Kimo C310 User manual

Configuration of C310 multifunction
transmitters by keypad


Table of contents
1. Introduction..........................................................................................................................................5
1.1. Description of the transmitter.......................................................................................................5
1.2. Description of the keys.................................................................................................................5
1.3. Protection tips of the sensor........................................................................................................6
2. Modbus................................................................................................................................................7
2.1. Confi uration of parameters........................................................................................................7
2.2. Functions......................................................................................................................................7
2.3. Access code to re ister................................................................................................................7
3. Confi ure the transmitter...................................................................................................................10
3.1. Enter the security code..............................................................................................................10
3.2. Set the parameters of the transmitter........................................................................................10
3.2.1 Set the screen......................................................................................................................11
3.2.2 Set the raphical period.......................................................................................................11
3.2.3 Set the lan ua e..................................................................................................................11
3.2.4 Set date and time.................................................................................................................11
3.2.5 Activate /deactivate the beep key........................................................................................11
3.2.6 Lock and unlock the keyboard.............................................................................................11
3.2.7 Modify the security code......................................................................................................12
3.2.8 Back to factory parameters..................................................................................................12
3.2.9 Set the Modbus (optional)...................................................................................................12
3.2.10 Set the Ethernet communication (optional).......................................................................12
3.2.11 Access to the transmitter information................................................................................13
3.3. Confi ure the measurement channels.......................................................................................13
3.4. Confi ure the analo ue outputs.................................................................................................13
3.4.1 Select the type of output......................................................................................................13
3.4.2 Set the minimum and maximum ran es..............................................................................14
3.4.3 Perform an output dia nostic...............................................................................................14
3.5. Confi ure the alarms..................................................................................................................15
3.5.1 Select the alarm mode.........................................................................................................15
3.5.2 Set the alarm in risin or fallin ed e..................................................................................16
3.5.3 Confi ure the alarm in monitorin mode.............................................................................16
3.5.4 Confi ure the alarm on Transmitter state mode..................................................................16
3.5.5 Confi ure the relays (optional)............................................................................................17
3.6. Confi ure the probes, boards and normative values.................................................................18
3.6.1 Confi ure an air velocity and airflow probe.........................................................................18
3.6.2 Confi ure a CO2 probe or a temperature/hy rometry probe..............................................19
3.6.3 Confi ure a board................................................................................................................19
3.7. Activate an option.......................................................................................................................22
4. Description of function and Modbus connections.............................................................................24
4.1. Device........................................................................................................................................24
4.2. Channels....................................................................................................................................24
4.3. Outputs.......................................................................................................................................25
4.4. Alarms........................................................................................................................................26
4.5. Measurement parameters..........................................................................................................26


1. Introduction
1.1. Description of the transmitter
C310 transmitters with display can be configured via keypad. It is possible to set the measurement units, to activate or not a
channel,...
Trend indicator : on the screen, a trend indicator represented by a rising arrow, a going down arrow or a stable arrow is
displayed above the measurement unit. It means the calculation of a moving average on the last hour ( 1) to compare with
a moving average on the last 5 minutes ( 2) :
•If 1 = 2, the trend is stable.
•If 1 < 2, the trend is upwards.
•If 1 > 2, the trend is downward.
This trend indicator is displayed after 5 minutes.
Indicator light : the indicator light indicates that the transmitter if powered on.
1.2. Description of the keys
–Up key : increments a value, allows to browse through the menus
–Down key : decrements a value, allows to browse through the menus
–Left key : allows to browse through the menus
–Right key : allows to browse through the menus
–OK key : validates an input, allows to enter in the main menu
–Esc key : cancels an input or backs to the previous step
5
Right key
Down key OK key
Left key Indicator light
C1 9
Pa
Graphique►
Mesures Alarm lights
Up key
Input probe 1 Input probe 2
C2 35
%RH
◄Min/max
05/02/14 9.53
Esc key

