Kane 905 User manual

KANE905
Hand-held Combustion
Analyser with Wireless
Communication Option
Stock No: 19711 July 2015
© Kane International Limited

KANE905 Operators Manual Page 2
CONTENTS
Page No:
1. ANALYSER LAYOUT AND FEATURES ................................. 4-8
KEY FEATURES ..................................................................................... 4
OPTIONS ................................................................................................ 4
INSTRUMENT FEATURES AND KEYPAD ............................................. 5
INSTRUMENT LAYOUT (REAR) ............................................................ 6
STANDARD PROBE CONFIGURATION ................................................ 7
ANALYSER CONNECTIONS ................................................................. 8
2. SAFETY WARNING ................................................................... 9
3. FIRST TIME USE ..................................................................... 10
4. NORMAL START UP SEQUENCE ..................................... 11-17
EVERY TIME YOU USE THE ANALYSER ........................................... 11
AUTOMATIC CALIBRATION ................................................................ 11
MAIN DISPLAYS ................................................................................. 12
4 PAGE MODE ..................................................................................... 13
LINE SCROLL MODE ........................................................................... 14
8 PAGE MODE ..................................................................................... 14
SAMPLING THE FLUE GAS ................................................................. 15
TAKING A PRESSURE READING ....................................................... 16
TAKING A FLOW READING ................................................................. 16
REGULAR CHECKS DURING SAMPLING .......................................... 16
NORMAL SHUTDOWN SEQUENCE .................................................... 17
5. MOVING THROUGH THE MENUS .................................... 18-19
THE MENU STRUCTURE .................................................................... 18
BASIC OPERATION ............................................................................. 18
6. MENU OPTIONS AND SETTINGS ..................................... 20-26
MAIN MENU ......................................................................................... 20
SUB MENU – 1. MENU ........................................................................ 20
SUB MENU – 2. UNITS ........................................................................ 22
SUB MENU – 3. DISPLAY .................................................................... 23
SUB MENU – 4. SELECT SET-UP ....................................................... 23

KANE905 Operators Manual Page 3
7. PRINTING INFORMATION ................................................. 27-28
OPTIONAL INFRA-RED THERMAL PRINTER ..................................... 27
OPTIONAL WIRELESS MODULE ........................................................ 27
PRINTING A ‘LIVE’ TEST ..................................................................... 27
STANDARD PRINTOUT ....................................................................... 28
SOFTWARE COMPATIBILITY.............................................................. 28
8. STORING AND RETRIEVING DATA.................................. 29-30
STORING A ‘LIVE’ TEST ...................................................................... 29
VIEWING AND PRINTING A ‘STORED’ TEST ..................................... 29
DELETING DATA ................................................................................. 30
9. AVERAGE OF THREE (ITALIAN VERSION ONLY) .......... 31-32
STORING ............................................................................................. 31
VIEWING .............................................................................................. 31
PRINTING ............................................................................................. 32
10. MAINTENANCE ....................................................................... 33
EMPTYING AND CLEANING THE IN-LINE WATER TRAP .................. 33
CHANGING THE PARTICLE FILTER ................................................... 33
11. PROBLEM SOLVING ............................................................... 34
12. ANNUAL SERVICE & RE-CERTIFY ................................... 35-37
RETURNING YOUR ANALYSER TO KANE ......................................... 36
PACKING YOUR ANALYSER .............................................................. 36
SENDING YOUR ANALYSER .............................................................. 36
WHEN WE RECEIVE YOUR ANALYSER ............................................ 36
SERVICE RETURNS ............................................................................ 37
13. PRODUCT SPECIFICATION ................................................... 39
APPENDICES
A. MAIN DISPLAY PARAMETERS .............................................. 40
B. COMBUSTION EFFICIENCY CALCULATION ........................ 42
C. CALCULATION OF FUEL DATA ............................................. 45
D. ELECTROMAGNETIC COMPATABILITY STATEMENT ........ 46
END OF LIFE DISPOSAL ............................................................. 47
PRODUCT REGISTRATION ..................................................... 49-50

KANE905 Operators Manual Page 4
1. ANALYSER LAYOUT AND FEATURES
KEY FEATURES
Measures Temperature, Pressure, O2 and CO as standard.
Stores 150 sets of test results.
Output to IR Printer (optional).
OPTIONS
(CO & any two other sensors)
High Range CO sensor
Low Range NO sensor
High Range NO sensor
NO2 sensor
SO2 sensor
Wireless upgrade (2.1)
Teflon Hose

