Interacoustics Affinity Compact User manual

Instructions for Use - EN
Affinity
Compact
Science made smarter
D-0123234-D – 2020/05


Table of Contents
1INTRODUCTION .......................................................................................................................... 1
1.1 About this manual ............................................................................................................... 1
1.2 Intended use........................................................................................................................ 1
1.3 Product description ............................................................................................................. 1
1.4 Included and optional parts and accessories...................................................................... 2
1.5 Warnings and precautions .................................................................................................. 3
2UNPACKING AND INSTALLATION............................................................................................ 4
2.1 Unpacking and inspection ................................................................................................... 4
2.2 Symbols............................................................................................................................... 5
2.3 Important safety instructions ............................................................................................... 6
2.3.1 Electrical system safety ............................................................................................ 6
2.3.2 Electrical safety......................................................................................................... 6
2.3.3 Explosion hazards..................................................................................................... 7
2.3.4 Electromagnetic compatibility (EMC)........................................................................ 7
2.3.5 Cautions – General ................................................................................................... 7
2.3.6 Environmental factors ............................................................................................... 8
2.3.7 NOTICE .................................................................................................................... 8
2.4 Connection panel dictionary.............................................................................................. 10
2.5 Affinity Compact Indicators ............................................................................................... 11
2.6 Software installation .......................................................................................................... 11
2.6.1 Software installation Windows®7 ........................................................................... 12
2.6.2 Software installation Windows®8.1 ........................................................................ 15
2.6.3 Software installation Windows®10 ......................................................................... 18
2.7 Driver installation............................................................................................................... 21
2.8 Using with databases ........................................................................................................ 21
2.8.1 Noah 4 .................................................................................................................... 21
2.9 Standalone version ........................................................................................................... 21
2.10 How to configure an alternative data recovery location .................................................... 21
2.11 License .............................................................................................................................. 22
2.12 About Affinity Suite............................................................................................................ 22
3OPERATING INSTRUCTIONS .................................................................................................. 23
3.1 Using the tone screen ....................................................................................................... 24
3.2 Using the speech screen................................................................................................... 30
3.2.1 Speech audiometry in graph mode......................................................................... 32
3.2.2 Speech audiometry in table mode .......................................................................... 33
3.2.3 PC keyboard shortcuts manager ............................................................................ 35
3.2.4 Technical specifications of the AC440 Software .................................................... 36
3.3 The REM440 screen ......................................................................................................... 38
3.3.1 REM software - Technical specifications ................................................................ 45
3.4 The HIT440 screen ........................................................................................................... 46
3.4.1 HIT440 Software - Technical Specifications ........................................................... 50
3.5 Using the print wizard........................................................................................................ 52
4MAINTENANCE ......................................................................................................................... 55
4.1 General maintenance procedures..................................................................................... 55
4.2 How to clean Interacoustics products ............................................................................... 55
4.3 Concerning repair.............................................................................................................. 56
4.4 Warranty............................................................................................................................ 56

5GENERAL TECHNICAL SPECIFICATIONS ............................................................................. 57
5.1 Affinity Compact Hardware - Technical Specifications...................................................... 57
5.2 Tone reference equivalent threshold values for transducers............................................ 59
5.3 Pin assignments................................................................................................................ 72
5.4 Electromagnetic compatibility (EMC) ................................................................................ 74

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 1
1Introduction
1.1 About this manual
This manual is valid for the Affinity Compact. This product is manufactured by:
Interacoustics A/S
Audiometer Allé 1
5500 Middelfart
Denmark
Tel.: +45 6371 3555
Fax: +45 6371 3522
E-mail: info@interacoustics.com
Web: www.interacoustics.com
1.2 Intended use
The Affinity Compact with AC440 is intended to be used for the detection and diagnosis of suspected hearing
loss.
The Affinity Compact with HIT440 is intended to be used as a means to get an objective indication of the
characteristics of a hearing aid and as a help for making the adjustments of the hearing aid to the patient. It
is used by manufacturers of hearing aids and in clinics for hearing aid fitting.
The Affinity Compact with REM440 is intended to be used for Real-Ear Measurement that takes care of all
clinical verification needs.
Intended operator
Trained operators like audiologist, hearing healthcare professional, or trained technician.
1.3 Product description
The Affinity Compact is a Hearing Aid Analyzer that interfaces with integrated audiologic software modules
on a PC. Depending on the installed software modules it can perform:
•Audiometry (AC440)
•Real Ear Measurements (REM440) including Visible Speech Mapping
•Hearing Instrument Testing (HIT440)

