5DIM‐2904‐DCI3/2015
DailyCleaningInstructions:KL20
Cleanthefollowingcomponentsatleastoncedaily.Cleanmorefrequentlyifyourpharmacy
dispensesmanydust‐generatingmedicationsorifyouexperiencecountingerrors.Followthe
instructionalvideodirectionsshownonthescreenandinthisguidetocleantheKL20.Yourwarranty
doesnotcoverdamageduetomisuseorabuse.
Materials:Distilledwateror70%isopropylalcohol,lint‐freeclothorpapertowels,and
computerscreencleaner.Donotusefacialtissue,cottonballsorcottonswabs.
Beforestarting,turnoffandunplugtheunit.
1.TopFunnel
A. Removethetopfunnelandchannelfromtheunit.
B. Cleanoutsideandinsidethoroughlywithmildsoapywateroralcohol.
2.GlassDetectorWindows
Theunit’scountingsensorsarelocatedbehindfourglasswindowsinsidethe
squarefunnelhole.Mostcountingerrorsarecausedbypilldustaccumulating
ontheedgesandcornersofthewindows.Cleanthewindowswell.
A. Wipeoffexcesspilldustontheglasswindowswithadrylint‐freecloth.
B. Wetacleanclothwithwateroralcoholandcleantheinsideofthefourglass
windowsthoroughlytoremoveallremainingdustandresidue.Carefullywipe
back‐and‐forth,notup‐and‐down.Donotpresshardasthismaybreaktheglass
windows.
C. Wipeglasswindowsdrywithcleanclothtoremovestreaks.
3.CountingTray
A. Carefullyremovetheclearplastictraycover.Cleanallpartsthoroughlywithawater‐
dampenedcloth(notalcohol).Wipedry.Replacethetraycover.
B. Turnthetrayover,andwipethesilversensorstripwithawater‐dampenedcloth
(notalcohol).
4.TraySensor
A. (Locatedonthebottomoftheunitinbackofthecountingtray.)Thissensordetects
thetray’ssilverstrip,tellingtheunitthatthetrayisproperlyseated.
B. Withthecountingtrayremoved,tiptheunitcarefullytoaccessthetraysensor.
Cleanwithawater‐dampenedcloth(notalcohol).
5.CustomizeOnscreenCleaningReminders
Youcancustomizetheautomaticcleaningreminderfunctionofyourunit.SeetheKL20
operatingmanualforfullinstructionsonmakingtheseadjustments:
Turnthecleaningchecklistoff
Changethefrequencyofthecleaningreminder(default150scripts)
Addmedicationsthatautomaticallytriggeracleaningreminder(e.g.,sulfa)
TraySensor
2‐piece
Funnel
4Detector
Windows