KERN ILT 12K0.1DM User manual

KERN & Sohn GmbH
Ziegelei 1
D-72336 Balingen
Tel: +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.kern-sohn.com
Operating instruction
Platform scale
KERN ILT
Version 1.2
04/2006
GB
ILT-BA-e-0612

ILT-BA-e-0612
2
GB KERN ILT
Version 1.2 04/2006
Operating instruction
Platform scale
Contents
1Technical data__________________________________________________ 4
2Declaration of conformity ________________________________________ 6
3Appliance overview _____________________________________________ 8
4Basic Information (General)______________________________________ 10
4.1 Proper use______________________________________________________ 10
4.2 Improper Use ___________________________________________________ 10
4.3 Warranty _______________________________________________________ 10
4.4 Monitoring of Test Resources ______________________________________ 10
5Basic Safety Precautions________________________________________ 11
5.1 Pay attention to the instructions in the Operation Manual_______________ 11
5.2 Personnel training _______________________________________________ 11
6Transport and storage __________________________________________ 11
6.1 Testing upon acceptance _________________________________________ 11
6.2 Packaging ______________________________________________________ 11
7Unpacking, Setup and Commissioning ____________________________ 11
7.1 Installation Site, Location of Use ___________________________________ 11
7.2 Unpacking ______________________________________________________ 12
7.2.1 Setup ______________________________________________________________ 13
7.2.2 Scope of delivery _____________________________________________________ 13
7.3 Mains connection ________________________________________________ 13
7.4 Connection of peripheral devices___________________________________ 13
7.5 Initial Commissioning ____________________________________________ 13

ILT-BA-e-0612 3
8User menu ____________________________________________________ 14
8.1 User principle of the menu control __________________________________ 17
8.2 Operation of the balance with PS/2-keyboard _________________________ 18
8.3 Menu selection __________________________________________________ 19
8.3.1 P1 Calibration (adjustment) _____________________________________________ 19
8.3.2 P2 GLP (Good laboratory practice) _______________________________________ 25
8.3.3 P3 Date/time_________________________________________________________ 27
8.3.4 P4 Readout__________________________________________________________ 29
8.3.5 P5 RS-232 __________________________________________________________ 32
8.3.6 P6 Printout __________________________________________________________ 34
8.3.7 P7 Units ____________________________________________________________ 40
8.3.8 P8 Operating modes___________________________________________________ 40
8.3.9 P9 Globals __________________________________________________________ 41
9Operation_____________________________________________________ 42
9.1 Operating modes (P8) ____________________________________________ 44
9.1.1 Simple weighing ______________________________________________________ 46
9.1.2 Counting ____________________________________________________________ 47
9.1.3 Checkweighing (weighing with tolerance range) _____________________________ 55
9.1.4 Filling ______________________________________________________________ 57
9.1.5 Percent weighing _____________________________________________________ 58
9.1.6 Animal weighing (dynamic weighing) ______________________________________ 60
9.1.7 Formulation__________________________________________________________ 63
9.1.8 Statistics: ___________________________________________________________ 67
10 Data output _________________________________________________ 70
10.1 Technical data___________________________________________________ 70
10.2 Pin allocation of the balance output plug (front view) __________________ 70
10.3 Remote control instructions _______________________________________ 71
11 Service, maintenance, disposal_________________________________ 72
11.1 Cleaning _______________________________________________________ 72
11.2 Service, maintenance_____________________________________________ 72
11.3 Disposal________________________________________________________ 72
12 Instant help _________________________________________________ 73

ILT-BA-e-0612
4
1 Technical data
KERN ILT 12K0.1DM ILT 30K0.2DM
Readability (d) 0.1/0.2 g 0.2/0.5 g
Weighing range (max) 6/12 kg 12/30 kg
Minimum load 5/10 g 10/25 g
Verification value 1/2 g 2/5 g
Verification class II II
Reproducibility 0.2/0.4 g 0.4/1 g
Linearity ± 0.2/0.4 g ± 0.4/1 g
Recommended adjusting
weight (not supplied) 10 kg (F2) 20 kg (F2)
Minimum piece weight for
counting function > 0.1 g > 0.2 g
Stabilization time 4 sec.
Weighing units
(calibrated appliances) g, kg
Weighing units
(non-calibrated appliances) g, kg, ct, lb, oz, ozt, dwt, t {h}, t {S}, t {t}, mom, GN, N, t,
freely programmable
Reference quantities at
piece counting 10, 20, 50,100, freely selectable
Weighing plate, stainless
steel 300 x 300 mm
Dimensions of the housing
(B x D x H) 300 x 300 x 75 (platform)
197 x 110 x 35 ( display)
Net weight (kg) 7,7
Permitted environmental
condition -15° C to 30° C
Humidity of air max. 80 % relative (not condensing)
Voltage 230 V/50Hz

