IET Labs HRRS 5kV Series Troubleshooting guide

♦PRECISION INSTRUMENTS FOR TEST AND MEASUREMENT ♦
Email: inf[email protected]
TEL: (516) 334-5959 • FAX: (516) 334-5988
www.ietlabs.com
IET LABS, INC.
HRRS 5kV AND 10kV SERIES
High Resistance, 5kV and 10kV
Decade Substituter
User and Service Manual
Copyright © 2017 IET Labs, Inc.
Visit www.ietlabs.com for manual revision updates
HRRS 5kV and 10kV im/Nov. 2017

i
HRRS 5kV and 10kV Series
WARRANTY
We warrant that this product is free from defects in material and workmanship and, when properly used,
will perform in accordance with applicable IET specifications. If within one year after original shipment,
it is found not to meet this standard, it will be repaired or, at the option of IET, replaced at no charge when
returned to IET. Changes in this product not approved by IET or application of voltages or currents greater
than those allowed by the specifications shall void this warranty. IET shall not be liable for any indirect,
special, or consequential damages, even if notice has been given to the possibility of such damages.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.

ii
HRRS 5kV and 10kV Series
WARNING
OBSERVE ALL SAFETY RULES
WHEN WORKING WITH HIGH VOLTAGES OR LINE VOLTAGES.
Dangerous voltages may be present inside this instrument. Do not open the case
Refer servicing to qualified personnel
HIGH VOLTAGES MAY BE PRESENTAT THE TERMINALS OF THIS INSTRUMENT
WHENEVER HAZARDOUS VOLTAGES (> 45 V) ARE USED, TAKE ALL MEASURES TO
AVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS.
USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARE
CONDUCTORS WHEN USING THIS INSTRUMENT.
Use extreme caution when working with bare conductors or bus bars.
WHEN WORKING WITH HIGH VOLTAGES, POST WARNING SIGNS AND
KEEP UNREQUIRED PERSONNEL SAFELY AWAY.
CAUTION
DO NOT APPLY ANY VOLTAGES OR CURRENTS TO THE TERMINALS OF THIS
INSTRUMENT IN EXCESS OF THE MAXIMUM LIMITS INDICATED ON
THE FRONT PANEL OR THE OPERATING GUIDE LABEL.

iii
HRRS 5kV and 10kV Series
Table of Contents
Chapter 1 Introduction
1.1 Overview............................................................................................................... 1
Chapter 2 Specifications
2.1 Specifications ........................................................................................................ 2
2.2 Ordering Information ............................................................................................ 3
Chapter 3 Operation
3.1 Initial Inspection and Setup .................................................................................. 5
3.2 Dial Setting ........................................................................................................... 5
3.3 Environmental Conditions .................................................................................... 5
3.4 Storage .................................................................................................................. 5
Chapter 4 Maintenance
4.1 Maintainability and Reliability ............................................................................. 6
4.2 Preventive Maintenance........................................................................................ 6
4.3 Verification of Performance .................................................................................. 6
4.3.1 Calibration Interval..................................................................................... 6
4.3.2 General Considerations............................................................................... 6
4.3.3 Calibration Procedure ................................................................................. 7
4.4 Schematic.............................................................................................................. 7
4.5 Replaceable Parts .................................................................................................. 8
Figures and Tables
Figure 1-1: Example of an HRRS-5kV unit with 7 decades ................................... 1
Figure 1-2: Example of an HRRS-10kV unit with 4 decades .................................. 1
Figure 2-1: Typical operating guide .........................................................................4
Figure 4-1: HRRS 5kV and 10kV Series Schematic Diagram ................................ 7
Table 4-1: Replaceable Parts ..................................................................................8
Figure 4-2: Replaceable Parts ................................................................................8

iv
HRRS 5kV and 10kV Series
This page is intentionally left blank.