1.3. Protection tips of the sensor
It's extremely unwise to remove the protection tip of our hygrometry probes as
the sensitive element is very fragile even to light contacts. However, if you
have to remove the protection tip, take all possible precautions and avoid any
contact with the sensitive element.
To remove the protection tip, unscrew it or unclip it.
6
Protection tip
to unscrew
Protection tip
to unscrew
Sensitive
element
Sensitive
element

2. Modbus
2.1. Configuration of parameters
•Comm nication speed : between 2400 and 115200 bauds, 19200 bauds by default
•Data bits : 8 bits
•Stop bit : 1 bit
•Parity : None
•Flow control : None
•Transmitter addressing : between 1 and 255 (automatically answers the requests from address 0)
2.2. Functions
•Register f nction : Function 03
•Register writing : Function 16
•Comm nication loop test : Function 08
2.3. Access code to register
•Registers type:
Type Size Detail
Serial number 8 octets Data organization
U8 1 octet 8 bits unsigned integer
U16 2 octets 16 bits unsigned integer
U32 4 octets 32 bits unsigned integer
Real 4 octets 32 bits real
Enumeration 1 octet See table “Enumerations”” pError: Reference source not found
Boolean 1 octet True = 1 ; False = 0
Date 4 octets Year (2 octets), onth (1 octet), Day (1 octet)
Time 5 octets Hour (1 octet), inute (1 octet), Second (1 octet), illisecond (2 octets)
•Alarms stat s – odbus code : 7000
b31 … b4 b3 b2 b1 b0
0 1 0 1
•Val es – odbus code : 7010 (channel 1)
7040 (channel 2)
7070 (channel 3)
7100 (channel 4)
•N mber of digits after the decimal point – odbus code : 7020 (channel 1)
7050 (channel 2)
7080 (channel 3)
7110 (channel 4)
7
Alarm channel 1
Alarm channel 2
Alarm channel 3
Alarm channel 4

•Unit – odbus code : 7030 (channel 1)
7060 (channel 2)
7090 (channel 3)
7120 (channel 4)
List of nits :
Field Unit Val e Field Unit Val e
None 0 None 0
Temperat re °C 16
Air velocity
m/s 64
°F 17 fpm 65
Hygrometry
%HR 32 km/h 66
G/Kg 33
Airflow
m3/h 80
Kj/KG 34 L/s 81
°C td 35 cfm 82
°F td 36 m3/s 83
m3/h 80
°C Tw 37 Comb stion ppm 112
°F Tw 38
Electricity
mV 160
Press re
kPa 50 V 161
inWg 51 µA 162
hPa 52 mA 163
mbar 53 A 164
mmHg 54
mmH2O 55
daPa 56
Pa 57
Example of a meas rement at 24.2 °C :
•Value : code 7010, real type ( ost Significant Byte / Least Significant Byte)
Byte 3 Byte 4 Byte 1 Byte 2
b15 b16 b31 b24 b7 b0 b15 b8
0100 0001 1100 0001 1001 1001 1001 1010
Val e : 24.2
•Number of digits after the decimal point : code 7020, U8 type
Byte 1
b7 b0
0000 0001
N mber of digits after the decimal point : 1
8

•Unit : code 7030, U8 type
Byte 1
b7 b0
0001 0000
Unit val e : 16 : °C (See table “List of nits” p8)
“En merations” table :
Corresponding val es 0 1 2 3 4 5 6 7
Backlight d ration Off 10s 30s 60s Permanent
Graphical period 3 mn 15 mn 30 mn 1 hour 3 hours 6 hours 12
hours
24
hours
Lang age French English Third language
Date Format dd-mm-yyyy mm-dd-yyyy yyyy-mm-dd
Time Format 24H 12H
Modb s Com speed 2400 4800 9600 19200 38400 115200
Channel x Unit See list of unit
Channel x Transmitter none probe 1 probe 2 module
O tp t x Type 4 - 20 mA 0 - 20 mA 0 - 10 V 0 - 5 V 0 - 1 V
O tp t x Diagnostic Deactivate 0% 50% 100%
Inp t x Type 4 - 20 mA 0 - 20 mA 0 - 10 V 0 - 5 V 0 - 1 V
Alarm x Mode Deactivate Rising edge Falling edge onitoring Transmitter
state
Alarm x Sec rity Negative Positive
Relay x Selection OFF ON Alarm 1 Alarm 2 Alarm 3 Alarm 4
Compensation Temperat re :
Unit °C °F
Compensation Temperat re :
Mode Value Thermocouple Probe N°1
Compensation Press re :
Unite hPa mbar mmHg m (alt)
Compensation Normative val e None DIN1343 ISO2533
Débit Press re nit
for airflow Pa mmH2O inWg mbar
Moy. Mes re Differential
press re device Pitot L Pitot S Debimo blades Factor
Section Type Rectangular Circular Coefficient
Section Unit mm inch
9

