1Technical data ............................................................................................ 4
2Appliance overview.................................................................................. 12
2.1 Overview of display ........................................................................................................... 15
2.2 Keyboard overview ............................................................................................................ 16
2.2.1 Numerical input via the navigation buttons .......................................................................... 17
3Basic Information (General).................................................................... 18
3.1 Proper use........................................................................................................................... 18
3.2 Improper Use...................................................................................................................... 18
3.3 Warranty.............................................................................................................................. 18
3.4 Monitoring of Test Equipment.......................................................................................... 19
4Basic Safety Precautions ........................................................................ 19
4.1 Pay attention to the instructions in the Operation Manual............................................ 19
4.2 Personnel training.............................................................................................................. 19
5Transport and storage............................................................................. 19
5.1 Testing upon acceptance.................................................................................................. 19
5.2 Packaging / return transport............................................................................................. 19
6Unpacking, Setup and Commissioning.................................................. 20
6.1 Installation Site, Location of Use ..................................................................................... 20
6.2 Unpacking........................................................................................................................... 20
6.2.1 Scope of delivery .................................................................................................................. 21
6.3 Mains connection............................................................................................................... 21
6.4 Rechargeable battery operation....................................................................................... 21
6.5 Connection of peripheral devices .................................................................................... 21
6.6 Initial Commissioning........................................................................................................ 21
6.7 Adjustment.......................................................................................................................... 21
6.8 Adjustment.......................................................................................................................... 22
6.9 Linearisation....................................................................................................................... 24

EOC-BA-e-1711 3
7Operation .................................................................................................. 25
7.1 Start-up................................................................................................................................ 25
7.2 Switching Off...................................................................................................................... 25
7.3 Zeroing................................................................................................................................ 25
7.4 Decimal point...................................................................................................................... 26
7.5 Simple weighing................................................................................................................. 27
7.6 Switch-over weighing unit................................................................................................. 27
7.7 Weighing with tare............................................................................................................. 28
7.8 Weighing with tolerance range......................................................................................... 29
7.8.1 Tolerance check for target weight ........................................................................................ 30
7.8.2 Tolerance check for target quantity ...................................................................................... 32
7.9 Manual totalizing................................................................................................................ 34
7.10 Automatic adding-up......................................................................................................... 36
7.11 Parts counting.................................................................................................................... 37
7.12 Percentage weighing ......................................................................................................... 38
7.13 Animal weighing................................................................................................................. 39
7.14 Lock keyboard.................................................................................................................... 39
7.15 Display background illumination ..................................................................................... 40
7.16 Automatic switch-off function „AUTO OFF“................................................................... 41
8The menu .................................................................................................. 42
8.1 Navigation in the menu...................................................................................................... 42
8.2 Menu overview ................................................................................................................... 43
9Servicing, maintenance, disposal........................................................... 46
9.1 Cleaning .............................................................................................................................. 46
9.2 Servicing, maintenance..................................................................................................... 46
9.3 Disposal .............................................................................................................................. 46
9.4 Error messages.................................................................................................................. 47
10 Data output RS232C................................................................................. 48
10.1 Technical data .................................................................................................................... 48
10.2 Printer mode....................................................................................................................... 48
10.3 Remote control instructions............................................................................................. 49
10.4 KERN Communications Protocol (KERN Interface Protocol)........................................ 50
11 Instant help............................................................................................... 51
12 Declaration of conformity........................................................................ 52

