LEGRAND Tynetec ZXT426 User manual

DocNo.FM0460issueJPage1
169MHzTelecareDevices
TemperatureExtremeSensor

__________________________________________________________________________________________________________


CompatiblewithReachat‐homealarms,Adventxtwardencall,
AltecResponseandTouchsafeProNursecallsystems
Solidstatetemperaturesensingelement
Loweralarmtemperature:5oC+/‐2.5oCfor30/60minutes
Upperalarmtemperature:30oC+/‐2.5oCfor5/10minutes
Rateofrisealarmtemperature:8oCperminute
Userreplaceable3.6VAALithiumbattery
Lowbatterywarning
Testbutton
Dimensions:84x84x36mm(HxWxD)
Weight:102grams
ProductCode:ZXT426


WhatareTemperatureExtremeSensorsusedfor?

Thetemperatureextremesensorwillprovideearlywarningofextremeloworhighhouseholdtemperatureswhichcouldresultin
theoccupiersufferinghypothermiaordehydration.Inaddition,thebuildingwillalsobeprotectedagainstburstpipesorpotential
firehazards.

WhereshouldTemperatureExtremeSensorsbefitted?

Thetemperatureextremesensorshouldbefittedinthehallwayorlivingroom.
Itshouldbewallmountedataround1.4metresabovefloorlevelandawayfromanydraughtsorheatsources.

TypicalToolsRequired

Adrillandscrewdriverarenotrequirediftheselfadhesivepadisused.
Alwaysmakesurethecorrectdrillbitsandscrewdriversareused.

Installation–UsingScrewFixings

1. Removethetopbypullingapartfromthebacksection.

2. Drillthebacksectionasrequiredandfixtothewallusing
thescrews&plugsprovided.

3. SettheDILswitches&links,fitthebatteryandtestas
describedonthefollowingpagebeforerefittingthetopsection.


Installation–UsingSelfAdhesivePads

1.Removethetopbypullingapartfromthebacksection.

2. Selectthebestlocationwherethewallisclean,dryandeven.

3. Sticktheadhesivepadtotherearfaceofthebacksectionand
pressfirmlytothewallensuringitisstucksecurely.

4. SettheDILswitchesandlinks,fitthebatteryandtestas
describedonthefollowingpagebeforerefittingthetop.












PULLAPART
TOREMOVE
TOP
SECTION
FIXBACK
SECTIONTO
WALL
Tynetec andAidcallare businessunitsofLegrandElectricLtd
Unit10CowleyRoad,BlythRiversideBusinessPark,Blyth
Northumberland,NE245TF.Tel:01670352371Fax:01670362807
Email:[email protected]Web:www.tynetec.co.uk
DocNo.FM0460issueJPage2
DILSwitchSettings

SetDILswitch1ONforMode1;
LowTemperatureAlarm=lessthan5oC+/‐2.5oCfor30minutes
HighTemperatureAlarm=greaterthan30oC+/‐2.5oCfor5minutes

SetDILswitch1OFFforMode2;
LowTemperatureAlarm=lessthan5oC+/‐2.5oCfor60minutes
HighTemperatureAlarm=greaterthan30oC+/‐2.5oCfor10minutes

SetDILswitch2OFFfor169MHztransmitterfrequency


LinkSettings

LK1–notused
SetLK2&LK3bothinposition“B”


FittingtheBattery

Onlyusethe3.7VAALithiumbatterysupplied.
Takecaretoensurethebatteryisfittedcorrectly,fitwiththered“+”endtotherighthandsideasshownabove.


Testing&RegisteringtheTemperatureExtremeSensor

TheTESTbuttonmustbepressedwhenlearningthetemperatureextreme
sensorontothewardencallsystem,at‐homealarmorNursecallsystem. 

Seethefollowingpagesforhowtoputeachofthese
productsinto“LearnMode”.

PressandholdtheTESTbuttonforabout3secondsuntil
theREDlightstartstoflash.

After9flashestherewillbea10secondpausefollowedby
another9flashes(thesensortransmitsahightemperature
signalfollowedbyalowtemperaturesignal).

RemovethebatterytoEXITTESTMODE
LeaveforafewsecondsthenrefitforNORMALOPERATION.

Therangeofallradiodevicescanbeaffectedbytheworkingenvironment‐alwaystakecareduringinstallationand
performseveraltestcalls.


Cleaning&Care

Temperatureextremesensorsshouldbecleanedperiodically.
Usethenozzleattachmentonavacuumcleanertoremovedust,insectsandcobwebsfromthecoverslots.
Thecasecanbewipedcleanwithadampcloth,neverusedetergentsorpolishetc.



Decorating

Neverpaintthetemperatureextremesensoritself.
Itisnotsusceptibletocontaminationfrompaintfumesthereforeno
specialprecautionsneedtobetaken.