3. Configure the transmitter
3.1. Enter the security code
To configure the transmitter, and for security purposes, a security code must be entered. The default code is 0101.
➢Connect the transmitter.
➢Wait until the initializing period is over.
➢Press OK.
“Code” is displayed with “0000” below.
➢Press OK to go to the second zero.
➢Press Up key to display 1 then press OK.
➢Press OK to go to the 4th zero.
➢Press Up key to display 1 then press OK.
The configuration screen is displayed.
This screen allows to set :
•the transmitter
•the channels
•the analogue outputs
•the alarms
•the relays
•the configurations
•the options
If a wrong code is entered, the following message is displayed : “Bad code”. Press OK to go back to the
measurement screen then press OK to enter the code again.
3.2. Set the parameters of the transmitter
This part allows to set :
•the screen : contrast and backlight
•the graphical period : during”X” last minutes, the transmitter displays the measured values as a graph
•the language : French or English
•the date and time
•the beep key
•the keypad locking
•the safety code : modify the access code
•back to factory configuration
•the odbus
•the Ethernet
•Information : get an access to information about the transmitter and the connected probes
➢Enter the security code then press OK.
➢Press OK on “Device” line.
The screen allowing to configure the transmitter is displayed.
10
Code ?
0101

3.2.1 Set the screen
➢Press OK on “Screen” line.
➢Press OK on “Contrast” line to set the contrast between 0 and 3.
➢Use Up and Down keys to set the contrast then press OK.
➢Go to the “Backlight settings” line with the Down key then press OK.
➢Press OK on “D ration” line then select the backlight duration with Up and Down keys :
•ON : continuous backlight
•10 sec, 30 sec or 1 min : backlight turns off as soon as the selected time is up.
•OFF : deactivated backlight
➢Press OK to validate.
➢Press Down key then OK on “Backlight” line.
➢Set the backlight between 0 and 10 with Up and Down keys.
➢Press OK to validate.
➢Press Esc to back to the configuration screen of the device.
3.2.2 Set the graphical period
➢Press OK “Graphical period” line.
➢Use Up and Down keys to select the requested duration then press OK to validate.
Example : if 1 min is selected, the transmitter will graphically display the measured values during the last 1
minutes.
3.2.3 Set the language
➢Press OK on “Lang age” line.
➢Select the language with Up and Down keys : Frencjh or English.
➢Press OK to validate.
3.2.4 Set date and time
➢Press OK on “Date-Time” line.
“Date-time” screen is displayed.
➢Press OK on “Date” line then select with Up and Down keys the date format :
•JJ/ /AAAA : day/month/year
•/JJ/AAAA : month/day/year
•AAAA/ /JJ : year/month/ day
➢Press OK.
➢Go the number to modify with Up/Down/left/right keys then press OK.
➢Enter the required number with Up/Down/left/right keys then press OK.
➢Once the settings are performed, press Esc key to validate the date and time and to back to the configuration
screen of the device.
3.2.5 ctivate /deactivate the beep key
➢Press OK on “Beep key” line.
➢Select “YES” to activate the beep key or “NO” to deactivate it.
➢Press OK to validate.
3.2.6 Lock and unlock the keyboard
➢Press OK on “Keyboard Lock” line.
➢Select “YES” to lock the keyboard then press OK.
A message is displayed indicating that the keyboard is locked and the device backs to measurement screen.
Press OK d ring 6 s to nlock the keyboard.
11