KANE905 Operators Manual Page 5
INSTRUMENT FEATURES AND KEYPAD
ON/OFF
UP
Scrolls up through options ie Fuel
MENU
Allows access to all menu functions
DOWN
Scrolls down through options
PUMP
Turns pump on and off
STORE
Enters data storage menu
ENTER
Accepts a command ie enters a menu
option
PRINT
Prints current data

KANE905 Operators Manual Page 6
INSTRUMENT LAYOUT (REAR)
NOTE: Do NOT cover exhaust port as this
will severely affect analyser operation

KANE905 Operators Manual Page 7
STANDARD PROBE CONFIGURATION

KANE905 Operators Manual Page 8
ANALYSER CONNECTIONS

KANE905 Operators Manual Page 9
2. SAFETY WARNING
This analyser extracts combustion gases that may be toxic in relatively
low concentrations. These gases are exhausted from the back of the
instrument. This analyser must only be used in well-ventilated
locations by trained and competent persons after due consideration
of all the potential hazards.
Users of portable gas detectors are recommended to conduct a “bump”
check before relying on the unit to verify an atmosphere is free from
hazard.
A “bump” test is a means of verifying that an instrument is working within
acceptable limits by briefly exposing to a known gas mixture formulated
to change the output of all the sensors present. (This is different from a
calibration where the instrument is also exposed to a known gas mixture
but is allowed to settle to a steady figure and the reading adjusted to the
stated gas concentration of the test gas).
Protection Against Electric Shock (in accordance with
EN 61010-1 : 2010)
This instrument is designated as Class III equipment and should only be
connected to SELV circuits.
The battery charger is designated as:
Class II equipment
Installation category II
Pollution degree 2
Indoor use only
Altitude to 2000m
Ambient temperature 0C-40C
Maximum relative humidity 80% for temperatures up to 31C decreasing
linearly to 50%RH at 40C
Mains supply fluctuations not to exceed 10% of the nominal voltage.

KANE905 Operators Manual Page 10
3. FIRST TIME USE
Charge the battery for 12 hours. Following this, an overnight charge
should be sufficient for an average 8 hour day. See Main Parameter
displays for Battery Indicator.
The KANE905 has a rechargeable NiMh battery which uses a different
charger than other Kane analysers.
Ensure the correct charger is
used or damage may occur to the instrument.
Check that you have all the items you have ordered.
Take time to read this manual fully.
Be aware that the analyser
configuration that you have purchased may not support all the
features detailed in this manual.
When using the analyser for the first time you will need to choose from:-
Language selection
Calibration countdown time
CO gas alarm
NOx percentage for calculation
Time and Date
Printed header name and telephone number
The SET UP MENU gives details of how to change the above settings.

KANE905 Operators Manual Page 11
4. NORMAL START UP SEQUENCE
EVERY TIME YOU USE THE ANALYSER
BEFORE SWITCH-ON CHECK THAT:
the particle filter is not dirty
the water trap and probe line are empty of water
all hose connections, etc, are properly made
the probe is sampling CLEAN AMBIENT air
the water trap is correctly fitted and the instrument upright
the flue temperature is connected
Switch ON the instrument by pressing
AUTOMATIC CALIBRATION
During this sequence the analyser pumps fresh air into the sensors to allow toxic
sensors (if fitted) to be set to zero and the Oxygen sensor to be set to 20.9 %.
After switch-on the analyser will briefly display header information :-
Kane International
KANE905
SW19604 Version: 1.02
And then show the countdown
screen :-
ZERO CAL
Time: 180
FRESH AIR PURGE
The calibration time will count down in seconds to zero. Calibration time
may be changed to 90, 120, 180, 300 seconds. See SET-UP MENU.
Note! Three minutes is recommended to allow the sensors to stabilise
fully. Anything less than this may result in drift of the toxic and
oxygen sensors in clean ambient air.