1
HRRS 5kV and 10kV Series
Introduction
Chapter 1
INTRODUCTION
1.1 Overview
The High Resistance, 5kV and 10 kV Decade
Substituters (Figure 1-1 and Figure 1-2) is a family
of instruments that answers the need of the calibration
and test community for decades and working stan-
dards that go up to 10 T, offering a broad choice of
high-range, excellent performance resistance sources.
High values of resistance are available for use at high
voltages, without sacrificing accuracy, stability, and
temperature and power coefficients of resistance.
The HRRS-5kV/HRRS-10kV units employ state-
of-the-art precision resistors. They are designed
specifically for operation at high voltage, with very
low leakage to minimize degradation of accuracy.
The standard models offer a choice of one to nine
decades. The panels are clearly labeled showing the
step size and maximum allowable applied input for
each decade. Custom requirements outside of the
standard models can be satisfied.
To ensure low leakage, the connection terminals are
insulated with Kel-F high-resistance, non-moisture-
absorbing material.
With a resolution as low as 10 and a maximum
available resistance of 10 T, these High-Resistance,
High-Voltage units may be used for exacting preci-
sion high-value resistance applications that require
voltages of up to 5kV or 10 kV.
Applications include calibration of meters and meg-
ohmmeters, and checking of electrochemical and bio-
medical sensors and instruments. These instruments
are useful wherever high resistances are required,
such as in testing high-impedance amplifiers and the
insulation of low-power circuits.
The HRRS 5kV and 10kV series complement the
HARS series which provides resistance steps as low
as 1 m. The units may be rack-mounted to serve
as components in measurement and control systems.
Figure 1-1: Example of an HRRS-5kV unit
with 7 decades
Figure 1-2: Example of an HRRS-10kV unit
with 4 decades

2
HRRS 5kV and 10kV Series
Specifications
*To apply up to 10 kV at 10 MΩ, set the 1 MΩdecade to its “10” position.
Zero resistance:
<10 mper decade
Environmental conditions:
Operating Conditions: 10°C to 40°C; <50% RH
Storage conditions: -40°C to 70°C
Terminals:
Two five-way binding posts on 2 special, low-
leakage, Kel-F insulating sockets, and one metal
ground binding post electrically connected to the
case
Resistance
per step
Total
decade
resistance
Accuracy *
18-28°C; <50% RH HRRS-5kV HRRS-10kV Temperature
coefficient
(±ppm/°C)
Voltage
coefficient
(±ppm/V)
Stability
(±ppm/
yr) Resistor type
Q B F Max voltage
per step
(V)
Maximum
voltage
(V)
Max voltage
per step
(V)
Maximum
voltage
(V)
10 Ω100 Ω±(0.01%+2 mΩ) ±(0.03%+2 mΩ) ±(0.10%+2 mΩ) 2.5 25 2.5 25 15 - 10
Wirewound,
non-inductive
100 Ω1 kΩ±(0.01%+2 mΩ±(0.03%+2 mΩ) ±(0.10%+2 mΩ) 8 80 8 80 5 - 10
1 kΩ10 kΩ±0.01% ±0.03% ±0.10% 23 230 23 230 5 - 10
10 kΩ100 kΩ±0.01% ±0.03% ±0.10% 70 700 70 700 5 - 10
100 kΩ1 MΩ±0.01% ±0.03% ±0.10% 230 2300 230 2300
5 - 10
1 MΩ10 MΩ±0.01% ±0.03% ±0.10% 1000 5000 1000 10,000 15 <1 25
10 MΩ100 MΩ±0.03% ±0.10% ±1% 5000 5000 5000
(*10,000) 10,000 25 <1 50
High-voltage
film
100 MΩ1 GΩ±0.10% ±0.20% ±1% 5000 5000 10,000 10,000 25 1 100
1 GΩ10 GΩ±0.20% ±0.50% ±1% 5000 5000 10,000 10,000 25 1 500
10 GΩ100 GΩ±0.50% ±1% ±1% 5000 5000 10,000 10,000 25 2 500
100 GΩ1 TΩ±0.50% ±1% ±1% 5000 5000 10,000 10,000 100 5 500
1 TΩ10 TΩ±3% ±5% ±10% 5000 5000 10,000 10,000 300 <20 (10 ppm
typical) 500
Tested at low voltages except 10 MΩat 20 V, 1 GΩat 100 V, ≥10 GΩat 1000 V
Chapter 2
SPECIFICATIONS
For convenience to the user, the pertinent specifications are given in an OPERATING GUIDE affixed to the
case of the instrument. Figure 2-1 shows a typical example.
2.1 Specifications
Mechanical:
Model Dimensions Weight
3-4 Decade
Units 43.2 cm W x 13.3 cm H x 13.5 cm D
(17" x 5.25" x 5.3")
3.4 kg
(7.5 lbs)
5-7 Decade
Units 48.2 cm W x 22.2 cm H x 33 cm D
(19" x 8.75" x 13")
5 kg
(11 lbs)
8-9 Decade
Units 43.7 cm W x 30.7 cm H x 20.3 cm D
(17.2" x 12.1" x 8")
7.7 kg
(17 lbs)