4 EOC-BA-e-1711
1 Technical data
KERN EOC 6K-3 EOC 6K-4A EOC 10K-3
Readability (d) 1 g / 2 g 0.5 g 2 g / 5 g
Weighing range (max) 3 kg / 6 kg 6 kg 6 kg / 12 kg
Minimum piece weight 0.25 g 0.25 g 0.5 g
Reproducibility 1 g / 2 g 0.5 g 2 g / 5 g
Linearity ± 3 g / 6 g 1.5 g ± 6 g / 15 g
Warm-up time 10 minutes 30 minutes 10 minutes
Reference unit weights at
piece count 10, 20, 50, 100, 200
Weighing Units Details „Weighing units“ chapter 7.6
Recommended adjustment
weight, not added (class)
Details for „Selection of
the Adjustment weight“
in chapter 7.6
6 kg
(M1)
6 kg
(F2)
12 kg
(M1)
Stabilization time (typical) 3 sec.
Electric Supply 100 V - 240 V, 50 / 60 Hz
Auto Off off, 3 min., 5 min., 15 min., 30 min.
Operating temperature - 10° C .... + 40° C
Humidity of air max. 80 % (not condensing)
Terminal (B x D x H) mm 268 x 115 x 80
Platform (B x D x H )mm 300 x 300 x 100 300 x 300 x 100 300 x 300 x 100
Weight kg (net) 5.2 5.2 5.2

EOC-BA-e-1711 5
KERN EOC 10K-3A EOC 10K-4 EOC 20K-3A
Readability (d) 1 g 0,2 g / 0,5 g 2 g
Weighing range (max) 12 kg 6 kg / 15 kg 24 kg
Minimum piece weight 0.5 g 0.5 g 1 g
Reproducibility 1 g 0.2 g / 0.5 g 2 g
Linearity ± 3 g ± 0.6 g / 1.5 g ± 6 g
Warm-up time 30 minutes 2 hours 30 minutes
Reference unit weights at
piece count 10, 20, 50, 100, 200
Weighing Units Details „Weighing units“ chapter 7.6
Recommended adjustment
weight, not added (class)
Details for „Selection of
the Adjustment weight“
in chapter 7.6
12 kg
(F2)
15 kg
(F2)
24 kg
(F2)
Stabilization time (typical) 3 sec.
Electric Supply 100 V - 240 V, 50 / 60 Hz
Auto Off off, 3 min., 5 min., 15 min., 30 min.
Operating temperature - 10° C .... + 40° C
Humidity of air max. 80 % (not condensing)
Terminal (B x D x H) mm 268 x 115 x 80
Platform (B x D x H )mm 300 x 300 x 100 300 x 300 x 100 300 x 300 x 100
Weight kg (net) 5.2 5.2 5.2

6 EOC-BA-e-1711
KERN EOC 30K-3 EOC 30K-3L EOC 30K-4
Readability (d) 5 g / 10 g 5 g / 10 g 0,5 g / 1 g
Weighing range (max) 15 kg / 35 kg 15 kg / 35 kg 15 kg / 35 kg
Minimum piece weight 1 g 1 g 1 g
Reproducibility 5 g / 10 g 5 g / 10 g 0.5 g / 1 g
Linearity ± 15 g / 30 g ± 15 g / 30 g ± 1.5 g / 3 g
Warm-up time 10 minutes 10 minutes 2 hours
Reference unit weights at
piece count 10, 20, 50, 100, 200
Weighing Units Details „Weighing units“ chapter 7.6
Recommended adjustment
weight, not added (class)
Details for „Selection of
the Adjustment weight“
in chapter 7.6
30 kg
(M1)
30 kg
(M1)
30 kg
(M1)
Stabilization time (typical) 2 sec.
Electric Supply 100 V - 240 V, 50 / 60 Hz
Auto Off off, 3 min., 5 min., 15 min., 30 min.
Operating temperature - 10° C .... + 40° C
Humidity of air max. 80 % (not condensing)
Terminal (B x D x H) mm 268 x 115 x 80
Platform (B x D x H )mm 300 x 300 x 110 500x400x120 500x400x120
Weight kg (net) 5.2 9.0 9.0

