Kane 458s User manual

Stock No: MAN00099 Rev: 0.00000 JULY 2020
© Kane International Ltd
KANE 458s
Flue Gas Analyser with direct CO₂ measurement
and CO sensor protection

2KANE 458s MANUAL
CONTENTS
Page No.
KANE 458s OVERVIEW 4-5
INSTRUMENT FEATURES & KEYPAD 6-9
FUNCTION KEYS 7
ANALYSER LAYOUT 8
BACK OF ANALYSER - PROBE ETC 9
BATTERIES 10
BATTERY TYPE 10
REPLACING BATTERIES 10
TIME AND DATE 10
CHARGING NIMH BATTERIES 10
BATTERY DISPOSAL 10
GENERAL SAFETY 11
USING YOUR KANE458s FOR THE FIRST TIME 12-13
USING YOUR KANE458S EVERY TIME 12
FRESH AIR OURGE ON SWITCH ON 12-13
USING THE MENU 14-16
MEASURING FLUE GASSES 16
CO PROTECTION PUMP OPERATION 17
USING YOUR ANALYSER 18-19
COMBUSTION TESTS 18
RATIO DISPLAY 18
02/EFF DISPLAY 19
AUX DISPLAY 19
TRANSFERING RESULTS 19-21
TO VIEW / PRINT A LOGGED RESULT 20
VIEWING / PRINTING A LOGGED COMBUSTION TEST 20
KANE INFRARED PRINTER 20
PRINTOUTS 20
COMMISSIONING TEST 22
PRESSURE & TEMPERATURE TESTING 23-24
PRESSURE MEASUREMENT GOOD PRACTICE 24

3KANE 458s MANUAL
LARGE BORE TUBING ISSUES 24
LET-BY & TIGHTNESS TESTING 25-26
ROOM CO TESTING 27-28
TEST TYPES 27
ROOM CO DISPLAY 28
USING YOUR KANE458s AS A THERMOMETER OR PRESSURE
METER 27-28
ANALYSER PROBLEM SOLVING 29
SPECIFICATIONS 30-31
EU DECLARATION OF CONFORMITY 32
SERVICE – CALIBRATE - RECERTIFY 33-37
KANE ASSET MANAGER (KAM) 34-35
GUARANTEED SAME DAY DESPATCH 36
WHERE TO SEND YOUR ANALYSER 37
COLD WEATHER PRECAUTIONS 37

4KANE 458s MANUAL
KANE 458s OVERVIEW
Your KANE458s combustion analyser measures:
● Carbon Monoxide (CO)
● Carbon Dioxide (CO₂)
● Oxygen (O2) if tted
● Nitric Oxide (NO) if tted
● Dierential Pressure
● Dierential temperature
Your KANE458s measures CO2 using KANE’s unique NDIR sensor.
Always use fresh outdoor air to “zero” your analyser during the initial
countdown or when prompted to “ZERO CO” as indoor CO2 levels are
increased by human activity.
This applies even if your analyser measures O2 instead of CO2 as it is
best practice to use fresh, outdoor air during initial countdown to protect
you from unsafe levels of CO and CO2 possibly present in the room.
Depending on your options these parameters are calculated:
● Oxygen (O2)
● CO/CO₂ratio
● Carbon Monoxide (CO₂)
● Combustion Eciency
● Losses
● Excess Air
● Dierential Temperature
Your analyser also measures CO levels in ambient air – useful when a
CO alarm is triggered – & performs a Room CO test for up to 30 minutes.
Your KANE458s has a structured Commissioning Test to help install boilers.
Your KANE458s has a protective rubber cover with magnets for
“hands-free” operation and is supplied with a ue probe with integral
temperature sensor.
Your analyser has a low ow detector to switch o the analyser’s pump
if it detects an over lled water trap.

5KANE 458s MANUAL
Your KANE458s has a large 6 line display showing data & test results
based on your actions.
Your KANE458s can send test results to our optional KANE IRP-2
infrared printer or KANE’s wireless printer App.
The memory can store up to:
● 60 combustion tests
● 20 AUX tests
● 20 let-by/tightness tests
● 20 temperature & pressure tests
● 20 room CO tests
● 20 commissioning Tests
You can personalise 2 lines of 16 characters on your test results header
- for example your company name & telephone number.