ILT-BA-e-0612 5
KERN ILT 60K0.5DM ILT 120K1DM
Readability (d) 0.5/1 g 1/2 g
Weighing range (max) 30/60 kg 60/120 kg
Minimum load 25/50 g 50/100 g
Verification value 5/10 g 10/20 g
Verification class II II
Reproducibility 1/2 g 2/4 g
Linearity ± 1/2 g ± 2/4 g
Recommended adjusting
weight (not supplied) 50 kg (F2) 100 kg (F2)
Minimum piece weight for
counting function > 1 g > 1 g
Stabilization time 4 sec.
Weighing units
(calibrated appliances) g, kg
Weighing units
(non-calibrated appliances) g, kg, ct, lb, oz, ozt, dwt, t {h}, t {S}, t {t}, mom, GN, N, t,
freely programmable
Reference quantities at
piece counting 10, 20, 50,100, freely selectable
Weighing plate, stainless
steel 410 x 410 mm
Dimensions of the housing
(B x D x H) 410 x 410 x 90 (platform)
197 x 110 x 35 ( display)
Net weight (kg) 13
Permitted environmental
condition -15° C to 30° C
Humidity of air max. 80 % relative (not condensing)
Voltage 230 V/50Hz

ILT-BA-e-0612
6
2 Declaration of conformity
KERN & Sohn GmbH
D-72322 Balingen-Frommern
Postfach 4052
Tel: 0049-[0]7433- 9933-0
Fax: 0049-[0]7433-9933-149
Internet: www.kern-sohn.de
Konformitätserklärung
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con marca CE
Dichiarazione di conformitá per apparecchi contrassegnati con la marcatura CE
English We hereby declare that the product to which this declaration refers conforms with the
following standards.
Deutsch Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den
nachstehenden Normen übereinstimmt.
Français Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente
déclaration, est conforme aux normes citées ci-après.
Español Manifestamos en la presente que el producto al que se refiere esta declaración est´´a de
acuerdo con las normas siguientes
Italiano Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è
conforme alle norme di seguito citate.
Electronic Balance: KERN ILT
Mark applied EU Directive Standards
89/336/EEC EMC
EN45501
EN55022
73/23/EEC Low Voltage EN60950
Date: 30.01.2006 Signature:
Gottl. KERN & Sohn GmbH
Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-[0]7433/9933-0,Fax +49-[0]7433/9933-149

ILT-BA-e-0612 7
KERN & Sohn GmbH
D-72322 Balingen-Frommern
Postfach 4052
Tel: 0049-[0]7433- 9933-0
Fax: 0049-[0]7433-9933-149
Internet: www.kern-sohn.de
Konformitätserklärung
Declaration of conformity for apparatus with CE mark
Konformitätserklärung für Geräte mit CE-Zeichen
Déclaration de conformité pour appareils portant la marque CE
Declaración de conformidad para aparatos con marca CE
Dichiarazione di conformitá per apparecchi contrassegnati con la marcatura CE
English We hereby declare that the product to which this declaration refers conforms with the
following standards.
This declaration is only valid with the certificate of conformity by a notified body.
Deutsch Wir erklären hiermit, dass das Produkt, auf das sich diese Erklärung bezieht, mit den
nachstehenden Normen übereinstimmt.
Diese Erklärung gilt nur in Verbindung mit der Konformitätsbescheinigung einer
benannten Stelle.
Français Nous déclarons avec cela responsabilité que le produit, auquel se rapporte la présente
déclaration, est conforme aux normes citées ci-après.
Cette déclaration est valide seulement avec un certificat de conformité d’un
organisme notifié.
Español Manifestamos en la presente que el producto al que se refiere esta declaración está de
acuerdo con las normas siguientes.
Esta declaración solo será válida acompañada del certificado de conformidad de
conformidad de la parte nominal.
Italiano Dichiariamo con ciò che il prodotto al quale la presente dichiarazione si riferisce è
conforme alle norme di seguito citate.
Questa dichiarazione sarà valida solo se accompagnata dal certificato di conformità
della parte nominale.
Model: KERN ILT
EU Directive Standards EC-type-approval certificate no. Issued by
90/384/EEC EN 45501 PL 06 001 GUM
Date: 30.01.2006 Signature:
Gottl. KERN & Sohn GmbH
Management
Gottl. KERN & Sohn GmbH, Ziegelei 1, D-72336 Balingen, Tel. +49-07433/9933-0,Fax +49-074433/9933-149