3.2.7 Modify the security code
➢Press OK on “Sec rity code” line.
The following message is displayed : “Modification Security code?”.
➢Go to “Contin e” then press OK.
➢Enter the current code.
➢Enter the new code.
➢Go to “YES” then press OK to confirm the new code.
The transmitter displays a message indicating that the code has been modified.
➢Press OK.
Keep the new created code beca se it will be not possible to access to the config ration of the transmitter
with the code 0101.
3.2.8 Back to factory parameters
➢Press OK on “Back to factory settings” line.
➢Go to YES then press OK to confirm.
3.2.9 Set the Modbus (optional)
➢Press OK on “Modb s” line.
➢Press OK on “Slave n m” line.
➢Enter the slave number with Up and Down keys.
➢Press OK to validate.
➢Go to “Comm nication speed” line then press OK.
➢Select the speed with Up and Down keys then press OK.
If the odbus option is selected when ordered the transmitter, the default slave number will be indicated on serial
number label and the default speed is 19 200 bds.
3.2.10 Set the Ethernet communication (optional)
➢Press OK on “Ethernet” line.
➢Press OK on “Address” line.
The address consists of 4 groups of numbers.
➢Set the 1st group with the arrows then press OK.
➢Repeat the procedure for the 3 following groups.
➢Go to “OK” then press OK to validate the address.
➢Press OK on “Mask” line.
The mask consists of 4 groups of numbers.
➢Set the 1st group with the arrows then press OK.
➢Repeat the procedure for the 3 following groups
➢Go to “OK” then press OK to validate the mask.
➢Press OK on “Gateway” line.
The gateway consists of 4 groups of numbers.
➢Set the 1st group with the arrows then press OK.
➢Repeat the procedure for the 3 following groups.
➢Go to “OK” then press OK to validate the gateway.
➢Press OK on “Port” line.
➢Use the arrows to set the port then press OK.
The Address ac can not be modified because it is linked to the Ethernet board.
12

3.2.11 ccess to the transmitter information
This part allows to access to the information about the transmitter, the boards and the probes.
➢Press OK on “Information” line.
A 1st screen is displayed with information about the transmitter (name, serial number, calibration date).
➢Press Right arrow.
The screen displays information about the outputs.
➢Press Right arrow.
The screen displays information about the connected boards.
➢Press Right arrow.
The screen displays information about the connected probes.
➢Press Esc key to quit the information screen.
3.3. Configure the measurement channels
This part allows to activate and assign a measurement channel to a probe or a board.
By default, channels are activated according to the supplied configuration of the transmitter. This configuration is indicated
on the label located on the side of the transmitter.
To activate a channel, at least one probe or one board m st be connected to the transmitter.
➢Press OK.
➢Enter the security code then press OK.
➢Go to the “Channel” line then press OK.
The transmitter displays the 4 available channels.
➢Go to the line of the channel to activate then press OK.
The transmitter displays “Probe – Unit OFF”.
➢Press OK.
➢Select with Right and Left arrows “Probe 1”, “Probe 2” or “Board” according to the required attribution for the
channel.
The available units of the selected probe or board appear.
➢Select the unit with Up and Down arrows then press OK.
The transmitter displays the feature of the channel.
•the type of probe (probe1, probe 2 or board) and its unit
•the coefficient
•the offset
➢odify if necessary the coefficient and offset.
3. . Configure the analogue outputs
One channel m st be config red at least (see chapter 3.3).
If no channel has been config red, the transmitter will indicate that the o tp ts config ration is
impossible.
➢Press OK.
➢Enter the security code.
3.4.1 Select the type of output
➢Go to “Analog e o tp ts” line then press OK.
The transmitter displays the available analogue outputs : by default, outputs 1 and 2 are available. Outputs 3 and 4
are available as option.
➢Go to the output to configure with Up and Down arrows then press OK.
➢Go to the “Type” line then press OK.
13

