
MODEL 149 MILLI-MICROVOLl'METI OPERATION
FIGURE 4
4. Thermal EMF - Extreme precautions have been taken in the input
circuit to minimize thermal EMF's so the residual RMF is usually less
than 0.5 uv. The material used in the input circuit is pure copper.
Any other metal will generate a thermocouple potential. Lead solder is
particularly troublesome. Where thermal W's are a problem, soldering
should be done with the cadmium-tin solder supplied with the instrument.
5. Input Noise - The noise at the input is a function of input
resistance and is approximately given by
E = 1.29 x 10-loum
where E is the RMS noise, and R is the source resistaixe. It is assumed
that the bandwidth of the instrument is about 1 cps and the temperature
is 80° F. If noise is observed, calculate the theoretical noise and
compare results. Also bear in mind that only wire-wound resistors
approach the ideal resistor. However, if Evanohm or Manganin resistors
are used, a considerable thermal EMF of the resistor material against
copper will be observed.
6. Checking the Zero Point - At low levels, spurious W's may be
generated
simply by contact between the input leads and the terminals
under test. If possible, always leave the instrument connected and
adjust the zero after establishing a zero reference in the apparatus
under test. For example, in bridge measurements, disconnect the bridge
exciting voltage; or with a phototube, shield the tube from light.
7. Overloads - The current applied to the input circuit should be less
than one milliampere dc steady state, 10 milliamperes dc short-term.
When the FUNCTION switch is on the MILLIVOLTS position, the off-scale
impedance can be as low as 1000 ohms. On the MICROVOLTS position, it may
approach one ohm.
D. MEASURINGVOLTAGE
1. Direct Voltage Measurements - Place the FUNCTION switch at MILLIVOLTS
or MICROVOLTS as necessary for the measurement to be taken. Then turn
the RANGEswitch to more sensitive ranges until the meter gives a usable
deflection,
2. Measuring Voltage Variations - To observe small variations in a large
steady signal, first set the FUNCTION and RANGE switches as described in
D 1. Then operate the ZERO SUPPRESSswitch and potentiometer to reduce
the meter deflection to zero. Increase the meter sensitivity with the range
0667R 2-3