ILT-BA-e-0612
8
3 Appliance overview
Models: ILT 12K0.1DM, ILT 30K0.2M
1. Mains connection
2. PS/2 keyboard
3. RS 232 interface
Air bubble
ON/OFF switch

ILT-BA-e-0612 9
Models: ILT 60k0.50M, ILT 120K1DM
1. ON/OFF switch
2. Mains connection
3. PS/2 keyboard
4. RS 232 interface
Air bubble (underneath weighing plate)

ILT-BA-e-0612
10
4 Basic Information (General)
4.1 Proper use
The balance you purchased is intended to determine the weighing value of material
to be weighed. It is intended to be used as a “non-automatic“ balance, i.e. the
material to be weighed is manually and carefully placed in the centre of the weighing
plate. As soon as a stable weighing value is reached the weighing value can be read.
4.2 Improper Use
Do not use balance for dynamic add-on weighing procedures, if small amounts of
goods to be weighed are removed or added. The “stability compensation“ installed in
the balance may result in displaying an incorrect measuring value! (Example: Slowly
draining fluids from a container on the balance.)
Do not leave permanent load on the weighing plate. This may damage the measuring
system.
Impacts and overloading exceeding the stated maximum load (max) of the balance,
minus a possibly existing tare load, must be strictly avoided. Balance may be
damage by this.
Never operate balance in explosive environment. The serial version is not explosion
protected.
The structure of the balance may not be modified. This may lead to incorrect
weighing results, safety-related faults and destruction of the balance.
The balance may only be used according to the described conditions. Other areas of
use must be released by KERN in writing.
4.3 Warranty
Warranty claims shall be voided in case
•Our conditions in the operation manual are ignored
•The appliance is used outside the described uses
•The appliance is modified or opened
•Mechanical dammage or dammage by media, liquids, natural wear and tear
•The appliance is improperly set up or incorrectly electrically connected
•The measuring system is overloaded
4.4 Monitoring of Test Resources
In the framework of quality assurance the measuring-related properties of the
balance and, if applicable, the testing weight, must be checked regularly. The
responsible user must define a suitable interval as well as type and scope of this test.
Information is available on KERN’s home page (www.kern-sohn.com with regard to
the monitoring of balance test substances and the test weights required for this. In
KERN’s accredited DKD calibration laboratory test weights and balances may be
calibrated (return to the national standard) fast and at moderate cost.

ILT-BA-e-0612 11
5 Basic Safety Precautions
5.1 Pay attention to the instructions in the Operation Manual
Carefully read this operation manual before setup and commissioning, even if you
are already familiar with KERN balances.
5.2 Personnel training
The appliance may only be operated and maintained by trained personnel.
6 Transport and storage
6.1 Testing upon acceptance
When receiving the appliance, please check packaging immediately, and the
appliance itself when unpacking for possible visible damage.
6.2 Packaging
Keep all parts of the original packaging in case you need to return the appliance.
Only use original packaging for returning.
Before sending, disconnect all connected cables and loose/movable parts.
Attach possibly existing transport safeguards. Secure all parts, e.g. weighing plate,
power unit etc., to prevent slipping and damage.
7 Unpacking, Setup and Commissioning
7.1 Installation Site, Location of Use
The balances are designed in a way that reliable weighing results are achieved in
common conditions of use.
You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
•Place the balance on a firm, level surface;
•Avoid extreme heat as well as temperature fluctuation caused by installing
next to a radiator or in the direct sunlight;
•Protect the balance against direct draughts due to open windows and doors;
•Avoid jarring during weighing;
•Protect the balance against high humidity, vapours and dust;
•Do not expose the device to extreme dampness for longer periods of time.
Non-permitted condensation (condensation of air humidity on the appliance)
may occur if a cold appliance is taken to a considerably warmer environment.
In this case, acclimatize the disconnected appliance for ca. 2 hours at room
temperature.
•Avoid static charge of goods to be weighed or weighing container.