The transmitter displays the different types of outputs : 4-20 mA, 0-20 mA, 0-10 V or 0- V.
➢Go to the type of required output then press OK.
The output number is equivalent to the number of the displayed channel (ex : output 1 = channel1)
3.4.2 Set the minimum and maximum ranges
➢Go to “Range min” then press OK.
➢Set the value of the minimum range : use Up and Down arrows to modify the digit and use Left and Right arrows to
go to one digit to an other.
➢Press OK.
➢Go to “Range max” line then press OK.
➢Set the value of the maximum range : use Up and Down arrows to modify the digit and use Left and Right arrows to
go to one digit to an other.
3.4.3 Perform an output diagnostic
This function allows to check on a measurement device (multimeter, regulator or automate) the proper functioning of the
outputs. The transmitter will generate a voltage (between 0 and 10 V) or a current (between 0 and 20 mA) according to the
setting of the type of output.
•For a 0-10 V output signal, the transmitter will generate 0 – 5 or 10 V.
•For a 0-5 V output signal, the transmitter will generate 0 – 2.5 or 5 V.
•For a 4-20 mA output signal, the transmitter will generate 4 – 12 or 20 mA.
•For a 0-20 mA output signal, the transmitter will generate 0 – 10 or 20 mA.
Before carrying o t the o tp t diagnostics, all connections and config rations of the transmitter m st be
enabled, to avoid any damage on the transmitter and the meas rement device !
➢Select a channel for the outputs diagnostic.
OUT1, OUT2, OUT3* or OUT4* indicated on the electronic board above the terminal blocks.
➢
➢Connect a measurement device on the output 1, 2, 3* or 4*.
* : optional outputs
Once the connection to the measurement device is complete, the output diagnostic on several check points can be
performed :
➢Go to “Diagnostic” line then press OK.
According to the type of output, the transmitter displays the outputs it will generate.
➢Use Up and Down arrows to go from one point to generate to an other.
If the deviations are too large (> 0,05V or > 0,05mA) between the signal iss ed and the val e displayed on the
m ltimeter, we recommend that yo ret rn the transmitter to o r factory.
14

3.5. Configure the alarms
This part allows to activate and define the alarms conditions on one or several channels.
One channel m st be config red at least (see chapter 3.3).
If no channel has been config red, the transmitter will indicate that the alarms config ration is
impossible.
➢Press OK.
➢Enter the security code then press OK.
➢Go to “Alarms” line then press OK.
The transmitter displays the 4 alarms : “Alarm ”, “Alarm 2”, “Alarm 3” or “Alarm 4”.
3.5.1 Select the alarm mode
➢Go to the line of the alarm to configure then press OK.
➢Press OK on “Mode” line.
The transmitter displays the different available modes :
•None
•Rising edge
•Falling edge
•onitoring
•Transmitter state
➢Go to the required mode with Up and Down arrows then press OK.
Details of the available alarms modes :
•Rising edge (1 threshold) : the alarm goes off when the measurement exceeds the threshold and stops when it is
below the threshold
•Falling edge (1 threshold) : the alarm goes off when the measurement is below the threshold and stops when it
exceeds the threshold.
•Monitoring (2 thresholds) : the alarm goes off when the measurement is outside the defined low and high
thresholds.
•Transmitter state : the alarm goes off according to defined conditions (high ambient temperature*, memory error...)
*only with the pressure board
15
Rising edge
easurement (m) > Threshold (S) during the time-delay T1 → Alarm
activation
easurement (m) < Threshold (S) - Hysteresis (H) during the time-delay
T2 → Alarm deactivation
Falling edge
easurement (m) < Threshold (S) during time-delay T1 →
Alarm activation.
easurement (m) > Threshold (S) + Hysteresis (H) during
time-delay T2 → Alarm deactivation