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 2
1.4 Included and optional parts and accessories
AC440
REM440/VSP440
HIT440
Standard parts:
•Affinity suite software
•DD45 Audiometric headset1
•Monitor Headset
•Talk backmicrophone
•B81 Bone conductor1
•APS3 Patient response
button1
•Standard USB cable
•Power supply -
UE60-240250SPA3
•Power cable
•Mouse pad
Optional parts:
•Earphone 3Ainsert
earphones 1/2
•IP30 insert earphones1
•B71 Bone Conductor1
•IP30 insert phone – single
sited
•Audiocup enclosures
•DD65 v21
•HDA300 Audiometric
headset1
•DD450 High frequency
headset1
•SP85A Loudspeaker
•SP90A Loudspeaker
•SP100 Loudspeaker
•EM400 Electret microphone
•Ambient noise microphone
•Accessory bracket
•Desktop bracket
•Table mount bracket
•Wall mount bracket
•Cable Extender Box
•Sound room kit
•OtoAccess® database
Standard parts:
•Affinity suite software
•IHM65 In-situ headset1/2
•Probe tubes, 50 pcs.
•SPL60 Transducer kit for
RECD measurement
including probes and ear
tips
•SP100 loudspeaker
•Standard USB cable
•Power supply -
UE60-240250SPA3
•Power cable
•Mouse pad
Optional parts:
•Accessory bracket
•Desktop bracket
•Table mount bracket
•Wall mount bracket
•Battery adapters BAA675,
BAA13, BAA312, BAA10,
BAA5
•Coupler and microphone kit
o½” microphone
oReference
microphone
o2cc coupler
o0.4cc coupler
oBTE Short
oBTE Long
oITE
oRubber tubes
•Coupler seal wax
•Ear simulator
•Monitor headset
•SP85A Loudspeaker
•SP90A Loudspeaker
•OtoAccess® database
Standardparts:
•Affinity suite software
•Coupler and microphone kit
o½” microphone
oReference
microphone
o2cc coupler
o0.4cc coupler
oBTE Short
oBTE Long
oITE
oRubber tubes
•Coupler seal wax
•Standard USB cable
•Power supply - UE60-
240250SPA3
•Power cable
•Mouse pad
Optional parts:
•Battery adapters BAA675,
BAA13, BAA312, BAA10,
BAA5
•Adaptor for Body Style HA
•Ear simulator
•SKS10 Skull Simulator with
power supply
•Tele coil
•Monitor headset
•OtoAccess® database
1Applied part according to IEC 60601-1
2This part is not certified according to IEC 60601-1

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 3

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 4
1.5 Warnings and precautions
Throughout this manual, the following definitions of warning, caution and notice are used:
The WARNING label identifies conditions or practices that may present danger
to the patient and/or user.
The CAUTION label identifies conditions or practices that could result in
damage to the equipment.
NOTICE NOTICE is used to address practices not related to personal injury.
Federal law restricts this device to sale by or on the order of a licensed medical practitioner
WARNING
CAUTION

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 5
2Unpacking and installation
2.1 Unpacking and inspection
Check box and contents for damage
When the instrument is received please check the shipping box for rough handling and damage. If the box is
damaged it should be kept until the contents of the shipment have been checked mechanically and
electrically. If the instrument is faulty please contact your local distributor. Keep the shipping material for the
carrier’s inspection and insurance claim.
Keep carton for future shipment
The Affinity Compact comes in its own shipping carton, which is specially designed for the Affinity Compact.
Please keep this carton. It will be needed if the instrument has to be returned for service. If service is
required please contact your local distributor.
Reporting Imperfections
Inspect before connection
Prior to connecting the product it should once more be inspected for damage. All of the cabinet and the
accessories should be checked visually for imperfections and missing parts.
Immediately report any faults
Any missing part or malfunction should be reported immediately to the supplier of the instrument together with
the invoice, serial number, and a detailed report of the problem. In the back of this manual you will find a
"Return Report" where you can describe the problem.
Please use "Return Report"
Please realise that if the service engineer does not know what problem to look for he may not find it, so using
the Return Report will be of great help to us and is your best guarantee that the correction of the problem will
be to your satisfaction.
Storage
If you need to store the Affinity Compact for a period, please ensure it is stored under the following
conditions:
Temperature; 0-50°C
Relative Humidity: 10-95% Non condensing

