
The iBlot™ 2 Gel Transfer Device complies with the TUV Rhineland North America
Inc. safety requirements, part 15 of the FCC rules, and the European Community
Safety requirements. Operation of the iBlot™ 2 Gel Transfer Device is subject to the
conditions described in this manual.
Operation of the iBlot™ 2 Gel Transfer Device is subject to the following conditions:
• Indoor use
• Altitude below 2,000 meters
• Temperature range: 5 to 40°C
• Maximum relative humidity: 80% (maximum relative humidity 80% for
temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40°C)
• Installation categories (over voltage categories) II; Pollution degree 2
• Mains supply voltage uctuations not to exceed 10% of the nominal voltage (100–
240 V, 50/60 Hz, 6.3 A)
• Mains plug is a disconnect device and must be easily accessible.
• Do not aempt to open the iBlot™ 2 Gel Transfer Device. To honor the
warranty,iBlot™ 2 Gel Transfer Device can only be opened and serviced by
Thermo Fisher Scientic.
• The protection provided by the equipment may be impaired if the equipment is
used in a manner not specied by Thermo Fisher Scientic.
• The device must be connected to a mains socket outlet with protective earthing
connections.
• Ventilation requirements: Room ventilation
• Do not use if device becomes cracked or broken in any area.
The product may be installed only under the conditions and in the positions specied
by Thermo Fisher Scientic.
Following are the required operating position and conditions:
• Do not place the product in an area where it will be subject to vibration.
• Do not place the product on surfaces, vehicles, cabinets or tables that for reasons
of weight or stability are unsuitable for this purpose.
• Do not place the product on heat-generating surface or near heat emiing devices
such equipment racks or heaters. Verify that there is sucient clearance between
the product and any other system that may exhaust warm air.
• The product's ventilation should not be obstructed. If proper ventilation is not
provided it can result in electric shock, re and/or serious personal injury or
death.
• The product is for indoor use only.
• Use only with suitably rated mains supply cord (having 3 conductors, min. 16
AWG or 1.5 mm2, min. 300V, Harmonized Type for Europe and UL Listed/CSA
Certied for North America, with molded plug rated min. 10A).
• A tolerance of ±10% shall apply to the nominal input voltage and ±3% to the
nominal frequency, over voltage category 2.
• Maximum operating altitude 2,000 m asl. Maximum transport altitude 4,500 m
asl.
Installing the
instrument
Chapter 1 Product information
Product contents
1
iBlot
™
2 Dry Blotting System User Guide
7