EOC-BA-e-1711 7
KERN EOC 60K-2 EOC 60K-2L EOC 60K-3
Readability (d) 10 g / 20 g 10 g / 20 g 1 g / 2 g
Weighing range (max) 30 kg / 60 kg 30 kg / 60 kg 30 kg / 60 kg
Minimum piece weight 2 g 2 g 2 g
Reproducibility 10 g / 20 g 10 g / 20 g 1 g / 2 g
Linearity ± 30 g / 60 g ± 30 g / 60 g ± 3 g / 6 g
Warm-up time 10 minutes 10 minutes 2 hours
Reference unit weights at
piece count 10, 20, 50, 100, 200
Weighing Units Details „Weighing units“ chapter 7.6
Recommended adjustment
weight, not added (class)
Details for „Selection of
the Adjustment weight“
in chapter 7.6
60 kg
(M1)
60 kg
(M1)
60 kg
(M1)
Stabilization time (typical) 2 sec.
Electric Supply 100 V - 240 V, 50 / 60 Hz
Auto Off off, 3 min., 5 min., 15 min., 30 min.
Operating temperature - 10° C .... + 40° C
Humidity of air max. 80 % (not condensing)
Terminal (B x D x H) mm 268 x 115 x 80
Platform (B x D x H )mm 300 x 300 x 110 500x400x120 300 x 300 x 110
Weight kg (net) 5.2 9.0 5.2

8 EOC-BA-e-1711
KERN EOC 60K-3A EOC 60K-3L EOC 100K-2
Readability (d) 5 g 1 g / 2 g 20 g / 50 g
Weighing range (max) 60 kg 30 kg / 60 kg 60 kg / 150 kg
Minimum piece weight 2 g 2 g 5 g
Reproducibility 5 g 1 g / 2 g 20 g / 50 g
Linearity ± 15 g ± 3 g / 6 g ± 60 / 150 g
Warm-up time 30 minutes 2 hours 10 minutes
Reference unit weights at
piece count 10, 20, 50, 100, 200
Weighing Units Details „Weighing units“ chapter 7.6
Recommended adjustment
weight, not added (class)
Details for „Selection of
the Adjustment weight“
in chapter 7.6
60 kg
(F2)
60 kg
(M1)
150 kg
(M1)
Stabilization time (typical) 2 sec.
Electric Supply 100 V - 240 V, 50 / 60 Hz
Auto Off off, 3 min., 5 min., 15 min., 30 min.
Operating temperature - 10° C .... + 40° C
Humidity of air max. 80 % (not condensing)
Terminal (B x D x H) mm 268 x 115 x 80
Platform (B x D x H )mm 300 x 300 x 110 500 x 400 x 120 300 x 300 x 110
Weight kg (net) 5.2 9.0 5.2

EOC-BA-e-1711 9
KERN EOC 100K-2L EOC 100K-2A EOC 100K-2XL
Readability (d) 20 g / 50 g 10 g 20 g / 50 g
Weighing range (max) 60 kg / 150 kg 120 kg 60 kg / 150 kg
Minimum piece weight 5 g 5 g 5 g
Reproducibility 20 g / 50 g 10 g 20 g / 50 g
Linearity ± 60 / 150 g ± 30 g ± 60 g / 150 g
Warm-up time 10 minutes 30 minutes 10 minutes
Reference unit weights at
piece count 10, 20, 50, 100, 200
Weighing Units Details „Weighing units“ chapter 7.6
Recommended
adjustment weight, not
added (class)
Details for „Selection of
the Adjustment weight“
in chapter 7.6
150 kg
(M1)
120 kg
(F2)
150 kg
(M1)
Stabilization time (typical) 2 sec.
Electric Supply 100 V - 240 V, 50 / 60 Hz
Auto Off off, 3 min., 5 min., 15 min., 30 min.
Operating temperature - 10° C .... + 40° C
Humidity of air max. 80 % (not condensing)
Terminal (B x D x H) mm 268 x 115 x 80
Platform (B x D x H )mm 500 x 400 x 120 500 x 400 x 120 600 x 500 x 150
Weight kg (net) 9.0 9.0 18.4

