5DIM‐3802‐DCI4/2016
DailyCleaningInstructions:KL1

Cleanthefollowingcomponentsatleastoncedaily.Cleanmorefrequentlyifyourpharmacy
dispensesmanydust‐generatingmedicationsorifyouexperiencecountingerrors.

Materials:70%isopropylalcoholormildsoapywater(e.g.,dishsoap),lint‐freeclothor
papertowels.Donotusefacialtissue,cottonballsorcottonswabs.

Beforestarting,turnofftheunitandunplugit.

1.Funnel
A. Removethefunnelfromtheunit.
B. Cleanoutsideandinsidethoroughlywithalcoholormildsoapywater.


2.GlassDetectorWindows
Theunit’scountingsensorsarelocatedbehindtheremovableinsidechannelassembly.
Mostcountingerrorsarecausedbypilldustaccumulatingontheedgesandcornersof
thewindowsonthechannelassembly.Cleanthewindowswell.

A. Unscrewthetworoundthumbscrewsontopoftheunit.Removethe
squareguard.
B. Liftouttheentireinsidechannelassembly.Donotremovetheindividualglasspanes
attachedtotheinsidechannelassembly.
C. Wipeoffexcesspilldustontheinsideofthefourcleardetectorwindowswith
adrycloth.
D. Wetacleanclothwithalcoholormildsoapywaterandcleantheinsideofthefour
glasswindowsthoroughlytoremoveallremainingdustandresidue.Wipeback‐and‐
forth,notup‐and‐down.Donotpresshardasthismaybreaktheglasswindows.Do
notremovetheindividualglasspanes.
E. Foldtheclothtoformanedgetocleananyedgedustcanaccumulate,especiallycorners.
F. Wipetheglasswindowsdrywithacleancloth.Holdthewindowsuptoalighttovisually
inspecttheglassforanydust,fingerprints,orstreaks.Re‐cleanifnecessary.Wipedry.
G. LookdowninsidetheKL1deviceandvisuallyinspectthefourinsideglasswindows
fordustorstreaks.Ifnecessary,cleanwithwaterandwipedry.
H. Replacetheinsidechannelassemblyandsquareguard.Tightentheroundthumb
screwsuntilsnug.Donotover‐tightenthethumbscrews.
StepA–Remove
SquareGuard
StepsC‐F,Inside
ChannelAssembly:
Cleaninsideand
outsideof4glass
windowswithout
removingor
damagingthepanes
5DIM‐3802‐DCI4/2016

I. Ifdesired,theglasswindowholdercanberemovedfromtheinsidechannel
assemblyforamorethoroughcleaningandvisualinspection.Gentlypushdownon
thesquareslidingtube,whichspreadsthetwoblackclipsholdingtheglasswindow
holder.Takespecialcarenottobreaktheblackclips.

3.CountingTray
Cleanthoroughlywithmildsoapywater.Donotusealcohol.


4.TraySensor
LocatedontheKL1behindthecountingtraywhenthetrayisfullyinserted.Thissensor
tellstheunitthatthetrayisproperlyseated.Withthetrayremoved,cleanthesensor
windowwithwater.Wipedrytoremovestreaks.




Noteabouttheuseofmildsoapywaterversusisopropylalcohol:Mildsoapywaterandalcoholboth
provideappropriatesanitarycleansing.WaterleavesfewerstreaksthanalcoholonaKirbyLester
device'sinsideglasssensorwindows.Ifstreaksarefoundaftercleaning,usejustwaterandwipedry.
Ingeneral,prolongedalcoholexposurecausesfasterwear‐and‐tearonplastic.

5.CleaningandTroubleshooting

ProblemorErrorCodeProbableCause PossibleSolution(s)
E‐2,E‐4,E‐9,O‐Sorunitcountsby
itself
Tabletdustorresidueonglassdetector
windows,orreplacecrackedglassdetector
window.
Cleanglassdetectorwindowsthoroughly.Or
replacecrackedglassdetectorwindow.
OvercountsTabletfragments,ortabletdustonglass
detectorwindows.
Checkcountedmedicationsfortablet
fragments.Cleanglassdetectorwindows
thoroughly.
UndercountsPouringtoofast,orpouringtabletsdirectly
intothefunnelhole.
Slowpouringspeedsothe“PourSlower”
indicatordoesnotactivate.
Displayreads"‐‐‐‐"insteadof
"0000"withtraypushedin
Traynotfullyinserted.Traysensorwindow
isdirty.
Cleantraysensorwindow.Fullyinserttray.

6.QuestionsorProblems
PleasecontactKirbyLesterServiceat800.243.2465or[email protected].


StepI–Visually
InspectWindows
TraySensor

Other Kirby Lester Laboratory Equipment manuals

Kirby Lester KL20 Configuration guide

Kirby Lester

Kirby Lester KL20 Configuration guide

Kirby Lester KL1 Configuration guide

Kirby Lester

Kirby Lester KL1 Configuration guide

Kirby Lester KL15 Series Configuration guide

Kirby Lester

Kirby Lester KL15 Series Configuration guide

Kirby Lester KL16 Series Configuration guide

Kirby Lester

Kirby Lester KL16 Series Configuration guide

Popular Laboratory Equipment manuals by other brands

Ele 38-1000 operating instructions

Ele

Ele 38-1000 operating instructions

Metrohm 940 Professional IC Vario ONE/ChS/LPG manual

Metrohm

Metrohm 940 Professional IC Vario ONE/ChS/LPG manual

IKA Mini MR standard operating instructions

IKA

IKA Mini MR standard operating instructions

SteriFeed S90 Installation and maintenance manual

SteriFeed

SteriFeed S90 Installation and maintenance manual

PASCO OS-9255A instruction manual

PASCO

PASCO OS-9255A instruction manual

Tobii Pro Fusion user manual

Tobii Pro

Tobii Pro Fusion user manual

Macherey-Nagel NANOCOLOR Advance quick start guide

Macherey-Nagel

Macherey-Nagel NANOCOLOR Advance quick start guide

3M Attest 290 Operator's manual

3M

3M Attest 290 Operator's manual

Agilent Technologies Varian 325-MS LC/MS Pre-installation instructions

Agilent Technologies

Agilent Technologies Varian 325-MS LC/MS Pre-installation instructions

SI Analytics TITRONIC 300 operating manual

SI Analytics

SI Analytics TITRONIC 300 operating manual

Thermo Scientific Thermo Scientific HyPerforma quick start guide

Thermo Scientific

Thermo Scientific Thermo Scientific HyPerforma quick start guide

Wasatch Photonics Cobra OCT Getting started guide

Wasatch Photonics

Wasatch Photonics Cobra OCT Getting started guide

Westermo PII-2G Quick installation guide

Westermo

Westermo PII-2G Quick installation guide

GE ImageQuant LAS 500 Site preparation guide

GE

GE ImageQuant LAS 500 Site preparation guide

Thermo Scientific Labofuge 200 Operating lnstructions

Thermo Scientific

Thermo Scientific Labofuge 200 Operating lnstructions

BioLAB BCFL-203 Operation manual

BioLAB

BioLAB BCFL-203 Operation manual

EDG Fieldscope quick start guide

EDG

EDG Fieldscope quick start guide

Fortex Etching Station 1 manual

Fortex

Fortex Etching Station 1 manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.