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 6
2.2 Symbols
The following symbols can be found on the instrument, accessories or packaging:
Symbol
Explanation
Type B applied parts
Follow instructions for use
WEEE (EU-directive)
This symbol indicates that when the end-user wishes to discard this product, it
must be sent to separate collection facilities for recycling.
The CE-mark indicates that Interacoustics A/S meets the requirements of Annex II
of the Medical Device Directive 93/42/EEC. TÜV Product Service, Identification No.
0123, has approved the quality system.
Year of manufacture
Manufacturer
Serial number
Reference number
Indicates a component is intended for one use, or for use on a single patient during
a single procedure. Cross contamination risk.
Stand by
Keep dry
Transport and storage temperature range
Transport and storage humidity limitations
ETL listing mark
Logo
REF

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 7
2.3 Important safety instructions
Read these instructions carefully and completely before using the product
2.3.1 Electrical system safety
When connecting the instrument to the computer, the following warnings must
be observed:
This equipment is intended to be connected to other equipment thus forming a
Medical Electrical System. External equipment intended for connection to
signal input, signal output or other connectors shall comply with the relevant
product standard e.g. IEC 60950-1 for IT equipment and the IEC 60601-series
for medical electrical equipment. In addition, all such combinations – Medical
Electrical Systems – shall comply with the safety requirements stated the
general standard IEC 60601-1, edition 3, clause 16. Any equipment not
complying with the leakage current requirements in IEC 60601-1 shall be kept
outside the patient environment i.e. at least 1.5 m from the patient support or
shall be supplied via a separation transformer to reduce the leakage currents.
Any person who connects external equipment to signal input, signal output or
other connectors has formed a Medical Electrical System and is therefore
responsible for the system to comply with the requirements. If in doubt, contact
qualified medical technician or your local representative. If the instrument is
connected to a PC (IT equipment forming a system) ensure not to touch the
patient while operating the PC.
A Separation Device (isolation device) is needed to isolate the equipment
located outside the patient environment from the equipment located inside the
patient environment. In particular such a Separation Device is required when a
network connection is made. The requirement for the Separation Device is
defined in IEC 60601-1 clause 16
2.3.2 Electrical safety
Do not modify this equipment without authorization of Interacoustics Do not
disassemble or modify the product as this may impact on the safety and/or
performance of the device. Refer servicing to qualified personnel.
For maximum electrical safety, turn off the power when it is left unused
The power plug shall be placed so it is easy to pull out the plug
Do not use any additional multiple socket-outlet or extension cord.
Do not use the equipment if it is showing visible signs of damage.
The instrument is not protected against ingress of water or other liquids. If any
spillage occurs, check the instrument carefully before use or return for service
No part of the equipment can be serviced or maintained while in use with the
patient.
To avoid the risk of electric shock, this equipment must only be connected to a
supply mains with protective earth.
WARNING
WARNING

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 8
2.3.3 Explosion hazards
Do NOT use in the presence of flammable gaseous mixtures. Users should
consider the possibility of explosions or fire when using this device in close
proximity to flammable anesthetic gases.
Do NOT use the instrument in a highly oxygen-enriched environment, such as
a hyperbaric chamber, oxygen tent, etc.
Before cleaning make sure to disconnect power source
2.3.4 Electromagnetic compatibility (EMC)
Although the instrument fulfills the relevant EMC requirements, precautions
should be taken to avoid unnecessary exposure to electromagnetic fields, e.g.,
from mobile phones, etc. If the device is used adjacent to other equipment it
must be observed that no mutual disturbance appears. Please also refer to the
appendix regarding EMC.
Use of accessories, transducers and cables other than specified, with the
exception of transducers and cables sold by Interacoustics or representatives,
may result in increased emission or decreased immunity of the equipment. For
a list of accessories, transducers and cables that fulfil the requirements please
also refer to the appendix regarding EMC.
2.3.5 Cautions – General
If the system is not functioning properly, do not operate it until all necessary
repairs are made and the unit is tested and calibrated for proper functioning in
accordance with Interacoustics’ specifications.
Do not drop or in any other way cause undue impact to this device. If the
instrument is damaged, return it to the manufacturer for repair and/or
calibration. Do not use the instrument if any damage is suspected.
This product and its components will perform reliably only when operated and
maintained in accordance with the instructions contained in this manual,
accompanying labels, and/or inserts. A defective product should not be used.
Make sure all connections to external accessories are secured properly. Parts
which may be broken or missing or are visibly worn, distorted, or contaminated
should be replaced immediately with clean, genuine replacement parts
manufactured by or available from Interacoustics.
Interacoustics will make available on request circuit diagrams, component part
lists, descriptions, calibration instructions, or other information that will assist
authorized service personnel to repair those parts of this instrument that are
designated by Interacoustics as repairable by service personnel.
No parts of the equipment can be serviced or maintained while in use with the
patient.
WARNING
CAUTION
CAUTION

