planned transmission.
§
Internet-Parameter
§
Internet Server: Enter the address or the ip number if the server to which the data should be sent.
§
Important: Always enter the address without 'www' at the beginning. For example, enter sensor manager.net
instead of www.sensormanager.net.
§
Script: Enter the location of the PHP script which receives the information on the server.
§
E-MAIL Format: When the data is transmitted to the sever, the server can send the data to an e-mail address. You
can choose between plain text e-mails or html formatted e-mails. The data is attached to the e-mail either as a
binary file or as a plain text file.
§
E-MAIL address: If you want to receive the data via e-mail, enter the e-mail dress to which the data should be sent
here.
§
GRPS Synchro.: When enabled, the logger will synchronize its clock with the server every time a transmission is
performed.
§
Auto Delete: When enabled, the logger will delete all logged data when 95% of the memory is used. This will
happen only if all data has been transmitted successfully.
§
GPRS Offset (sec): Time difference between the logger and the server in seconds. The server’s clock is mostly
operated in GMT mode (Greenwich without summer time). When using a logger in another time zone (e.g. Berlin,
Rome = GMT+1) the value to be added to the time must be entered here. For example, GMT+1 requires a value of
3600 seconds to be entered here.
§
Alarm Receiver
§
Alarm: E-Mail Address #1 - #4: You can enter up to four e-mail addresses to which a mail will be sent in case if an
alarm. Important: The alarm set here is performed by the logger via a script on the server.
§
OK: Confirm the new settings by pressing the OK button and then transfer it by clicking the Transfer button.
§
Cancel: Close the current dialog without saving any changes.
3.3 Two Point Calibration
Perform a linear calibration of a sensor. Therefore, 2 points need to be calibrated. The points can be calibrated one after
one or both at on time.
§
Y1,Y2 (nominal): Enter the measured value of the sensor
§
Y1,Y2 (actual): Enter a measured value of a reference.
§
Calculate and OK: Save the settings to the parameter set.
§
Reset Coefficients: Restore the original parameters of the sensor.

© 2022 HyQuest Solutions 16
Configuration
3.4 Tare
Choose, if tare or tapping should be indicated
§
Measure (Actual): If a current measured value is available, it will be shown here. If not, it will display “No value”. In
that case, the current value can be measured and displayed by clicking the button “Measure...”.
§
Set Point (Nominal): Enter the required current value. This can be a currently measured tapping or the water level
taking the main sea level into account.
§
Calculate and OK: If all settings are completed, click “Calculate and OK”. The software now calculates the values for the
data fields “Offset” and “Multi”.
§
Reset calibration: Set the values for Multi and Offset to the default settings (Offset = 0.00, Multi = 1.00)
§
Cancel: Close the windows without changing any settings.

© 2022 HyQuest Solutions 17
Maintenance
4Maintenance
This chapter contains the following subsections:
§
HyQuest Solutions GP-Shell – local Software
§
Loggers
4.1 HyQuest Solutions GP-Shell – local Software
The software “HyQuest Solutions GP-Shell” does not require any maintenance by the customer. If an updated version is
released.
4.2 Loggers
Our loggers only require very low maintenance. Do not use any sharp tool or any aggressive cleaning agent to clean the
hardware. Replace batteries only if necessary. Do not open the battery case during wet weather conditions. Humidity can
cause system failure.
17
17

© 2022 HyQuest Solutions 18
Repair
5Repair
HyQuest Solutions precision instruments and data loggers are produced in quality-controlled processes. All HyQuest
Solutions production and assembly sites in Australia, New Zealand and Europe are ISO 90001 certified. All equipment is
factory tested and/or factory calibrated before it is shipped to the client. This ensures that HyQuest Solutions products
perform to their fullest capacity when delivered.
Despite HyQuest Solutions most rigorous quality assurance (QA), malfunction may occur within or outside of the warranty
period. In rare cases, a product may not be delivered in accordance with your order.
In such cases HyQuest Solutions’ return and repair policy applies. For you as a customer, this means the following:
1. Contact HyQuest Solutions using the Repair Request Form made available online:
https://www.hyquestsolutions.eu/fileadmin/Form/Telemetry_Solution_app.jpg.pdf
In response you will receive a reference number that must be referenced on all further correspondence and on the
freight documents accompanying your return shipment.
2. Please provide as much information and/or clear instructions within the return paperwork. This will assist our test
engineers with their diagnosis.
3. Please do not ship the goods prior to obtaining the reference number. HyQuest Solutions will not reject any
equipment that arrives without reference number; however, it may take us longer to process.
Custom requirements for items sent to HyQuest Solutions for warranty or non-warranty repairs: Check with your
national customs/tax authorities for details, processes and paperwork regarding tax exempt return of products. Typically,
special custom tariff codes are available (such as HS Code = 9802.00) that verify the item is being returned for repair and
has no commercial value. Please note that the customs invoice / dispatch documents should also clearly state: “Goods
being returned to manufacturer for repair – No Commercial value”. It is mandatory to have any returned goods
accompanied by a commercial invoice on headed paper. HyQuest Solutions reserves the right to charge the customer for
time spent rectifying incorrect customs documents.
Note: Please ensure that your goods are packed carefully and securely. Damage that occurs during transit is not covered
by our warranty and may be chargeable.

