Physik Chemie ⋅Biologie Technik LEYBOLD DIDACTIC GMBH
9/95-Ba/Ne-
Gebrauchsanweisung 666 413
Instruction Sheet
The CPS-metering unit meters liquids via a solenoid valve,
either manually or computer-controlled, e.g. for pH-value
control in fermentation experiments or titrations. The metering
unit is suitable for mounting in a CPS-panel frame.
1 Safety notes
2 Description, scope of supply
Metering unit on chemistry panel, consisting of:
11
4-mm sockets for connecting an external timing device
(e.g. relay output of CASSY interface)
22
LED for indicating switching state
33
Screw cover
44
Reservoir
55
Clamp
66
GL-screw connection for solenoid valve
77
Solenoid valve
88
Silicone tubing
99
Pushbutton for manually actuating the valve
AA
Circuit diagram
BB
Socket for plug-in power supply (on rear of unit)
2.1 Technical data
Operating voltage: 12 V AC/580 mA via plug-in power supply
(230 V)
Dimensions: 150 x 297 x 175 mm
Weight: 1.1 kg
Die CPS-Dosiereinrichtung dient der Dosierung von Flüssigkei-
ten über ein Magnetventil, entweder manuell oder per Compu-
ter, z.B. zur pH-Wert-Regelung bei Fermentationen oder Titra-
tionen. Die Dosiereinrichtung eignet sich zum Einbau in einen
CPS-Profilrahmen.
1 Sicherheitshinweise
2 Beschreibung, Lieferumfang
Dosiereinrichtung auf Experimentierplatte, bestehend aus:
11
4-mm-Buchsen zum Anschluß eines externen Taktgebers
(z.B. Relaisausgang des CASSY-Interfaces)
22
LED zur Anzeige des Schaltzustands
33
Schraubdeckel
44
Vorratsgefäß
55
Befestigungsklammer
66
GL-Schraubverbindung zum Magnetventil
77
Magnetventil
88
Silikonschlauch
99
Handtaster zur manuellen Betätigung des Ventils
AA
Schaltbild
BB
Buchse für Steckernetzgerät (auf der Geräterückseite)
2.1 Technische Daten
Betriebsspannung: 12 VAC/580 mA über Steckernetzgerät
(230 V)
Abmessungen 150 x 297 x 175 mm
Gewicht: 1,1 kg
11
666 413
9
2
10
3
4
5
6
7
8
1
CPS-Dosiereinrichtung
CPS-metering unit
Fig. 1
Use caution when experimenting with aggressive liquids.
Max. voltage at sockets
11
: 30 V, 1A
Vorsicht beim Experimentieren mit aggressiven Flüssigkei-
ten.
Max. Spannung an Buchsen
11
: 30 V,1A
LEYBOLD DIDACTIC GMBH ⋅Leyboldstrasse 1 ⋅D-50354 Hürth ⋅Phone (02233) 604-0 ⋅Telefax (02233) 604-222 ⋅ Telex 17 223 332 LHPCGN D
©byLeyboldDidacticGmbH, PrintedintheFederalRepublicofGermany
Technical alterations reserved
3 Bedienung
3.1 Montage
CPS-Profilrahmen montieren, dann CPS-Dosiereinrichtung von
der Rückseite des Profilrahmens in die obere Profilschiene ein-
führen und nach oben schieben (ca. 1 cm); anschließend in die
untere Profilschiene einführen und nach unten schieben (ca.
1 cm). Ausbau der CPS-Dosiereinrichtung aus dem Profilrah-
men in umgekehrter Reihenfolge: Nach oben bis zum Anschlag
schieben; dann nach hinten aus der oberen Profilschiene her-
ausnehmen.
3.2 Inbetriebnahme
Experimentierplatte so im Versuchsaufbau anbringen, daß sich
die über dem Gefäß befindet, in das die Flüssigkeit zudosiert
werden soll. Klinkenstecker des Steckernetzgeräts in die Buch-
se
B
auf der Geräterückseite stecken. Silikonschlauch mit dem
darunterbefindlichen Gerät verbinden (z. B. Bioreaktor
[666 410]). Vorratsgefäß
4
mit der zuzugebenden Flüssigkeit
füllen. Steckernetzgerät in die Steckdose (230 V) stecken.
Bei manuellem Betrieb:
Durch Druck auf den Taster
9
das Magnetventil
7
öffnen. Die
grüne LED
2
leuchtet. Soviel Flüssigkeit wie gewünscht zudo-
sieren. Magnetventil
7
durch Loslassen des Tasters
9
schließen.
Bei externem Betrieb:
Zur automatischen Steuerung der Dosiereinrichtung wird das
Programm: "Universelle Meßwerterfassung" (525 031) einge-
setzt. Zur Benutzung des Programms bitte die zugehörige An-
leitung beachten.
Relaisausgang des CASSY-Interfaces (z. B. CASSYpack-P
728 307 bzw. CASSYpack-E 524 007 oder CASSY-C-Modul
667 817) mit 4-mm-Experimentierkabeln mit den beiden Buch-
sen
1
verbinden. Die gewünschte zu regelnde Größe (z.B. pH-
Wert) mit entsprechendem Meßsystem und Sensor erfassen
(z. B. pH-Box (524 035) und pH-Elektrode (667 424)). Den ge-
wünschten pH-Wert durch Eingabe in das Programm vorgeben
(Relais - Einstellungen). Nach Meßbeginn wird über die Soft-
ware der Ist-Wert des pH-Wertes mit dem eingegebenen Soll-
Wert verglichen und das Dosierventil solange geöffnet, bzw.
geschlossen bis Ist-Wert und Soll-Wert gleich sind.
Nach Gebrauch Vorratsgefäß und Schlauch entleeren und mit
Wasser reinigen.
3Operation
3.1 Setup
Assemble the CPS-panel frame, then insert the CPS-metering
unit in the top profile rail of the panel frame from the rear and
slide it upwards (approx. 1 cm); next, insert it in thelower profile
rail and slide it downward (approx. 1 cm). Remove the CPS-
metering unit from the panel frame in reverse order: slide the
panel up until it stops, then remove it from the top profile rail
from the rear.
3.2 Putting into operation.
Set up the chemistry panel in the experiment setup so that the
metering unit is above the vessel into which the liquid is to be
metered. Insert the jack plug of the plug-in power supply in
socket
B
on the rear of the unit. Attach the silicone tubing to the
device positioned below (e.g. biological reactor [666 410]). Fill
reservoir
4
with the desired liquid. Plug the plug-in power
supply into the mains socket (230 V).
Manual operation:
Open solenoid valve
7
by pressing pushbutton
9
. The green
LED
2
lights up. Meter the desired liquid quantity. Close the
solenoid valve
7
by releasing pushbutton
9
.
External operation:
The software "Universal Data Acquisition" (525 032) is used to
control the metering unit. For information on using this software,
please refer to the corresponding manual and the on-line help.
Connect the relay output of the CASSY interface (e.g.
CASSYpack-P 728 307 or CASSYpack-E 524 007 or CASSY-C
module 667 817) with the two sockets
1
using 4-mm con-
necting leads. Record the variable to be controlled (e.g.
pH value) using an appropriate measuring system and sensor
(e.g. pH-box (524 035) and pH-electrode (667 424)). Set the
desired pH value by entering the corresponding value in the
software (relay settings). After the start of measurement, the
software compares the actual pH value with the pH value you
set, and the solenoid valve is opened or closed until the actual
and set values are identical.
Drain the reservoir and tubing and clean them with water after
each use.