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 9
Connect only accessories purchased from Interacoustics to the instrument.
Only accessories which have been stated by Interacoustics to be compatible
are allowed to be connected to the device.
Never insert, or in any way use, the insert headset without a new clean and
non-defective ear-tip. Always make sure that the foam or ear-tip is mounted
correctly. Ear-tips and foam are for single use only.
The instrument is not intended for use in environments exposed to fluid spills.
Check calibration if any parts of the equipment are exposed to shock or rough
handling.
Components marked for ‘single use’ are intended for a single patient during a
single procedure, and there is a risk of contamination if the component is re-
used. Components marked for ‘single use’ are not intended to be reprocessed.
Use only transducers calibrated with the actual instrument.
In case of a serious incident with serious health impact for the patient or user
Interacoustics shall be informed. Beside that the competent authority in
patient’s home country shall be informed. Interacoustics has a vigilance
system to help with this.
2.3.6 Environmental factors
Storage outside temperature range as specified in Section 2.1 may cause
permanent damage to the instrument and its accessories.
Do not use the device in the presence of fluid that can come into contact with
any of the electronic components or wiring. Should the user suspect fluids have
contacted the system components or accessories, the unit should not be used
until deemed safe by an authorized service technician.
Do not place the instrument next to a heat source of any kind, and allow
sufficient space around the instrument to ensure proper ventilation.
2.3.7 NOTICE
To prevent system faults take appropriate precautions to avoid PC viruses and
similar.
Using operating systems where Microsoft have discontinued software and
security support will increase the risk for viruses and malware, which may result
in breakdowns, data loss and data theft and misuse.
Interacoustics A/S cannot be held liable for your data. Some Interacoustics A/S
products support or may work with operating systems unsupported by Microsoft.
Interacoustics A/S recommends you to always use Microsoft supported operating
systems that are kept fully security updated.
CAUTION

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 10
Within the European Union, it is illegal to dispose of electric and electronic items
in unsorted municipal waste. Electric and electronic waste may contain
hazardous substances and therefore has to be collected separately. Such
products will be marked with the crossed-out wheeled bin symbol, shown below.
The cooperation of the user is important in order to ensure a high level of reuse
and recycling of electric and electronic waste. Failing to recycle such waste
products in an appropriate way may endanger the environment and consequently
the health of human beings.
Outside the European Union, local regulations should be followed when disposing
of the product after end of life.

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 11
2.4 Connection panel dictionary
Position:
Symbol:
Function:
1
Headset 2 Left and Ins.
Masker
Socket for Insert phone or HF Phone or Insert Masker
2
Headset 2 Right
Socket for Insert phone or HF Phone
3
Bone
Socket for Bone headset
4
Pat. Resp.
Socket for Patient Response Button
5
Monitor
Socket for Monitor Headset
6
Mic. –Talk F.
Socket for Talk forward Microphone
7
Ambient- Cal. Mic.
Socket for Ambient Noise Microphone or Automatic FF Verification
Microphone
8
AUX
Socket for line in from External Sound Source
9
Talk B.
Socket for Talk back Microphone
10
AC Headset Left
Socket for Left AC Phone or HF Phone
11
AC Headset Right
Socket for Right AC Phone or HF Phone
12
FF1 Power
Socket for power out to FF loudspeaker
13
FF2 Power
Socket for power out to FF loudspeaker
14
FF 1-2 Line
Socket for line output to FF loudspeaker
15
Insitu Headset
Socket for Insitu REM headset
16
UE60-240250SPA3
Socket for external power supply
17
USB-PC
Socket for USB connection to PC
Position:
Symbol:
Function:
1
Reference
Socket for Reference Microphone
2
Coupler
Socket for Coupler Microphone
3
Battery
Socket for Battery Simulator power out
4
Telecoil
Socket for Telecoil out