6KANE 458s MANUAL
ANALYSER FEATURES AND KEYPAD

7KANE 458s MANUAL
KANE458s BUTTONS
ICON DESCRIPTION ICONS
Save log – long press to store data
Print report – short press to enter a print or wireless
data transfer
Navigate up – short press to scroll up
Enter key – used to select current option
Navigate down – short press to scroll down
Data hold – short press to hold current data on screen
Pump toggle – long press to switch pump on & o
Function keys

8KANE 458s MANUAL
ANALYSER LAYOUT
Battery
charger
Watertrap drain with
red plug
Flue probe
temperature T1
Inlet temperature T2
Pressure
connection P1
Pressure
connection P2 –
differential
Flue gas
inlet
On / Off
Function keys
– see below
Water trap &
particle filter
Rotary switch
Red connector
– plugs into Flue
Gas Inlet
Flue Probe Temperature Plug –
into T1
Narrow (+ve) pin must be on
right side Flue Probe Hose

9KANE 458s MANUAL
Watertrap drain
with red plug
Thermocouple
Stainless Steel
Shaft
Depth Stop Cone
Gas Path
Magnets x2
Thermocouple
wire

10 KANE 458s MANUAL
BATTERIES
BATTERY TYPE
Your KANE458s uses rechargeable Nickel Metal Hydride (NiMH)
batteries - using other battery types may void your KANE458s’ warranty.
WARNING
Although you can use Alkaline batteries you must not charge your
analyser with Alkaline batteries tted.
Do not mix NiMH batteries of dierent capacities or from dierent
manufacturers – all must be identical.
REPLACING BATTERIES
Turn over your analyser and remove its protective rubber cover. Open
the battery compartment & t 3 NiMH “AA” rechargeable batteries –
check battery polarity is correct. Replace battery cover & protective
rubber cover.
TIME AND DATE
After changing the batteries reset your analyser’s time & date.
CHARGING NIMH BATTERIES
Your KANE458s uses a standard Micro USB connector for charging
(KANE part No: USB1).
For best results turn o your analyser then connect your charger – the
charging indicator illuminates and turns o when charging nishes.
Your rst charge should be for 8 hours continuously – thereafter NiMH
batteries can be topped up at any time, even for short periods.
If the analysers batteries discharge & the analyser enters a ‘low power
shut down’ a 1 hour charge will provide approx 2 hours continuous use.
BATTERY DISPOSAL
Always dispose of depleted batteries using approved disposal methods
that protect the environment.

11KANE 458s MANUAL
GENERAL SAFETY
WARNING
This analyser extracts combustion gases that may be toxic in relativity
low concentrations. These gases are exhausted from the bottom of the
analyser. This analyser must only be used in well-ventilated locations by
trained and competent persons after due consideration of all the
potential hazards.
Portable gas detector users should conduct “bump” test before relying
on the unit to verify an atmosphere is free from hazard.
A “bump” test is a way to test an instrument works within acceptable
limits by briey exposing to a known gas mixture that change the output
of all the sensors present.
Note: This is dierent from a calibration where your analyser is also
exposed to a known gas mixture but is allowed to settle to a steady
gure and the reading adjusted to the stated gas concentration of the
test gas.
Protection Against Electric Shock (In accordance with EN 61010-1: 2010):
This analyser is designed as Class III equipment and should only be
connected to SELV circuits. The battery charger is designated as:
● Class II equipment
● Installation category II
● Pollution degree 2
● Indoor use only
● Altitude to 2000m
● Ambient temperature 0°C-40°C
● Maximum relative humidity 80% for temperatures up to 31°C
decreasing linearly to 50%RH at 40°C
● Mains supply uctuations not to exceed 10% of the nominal voltage
WARNING
Your analyser’s protective cover is tted with strong magnets – magnetic
elds can cause damage to magnetic storage media. Certain electronic
devices are sensitive to magnetic elds and may be damaged
permanently if exposed to a strong magnetic eld.

