4. Data・Address list
Category Address Name Description of set values
00001 (P-01) Low cut
0: Instantaneous flow rates below 0.5% F.S. are indicated. Negative values are
also indicated.
1: Instantaneous flow rates below 0.5% F.S. are indicated as 0. Negative values
00002 (P-04)Integrated flow rate hold
0: The integrated flow rate value is reset when the power is turned off.
1: The integrated flow rate value is held after the power is turned off. The
integrated flow rate value hold cycle is 1 minute.
Input Status 10001 Flow rate unit 0: cc, 1: L
30001 Full scale flow rate [significand] Mantissa portion of the full scale flow rate currently set for operation 0001 ~
30002 Flow rate decimal point position
[number of decimal places] 0: none, 1: digit, 2: digit, 3: digit
30003 Instantaneous flow rate
[significand] -9999 ~9999 (As per the specifications)
40001 (P-30)Applicable gas 0: CF value, 1: N2(nitrogen), 2: Air, 3: H2(hydrogen), 4: He(helium), 5:
Ar(argon), 6: O2(oxygen), 7: CO2(carbon dioxide), 8: CH4(methane)
40002 (P-31)CF value 0500 ~1500 (0.500 ~1.500)
40003 (P-00)Reference condition on
0: 20℃(1atm), 1: 0℃(1atm), 2: 25℃(1atm)
40004 (P-02)PV filter
0: No
1: Moving average of sampling 2 times
2: Moving average of sampling 4 times
3: Moving average of sampling 8 times
4:Moving average of sampling 16 times
5: Moving average of sampling 32 times
40005 (P-03) Instantaneous flow rate
0: 25(ms), 1: 50(ms), 2: 100(ms), 3: 200(ms), 4: 500(ms), 5: 1(sec)
40006 Integrated flow rate reach preset
0000 ~9999
40007 Integrated flow rate reach preset
0000 ~9999 (Mantissa portion not including the decimal point)
40008 (P-10)Event output 1
10: Instantaneous flow rate upper limit : ON
11: Instantaneous flow rate upper limit : OFF
20: Instantaneous flow rate lower limit : ON
21: Instantaneous flow rate lower limit : OFF
30: Instantaneous flow rate upper/lower limit : ON
31: Instantaneous flow rate upper/lower limit : OFF
40: Integrated flow rate reach : ON
41: Integrated flow rate reach : OFF
50: Integrating pulse : ON
51: Integrating pulse : OFF
※
ON: Normal output, OFF: Inverted output
40009 (P-11)Instantaneous flow rate
upper limit preset setting
0000 ~9999
40010 (P-12)Instantaneous flow rate
lower limit preset setting
0000 ~9999
40011 (P-13)Dead time 00 ~30(sec)
40012 (P-14)Integrated flow rate reach
preset setting upper 4 digits 0000 ~9999
40013 (P-15)Integrated flow rate reach
preset setting lower 4 digits 0000 ~9999
Coil
Input
Register
Holding
Register