3
HRRS 5kV and 10kV Series
Specifications
Model Total resistance Number of
decades Resolution
HRRS-*-3-10k-** 11.1 MΩ3 10 kΩ
HRRS-*-3-100k-** 111 MΩ3 100 kΩ
HRRS-*-3-1M-** 1.11 GΩ3 1 MΩ
HRRS-*-3-10M-** 11.1 GΩ3 10 MΩ
HRRS-*-3-100M-** 111 GΩ3 100 MΩ
HRRS-*-3-1G-** 1.11 TΩ3 1 GΩ
HRRS-*-3-10G-** 11.1 TΩ3 10 GΩ
HRRS-*-4-1k-** 11.11 MΩ4 1 kΩ
HRRS-*-4-10k-** 111.1 MΩ4 10 kΩ
HRRS-*-4-100k-** 1.111 GΩ4 100 kΩ
HRRS-*-4-1M-** 11.11 GΩ4 1 MΩ
HRRS-*-4-10M-** 111.1 GΩ4 10 MΩ
HRRS-*-4-100M-** 1.111 TΩ4 100 MΩ
HRRS-*-4-1G-** 11.11 TΩ4 1 GΩ
HRRS-*-5-100-** 11.111 MΩ5 100 Ω
HRRS-*-5-1k-** 111.11 MΩ5 1 kΩ
HRRS-*-5-10k-** 1.111 1 GΩ5 10 kΩ
HRRS-*-5-100k-** 11.111 GΩ5 100 kΩ
HRRS-*-5-1M-** 111.11 GΩ5 1 MΩ
HRRS-*-5-10M-** 1.111 1 TΩ5 10 MΩ
HRRS-*-5-100M-** 11.111 TΩ5 100 MΩ
HRRS-*-6-10-** 11.111 1 MΩ6 10 Ω
HRRS-*-6-100-** 111.111 MΩ6 100 Ω
HRRS-*-6-1k-** 1.111 11 GΩ6 1 kΩ
HRRS-*-6-10k-** 11.111 1 GΩ6 10 kΩ
HRRS-*-6-100k-** 111.111 GΩ6 100 kΩ
HRRS-*-6-1M-** 1.111 11 TΩ6 1 MΩ
HRRS-*-6-10M-** 11.111 1 TΩ6 10 MΩ
Model Total resistance Number of
decades Resolution
HRRS-*-7-10-** 111.111 1 MΩ7 10 Ω
HRRS-*-7-100-** 1.111 111 GΩ7 100 Ω
HRRS-*-7-1k-** 11.111 11 GΩ7 1 kΩ
HRRS-*-7-10k-** 111.111 1 GΩ7 10 kΩ
HRRS-*-7-100k-** 1.111 111 TΩ7 100 kΩ
HRRS-*-7-1M-** 11.111 11 TΩ7 1 MΩ
HRRS-*-8-10-** 1,111.111 1 MΩ8 10 Ω
HRRS-*-8-100-** 11.111 111 GΩ8 100 Ω
HRRS-*-8-1k-** 111.111 11 GΩ8 1 kΩ
HRRS-*-8-10k-** 1,111.111 1 GΩ8 10 kΩ
HRRS-*-8-100k-** 11.111 111 TΩ8 100 kΩ
HRRS-*-9-10-** 11,111.111 1 MΩ9 10 Ω
HRRS-*-9-100-** 111.111 111 GΩ9 100 Ω
HRRS-*-9-1k-** 1,111.111 11 GΩ9 1 kΩ
HRRS-*-9-10k-** 11,111.111 1 GΩ9 10 kΩ
*To specify accuracy grade, replace * with “Q,” “B,” or “F” as required.
**To specify voltage, replace ** with either “5kV” or “10kV.”
Options:
-RM: Rack mountable case for standard 19” rack
-K: Kelvin-type 4-terminal posts
-RO: Rear output
2.2 Ordering Information