Other LEYBOLD Measuring Instrument manuals

LEYBOLD LEYSPEC view Series User manual

LEYBOLD

LEYBOLD LEYSPEC view Series User manual

LEYBOLD I-Measuring Amplifier D User manual

LEYBOLD

LEYBOLD I-Measuring Amplifier D User manual

LEYBOLD IONIVAC ITR 90 User manual

LEYBOLD

LEYBOLD IONIVAC ITR 90 User manual

LEYBOLD LEYSPEC view Series Instruction manual

LEYBOLD

LEYBOLD LEYSPEC view Series Instruction manual

LEYBOLD COMBIVAC CM 31 User manual

LEYBOLD

LEYBOLD COMBIVAC CM 31 User manual

LEYBOLD 332 101 User manual

LEYBOLD

LEYBOLD 332 101 User manual

LEYBOLD 531 94 User manual

LEYBOLD

LEYBOLD 531 94 User manual

LEYBOLD Phoenix Mobilis User manual

LEYBOLD

LEYBOLD Phoenix Mobilis User manual

LEYBOLD 555 88 User manual

LEYBOLD

LEYBOLD 555 88 User manual

Popular Measuring Instrument manuals by other brands

ABB FV4000 Operating	 instruction

ABB

ABB FV4000 Operating instruction

Kimo SL200 user manual

Kimo

Kimo SL200 user manual

Spirax Sarco R1C Installation and maintenance instructions

Spirax Sarco

Spirax Sarco R1C Installation and maintenance instructions

Kobold DRF34 Series Installation & operation instructions

Kobold

Kobold DRF34 Series Installation & operation instructions

ICP DAS USA iWSN-110X-RCT Series quick start

ICP DAS USA

ICP DAS USA iWSN-110X-RCT Series quick start

Flowmetrics 926-ST1LE manual

Flowmetrics

Flowmetrics 926-ST1LE manual

Siko AP10S user manual

Siko

Siko AP10S user manual

RTI VMX-8 reference guide

RTI

RTI VMX-8 reference guide

Emerson Micro Motion MVD Direct Connect installation manual

Emerson

Emerson Micro Motion MVD Direct Connect installation manual

PAC R6000S Operation manual

PAC

PAC R6000S Operation manual

Geosense LS-G6-VW user guide

Geosense

Geosense LS-G6-VW user guide

Aeroflex ATC-600A-2 Maintenance manual

Aeroflex

Aeroflex ATC-600A-2 Maintenance manual

GPI MR 5-30 Series owner's manual

GPI

GPI MR 5-30 Series owner's manual

Hameg HZ350 manual

Hameg

Hameg HZ350 manual

Yamaha MZ-N10 manual

Yamaha

Yamaha MZ-N10 manual

Renishaw REVO-2 user guide

Renishaw

Renishaw REVO-2 user guide

Gossen MetraWatt METRACLIP EARTH quick start guide

Gossen MetraWatt

Gossen MetraWatt METRACLIP EARTH quick start guide

Sadelco DisplayMax 600 Operator's manual

Sadelco

Sadelco DisplayMax 600 Operator's manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.