12 KANE 458s MANUAL
USING YOUR KANE458s FOR THE FIRST TIME
Charge your analyser’s batteries for 8 hours – an overnight charge
should be sucient for an average 8-hour day.
Take time to read this manual fully & be aware your analyser’s
conguration may not support all features detailed in this manual.
Take time to set up your requirements before using your analyser.
USING YOUR KANE458s EVERY TIME:
Check your analyser’s water trap is empty & particle lter is not dirty.
To empty, pull out the red plug, shake out water & replace red plug.
To change the particle lter, remove protective rubber cover, slide the
water trap from analyser, remove particle lter from its spigot &
replace.
Reconnect your analyser’s water trap & rubber protective cover.
Connect your ue probe hose to your analyser’s ue gas inlet – see p.8.
Connect your ue probes temperature plug to your analyser’s T1 socket
– check plug orientation is correct – See p.8.
FRESH AIR PURGE ON SWITCH ON
Your analyser starts a zero countdown to auto calibrate itself.
When nished, rotate dial to “RATIO” – In fresh air CO reading should be
zero.
Rotate dial to “O2/EFF” – In fresh air O2 readings should be 20.9% +
0.3%.

13KANE 458s MANUAL
WARNING
This message says your analyser must be reset in fresh air.
Place your analyser in outdoor fresh air, press .
You can perform a manual “ZERO” any time – rotate dial to “RATIO” &
hold down the button to see the message above.

14 KANE 458s MANUAL
USING THE MENU
Rotate dial to “Menu” and navigate using the function buttons:
= SCROLL UP = SCROLL DOWN = ENTER
NOTE: To exit MENU rotate dial to any position – any changes not
entered will not store.
As you scroll up or down, the side LEDs illuminate to point to the active line
MAIN
MENU SUB MENU OPTIONS/COMMENTS
SETUP
LANGUAGE ENGLISH
SET TIME HH:MM:SS format
e.g. 7 am = 07:00:00, 7pm = 19:00:00
SET DATE DD/MM/YY format
REPORTS
PRINTER
KM IRP
KANE IRP-2
WIRELESS
SERIAL
BACK

15KANE 458s MANUAL
MAIN MENU SUB MENU OPTIONS/COMMENTS
PRESSURE
FILTER O F F = n o r m a l r e s p o n s e .
ON = slower (damped) response
RESOLUTION LOW = e.g. 0.01mbar resolution.
HIGH = displays to an extra decimal place
UNITS mbar, Pa, PSI, mmHg, hPa, inH2O,
mmH20, kPa, psi
TIME LET BY = Set duration of let-by test in
minutes. Default = 1 minute
PRINTER STABIL’N = Set duration of stabilisation in
minutes. Default = 1 minute
PASS KEY TIGHTN’S = Set duration of tightness test
in minutes. Default = 2 minute
BACK
MAIN MENU SUB MENU OPTIONS/COMMENTS
UNITS
FUEL TYPE
NAT GAS, TOWN GAS, COKE GAS,
PROPANE, BUTANE, LPG, LIGHT OIL,
BIO OIL, WOOD PELLETS, BIO GAS,
USER FUEL 1 TO 5
FUEL ORIGIN UK, FRANCE, SPAIN, N AMERICA,
BELGIUM, NETHERLAND
EFFICIENCY GROSS, NET, GROSS COND, NET COND
PRESSURE SEE NEXT TABLE BELOW
GAS PPM, PPM(N), MG/M3, MG/M3(N), MG/
KWH, MG/KWH(N)
TEMP C , F
02 REF UP/DOWN TO SET VALUE
(3% DEFAULT)
NOX CALC UP/DOWN TO SET VALUE
(5% DEFAULT)
BACK