ILT-BA-e-0612
12
If electro-magnetic fields or static charge occur, or if the power supply is unstable
major deviations on the display (incorrect weighing results) are possible. In that case,
the location must be changed.
7.2 Unpacking
Carefully remove the balance from the packaging and setup balance at the intended
workstation.
Remove the transport arrestment (models ILT 60k0.50M, ILT 120K1DM):
•Screw in the 4 foot screws
•Remove the 4 plastic inserts (2)
•Place the weighing pan (3) on the balance

ILT-BA-e-0612 13
7.2.1 Setup
Level balance with foot screws until the air bubble of the water balance is in the
prescribed circle.
7.2.2 Scope of delivery
Serial accessories:
•Balance
•Weighing plate
•Mains power supply
•Operating Manual
•Safety hood above control panel
7.3 Mains connection
Power is supplied via the external mains power supply. The stated voltage value
must be the same as the local voltage.
Only use original KERN mains power supplies. Using other makes requires consent
by KERN.
7.4 Connection of peripheral devices
Before connecting or disconnecting of additional devices (printer, PC) to the data
interface, always disconnect the balance from the power supply.
With your balance, only use accessories and peripheral devices by KERN, as they
are ideally tuned to your balance.
7.5 Initial Commissioning
A warming up time of 15 minutes after switching on stabilizes the measuring values.
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter Adjustment.

ILT-BA-e-0612
14
8 User menu
The user menu has nine main menus (P1 - P9), arranged in the following sub-menus:
P1 Calibration
01 External calibration * * * * * * * * Function
02 User calibration * * * * * * * * Function
03 Calibration test * * * * * * * * Function
04 Print report 1 on
P2 GLP
01 User
02 Project
03 Time print 1 on
04 Date print 1 on
05 User print 1 on
06 Project print 1 on
07 Id printout 1 on
08 Last Cal. print 1 on
P3 Date/time
01 Date format 0 DA/MO/YR
02 Time format 0 24 hour
03 Time * * * * * * * * Function
04 Date * * * * * * * * Function
05 Display time 1 on
06 Display date 1 on
P4 Readout
01 Filter 3 normal
02 Med. filter 1 active
03 Disp.refresh 1
04 Auto zero 1 on
05 Last digit 1 always
06 Negative 0 disabled

ILT-BA-e-0612 15
P5 RS-232
01 Baud rate 1 4800
02 Parity 0 None
03 Data bits 2 8 bit
04 Stop bits 1 1 Stop bit
05 Handshake 0 none
06 Auto. Print 0 None
07 Interval 1 * 0.04 s
08 Min. mass 4 10 d
09 Print on stab 1 enabled
10 Pause 1 * 0.1 s
P6 Printouts
01 Prontout men. * * * * * * * * Function
02 1. Printout dest. 0 RS+Mem
03 Printout No. 0 Standard
04 Printout Mem. No. 0 Standard
05 1. Printout start 1
06 1. Printout stop 1
07 2. Printout start 1
08 2. Printout stop 1
.. ………………….
13 1. String
14 2. String
.. ………………….
92 80. String
P7 Units
01 Gram 1 enabled
02 Kilogram 0 disabled
03 Pound 0 disabled
04 Ounce 0 disabled
05 Troy ounce 0 disabled
06 Carat 0 disabled
07 Dwt 0 disabled
08 Teal Hk. 0 disabled
09 Teal S. 0 disabled
10 T.eal T 0 disabled
11 Momme 0 disabled
12 Grain 0 disabled
13 Newton 0 disabled
14 Tical 0 disabled
15 User defined 0 disabled
16 User factor 0.1
P8 Operating modes
01 Parts Count 1 enabled
02 Checkweighing 1 enabled
03 Filling 1 enabled
04 Percent 1 enabled
05 Animal 1 enabled
06 Density 1 enabled
07 Formulation 1 enabled
08 Statistics 1 enabled

ILT-BA-e-0612
16
P9 Globals
01 ID setting * * * * * * * * Function
02 ID autoprint 0 no
03 Beep 1 on
04 Language * * * * * * * * Function
05 Backlight 1 on
06 Contrast * * * * * * * * Function
07 Screensaver 0 disabled
08 Balance Id WL141475
09 Software rev. MBT 1b24
10 Parameter printout * * * * * * * * Function
11 Parameter receive. * * * * * * * * Function
Main menu:
1. Menu number (P1 - P9)
2. Menu designation
3. Cursor ( ►) indicates the current
menu item
Submenu:
1. Main menu number (P1 – P9)
2. Sub-menu number and designation
3. Parameter value or ****** i.e. this
menu item is stored as function
4. Description of function
5. Cursor ( ►) shows the current
menu item (e.g. P1).
6. Cursor ( ►) shows the current
setting (e.g. 06).
7. Status
8. Use the arrow key to modify the
parameter value, the active
parameter blinks.