10 EOC-BA-e-1711
KERN EOC 100K-2XXL EOC 100K-3 EOC 100K-3L
Readability (d) 20 g / 50 g 2 g / 5 g 2 g / 5 g
Weighing range (max) 60 kg / 150 kg 60 kg / 150 kg 60 kg / 150 kg
Minimum piece weight 10 g 5 g 5 g
Reproducibility 20 g / 50 g 2 g / 5 g 2 g / 5 g
Linearity ± 60 / 150 g ± 6 g / 15 g ± 6 g / 15 g
Warm-up time 10 minutes 2 hours 2 hours
Reference unit weights at
piece count 10, 20, 50, 100, 200
Weighing Units Details „Weighing units“ chapter 7.6
Recommended
adjustment weight, not
added (class)
Details for „Selection of
the Adjustment weight“
in chapter 7.6
150 kg
(M1)
150 kg
(F2)
150 kg
(F2)
Stabilization time (typical) 2 sec. 3 sec. 3 sec.
Electric Supply 100 V - 240 V, 50 / 60 Hz
Auto Off off, 3 min., 5 min., 15 min., 30 min.
Operating temperature - 10° C .... + 40° C
Humidity of air max. 80 % (not condensing)
Terminal (B x D x H) mm 268 x 115 x 80
Platform (B x D x H )mm 950 x 500 x 60 300 x 300 x 110 mm 500 x 400 x 120
Weight kg (net) 15.7 5.2 9.0

EOC-BA-e-1711 11
KERN EOC 300K-2 EOC 300K-2L EOC 300K-3
Readability (d) 50 g / 100 g 50 g / 100 g 5 g / 10 g
Weighing range (max) 150 kg / 300 kg 150 kg / 300 kg 150 kg / 300 kg
Minimum piece weight 10 g 10 g 10 g
Reproducibility 50 g / 100 g 50 g / 100 g 5 g / 10 g
Linearity ± 150 / 300 g ± 150 / 300 g ± 15 g / 30 g
Warm-up time 10 minutes 10 minutes 2 hours
Reference unit weights at
piece count 10, 20, 50, 100, 200
Weighing Units Details „Weighing units“ chapter 7.6
Recommended
adjustment weight, not
added (class)
Details for „Selection of
the Adjustment weight“
in chapter 7.6
300 kg
(M1)
300 kg
(M1)
300 kg
(F2)
Stabilization time (typical) 2 sec. 2 sec. 2 sec.
Electric Supply 100 V - 240 V, 50 / 60 Hz
Auto Off off, 3 min., 5 min., 15 min., 30 min.
Operating temperature - 10° C .... + 40° C
Humidity of air max. 80 % (not condensing)
Terminal (B x D x H) mm 268 x 115 x 80
Platform (B x D x H )mm 500 x 400 x 120 600 x 500 x 150 mm 500 x 400 x 120
Weight kg (net) 9.0 18.4 9.0

12 EOC-BA-e-1711
2 Appliance overview
1. Weight display
2. Weighing unit
3. Keyboard
4. Connection of mains adapter
5. RS232
6. Input connection load cell cable
7. Battery compartment
8. Guide rail support base / stand

EOC-BA-e-1711 13
Support base/Wall fixture
Balance EOC with assembly plate EOC-A03 (optional):
1. Platform
2. Fitting panel
3. Display Unit

14 EOC-BA-e-1711
Balance with tripod EOC-A05 (optional):
Flip/Flop evaluation device:
May be positioned in many ways, e.g. self-supporting or screwed into the wall
(optional). By turning the upper housing shell, the angle of the display as well as the
exit position of the cables can be changed.

EOC-BA-e-1711 15
2.1 Overview of display
Display Significance
W1 Weighing range 1
W2 Weighing range 2
Battery very low
Stability display
Zero indicator
GROSS Gross weight
NET Net weight
PT Pre-Tare
hold Hold function
Pcs Parts counting
Kg Weighing unit
Totalization
Indicators for weighing with tolerance range

16 EOC-BA-e-1711
2.2 Keyboard overview
Buttons Function
• Turn on/off
• Zeroing
Navigation button • Confirm entry
• Taring
Navigation button
• At numeric input increase flashing digit
• Scroll forward in menu
• Display sum total
Navigation button • Digit selection to the right
• Add weighing value to summation memory
Navigation button • Digit selection to the left
• Calculate weighing data via interface
C • Delete
• Change between gross and net weight
ESC • Back to menu/weighing mode
+
• Call up mean value function
+
• Call up weighing with tolerance range
+
• Delete total added memory

