Cleaning Guide
Small Battery Drive II
1 Remove debris.
Rinse the device under running cold tap water for a
minimum of 2 minutes. Use a sponge, soft lint-free cloth,
or soft-bristled brush to assist in removing gross soil.
For cannulations of the handpiece and attachments, the
cleaning brush (519.40) should be used.
Note: Do not use pointed objects for cleaning.
2 Manipulate moving parts.
Manipulate all moving parts such as the triggers, sleeves,
and switches under running tap water to loosen and
remove gross debris.
Manual Cleaning Instructions
3 Spray and wipe.
Spray and wipe the device using a neutral pH enzymatic
solution for a minimum of 2minutes.
Follow the enzymatic detergent manufacturer’s
directions for correct temperature, water quality
(i.e., pH, hardness) and concentration/dilution.
4 Clean with detergent.
Clean the device manually under running warm water
using an enzymatic cleaner or detergent for a minimum
of 5minutes. Manipulate all moving parts under running
water. Use a soft-bristled brush and/or soft lint-free cloth
to remove all visible soil and debris.
Follow the enzymatic cleaner or detergent manufacturer’s
instructions for use for correct temperature, water
quality, and concentration/dilution.
Handpieces, battery casings, sterile covers and
attachments must be processed using
a) manual cleaning and / or
b) automated cleaning with manual pre-cleaning.
(78 mm) (61 mm)
(265 mm)
(B3,6 mm)
Cleaning Brush (519.400)