SomnoSuite® Troubleshooting
• Increase the flow rate in Anest Setup >
Nose Cone Min Vol or Induction Min Vol
• Verify that the syringe size setting is
correctly entered.
• Check that the Anesthetic Delivery Tube
is tightly connected to the Syringe.
• Try using a dierent syringe to see if
the problem improves.
Problem Possible Cause Solutions
Anesthetic depth too
light, or the animal
wakes up
Anesthetic depth too
deep
Minute volume displays
0mL/min.
Pusher block isn’t
moving and Anest%
shows “--%”.
Anesthetic agent
remains in syringe
although SomnoSuite
shows 0mL VolLeft.
• Flow rate is too low
• Incorrect syringe size
• Anesthetic Delivery Tube
or Syringe is leaking
• SomnoSuite was
incorrectly primed
• Flow rate is too low
• Incorrect syringe size
• Incorrect Air Source setting
• Blockage in tubing connection
• Compressed gas source not
turned on or turned on at
incorrect pressure
• Airflow not turned on
• Anesthesia is disabled
• Syringe Empty point
does not match syringe
dimensions
• Using incorrect syringe
• Take care not to overprime the SomnoSuite.
Liquid isoflurane should never be injected
into the vaporizer block during priming.
• Increase the flow rate by pressing and then
turning Dial A.
• Verify that the syringe size setting is
correctly entered.
• Check that Air Source is set correctly, either
to “Room air” or “Compressed Gas”.
• Check all connections and ports for
blockages.
• If using compressed gas, be sure the gas
source is turned on and flowing properly.
• Pressure should be between 5 and 10PSI.
• Check that airflow is turned on, indicated
by the word “Running” on the bottom right
corner of the screen.
• Be sure that the Deliver setting is selected
from the Anest Run Screen.
• Ensure that you are using the glass syringe
included with your SomnoSuite. No other
syringe should be used.
• Ensure that the Syringe Size setting is
correct.
• Re-calibrate the Syringe Empty Position