EOC-BA-e-1711 17
2.2.1 Numerical input via the navigation buttons
Press , the current setting will be displayed. The first digit will be flashing
and is ready for changing.
If you do not wish to change the first digit, press , the second digit will start
flashing.
Each time you press , the display will move to the subsequent digit, after
the last digit the display will return to the first digit.
To change the selected (flashing) digit, press repeatedly until the desired
value is displayed. Then press
to access further digits and change them by
.
Complete your entry by .

18 EOC-BA-e-1711
3 Basic Information (General)
3.1 Proper use
The balance you purchased is intended to determine the weighing value of a load. It
is intended to be used as a “non-automatic balance”, i.e. the material to be weighed
is manually and carefully placed in the centre of the weighing pan. As soon as a
stable weighing value is reached the weighing value can be read.
3.2 Improper Use
Do not use balance for dynamic weighing. In the event that small quantities are
removed or added to the load, incorrect weighing results can be displayed due to the
“stability compensation“. (Example: Slowly draining fluids from a container on the
balance.)
Do not leave permanent load on the weighing pan. This may damage the measuring
system.
Impacts and overloading exceeding the stated maximum load (max) of the balance,
minus a possibly existing tare load, must be strictly avoided. Balance may be
damage by this.
Never operate balance in explosive environment. The serial version is not explosion
protected.
The structure of the balance may not be modified. This may lead to incorrect
weighing results, safety-related faults and destruction of the balance.
The balance may only be used according to the described conditions. Other areas of
use must be released by KERN in writing.
3.3 Warranty
Warranty claims shall be voided in case
• Our conditions in the operation manual are ignored
• The appliance is used outside the described uses
• The appliance is modified or opened
• Mechanical damage and damage caused by media, liquids
• Natural wear and tear
• The appliance is improperly set up or incorrectly electrically connected
• The measuring system is overloaded

EOC-BA-e-1711 19
3.4 Monitoring of Test Equipment
In the framework of quality assurance the measuring-related properties of the
balance and, if applicable, the testing weight, must be checked regularly. The
responsible user must define a suitable interval as well as type and scope of this
test. In KERN’s accredited DKD calibration laboratory test weights and
balances may be calibrated (return to the national standard) fast and at moderate
cost.
4 Basic Safety Precautions
4.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and
commissioning, even if you are already familiar with KERN
balances.
4.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
5 Transport and storage
5.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the
appliance itself when unpacking for possible visible damage.
5.2 Packaging / return transport
Keep all parts of the original packaging for a possibly
required return.
Only use original packaging for returning.
Prior to dispatch disconnect all cables and remove
loose/mobile parts.
Reattach possibly supplied transport securing devices.
Secure all parts such as the glass wind screen, the
weighing platform, power unit etc. against shifting and
damage.

20 EOC-BA-e-1711
6 Unpacking, Setup and Commissioning
6.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in
common conditions of use.
You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
• Place the balance on a firm, level surface;
• Avoid extreme heat as well as temperature fluctuation caused by installing next to
a radiator or in the direct sunlight;
• Protect the balance against direct draughts due to open windows and doors;
• Avoid jarring during weighing;
• Protect the balance against high humidity, vapours and dust;
• Do not expose the device to extreme dampness for longer periods of time. Non-
permitted condensation (condensation of air humidity on the appliance) may
occur if a cold appliance is taken to a considerably warmer environment. In this
case, acclimatize the disconnected appliance for ca. 2 hours at room
temperature.
• Avoid static charge of loads and weighing container.
Major display deviations (incorrect weighing results) may be experienced should
electromagnetic fields (e.g. due to mobile phones or radio equipment), static
electricity accumulations or instable power supply occur. Change location or remove
source of interference.
6.2 Unpacking
Carefully remove the balance from the packaging, remove plastic cover and setup
balance at the intended workstation.