When an audible alarm goes off, it is possible to acknowledge it pressing OK key on the transmitter.
3.5.2 Set the alarm in rising or falling edge
The alarm mode “Rising edge” or “Falling edge” is selected.
➢Go to the “Config ration” line then press OK.
➢Press OK on the “Channel” line and select the channel where the alarm will be active.
➢Press OK to validate.
➢Go to the “Threshold” line then press OK.
➢Use Up/Down/Left/Right arrows to enter a threshold then press OK.
➢Go to the “Hysteresis” line then press OK.
➢Use Up/Down/Left/Right arrows to enter the hysteresis value then press OK.
➢Go to the line “Threshold delay” then press OK.
➢Use Up/Down/Left/Right arrows to enter the value of threshold delay in second then press OK (this value is
between 0 and 600 s).
➢Go to the “Hysteresis threshold” then press OK.
➢Use Up/Down/Left/Right arrows to enter the value of the hysteresis delay in second then press OK (this value is
between 0 and 600 s).
➢Go to the “So nd alarm” line then press OK.
➢Select “YES” to activate the audible alarm or “NO” not to activate it then press OK.
3.5.3 Configure the alarm in monitoring mode
The alarm mode “Monitoring” is selected.
➢Go to the “Config ration” line then press OK.
➢Press OK on the “Channel” line and select the channel where the alarm will be active.
➢Go to the “Threshold 1” line then press OK.
➢Use Up/Down/Left/Right arrows to enter the threshold 1 then press OK.
➢Go to the “Threshold 2” line then press OK.
➢Use Up/Down/Left/Right arrows to enter the threshold 2 then press OK.
➢Go to the “Delay 1” line then press OK.
➢Use Up/Down/Left/Right arrows to enter the value of the delay 1 in second then press OK (this value is between 0
and 600 s).
➢Go to the “Delay 2” line then press OK.
➢Use Up/Down/Left/Right arrows to enter the value of the delay 2 in second then press OK (this value is between 0
and 600 s).
➢Go to the “So nd alarm” line then press OK.
➢Select “YES” to activate the audible alarm or “NO” not to activate it then press OK.
3.5.4 Configure the alarm on Transmitter state mode
16
Monitoring
The alarm goes off when the measurement is outside the low and
high thresholds.

The alarm mode “Transmitter state” is selected.
➢Go to the “Config ration” line then press OK.
➢Press OK on the “Condition selection” line.
The transmitter displays the list of alarms conditions :
•Ambient temperature too high*
•Ambient temperature too low*
•Memory error
•Board error*
•Probe 1 error
•Probe 2 error
•Pressure value too high*
➢Go to the alarm condition with Up and Down arrows then press OK to select the condition.
The alarm condition is ticked.
Use the right arrow to go to the 2nd page of the conditions and use the Left arrow to back to the 1st page.
➢Press Esc key when the required conditions are selected.
The transmitter displays a message asking to save.
➢Go to YES then press OK.
➢Then go to the “Audible alarm” line then press OK.
➢Select “YES” to activate the audible alarm or “NO” not to activate it then press OK.
*Only available with the pressure board
The alarm must be completely set to be active (for example, do not keep “---” in front of a threshold, a delay,...).
3.5.5 Configure the relays (optional)
➢Press OK.
➢Enter the security code.
➢Go to the “Relays” line then press OK.
The transmitter displays the 4 relays : “Relay ”, “Relay 2”, “Relay 3” or “Relay 4”.
➢Select the relay to configure then press OK.
➢Press OK on “Selection” line.
➢Select “ON”, “OFF” or “Alarm 1, 2, 3 or 4” with Up and Down arrows then press OK :
•“ON” is selected : the relay is connected. It also allows to perform a test of the relay.
•“OFF” is selected : the relay is deactivated.
•“Alarm 1, 2, 3 or 4” is selected : the activation of the relay is assigned to an alarm previously configured.
➢Go to “Sec rity” line then press OK.
The outputs relays are by default in negative security : the relay is energized during an alarm condition. Via the
keyboard, it is possible to switch the relays in positive security : the relay is de-energized during an alarm
condition or a power cut.
➢Select “Positive” or “Negative” with Up and Down arrows then press OK.
To activate the relays according to the alarms, the alarms must be completely configured.
17

