PAGE 4 OF 27 0417 IH-1649
OPERATION CONTINUED
PARTS COUNTING CONTINUED
5. Press the ZERO/ENTER key Zero when the displayed
number matches the number of items used in the
sample. As more weight is added, the display shows
the number of pieces (pcs).
6. Press the %key %to display the unit weight (g/pcs),
total weight (k g) or the count (pcs).
7. Press the FUNC key Func to return to weighing mode.
NOTE: See uline.com for a parts counting
instructional video.
CHECK WEIGHING
Check weighing sounds an alarm when the weight on the
scale exceeds limits stored in the memory. High and low
limits can be set and can be used individually or together.
Setting the High Limit:
1. Press the LIMIT key Limit . Display shows the current
high limit. left most number is flashing and the HI
icon is on.
2. To change the high limit, press the %/< %and the
LIMIT/>Limit keys to select which number to change.
3. Press the TARE/^key Tare to change the flashing
number.
4. Press ZERO/ENTER key Zero to set the number.
5. If you need to reset the number to zero, press the
FUNC/C Func key.
Setting the Low Limit:
1. After pressing ZERO/ENTER key Zero , display then
shows the low limit, LO will be on (left side of display).
2. To change the low limit, press the %/< %and the
LIMIT/> Limit to select which number to change.
3. Press the TARE/^key Tare to change the flashing
number.
4. Press ZERO/ENTER key Zero to set the number.
5. If you need to reset the number to zero, press the
FUNC/C key Func
.
6. After pressing the ZERO/ENTER key Zero , the scale
returns to weighing mode with the check-weighing
function turned on.
7. When a weight is placed on the scale, arrows
indicate if the weight is over or under the limits and
a beeper sounds.
Both Limits Set:
• The display shows OK when the weight is within
the limits.
Low Limit Set, High Limit Set to Zero:
• Display shows OK when weight is under the limit.
Shows HI when above the Low Limit.
High Limit Set, Low Limit Set to Zero:
• Display shows LO when weight is less than the high
limit. Above the high limit display shows OK.
Both Limits Set, Low is Set Greater than High:
• The beeper will never sound and display shows LO
if the weight is less than the LO limit. Display shows
HI if the weight is greater than the LO Limit.
NOTE: The weight must be greater than 20
scale divisions for check weighing to operate.
To Disable Check Weighing:
1. Press the LIMIT key Limit to display current limits.
2. Press the FUNC/C key Func to enter zero into both
limits when the current limits are shown, then press
the ZERO/ENTER key Zero to store the zero values.
ACCUMULATED TOTAL
Scale can be set to accumulate automatically or
manually.
To set automatic or manual accumulation:
1. Press FUNC Func and ACC Acc keys simultaneously.
2. Press LIMIT/> key Limit to scroll through options until
AU ON or AU OFF is displayed.
3. AU ON means automatic accumulation is on.
4. AU OFF means scale is set for manual
accumulation.
5. Press TARE/^ key Tare to toggle between AU ON
and AU OFF.
6. Press ZERO/ENTER key Zero to set your selection.
7. Press FUNC key Func to exit.
NOTE: Even when the scale is set to display
grams or ounces, the accumulate function
still keeps weight in kilograms or pounds.