Page 19 BGN K01/0908
and float with hard objects (see Removing/installing the
cone/float).
- The switching points of the limit transducers are
adjustable. To do this, remove the indicator cover,
unfasten the contact point indicator located on the scale
and readjust it. After the adjustment, reattach the bolts
of the contact point indicator. Reinstall and tighten the
indicator cover.
- The calibration of the KINAX signal output is firmly set
and not adjustable. Do not adjust the potentiometer of
the transmitter.
- The parameterization of the ES is possible and is done
via HART. Please refer to the separate Operating
Instructions for the ES.
- The gas and viscous damping cylinders can be
checked for dirt (see Installing/removing the damping
sets).
17 Installing and removing cone,
float, damping sets/spring stop
To remove the float (cone/float), detach the device from the
pipe. Then clamp the device horizontally in a vise, making
sure that fitting is not damaged.
Device with measuring ring: By using a suitable tool on
the guide star secure the float from contortion from above in
the fitting, and remove the lower guide star's retaining screw
identifiable from below in the fitting and the guide star. The
float can then be taken out from above.
Device with cone: Unscrew the cone with the float using a
suitable tool through the lower cone thread, and take out the
float after removing the top cone stabilizer.
Important! Avoid damaging the float/measuring ring and
cone. To change the measuring range, the float can be
replaced once the measured variable reaches 5-50 l/h of
water. In S-type devices the cone and float can be replaced
at up to 40 l/h water.
The gas and viscous damping cylinders as well as the spring
stop for the float can be removed from the top after the
device has been removed. Proceed in reverse order when
reinstalling thesed parts.
18 Trouble shooting
- Indicator window clouds over: Water in the indicator
unit.
- Indicator cover is not tight enough: Adjust the cover
seal, tighten the cover.
- Window is opaque: Corrosive atmosphere, ventilate.
- Window ices over due to cold and damp
atmosphere: The device can be equipped at the
factory with an air/nitrogen flush.
- Window ices over due to very cold medium and
damp atmosphere: The device can be equipped at
the factory with a pulled-forward indicator unit.
- Device shows incorrect values: Compare process
data, density, viscosity, temperature and pressure with
the values on the scale. If they deviate, convert the
scale values using the VDE/VDI 3513 standard; for the
signal output 2 option, carry out reparameterization.
- Pointer does not react in spite of varying flow: The
pointer may have gotten stuck; remove the cover and
move the pointer; if the pointer can be moved easily,
the float cannot move. If the pointer is unable to move
further, send the device to the head office for servicing.
- The float is stuck at one place due to dirt:
Disassemble the device. If necessary, dismantle and
clean the float. Install a magnetic filter if there are
magnetic contaminants.
- Float is stuck at one place due to deformation of
the measuring ring and float guide because the
float has been pushed upwards: Send the device to
the head office for servicing, replace the fitting, equip
with additional spring stop.
- Gas/viscous damping sets have gotten stuck due to
dirt: Disassemble the device, pull the gas/viscous
damping cylinder to the top and clean it. If the parts are
damaged or worn, send the device to the head office for
servicing. Install precautionary features such as filters
to protect against dirt.
- Scale pointer pulsates: Increase the supply pressure
when measuring gas. Alternatively, mount an aperture
plate behind the device and check whether the valve is
mounted behind the device. Select the device with the
smallest possible pressure loss. Adjust to the desired
flow rate, run at 100 %. Select a short distance
between the flowmeter and the downstream throttle
point (valve). When dealing with liquids, prevent
pulsation of the product by using a volume receptacle.
If necessary, have Kobold Messring install a viscous
damping set or a double eddy-current damping set.
- Electrical equipment are not functioning: Check the
auxiliary power. Are suitable power supply equipment
connected, have the terminals been selected correctly,
has the parameterization carried out correctly?
19 Returning devices for repair and
service
Note: In accordance with the applicable German waste
disposal legislation, the owner/client is responsible for the
disposal of special waste and hazardous materials.
Consequently, all devices sent to us for repair must be free
of any hazardous materials. This also applies to possible
hollow spaces and fissures in the devices. If repair is
necessary, confirm the above-mentioned item in writing
(please use the form in the Appendix).
If hazardous materials remain in or on the device after it has
been returned, Kobold Messring shall be authorized to
remove them at the client’s expense without further inquiry.
20 Replacement parts
The following parts can be ordered as replacement parts:
1 ) Indicator cover with
window/gasket/screws
2 ) Scale with standard scaling
3 ) Pointer
4 ) Limit value indicator
5 ) Pointer stop
6 ) Float with guides and safety screw
7 ) Spring stop
8 ) Complete gas damping set with float
9 ) Complete viscous damping set with float
10) Screwed cone set with float for
small measuring range up to 40 l/h
11 ) Limit value initiatior