3.6. Configure the probes, boards and normative values
This part allows to set the following parameters :
•for an air velocity and airflow probe : (hotwire or vane) : the type of section, the correction factor, the in air
velocity and the compensation in atmospheric pressure (only available for hotwire probes)
•For a press re board : air velocity and airflow (if the option is activated), the integration in pressure, the purge
mode and autozero duration.
•For a CO2 probe : the compensation in atmospheric pressure,
•For an hygro/temperat re probe : the compensation in atmospheric pressure for the psychrometry calculations.
3.6.1 Configure an air velocity and airflow probe
An air velocity/airflow probe m st be connected.
Select the type of section :
➢Press OK.
➢Enter the security code then press OK.
➢Go to the “Parameters” line then press OK.
➢Go to “Probe 1” or “Probe 2” line (according to the channel of the air velocity/airflow probe) then press OK.
➢Go to “Type of section” line then press OK.
For a rectangular section :
➢Select “Rectangular” with Up and Down keys then press OK.
The transmitter displays the necessary settings to perform for a rectangular section.
➢Press OK on “Unit” line then select “mm” or “inch” whith Up and Down keys.
➢Press OK.
➢Go to “Length” line then press OK.
➢Use Up/Down/Left/Right arrows to enter the length of the section then press OK.
➢Go to “Width” line then press OK.
➢Use Up/Down/Left/Right arrows to enter the width of the section then press OK.
Length and width of the section must be between 1 and 3000 mm.
For a circular section :
➢Select “Circ lar” with Up and Down arrows then press OK.
The transmitter displays the necessary settings to perform for a circular section.
➢Press OK on “Unit” line then select “mm” or “inch” with Up and Down keys.
➢Press OK.
➢Go to “Diameter” line then press OK.
➢Use Up/Down/Left/Right arrows to enter the diameter of the section then press OK.
The diameter of the section must be between 1 and 3000 mm.
Enter a correction factor : this correction coefficient allows to adjust the transmitter according to data in air velocity of the
installation.
How to calc late it ? For example, the air velocity in your section is 17 m/s and the transmitter displays 16.6 m/s. The
coefficient to apply is 17 / 16,6 it means 1.024.
➢Press OK.
➢Enter the security code then press OK.
➢Go to “Parameters” line then press OK.
18

➢Go to “Probe 1” or “Probe 2” line (according to the channel of the air velocity/airflow probe) then press OK.
➢Use Up/Down/Left/Right arrows to enter a correction factor then press OK.
The correction factor must be between 0.2 and 2.
Enter an integration in air velocity : The air velocity measurement element of the transmitter is very sensitive and reacts
to air velocity changes. When making measurements in unstable air systems, the air velocity measurement may fluctuate.
The integration coefficient (from 0 to 9) makes an average of the measurement ; this helps to avoid any excessive
variations and guarantees a stable measurement.
New displayed val e = [((10 - Coef.) x New Value) + (Coef. x Old value)] /10
Example : C310 with hotwire (0-30 m/s) – Current measurement : 2 m/s – New measurement : 8 m/s
The air velocity source being unstable, the user selects a strong integration. Integration : 7. The variation is lower than 10
m/s, it is possible to apply the integration calculation formula.
Next displayed val e : ((2 * 8) + (7 *2 ))/10 = 3 m/s.
➢Press OK.
➢Enter the security code then press OK.
➢Go to the “Parameters” line then press OK.
➢Go to “Probe 1” or “Probe 2” line (according to the channel of the air velocity/airflow probe) then press OK.
➢Go to “Air velocity integration” line then press OK.
➢Use Up and Down arrows to select an air velocity integration then press OK.
The air velocity integration must be between 0 and 9.
3.6.2 Configure a CO2 probe or a temperature/hygrometry probe
A CO2 probe or a temperat re/hygrometry probe m st be connected to the transmitter.
Set the compensation in atmospheric press re
➢Press OK.
➢Enter the security code then press OK.
➢Go to the “Parameters” line then press OK.
➢Go to “Probe 1” or “Probe 2” (according to the channel of the CO2 probe) then press OK.
➢Go to “Compensation” line then press OK.
➢Go to “Atmo pressure” line then select the unit with the Up and Down arrows.
➢Press OK.
➢Enter a value. This value must be between :
•0 and 4000 hPa
•0 and 4000 mbar
•0 and 3000.24 mmHg
•0 and 10 000 m (altitude)
3.6.3 Configure a board
A press re board m st be connected to the transmitter.
For the air velocity/airflow settings, the “Air velocity/airflow” option m st be activated.
Select the meas ring device :
➢Press OK.
➢Enter the security code then press OK.
➢Go to the “Parameters” line then press OK.
19

