T.O. 33K3-4-3760-1
10
4.3.3 Set the Oscilloscope Calibrator for a 600 mV p-p leveled sine wave at 50 kHz into 50 . Set the OUTPUT to
ON.
4.3.4 Adjust TI Trigger Level as necessary for a stable display.
4.3.5 Adjust the Oscilloscope Calibrator output controls for a TI P1:pkpk(C1) indication of as close as possible to
600 mV.
4.3.6 Press TI Clear Sweeps and allow at least 100 sweeps as indicated on the TI display.
4.3.7 Set the Oscilloscope Calibrator Frequency to 400 MHz.
4.3.8 Adjust the TI Horizontal controls as necessary for several cycles of display. Touch TI Timebase, Horizontal
Setup, set the Sampling Mode to RIS, then touch Close.
4.3.9 Press TI Clear Sweeps and allow at least 100 sweeps as indicated on the TI display.
4.3.10 Verify the TI P1:pkpk(C1) average indication is 425 mV.
4.3.11 Set the Oscilloscope Calibrator OUTPUT to OFF.
4.3.12 Disconnect the Oscilloscope Calibrator from the TI 1 connector and connect to the TI 2 connector.
4.3.13 Press TI Vertical 1, then Vertical 2to turn C1 off and C2 on.
4.3.14 Touch TI Timebase, Horizontal Setup, RealTime, then set the Time/Division to 20.0 µs.
4.3.15 Touch TI Trigger, Trigger Setup, Source, C2,then Close.
4.3.16 Repeat steps 4.3.3 through 4.3.11 for TI C2, using C2 controls, settings and readouts.
4.3.17 Disconnect the Oscilloscope Calibrator from the TI 2 connector and connect to the TI 3 connector.
4.3.18 Press TI Vertical 2, then Vertical 3to turn C2 off and C3 on.
4.3.19 Touch TI Timebase, Horizontal Setup, RealTime, then set the Time/Division to 20.0 µs.
4.3.20 Touch TI Trigger, Trigger Setup, Source, C3, then Close.
4.3.21 Repeat steps 4.3.3 through 4.3.11 for TI C3, using C3 controls, settings and readouts.
4.3.22 Disconnect the Oscilloscope Calibrator from the TI 3 connector and connect to the TI 4 connector.
4.3.23 Press TI Vertical 3, then Vertical 4to turn C3 off and C4 on.
4.3.24 Touch TI Timebase, Horizontal Setup, RealTime, then set the Time/Division to 20.0 µs.
4.3.25 Touch TI Trigger, Trigger Setup, Source, C4, then Close.
4.3.26 Repeat steps 4.3.3 through 4.3.11 for TI C4, using C4 controls, settings and readouts.
4.3.27 Disconnect the test setup.