5DIM‐3802‐DCI4/2016
I. Ifdesired,theglasswindowholdercanberemovedfromtheinsidechannel
assemblyforamorethoroughcleaningandvisualinspection.Gentlypushdownon
thesquareslidingtube,whichspreadsthetwoblackclipsholdingtheglasswindow
holder.Takespecialcarenottobreaktheblackclips.
3.CountingTray
Cleanthoroughlywithmildsoapywater.Donotusealcohol.
4.TraySensor
LocatedontheKL1behindthecountingtraywhenthetrayisfullyinserted.Thissensor
tellstheunitthatthetrayisproperlyseated.Withthetrayremoved,cleanthesensor
windowwithwater.Wipedrytoremovestreaks.
Noteabouttheuseofmildsoapywaterversusisopropylalcohol:Mildsoapywaterandalcoholboth
provideappropriatesanitarycleansing.WaterleavesfewerstreaksthanalcoholonaKirbyLester
device'sinsideglasssensorwindows.Ifstreaksarefoundaftercleaning,usejustwaterandwipedry.
Ingeneral,prolongedalcoholexposurecausesfasterwear‐and‐tearonplastic.
5.CleaningandTroubleshooting
ProblemorErrorCodeProbableCause PossibleSolution(s)
E‐2,E‐4,E‐9,O‐Sorunitcountsby
itself
Tabletdustorresidueonglassdetector
windows,orreplacecrackedglassdetector
window.
Cleanglassdetectorwindowsthoroughly.Or
replacecrackedglassdetectorwindow.
OvercountsTabletfragments,ortabletdustonglass
detectorwindows.
Checkcountedmedicationsfortablet
fragments.Cleanglassdetectorwindows
thoroughly.
UndercountsPouringtoofast,orpouringtabletsdirectly
intothefunnelhole.
Slowpouringspeedsothe“PourSlower”
indicatordoesnotactivate.
Displayreads"‐‐‐‐"insteadof
"0000"withtraypushedin
Traynotfullyinserted.Traysensorwindow
isdirty.
Cleantraysensorwindow.Fullyinserttray.
6.QuestionsorProblems
StepI–Visually
InspectWindows
TraySensor