KANE905 Operators Manual Page 12
To obtain the quoted specification an instrument should be calibrated
with clean ambient air at standard temperature and pressure (STP).
Once the time has reached zero
an audible beep will be heard and
will show the selected fuel on the
following display:-
NATURAL GAS
* PRESS – MENU – KEY *
Press
This zeros the toxic sensor and
sets Oxygen to 20.9%. The next
screen is the MAIN DISPLAY of
the analyser:
NETT
C
0.0
O2
%
20.9
CO
ppm
0000
EFF (G)
%
0.0
Use and to change the display.
CO2
%
0.0
FLUE
C
0.0
INLT
NOT FITTED
AMBIENT
C
21.5
All parameters are detailed in APPENDIX A - MAIN DISPLAY
PARAMETERS.
MAIN DISPLAYS
The main display can be changed to show either 4 or 8 parameters at
one time.
Two options are available when 4 parameters are selected.
4 Page Mode displays 4 lines of data in set format, each page is
predefined.
Line scroll mode allows you to customise the display to show the
data you require.
8 Page Mode displays 8 parameters on 4 lines in set format, the
bottom two can be changed.
Changing between the different modes is detailed in DISPLAY MENU .

KANE905 Operators Manual Page 13
4 PAGE MODE
Use the and keys to change the information that is
displayed on the screen. The following pages are available:
1.
NATURAL GAS
DATE
23-05-15
TIME
12:31:35
BATTERY
%
54
2.
NETT
C
0.0
O2
%
20.9
CO
ppm
0000
EFF (G)
%
0.0
3.
CO2
%
0.0
FLUE
C
0.0
INLT
NOT FITTED
AMBIENT
C
21.5
4.
CO/CO2
R
0.0001
P INDEX
%
0.01
XAIR
%
0.0
Prs
mbar
0.00
Screens 5 and 6 will vary dependent on sensors fitted:
5.
COn
LOSSES
NO2
NO2n
6.
NO
NOx
NOn
NOxn
TIP - In 4 page mode only turns the backlight ON and OFF.

KANE905 Operators Manual Page 14
LINE SCROLL MODE
Line scroll mode allows you to customise the display.
Use the and keys to change the bottom line of the display.
Once the correct line is displayed press to confirm and move the
line up. Select the next parameter and repeat until all lines display the
desired parameters.
Change bottom line using
and
NETT
C
0.0
O2
%
20.9
CO
ppm
0000
CO2
%
0.0
to select and move
parameter up
O2
%
…
20.9
CO
ppm
…
0000
CO2
%
…
0.0
CO2
%
…
0.0
Select next parameter.
Repeat above until
display reads desired data
O2:
%
…
20.9
CO:
ppm
…
0000
CO2:
%
…
0.0
CO/CO2
R
…
0.0001
8 PAGE MODE
Displays 8 parameters on the screen at one time. Symbols used in this
mode are different from those used in 4 page and line scroll modes and
are detailed in APPENDIX A - MAIN DISPLAY PARAMETERS.
O2 :
20.9%
CO2:
20.9
CO :
0ppm
Eff:
0000
PI :
T:
0.0
:
Tf:
0.0001
The bottom line of the display can be changed to display other
parameters.
Use the and keys to change this line.

KANE905 Operators Manual Page 15
SAMPLING THE FLUE GAS
Once the automatic calibration procedure has been completed and the
specific fuel has been selected (See SELECT menu) the probe can be
inserted into the desired sampling point.
It is recommended that the sampling point be located at least two flue
diameters downstream of any bend and that the probe tip is in the centre
of the flue. With balanced flues and other domestic units the probe
should be positioned far enough into the flue so that no air can ‘back
flush’ into the probe. This will be indicated by a low oxygen reading
and/or a low ‘Poison Index’ reading.
The probe depth stop cone provided with the instrument allows the probe
to be used in holes whose diameters range from 8 mm to 21 mm (5/16 to
13/16 inch).
The standard probe is rated at 650C (1202oF). Temperatures of up to
1200C (2200F) can be accommodated using an optional high
temperature probe.
TIP: To conserve battery power, switch off the pump when you are not
taking a measurement. Use the key to turn ON and OFF the
pump.