➢Go to “Board” line then press OK.
➢Go to “Air velocity/airflow” line then press OK.
➢Press OK on “ easure mean” line then use Up and Down arrows to select :
•Pitot tube L (coefficient : 1.0015)
•Pitot tube S (coefficient : 0.84)
•Debimo blades (coefficient : 0.8165)
•le coefficient of the element (free coefficient)
➢Press OK.
Select the type of section : 3 selections are possible : rectangular or circular section or airflow coefficient.
•For the rectangular and circular sections : see chapter 3.6.1
•For the airflow coefficient :
➢Press OK on “Section type” line then select “Flow coeff”.
➢Press OK on “Coefficient” line then enter the coefficient with Up/Down/Left/Right arrows.
➢Press OK.
➢Press OK on “Reference unit” line then select the unit with Up and Down arrows.
➢Press OK.
This airflow coefficient allows to calculate an airflow from the pressure. It is indicated by the manufacturer of the devices
supplied with pressure connections (+ and -). From the square root of the measured pressure (Delta P) and from this
coefficient, you get the airflow. Airflow = CD x
√
Δpressure
The coefficient must be between 0.1 and 9999.9.
The available units for the reference unit are : Pa, mmH2O, inWg and mbar.
Enter a correction factor : this correction coefficient allows to adjust the transmitter according to data in air velocity of the
installation.
How to calc late it ? For example, the air velocity in your section is 17 m/s and the transmitter displays 16.6 m/s. The
coefficient to apply is 17 / 16,6 it means 1.024.
➢Press OK.
➢Enter the security code then press OK.
➢Go to the “Parameters” line then press OK.
➢Go to “Board” line then press OK.
➢Go to “Correction factor” line then press OK.
➢Use Up/Down/Left/Right arrows to enter a correction factor then press OK.
The correction factor must be between 0.2 and 2.
Set the compensation in temperat re and in atmospheric press re : it is possible to modify the compensation
temperat re val e. Indeed, measured air velocity and airflow with a Pitot tube or Debimo blades (or other differential
pressure element) depends on operating temperature. It is required to enter the operating temperat re in order to get
more accurate results. It is possible to enter the value manually or to use a thermocouple K probe for an automatic
temperature compensation.
➢Press OK.
➢Enter the security code then press OK.
➢Go to the “Parameters” line then press OK.
➢Go to “Board” line then press OK.
➢Go to “Compensation” line then press OK.
➢Press OK on “Temperat re” line then select :
•anual entry to enter manually a temperature : press OK then select the unit (°C or °F) then enter a value
between -50 and +50 °C.
•Auto (thermocouple) : the transmitter will take into account the measured temperature by a thermocuple probe
connected to the pressure board.
20
Table of contents
Other Kimo Transmitter manuals

Kimo
Kimo 210-R Series User manual

Kimo
Kimo ATE 310 User manual

Kimo
Kimo CPE 311-S User manual

Kimo
Kimo HM 50 Instruction Manual

Kimo
Kimo TG 100 Instruction Manual

Kimo
Kimo CA 310 User manual

Kimo
Kimo Class 300 User manual

Kimo
Kimo CPE 300 User manual

Kimo
Kimo CP 111 Instruction Manual

Kimo
Kimo Class 200 User manual