BatteryLife

Thetemperatureextremesensorbatteryischeckedonceaday,ifthevoltagefallsandstaysbelowapresetlevelfor7consecutive
daysalowbatteryalertisautomaticallytransmitted.Oncealowbatterycallisreceivedthebatteryshouldbereplacedwithin7
days.DetailedinformationonbatterymanagementforallTynetec/Aidcallproductsisavailable‐requestDocNo.FM0630.








LINK
SETTINGS
DIL
SWITCH
SETTINGS
FITBATTERYASSHOWN
REDLIGHT
TESTBUTTON
DocNo.FM0460issueJPage3
RegisteringaTemperatureExtremeSensorontoaReachat‐homealarmunit

1.PuttheReachunitintoradiodevicelearnmode…

     









2.Activatethetemperatureextremesensor…


  









3.Exitradiodevicelearnmode…


     









4.CheckwiththeControlCentrethatitisconvenienttoperformatestcall,presstheTESTbuttonandconfirmtheyreceiveahigh
temperaturealarmcallfollowedbyalowtemperaturealarmcall.Remembertoremoveandrefitthebatteryafterthetestcall.


DeletingaTemperatureExtremeSensorfromaReachat‐homealarmunit

Enter1670usingthekeypadontherearoftheReachunit–thefrontlightwillFLASHFASTGREENandtheunitwillannounce
“ProgramMode‐Ready”.
Enter#041andtheReachwillannounce“DeleteRadioDevice”followedbya“Beep”.
Enter41(hightemperature)followedbythe6digitsensorIDtobedeletedandwaitforthe“Beep”
Enter42(lowtemperature)followedbythe6digitsensorIDtobedeleted,afterthe“Beep”press#toexit.
PressthekeytoexitProgramMode;theunitwillsounda“Double‐Beep”andthefrontlightwillreturntoSTEADYGREEN.


The6digitIDcanbefoundonalabelonthebackofthetemperatureextremesensor.





TheinstructionsgivenaboveforthestandardReachproductalsoappliestotheReachPlus,ReachPlusGSM
andReachUltramodels.Refertotheinstructionbookletsprovidedwiththeseproductsformoreinformation
onprogrammingandtestingradiodevices.












“Trigger
Radio
Device”
PressandHOLDtheGREENbuttonontheReachunit.

WhenthefrontlightFLASHESFASTGREENreleasethe
buttonandtheReachwillannounce“TriggerRadioDevice”.

Younowhave1minutetoactivatethetemperatureextreme
sensor.

PresstheGREENbuttonontheReachunit.

ThefrontlightwillreturntoSTEADYGREENandtheReach
isbackinnormaloperatingmode.

PressandholdtheTESTbuttonforabout3secondsuntilthe
REDlightstartstoflashandtheReachwillannounce
“HighTemperature”.

WhentheREDlightstopsflashingwaitforabout10seconds
anditwillstartflashingagainandtheReachwillannounce
“LowTemperature”.

IMPORTANT:removethebatterytoEXITTESTMODE,leave
forafewsecondsthenrefit.

DocNo.FM0460issueJPage4
RegisteringaTemperatureExtremeSensorontoanAdventxtwardencallsystem

1.PuttheAdventxtsystemintoradiodevicelearnmode…
 
     
 












2.Activatethetemperatureextremesensor…


  













3.ExittheAdventxtsystemradiodevicelearnmode…















4.PresstheTESTbuttonandcheckahightemperaturealarmcallisreportedonthewardencallsystemfollowedbyalow
temperaturealarmcallabout20secondslater.Remembertoremoveandrefitthebatteryafterthetestcall.


DeletingaTemperatureExtremeSensorfromanAdventxtwardencallsystem

PresstheLINEkeyandwaitforthe“Ready”prompt.
Enter5001followedbythekey.
Entertheflatnumberthetemperatureextremesensorisbeingdeletedfromfollowedbythekey.
PressTESTbuttononthetemperatureextremesensortobedeleted.
Waitforthe“PendantDisabled–EnterFlat”prompt.
Waitforabout20secondsforthesecond“PendantDisabled–EnterFlat”prompt.
Pressthe#keyandwaitforthe“Ready”prompt,thenpresstheLINEkeytocleartheconnection.





RefertotheAdventxtoperatorsmanual(DocNo.FM0420)providedwiththewardencallsystemformore
informationonprogrammingandtestingradiodevices.





UsingtheManager’sDECTtelephone;

PresstheLINEkeyandwaitforthe“Ready”prompt.

Enter5000followedbythekey.

Entertheflatnumberthetemperatureextremesensorisbeinginstalled
infollowedbythekey.

Younowhave1minutetoactivatethetemperatureextremesensor.

“Ready”
UsingtheManager’sDECTtelephone;

Pressthe#keyandwaitforthe“Ready”prompt.