KANE905 Operators Manual Page 16
TAKING A PRESSURE READING
A flue draught measurement can be made at any time.
Connect the standard probe to the pressure sensor inlet and the probe in
the flue.
The pressure reading will be
displayed :
CO/CO2
R
0.0000
P INDEX
%
0.00
XAIR
%
0.0
Prs
mbar
2.00
To perform a combustion test and display draught pressure at the same
time a special probe is required. Contact Kane International or
Authorised Distributor for details.
TAKING A FLOW READING
In the UNITS menu set the pressure units to metres/sec (m/sec). These
are the only units available for flow measurement.
This also set the pressure display to Pascal (Pa). There is a range limit
for the Pitot calculation of15Pa to 4600Pa (0.15mbar to 46mbar).
For most accurate flow measurement the flue gas temperature should
also be measured. If a flue temperature probe is not fitted then the
internal ambient measurement is used. ‘Flue’ temperature must be
between -10oC to +650oC.
REGULAR CHECKS DURING SAMPLING
Care must be taken at all times not to exceed the analysers operating
specifications, in particular ensure the following :-
Do not exceed the maximum temperature of the flue probe.
The analyser internal temperature does not exceed normal operating
range, typically 0-50C.
DO NOT PLACE THE INSTRUMENT ON A HOT SURFACE.
The water trap is vertical at all times. Water condenses in the probe
line and can quickly fill the water trap when the probe is moved. Take
care and watch the water trap closely.

KANE905 Operators Manual Page 17
The in-line particle filter is clean and does not become blocked.
NORMAL SHUTDOWN SEQUENCE
DO THIS EVERY TIME YOU USE THE ANALYSER
Remove the probe from the flue - TAKE CARE! THE PROBE WILL BE
HOT - and allow it to cool naturally. Do not immerse the probe in water
as this will be drawn into the analyser and damage the pump and
sensors.
Once the probe is removed from the flue press and the analyser
will count down from 30 to switch off.
OFF 30
MENU TO ESCAPE
If you have not finished but press by mistake, you can press
to return to normal operation and not switch OFF.

KANE905 Operators Manual Page 18
5. MOVING THROUGH THE MENUS
THE MENU STRUCTURE
MENU:
SELECT
FUEL
O2 REF
SMOKE
RESET
PITOT
UNITS
TEMP
GAS
PRS
EFF
DISPLAY
LIGHT
MODE
CONTRAST
SETUP
LANG
CO MENU
CALENDAR
ZERO
NOX %
HEADER
PRINT
BASIC OPERATION
From the MAIN DISPLAY
NETT
C
0.0
O2
%
20.9
CO
ppm
0000
EFF(G)
%
0.0
Press
to access the MAIN MENU
MAIN MENU
1. SELECT
3. DISPLAY
2. UNITS
4. SETUP
Press and to move
cursor up and down
MAIN MENU
1. SELECT
3. DISPLAY
2. UNITS
4. SETUP

KANE905 Operators Manual Page 19
Press to access selected
Menu
MAIN MENU
1. SELECT
3. DISPLAY
2. UNITS
4. SETUP
Press to select parameter
FUEL
: LIGHT OIL
O2 Ref
: OFF
SMOKE
: OFF
RESET
: NO
Use and to change
setting i.e. fuel selected
FUEL
: NATURAL GAS
O2 Ref
: OFF
SMOKE
: OFF
RESET
: NO
Press to enter value and
move to next parameter
FUEL
: LIGHT OIL
O2 Ref
: OFF
SMOKE
: OFF
RESET
: NO
Press to save settings and
return to the MAIN MENU
MAIN MENU
1. SELECT
3. DISPLAY
2. UNITS
4. SETUP
Press to return to the MAIN DISPLAY.

KANE905 Operators Manual Page 20
6. MENU OPTIONS AND SETTINGS
MAIN MENU
The MAIN MENU consists of 4
sub menus:
MAIN MENU
1. SELECT
3. DISPLAY
2. UNITS
4. SETUP
All sub-menus are accessed using and exited using .
The and keys move the cursor within a menu and allow
parameters to be changed.
TIP: Holding down one of these keys scrolls through the data quicker.
SUB MENU - 1. SELECT
Page 1:
FUEL
: NATURAL GAS
O2 Ref
: OFF
SMOKE
: OFF
RESET
: NO
Page 2:
PITOT
: 1.00
This menu allows selections to be made for the parameters detailed
below.
FUEL : Select the fuel being used by the boiler from either a standard
fuel stored in the analyser or by entering the user fuel. Once
the correct fuel has been selected press to view the
fuel constants.
NATURAL GAS
K1g
: 0.350
K1n
: 0.390
K_2
: 11.89
K_3
: 9.83
K_4
: 32
O2r
: 3.0
This manual suits for next models
1
Table of contents
Other Kane Measuring Instrument manuals