Introduction toGPIBOperation— 2756PProgrammers
TheREMOTEindicatorisnotlitwhentheinstrument
isunderlocal,operatorcontrol.Whileunderlocalcon-
trol,theinstrumentdoesnotexecuteGPIBmessages
thatwouldconflictwithfront-panelcontrols,anditdoes
notaccepttheCURVEinputcommand.
Whentheinstrumentisunderremotecontroland
RESETTOLOCALispressed,localcontrolisrestoredto
theoperatorunlessthecontrollerpreventsthiswiththe
locallockoutmessage.Programmablefunctionsdonot
changewhenswitchingfromremotetolocalcontrol
exceptasnecessarytomatchthesettingsoffront-panel
controlsforTIME/DIV,MINRFATTENdB,and
PEAK/AVERAGE.
Thespectrumanalyzerflashestheinstrumentname,
firmwareversionnumber,theGPIB'address,andthe
macrostatusmessageonthecrtwhenRESETTO
LOCALispressed.Ifthereisnomacrorunning,themes-
sagewillbeMACRO—OFF(#)where# indicatesthe
menulocationofthelastmacrothatwasrun.Ifthereisa
macrorunning,themessagewillbeMACRO-RUN(#)
where# indicatesthemenulocationofthemacrothatis
running.Ifthereisa macroinprocessbutstopped,the
messagewillbeMACRO-STOP(#)where# indicates
themenulocationofthemacrothatisstopped.Ifa
macroisbeingstoredinmemory,themessagewillbe
MACRO—STORE(#)where# indicatesthemenulocation
wherethemacroisbeingstored.
PLOT
Pressthispushbuttonwhenthespectrumanalyzeris
inthetalk-onlymodefortheinstrumenttosendthe
appropriatecommandsovertheGPIBtoa plotter,which
mustbeinthelisten-onlymode,connectedtothebus
(seetheTALKONLY,LISTENONLYswitchdescriptions
laterinthissection).Thespectrumanalyzerdisplay
(waveform,marker(s),graticule,andcrtreadout)canbe
recreatedona TEKTRONIX4662Option01or4662
Option31InteractiveDigitalPlotter(ora 4663inthe
4662emulationmode);ora Hewlett-PackardHP7470Aor
HP7475A,HP7580B,HP7585B,orHP7586B;ora Gould
6310or6320plotter.Selectplottertypewith<SHIFT>
SELECTTYPE,describedinSection4 ofthismanual.A
buscontrollerisnotrequired.
<SHIFT>SRQ
ThisSRQ(servicerequest)pushbuttonsequence
getsthecontroller'sattentionsoitwilllisten/respondto
thespectrumanalyzer.Forexample,ifthecontrollerput
instructionsonthescreen,intheTEXTLONGmode,to
setuptestequipment,etc.,thelastlineoftheinstruc-
tionsmightsayTRESS<SHIFT>SRQWHENREADY".
Thiswouldinstructthecontrollertogoontothenext
step.AnSRQwillonlybeissuedifRQSison.
ADDRESSED
Thisindicatorislitwhenthespectrumanalyzeris
addressedtolistenortalk.
GPIBFunctionReadout
Asinglecharacterappearsinthelowercrtreadout
whenthespectrumanalyzeristalking(T),orlistening(L);
seeFigure1-2.Twocharacterswillappearinthisloca-
tioniftheinstrumentistalkingorlisteningandalso
requestingservice(S),oriftheinstrumentisinboththe
talk-onlyandlisten-onlymodes.
Figure1-2.StatusofactiveOPIBfunctions.
SettingtheGPIBADDRESSSwitches
Therear-panelGPIBADDRESSswitchesshownin
Figure1-3setthevalueoftheGPIBaddress(referto
Table1-1).Theinstrument'sprimaryaddress(0through
31)isthevalueofthelowerfivebits,whicharelabeled4
through8 inFigure1-3.Theseswitchesarereadeach
timepoweristurnedontotheinstrumentandagaineach
timetheRESETTOLOCALorPLOTpushbuttonis
pressed.
1-2