4
HRRS 5kV and 10kV Series
Specifications
Figure 2-1: Typical operating guide
IET LABS, INC. 534 Main Street, Westbury, NY 11590 • (516) 334-5959• (800) 899-8438 • Fax (516) 334-5988
HRRS-5kV lbl/01-17-2013
Observe all safety rules when working with high voltages or line voltages. Connect the (G) terminal to earth ground in order to maintain the case at a safe voltage. Whenever hazardous
voltages (> 45 V) are used, take all measures to avoid accidental contact with any live components: a) Use maximum insulation and minimize the use of bare conductors. b) Remove
power when adjusting switches. c) Post warning signs and keep personnel safely away.
+556N96(5,(6+,*+5(6,67$1&(N9'(&$'(68%67,787(523(5$7,1**8,'(
CONSULT INSTRUCTION MANUAL FOR PROPER INSTRUMENT OPERATION
WARNING
.HHSXQLWLQDVHDOHGHQYLURQPHQWZKHQ
QRWLQXVH0DLQWDLQELQGLQJSRVWDUHD
FOHDQIRUPLQLPXPHOHFWULFDOOHDNDJH
CAGE CODE: 62015 www.ietlabs.com
MODEL: HRRS-B-9-1K-5kV SN: C1-1250517
5HVLVWDQFH
SHUVWHS
7RWDO
GHFDGH
UHVLVWDQFH
$FFXUDF\
&5+ 0D[YROWDJH
SHUVWHS
9
0D[LPXP
YROWDJH
9
7HPSHUDWXUH
FRHI¿FLHQW
SSP&
9ROWDJH
FRHI¿FLHQW
SSP9
6WDELOLW\
SSP
\U
5HVLVWRU
W\SH
4% )
ȍ ȍ Pȍ Pȍ Pȍ
:LUHZRXQG
QRQLQGXFWLYH
ȍ Nȍ Pȍ Pȍ Pȍ
Nȍ Nȍ
Nȍ Nȍ
Nȍ 0ȍ
0ȍ 0ȍ
0ȍ 0ȍ
+LJKYROWDJH
¿OP
0ȍ *ȍ
*ȍ *ȍ
*ȍ *ȍ
*ȍ 7ȍ
7ȍ 7ȍ SSP
W\SLFDO
7HVWHGDWORZYROWDJHVH[FHSW0ȍDW9*ȍDW9*ȍDW9
=HURUHVLVWDQFH
PȍSHUGHFDGH
(QYLURQPHQWDOFRQGLWLRQV
2SHUDWLQJ&RQGLWLRQV&WR&5+
6WRUDJHFRQGLWLRQV&WR&
7HUPLQDOV
7ZR¿YHZD\ELQGLQJSRVWVRQVSHFLDO
ORZOHDNDJH.HO)LQVXODWLQJVRFNHWVDQG
RQHPHWDOJURXQGELQGLQJSRVWHOHFWULFDOO\
FRQQHFWHGWRWKHFDVH

5
HRRS 5kV and 10kV Series
Operation
Chapter 3
OPERATION
and a setting of 10-10-10-l0-10.10 Ωbecomes:
10 1 0 0 0 0 0.0
10 1 0 0 0 0.0
10 1 0 0 0.0
10 1 0 0.0
10 1 0.0
.10 1.0
TOT 1 1 1 1 1 1.0
3.3 Environmental Conditions
For optimal accuracy, the decade box should be used
in an environment of 23°C. It should be allowed to
stabilize at that temperature after any significant
temperature variation.
Humidity should be maintained at <50% RH. This is
especially important if high resistances are involved.
3.4 Storage
If this instrument is to be stored for any lengthy period
of time, it should be sealed in plastic and stored in a
dry location. It should not be subjected to temperature
extremes beyond the specifications. Extended expo-
sure to such temperatures can result in an irreversible
change in resistance, and require recalibration.
3.1 Initial Inspection and Setup
This instrument was carefully inspected before ship-
ment. It should be in proper electrical and mechanical
order upon receipt.
An OPERATING GUIDE is attached to the case
of the instrument to provide ready reference to
specifications.
3.2 Dial Setting
Whenever the dials are used in positions 0-9, the
resulting resistance is read directly. Both the decimal
point and the steps are clearly marked on the panel.
For additional flexibility and range, each decade
provides a “10” position setting. This “10” position
on any one decade equals the “1” position on the next
higher decade. It adds about 11% to the nominal total
decade resistance.
To determine the resistance obtained when one or
more “10” settings are used, simply add “1” to the
next higher decade. For example, a setting of 3-6-
10-0-10 becomes:
3 3 0 0 0 0
6 6 0 0 0
10 1 0 0 0
0 0 0
10 1 0
TOT 3 7 0 1 0