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 12
2.5 Affinity Compact Indicators
The Affinity Compact hardware has an LED light indicator which changes status during different operations
of the Affinity Compact Suite and hardware. These different colours and their statuses are listed and shown
below.
The LED light is visible from both the front and the top of the Affinity Compact.
GREEN-Light: Ready
RED-Light: Indicates right ear selected in REM and HIT module
BLUE-Light: Indicates left ear selected in REM and HIT module
PURPLE-Light: Indicates both ears selected in REM and HIT module
LIGHT BLUE-Light: Indicates that the Affinity Compact is not correctly connected to the Affinity
Compact Suite
A dimmed light indicates that the Affinity Compact has entered power saving mode. This can happen in any of
the colours mentioned above.
2.6 Software installation
To know before you Start the Installation
You must have administrative rights to the computer on which you are installing the Affinity Compact Suite.
NOTICE
1. DO NOT connect the Affinity Compact hardware to the computer before the software has been
installed.
2. Interacoustics will not make any guarantee to the functionality of the system if any other software is
installed, with exception of the Interacoustics measurement (AC440/REM440) modules and
AuditBase system4, Otoaccess®, or Noah4 compatible Office Systems or later releases.
What you will need:
1. Affinity Compact Suite Installation USB Drive
2. USB Cable
3. Affinity Compact Hardware
Supported Noah Office Systems
We are compatible with all Noah-integrated office systems which are running on Noah and Noah engine.
To use the software in conjunction with a database, make sure the database is installed prior to the Affinity
Compact Suite installation. Follow the manufacturer’s installation instructions provided to install the relevant
database.
Note that if you are using AuditBase you must ensure to launch this office system prior to this installation of
the Affinity Compact Suite.
Installation on various Windows® versions
Windows® 7 (32 and 64 bit), Windows®8 and Windows®10 systems are supported.
Software Installation on Windows® 7
Insert the installation USB and follow the steps below to install the Affinity Compact Suite software. To find
the installation file; click “Start”, then go to “My Computer” and double click the USB drive to view the
contents of the installation USB. Double click the “AffinitySuiteSetup.exe” file to initiate the installation.

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 13
2.6.1 Software installation Windows®7
Insert the installation USB and follow the steps below to install the Affinity Compact Suite software. If the
installation procedure does not start automatically, click “Start”, then go to “My Computer” and double click
the USB drive to view the contents of the installation USB. Double click the “AffinitySuiteSetup.exe” file to
initiate the installation.
1. Wait for the dialog shown below to appear, please specify which hardware you wish to install the
software for. You must also accept the license terms and conditions ahead of installing. On checking
the box to accept this, the Install button will become available, click “Install” to begin the installation.

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 14
Should you wish to install the software to a different location than the default then please click on
’Options’ ahead of ’Install’
2. User Account Control may ask if you want to allow the program to make changes to your computer.
Click Yes if this happens.

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 15
3. The installer will now copy all the necessary files to the PC. This process may take several minutes.
4. When the installation is complete, the dialog box below is shown.
5. Click “Close” to finish the installation. The Affinity Compact/Equinox Suite is now installed.

D-0123234-D – 2020/05
Affinity Compact - Instructions for Use - EN Page 16
2.6.2 Software installation Windows®8.1
Insert the installation USB and follow the steps below to install the Affinity /Equinox Suite software. If the
installation procedure does not start automatically, click “Start”, then go to “My Computer” and double click
the USB/CD-RW drive to view the contents of the installation USB. Double click the “Affinity
CompactSuiteSetup.exe” file to initiate the installation.
1. Wait for the dialog shown below to appear, please specify which hardware you wish to install the
software for. You must also accept the license terms and conditions ahead of installing. On checking
the box to accept this, the Install button will become available, click “Install” to begin the installation.
Other manuals for Affinity Compact
2
This manual suits for next models
5
Table of contents
Other Interacoustics Test Equipment manuals