16 KANE 458s MANUAL
Make sure you do not exceed the analyser’s operating specications. In
particular:
• Do not exceed the ue probe’s maximum temperature (600°C)
• Do not exceed the analyser’s internal temperature operating range
MEASURING FLUE GASSES
After the countdown is nished and the analyser is correctly set up, put
its ue probe into the appliance’s sampling point. The tip of the probe
should be at the centre of the ue. Use the ue probe’s depth stop cone
to set the position.
With balanced ues, make sure the probe is positioned far enough into
the ue so no air can “back ush” into the probe.
WARNING
Ensure the ue probe handle does not get hot!
MAIN MENU SUB MENU OPTIONS/COMMENTS
SCREEN
CONTRAST Factory setting is 14
BACKLIGHT
0 to 300 secs
AUX
Enables users to customise the
parameters on the AUX display: LINE 1,
LINE 2, LINE 3, LINE 4, LINE 5, LINE 6,
BACK
BACK

17KANE 458s MANUAL
• Do not put the analyser on a hot surface
• Do not exceed the water trap’s levels
• Do not let the analyser’s particle lter become dirty and blocked
Look at your analyser’s displayed data to ensure stable operating
conditions are achieved & readings are within expected range.
CO PROTECTION PUMP OPERATION
Your analyser’s CO sensor is automatically protected from high levels of
CO. When levels of CO rise above 2000ppm the analyser’s pump stops
and its CO purge pump starts.
Your analyser displays P-OFF until CO level falls below 2000ppm.
Fuel type can be changed via “Menu”
or STATUS setting
CO/CO2 ratio
Pump Off, purge
Pump Off, purge
Excess air %
Pressure reading

18 KANE 458s MANUAL
USING YOUR ANALYSER
COMBUSTION TESTS
Insert your ue probe tip into the centre of the ue. Readings will settle
within 60 seconds assuming boiler conditions are stable.
Rotate dial to “Ratio” to display:
RATIO DISPLAY
Fuel type – can be changed via “Menu”
or STATUS setting
CO/CO2 ratio
Carbon monoxide (ppm)
Carbon dioxide (%)
Excess air %
Pressure reading
Press button to send a full combustion test to our optional KANE
IRP-2 printer or KANE’s App via Wireless module.
Press & hold button for 2 seconds to log a full combustion report.
Page 21 explains how to view & print stored reports.

19KANE 458s MANUAL
O2/EFF DISPLAY
AUX DISPLAY
Fuel type
AUX (auxiliary) display can be
customised via MENU / SCREEN / AUX
Oxygen (%) left after combustion.
Should be 20.9% ± 0.3% in fresh air
Parameters displayed on lines 1, 2, 3, 4,
5 & 6 can be set by the user
Gross efficiency
AUX parameters until changed by user
Flue temperature
Inlet temperature (oC) – normally set by
flue probe during fresh air purge
NO sensor not fitted
Differential temperature
NO sensor not fitted
TRANSFERING RESULTS
To print, simply press and release the button to send results to our
optional KANE IRP-2 printer or KANE’s App via wireless module – you
can stop printing by pressing again.
Your analyser’s side lights highlight the displays the active line.
Use or to change the pointer.
Press to select a line. The side lights now ash.
Use or to scroll or change the selected line.
Press to exit a line.

20 KANE 458s MANUAL
TO VIEW / PRINT A LOGGED REPORT
Select MENU / REPORT / COMBUSTION / VIEW.
Your analyser’s side lights highlight the top line of its display.
Press to select this line – side lights will ash.
Use or to scroll or change the Log No – if only 1 report is logged,
number will not change.
Press to conrm a Log No. – the side lights will stop ashing.
To view logged data press or to highlight another line.
Press – sidelights will ash on that line.
Use or to scroll through data.
To nish, press . Sidelights stop ashing.
Use or to scroll down to “PRINT”.
Press to print.
VIEWING / PRINTING A LOGGED COMBUSTION TEST
Press to send tests to our optional KANE IRP-2 printer or KANE’s
App via wireless module.
KANE INFRARED PRINTER
Switch on your printer and place its infrared receiver in line with the
emitter on top of your analyser – allow a 15cm gap between the printer
and analyser.
Table of contents
Other Kane Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Klein Tools
Klein Tools ET200 instruction manual

GESTRA
GESTRA NRG 16-51 Original Installation & Operating Manual

Veris Industries
Veris Industries E51H2A installation guide

PSI
PSI PlantPen PRI 200 instruction manual

TPI
TPI The Value Leader 505 instruction manual

Electronic Technologies
Electronic Technologies ES2010E user manual