24/7 Technical Support (877) 484-9119
Xpress™ Quick Start Guide
Adjusting the Xpress Mounting Arm
2. Set
platform
location
Platform
1. Secure
Arm
Clamp
3. Lock Release
lever (Pull
upwards until it
snaps into
locking position)
WARNING:
10” (25.4 cm) maximum
Horizontal extension
10” (25.4 cm)
1. Secure the arm clamp to an appropriate support.
2. Hold the platform (or Xpress handle) and unlock the
Lock Release lever to set the location of the
mounting platform. The mounting platform should
be parallel to the ground for best results.
CAUTION: When the Lock Release lever is unlocked, all
three pivot points are freed to move. Support the Xpress
with its handle when repositioning so that it does not fall.
3. Lock the release lever.
WARNING: The Xpress mounting arm is adjusted at the
factory to securely hold the Xpress steady under a vertical
acceleration of 3 gravities with a moment arm of 10 inches
(25.4 cm). To reduce the risk of injury, do not extend the
mounting arm platform further than 10 inches (25.4 cm)
from the base mounting point.
WARNING: To prevent the possibility of injury, do not
secure the Xpress Mounting Arm to any support structure
within 36 inches (91.4 cm) of a patient’s head or critical
areas.
1. Insert the front edge of the Xpress mounting plate
into the platform.
2. Lower the rear edge of the Xpress mounting plate
on to the latch button. This will release the locking
lever.
3. Press the locking lever tight. Locking lever
Quick Release
enabled
Quick Release
disabled
Latch button
Installing the Xpress on Mounting Platform Insert front edge
of Xpress
NOTE: For best results, mount arm on at, parallel surfaces.