6
HRRS 5kV and 10kV Series
Maintenance
Chapter 4
MAINTENANCE
4.1 Maintainability and Reliability
It is possible to maintain models HRRS-5kV and
HRRS-10kV indefinitely. They are reliable due to
their closed design and sealed switches and resistors.
The units are resistant to electromagnetic interference
(EMI) because of their metal enclosure.
4.2 Preventive Maintenance
Keep the units in a clean environment. This will help
prevent possible contamination.
The front panel should be periodically cleaned to
eliminate any leakage paths from near or around the
binding posts. To clean the front panel:
Wipe the front panel clean using alcohol and a
lint-free cloth.
4.3 Verification of Performance
4.3.1 Calibration Interval
The HRRS 5kV and 10kV Series instruments should
be verified for performance at a calibration interval of
twelve (12) months. This procedure may be carried
out by the user if a calibration capability is available,
by IET Labs, or by a certified calibration laboratory.
If the user should choose to perform this procedure,
then the considerations below should be observed.
4.3.2 General Considerations
It is important, whenever testing the HRRS 5kV and
10kV Series units, to be aware of the capabilities and
limitations of the test instruments used. A resistance
bridge may be employed, and there are direct-reading
resistance meters or digital multimeters available
that can verify the accuracy of these units, especially
when used in conjunction with standards that can
serve to confirm or improve the accuracy of the test-
ing instrument
Such test instruments must have a 4:1 TUR capabil-
ity for each value calibrated. Note that the accuracy
varies for different decades.A number of commercial
bridges and meters exist that can perform this task;
consult IET Labs.
It is important to allow both the testing instrument
and the HRRS-5kV/10kV Substituter to stabilize for
a number of hours at the nominal operating tempera-
ture of 23°C, and at nominal laboratory conditions of
humidity. There should be no temperature gradients
across the unit under test.

7
HRRS 5kV and 10kV Series
Maintenance
4.3.3 Calibration Procedure
To calibrate the HRRS-5kV/10kV unit, proceed as
follows
1. Confirm that the leakage resistance between
either binding post and the case ground is
>106times the highest unit resistance, but
may be as low at 1 T.
If this is not obtained, clean the area
around the binding posts as described in
Section 4.2.
2. Determine the allowable upper and lower
limits for each resistance setting of each
decade based on the specified accuracy (See
Specifications on page 2).
3. Confirm that the resistances fall within these
limits.
If any resistances fall outside thee limits,
the associated switch assembly may
require service or replacement.
WARNING
Connect the GND terminal to earth
or other suitable ground in order to
maintain the case at a safe voltage.
Whenever hazardous voltages
(>45 v) are used, take all measures to
avoid accidental contact with any live
components.
Use maximum insulation and
minimize the use of bare conductors.
REMOVE POWER WHEN SETTING
SWITCHES.
Post warning signs and keep
personnel safely away.
4.4 Schematic
Figure 4-1 shows a schematic of an HRRS-5kV/10kV
unit.
Figure 4-1: HRRS 5kV and 10kV Series
Schematic Diagram

8
HRRS 5kV and 10kV Series
Maintenance
4.5 Replaceable Parts
Model Ref IET part number Description
1 BP-1000-RD Binding Post, Red
2 BP-1000-BK Binding Post, Black
3 BP-1000-BL Binding Post, Blue
4 BP-1000-GN Binding Post, Green
5 HARS-X-4300-KNB Knob Assembly
Not Shown HARS-X-3100 Foot
Not Shown HRRS-*-SW-10-** 10 /step Decade Switch Assembly
Not Shown HRRS-*-SW-100-** 100 /step Decade Switch Assembly
Not Shown HRRS-*-SW-1k-** 1 k/step Decade Switch Assembly
Not Shown HRRS-*-SW-10k-** 10 k/step Decade Switch Assembly
Not Shown HRRS-*-SW-100k-** 100 k/step Decade Switch Assembly
Not Shown HRRS-*-SW-1M-** 1 M/step Decade Switch Assembly
Not Shown HRRS-*-SW-10M-** 10 M/step Decade Switch Assembly
Not Shown HRRS-*-SW-100M-** 100 M/step Decade Switch Assembly
Not Shown HRRS-*-SW-1G-** 1 G/step Decade Switch Assembly
Not Shown HRRS-*-SW-10G-** 10 G/step Decade Switch Assembly
Not Shown HRRS-*-SW-100G-** 100 G/step Decade Switch Assembly
Not Shown HRRS-*-SW-1T-** 1 T/step Decade Switch Assembly
*To specify accuracy grade, replace * with “Q,” “B,” or “F” as required.
**To specify voltage, replace ** with either “5kV” or “10kV.”
Table 4-1: Replaceable Parts