PresstheLINEkeytocleartheconnectionandthewardencallsystemis
nowbackinnormaloperatingmode.

PressandholdtheTESTbuttonforabout3secondsuntiltheRED
lightstartstoflash.

Waitforthe“EnterLocationCode”promptthenpress0
(nolocation)followedbythekey.

Waitforthe“PendantEnabled–EnterFlat”prompt.

Afterabout20secondsyouwillhearthe“EnterLocationCode”
promptagain‐press0(nolocation)followedbythekey.

Waitforthe“PendantEnabled–EnterFlat”prompt.

IMPORTANT:removethebatterytoEXITTESTMODE,leavefora
fewsecondsthenrefit.

“Ready”

DocNo.FM0460issueJPage5
RegisteringaTemperatureExtremeSensorontoanAltecResponselocalcareralarmunit

1.PuttheAltecResponseunitintoradiodevicelearnmode…


     
 











2.Activatethetemperatureextremesensor…


  








 




3.ExittheAltecResponseradiodevicelearnmode…















4.PresstheTESTbuttonandcheckahightemperaturealarmcallisreportedontheAltecResponseunitfollowedbyalow
temperaturealarmcallabout20secondslater.Remembertoremoveandrefitthebatteryafterthetestcall.


DeletingaTemperatureExtremeSensorfromanAltecResponselocalcareralarmunit

PressENTERtoactivatethewake‐updisplaythenENTERagaintodisplaythemainmenu
PressENTERtoselectRadioDevices
PressthekeytoselectDeleteDevicethenpressENTER
Thedisplaywillpromptyouto“TriggertheSensor”
PresstheTESTbuttononthetemperatureextremesensortobedeleted‐thedisplaywillshowhightempandits41xxxxxxID
IfthisisthedeviceyouactivatedpressthekeytoselectDeletethenpressENTER
Afterabout20secondsthedisplaywillshowthelowtempandits42xxxxxxID
IfthisisthedeviceyouactivatedpressthekeytoselectDeletethenpressENTER
Note:iftheIDisNOTthedeviceyouactivatedpresstheENTERkeytoselectRejectthentryagain
ToexitthemodeselectExitthenpresstheENTERkey.





RefertotheAltecResponseinstallationguide(DocNo.FM0548)providedwiththeportablealarmunitfor
moreinformationonprogrammingandtestingradiodevices.



PressandholdtheTESTbuttonforabout3secondsuntiltheREDlightstartstoflash.

ThedisplaywillshowHighTemperatureandits41xxxxxIDnumber.
ChecktheIDiscorrect,pressthekeytoselectAcceptthenpressENTER.
Note:ifanotherdevicehasbeenpicked‐uppressENTERtoRejectandtryagain.
ThedisplaywillshowFlat,usethe&keystochoosethedwellingtype.
PressthekeytoselectNo:000000thenpresstheENTERkey.
Usethe&keystoselecteachdigitoftheflatnumberandthe
&keystomovebetweendigitsthenpresstheENTERkey.
PresskeyseveraltimestoselectExitthenpresstheENTERkey.
Thedisplaywillpromptyouto“TriggertheSensor”again.

Afterabout20secondsthedisplaywillshowLowTemperatureandits42xxxxxID
number,repeatasabove.

IMPORTANT:removethebatterytoEXITTESTMODE,leaveforafewsecondsthenrefit.
PressENTERtoactivatethewake‐updisplaythenENTERagaintodisplay
themainmenu.

PressENTERtoselectRadioDevices

PressENTERtoselectLearnDevice

Thedisplaywillpromptyouto“TriggertheSensor”.

Younowhave20secondstoactivatethetemperatureextremesensor.
FromtheLearnDevicemenuselectExitthenpressENTER.

FromtheRadioDevicesmenuselectExitthenpressENTER.

Thedisplaywillreturntothewake‐upscreenandgoblankafter10seconds.

TheAltecResponseisnowbackinnormaloperatingmode.

tynetec
tynetec

DocNo.FM0460issueJPage6
RegisteringaTemperatureExtremeSensorontoaTouchsafeProNursecallsystem

1.PuttheTouchsafeProMasterPanelintoTelecareregistrationmode;touchthescreenandLogonasanEngineeroraManager
(passwordrequired),thenfromthemainmenupagetouchtheTelecarebuttontoopenthescreenshownbelow…























2.Enterthetemperatureextremesensor8digitID’s…











3.PresstheClosebuttontoexitTelecaremodeandtheNursecallpanelisnowbackinnormaloperatingmode.

4.PresstheTESTbuttonandcheckahightemperaturealarmcallisdisplayedontheNursecallpanellikebelowfollowedbyalow
temperaturealarmcallabout20secondslater.Remembertoremoveandrefitthebatteryafterthetestcall.