ILT-BA-e-0612 17
8.1 User principle of the menu control
In the menu the settings of the balance can be modified and functions can be
activated.
This way, the balance can be adjusted to individual weighing requirements. Proceed
as follows:
ÖSwitch on the balance using the -key
ÖAfter actuating the key the display shows the main menus that can be set
P1 Calibration
P2 GLP
P3 Date/Time
P4 Readout
P5 RS-232
P6 Printouts
P7 Units
P8 Modes
P9 Global
s
Setup2.02.05
ÖUse the cursor ( ►) to select your setting
Use the key to move the cursor ( ►) down, the key to move it up
ÖThe key confirms your selected setting, the display shows the sub-menu or the
active menu item blinks:
Setup
P1
01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
function
function
function
off
on
function
31.01.05
11 hour
ÖUse the arrow keys to move within and enter into the menu:
•Cursor control up
•Parameter selection within function
•Increasing the value for selected numeral or number
•Access to main menu
•Access to sub-menu
•Move to next menu item
on the right inside menu
•Selection of the number
to be changed to the
right (the respective
active position flashes)
•Return to main menu
•Selection of the number
to be changed to the left
(the respective active
position flashes)
•Cursor control down
•Parameter selection within function
•Decreasing the value for selected numeral or number

ILT-BA-e-0612
18
ÖActuating the -key will confirm your setting; to interrupt use the -key
ÖPress -key to return to menu
ÖActuate -key again, a message will inquire whether you want to save the
changes you made.
ÖActuating the key will save the selected setting.
ÖIf you do not want to save your selected setting, press the key
ÖThe balance returns automatically into weighing mode.
Weighing
0.0000g
0% 100%
0
27.01.05
8.2 Operation of the balance with PS/2-keyboard
Using a PS/2 keyboard will help you to enter numbers and texts faster and easier.
Key allocation:
Balance keyboard
PS- keyboard

ILT-BA-e-0612 19
8.3 Menu selection
8.3.1 P1 Calibration (adjustment)
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
Procedure when adjusting:
Observe stable environmental conditions. A warming up time of ca. 1 hour is required
for stabilization.
Note!
During the adjustment process no objects should be on the weighing plate.
In calibrated balances the adjustment function is switch locked. To carry out an
adjustment, the switch must be unlocked, see chpt. 8.3.1.2
Operation Display
ÖCall up menu item "P1 Calibration" (see chapt. 8.1)
ÖPress -button
P1 Calibration
P2 GLP
P3 Date/Time
P4 Readout
P5 RS-232
P6 Printouts
P7 Units
P8 Modes
P9 Globals
Setup2.02.05
01 Ext. Cal - adjustment with external weight (locked for verifiable devices)
ÖUse the cursor ( ►) to select "01 Ext. cal."
Einstellungen
P1 01
02
03
04
Ext.
test
Pr
User calibr.
C
int report
calibr.
alibr.
********
********
********
1
function
function
on
function
29.12.04
ÖPress -button
Ensure that there are no objects on the weighing
plate.
ÖPress -key, on the display appears the weight
value of the adjusting weight.

ILT-BA-e-0612
20
ÖNow move the adjusting weight to the centre of
the weighing plate, press -key. If the
adjustment was successful, this will be followed
by automatic reset to the menu.
The -key may be used to cancel the
adjustment. Errors during adjustment or the use
of an incorrect adjusting weight will result in an
error message. Repeat adjustment.
Einstellungen
P1 01
02
03
04
Ext.
test
Pr
User calibr.
C
int report
calibr.
alibr.
********
********
********
1
function
function
on
function
29.12.04
ÖPress -key until you enter the menu/weighing
mode (see chapt. 8.1)
02 User cal. (locked in calibrated appliances)
ÖUse the cursor ( ►) to select „02 User cal.“
Einstellungen
P1 01
02
03
04
Ext.
test
Pr
calibr.
alibr.
User calibr.
C
int report
********
********
********
1
function
function
on
function
29.12.04
ÖPress -key, on the display appears the weight
value of the adjusting weight. The first digit of the
displayed value is flashing. Now you can change
the weight value by pressing the arrow keys.
After adjusting, confirm your new adjusting weight
by pressing the -key.
Einstellungen
Benutzerk ierungalibr
Masse eingeben g :
12000.00
[]
ÖEnsure that there are no objects on the weighing
plate. Press -button
ÖPress -key, the display shows the weight
value of the adjusting value set by you
This manual suits for next models
3
Table of contents
Other KERN Scale manuals

KERN
KERN EMS 300-3 User manual

KERN
KERN MBB 15K5M User manual

KERN
KERN EOB Series User manual

KERN
KERN EOA 10K-3 User manual

KERN
KERN EMS 300-3 User manual

KERN
KERN MPL 200K-1P User manual

KERN
KERN PKS 200-3 User manual

KERN
KERN ADJ Series User manual

KERN
KERN ALS Series User manual

KERN
KERN PCB 200-3 User manual