© 2022 HyQuest Solutions 19
Technical Data
6Technical Data
Technical Specifications Data Logger
Input Channels
24 channels, high-resolution
Communications
§
Remote: GPRS, 2G or optionally 3G; bi-directional
§
Local: 433 MHz, license-free ISM radio
Power Supply
§
Lithium batteries 3.6 V (up to 10 years of operation or up to 20.000 internet transfers and
200.000 data logs per unit)
§
Optional: External 12 V solar panel power supply
Data Storage
§
Internal memory for up to 250k measurement entries, non-volatile
§
Storage cycle: Freely-definable, minimum 2”
Material and
Environmental
Conditions
§
Enclosure: stainless steel, ABS
§
Operating temperature range:
§
Modem -30 °C to +85 °C
§
Logger -40 °C to +85 °C
Dimensions: Diameter
Ø, Length, Mass
48.5 mm Ø, 411.5 mm length, 1,5 kg (without sensor)
Conformity/Complian
ce
CE
Technical Specifications Sensor
Measuring Range
§
§
Pressure: 0 - 10 mH2O with max. load 3 bar (further options on demand)
Resolution
Pressure 0.01 % FS; Water Temperature 0.05 °C
Material
§
Sensor: Stainless steel 316 L / 1.4435 (titanium (GR. 2) optional)
§
Seals: Viton (EPDM or Kalrez optional)
§
Cable: PUR (standard), PE (drinking water) or FEP (chemical resistant, Teflon-impregnated)
Dimensions: Diameter
Ø, Length, Mass
Typically 24 mm Ø, 157 mm length, 200 g (without cable)
Conformity/Complian
ce
CE

© 2022 HyQuest Solutions 20
Obligations of the Operator and Disposal
7Obligations of the Operator and Disposal
This chapter contains the following subsections:
§
Obligations of the Operator
§
Dismantling / Disposal
7.1 Obligations of the Operator
European Union
In the Single European Market it is the responsibility of the operator to ensure that the following legal regulations are
observed and complied with: national implementation of the framework directive (89/391/EEC) and the associated
individual directives, in particular 2009/104/EC, on minimum safety and health requirements for the use of work
equipment by employees at work.
Worldwide
Regulations: If and where required, operating licences must be obtained by the operator. In addition, national or regional
environmental protection requirements must be complied with, regardless of local legal provisions regarding the
following topics:
§
Occupational safety
§
Product disposal
Connections: Local regulations for electrical installation and connections must be observed.
7.2 Dismantling / Disposal
When disposing of the units and their accessories, the applicable local regulations regarding environment, disposal and
occupational safety must be observed.
Before dismantling
§
Electrical Devices:
§
Switch off the units.
§
Disconnect electrical appliances from the power supply, regardless of whether the appliances are connected to the
mains or to another power source.
§
Mechanical devices:
§
Fix all loose components. Prevent the device from moving independently or unintentionally.
§
Loosen mechanical fastenings: Please note that appliances can be heavy and that loosening the fastenings may cause
them to become mechanically unstable.
Disposal
Operators of old appliances must recycle them separately from unsorted municipal waste. This applies in particular to
electrical waste and old electronic equipment.
Electrical waste and electronic equipment must not be disposed of as household waste!
Instead, these old appliances must be collected separately and disposed of via the local collection and return systems.
Integrated or provided batteries and accumulators must be separated from the appliances and disposed of at the
designated collection point.
20
20
Table of contents
Other HyQuest Solutions Data Logger manuals

HyQuest Solutions
HyQuest Solutions iRIS 270 User manual

HyQuest Solutions
HyQuest Solutions ML-IoT/C User manual

HyQuest Solutions
HyQuest Solutions iRIS Under Cover User manual

HyQuest Solutions
HyQuest Solutions iRIS 150FX User manual

HyQuest Solutions
HyQuest Solutions WDC100 User manual

HyQuest Solutions
HyQuest Solutions iRIS 350FX User manual

HyQuest Solutions
HyQuest Solutions iRIS 350FX User manual