Thescreenbannerisredforhigh/lowtempalarmcalls(Touchthebannertodisplaynotes).
Ahighorlowtemperaturealarmcallcannotbeacceptedorsilencedfromthescreen.
YoumustvisittheappropriateNursecallCallPointtocancelthecall.

DeletingaTemperatureExtremeSensorfromaTouchsafeProNursecallsystem

SelectthehightemperaturesensorIDtobedeletedfromtheDeleteTelecareDevicepull‐downlistbox(prefixed“T41”).
DetailsofthesensorselectedwillbedisplayedintheRoom,HouseCodeandIDboxesbelow.
IfthedetailsshownarecorrectpresstheDeletebutton.
SelectthelowtemperaturesensorIDtobedeletedfromtheDeleteTelecareDevicepull‐downlistbox(prefixed“T42”)
RepeatasaboveandpresstheClosebuttontoexit.




RefertotheTouchsafeProDisplayPaneluserguide(DocNo.FM5505)providedwiththeNursecallsystemfor
moreinformation.



TynetecandAidcalloperateapolicyofcontinuousimprovementandreservetherighttochangeproductspecificationswithoutnotice.
IfanyvariationtothedetailscontainedinthisdocumentaresuspectedpleasecontactTynetec’scustomersupportdepton01670352371.
The8digitID’saremadeupofthe2digitdevicetypecode(HighTemp=41&LowTemp=42)
followedbythe6digitunitIDwhichcanbefoundonalabelonthebackofthesensor.

TouchtheAddTelecareDeviceboxthenusetheon‐screenkeypadtoenterthehightempID.

Hint:enterazoneintheFilterboxtonarrowthesearchfortheroom(seebelow).

SelecttheusersCallPointIDfromtheRoompull‐downlistboxthenpressSave.
Note:thisistheroomunitthatwillbewillbeusedtocancelthecall.

TouchtheAddTelecareDeviceboxagainandrepeatasabovebutthistimeenterthelow
tempIDi.e.42followedbythesame6digitIDfromthelabel.
Usekeypadtoenter
8digitID’s*here
Selecttherelevant
CallPointIDhere
Savewhendone
Closetoexitmode
On‐screenkeypad
CALLING
CANCEL
HC345ID001
PresstheLIGHTGREY
buttontocancelthecall
*Note:bothhighandlow
temperatureID’smust
beentered;
Prefix41=HighTemp
andPrefix42=LowTemp
ZXT426W66265
Ser:10151997
ID:037233
EE1770V1.05A
DR


Other LEGRAND Temperature Controllers manuals

LEGRAND 674 56 User manual

LEGRAND

LEGRAND 674 56 User manual

LEGRAND 5739 18/19 User manual

LEGRAND

LEGRAND 5739 18/19 User manual

LEGRAND 5739 18 User manual

LEGRAND

LEGRAND 5739 18 User manual

LEGRAND ZigBee 883 30 User manual

LEGRAND

LEGRAND ZigBee 883 30 User manual

Popular Temperature Controllers manuals by other brands

Newport 300 Series operating manual

Newport

Newport 300 Series operating manual

Elvaco CMa11L user manual

Elvaco

Elvaco CMa11L user manual

Cumberland Integra EXPERT 64EU user manual

Cumberland

Cumberland Integra EXPERT 64EU user manual

Drayton LP20RF Installation & user guide

Drayton

Drayton LP20RF Installation & user guide

Autonics TK Series user manual

Autonics

Autonics TK Series user manual

JRI SPY RF T+ user manual

JRI

JRI SPY RF T+ user manual

Avtech RMA-DET-SEN Installation notes

Avtech

Avtech RMA-DET-SEN Installation notes

Novus N321R operating manual

Novus

Novus N321R operating manual

Conotec FOX-2000TT quick guide

Conotec

Conotec FOX-2000TT quick guide

BMW E46 manual

BMW

BMW E46 manual

ATC M910165 Operator instructions

ATC

ATC M910165 Operator instructions

Full Gauge Controls TC-940Ri manual

Full Gauge Controls

Full Gauge Controls TC-940Ri manual

HIOTH TECHNOLOGY CT-EDEA4007-ID user manual

HIOTH TECHNOLOGY

HIOTH TECHNOLOGY CT-EDEA4007-ID user manual

Daikin OM-SLM-0501 Operation manual

Daikin

Daikin OM-SLM-0501 Operation manual

Crestron C2N-RTHS Operations & installation guide

Crestron

Crestron C2N-RTHS Operations & installation guide

YOKOGAWA Green Series user manual

YOKOGAWA

YOKOGAWA Green Series user manual

Warmworld Dataterm IHC user guide

Warmworld

Warmworld Dataterm IHC user guide

THORLABS TED4015 Operation manual

THORLABS

THORLABS TED4015 Operation manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.