Kirby Lester KL1 Configuration guide

5DIM‐3802‐DCI5/2016
DailyCleaningInstructions:KL1

Cleanthefollowingcomponentsatleastoncedaily.Cleanmorefrequentlyifyourpharmacy
dispensesmanydust‐generatingmedicationsorifyouexperiencecountingerrors.

Materials:70%isopropylalcoholormildsoapywater(e.g.,dishsoap),lint‐freeclothor
papertowels.Donotusefacialtissue,cottonballsorcottonswabs.

Beforestarting,turnofftheunitandunplugit.

1.Funnel
A. Removethefunnelfromtheunit.
B. Cleanoutsideandinsidethoroughlywithalcoholormildsoapywater.


2.GlassDetectorWindows
Theunit’scountingsensorsarelocatedbehindtheremovableinsidechannelassembly.
Mostcountingerrorsarecausedbypilldustaccumulatingontheedgesandcornersof
theglasswindowsontheinsidechannelassembly.Cleanthewindowswell.

A. Unscrewthetworoundthumbscrewsontopoftheunit.Removethe
squareguard.
B. Liftouttheentireinsidechannelassembly.
C. Wipeoffexcesspilldustontheinsideofthefourglasswindowswithadrycloth.
D. Wetacleanclothwithalcoholormildsoapywaterandcleantheinsideofthefour
glasswindowsthoroughlytoremoveallremainingdustandresidue.Wipeback‐and‐
forth,notup‐and‐down.Donotpresshardasthismaybreaktheglasswindows.Do
notremovetheindividualglasspanesunlessServiceinstructsyouto.
E. Foldtheclothtoforma“point”tocarefullycleananyedgeorcornerwheredustcan
accumulate.Payspecialattentiontocorners.
F. Wipetheglasswindowsdrywithacleancloth.Holdtheinsidechannelassemblyup
toalighttovisuallyinspecttheglasswindowsforanydust,fingerprints,orstreaks.
Re‐cleanifnecessary.Wipedry.
G. Lookdowninsidethedeviceandvisuallyinspectthefourinsideglasswindows
fordustorstreaks.Ifnecessary,cleanwithwaterandwipedry.
H. Replacetheinsidechannelassemblyandsquareguard.Tightentheroundthumb
screwsuntilsnug.Donotover‐tightenthethumbscrews.
Step1A–Remove
SquareGuard
Steps2C‐F,Inside
ChannelAssembly:
Cleaninsideand
outsideof4glass
windowswithout
removingor
damagingthepanes
5DIM‐3802‐DCI5/2016

I. Ifdesired,theglasswindowholdercanberemovedfromtheinsidechannel
assemblyforamorethoroughcleaningandvisualinspection.Gentlypushdownon
thesquareslidingring,whichspreadsthetwoblackclipsholdingtheglasswindow
holder.Takespecialcarenottobreaktheblackclips.

3.CountingTray
Cleanthoroughlywithmildsoapywater.Donotusealcohol.


4.TraySensor
LocatedontheKL1behindthecountingtraywhenthetrayisfullyinserted.Thissensor
tellstheunitthatthetrayisproperlyseated.Withthetrayremoved,cleanthesensor
windowwithwater.Wipedrytoremovestreaks.




Noteabouttheuseofmildsoapywaterversusisopropylalcohol:Mildsoapywaterandalcohol
bothprovideappropriatesanitarycleansing.WaterleavesfewerstreaksthanalcoholonaKirby
Lesterdevice'sinsideglasssensorwindows.Ifstreaksarefoundaftercleaning,usejustwaterand
wipedry.Ingeneral,prolongedalcoholexposurecausesfasterwear‐and‐tearonplastic.

5.CleaningandTroubleshooting

ProblemorErrorCodeProbableCause PossibleSolution(s)
E‐2,E‐4,E‐9,O‐Sorunitcountsby
itself
Tabletdustorresidueonglasswindows,or
replacecrackedglasswindow.
Cleanglasswindowsthoroughly.Orreplace
crackedglasswindow.
Overcounts(morecountedthan
poured)
Tabletfragments,ortabletdustonglass
windows.
Checkcountedmedicationsfortablet
fragments.Cleanglasswindowsthoroughly.
Undercounts(fewercountedthan
poured)
Pouringtoofast,orpouringtabletsdirectly
intothefunnelhole.
Slowpouringspeedsothe“PourSlower”
indicatordoesnotactivate.
Displayflashes"‐‐‐‐"insteadof
"0000"withtraypushedout
Traynotfullyinserted.Traysensorwindow
isdirty.
Cleantraysensorwindow.Fullyinserttray.

6.QuestionsorProblems
PleasecontactKirbyLesterServiceat800.243.2465or[email protected].


Step2I–Visually
InspectGlass
Windows
Step4–
TraySensor

Other Kirby Lester Laboratory Equipment manuals

Kirby Lester KL16 Series Configuration guide

Kirby Lester

Kirby Lester KL16 Series Configuration guide

Kirby Lester KL1 Configuration guide

Kirby Lester

Kirby Lester KL1 Configuration guide

Kirby Lester KL15 Series Configuration guide

Kirby Lester

Kirby Lester KL15 Series Configuration guide

Kirby Lester KL20 Configuration guide

Kirby Lester

Kirby Lester KL20 Configuration guide

Popular Laboratory Equipment manuals by other brands

Wheaton CRIMPenstein Quick start card

Wheaton

Wheaton CRIMPenstein Quick start card

Bruker NMR AVANCE NEO Systems user manual

Bruker

Bruker NMR AVANCE NEO Systems user manual

BioLAB BIFL-207 Operation manual

BioLAB

BioLAB BIFL-207 Operation manual

Wassermann Wapo-Ex 8 plus manual

Wassermann

Wassermann Wapo-Ex 8 plus manual

All American 25X-120 operating instructions

All American

All American 25X-120 operating instructions

VS Security Products SV91M Operating and maintenance manual

VS Security Products

VS Security Products SV91M Operating and maintenance manual

DIGISYSTEM DSB-500D instruction manual

DIGISYSTEM

DIGISYSTEM DSB-500D instruction manual

Showa Denko Shodex Asahipak NH2P-50 10E Operation manual

Showa Denko

Showa Denko Shodex Asahipak NH2P-50 10E Operation manual

Showa Denko Shodex IC NI-424 Operation manual

Showa Denko

Showa Denko Shodex IC NI-424 Operation manual

Promega Maxwell CSC 48 operating manual

Promega

Promega Maxwell CSC 48 operating manual

Hitachi CR22N instruction manual

Hitachi

Hitachi CR22N instruction manual

PASCO Brolight SE-9629 Product guide

PASCO

PASCO Brolight SE-9629 Product guide

Flexim FLUXUS G722ST-HT Operating	 instruction

Flexim

Flexim FLUXUS G722ST-HT Operating instruction

Thermo Scientific 3767A Installation and operation manual

Thermo Scientific

Thermo Scientific 3767A Installation and operation manual

Allsheng Auto-Pure Mini Operation manual

Allsheng

Allsheng Auto-Pure Mini Operation manual

Thermo Scientific mySPIN 12 instruction manual

Thermo Scientific

Thermo Scientific mySPIN 12 instruction manual

Beckman Coulter Avanti J-26S XPI Instructions for use

Beckman Coulter

Beckman Coulter Avanti J-26S XPI Instructions for use

IKA C 7010 operating instructions

IKA

IKA C 7010 operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.