Conducting a Manual Override Test
Note: This feature allows the completion of a test in the occasional instance when the
subject may have diminished lung capacity and cannot activate the Auto Test.
Turn the unit on.
Attach a mouthpiece to the top of the unit and verify the display reads “AUTO TEST” with
the Blowing Man icon.
Instruct the subject to blow into the mouthpiece firmly and steadily for as long as they
can.
While they are blowing and when they are near the end of their breath, press the Execute
button to take the sample.
Read the result.
Note: Manual Override testing is pressure activated. If the subject does not blow air into
the mouthpiece, the test cannot be conducted.
Instruct the subject to blow into the mouthpiece firmly and steadily for as long as they
can. (But not necessarily as hard as they can.) The unit calculates volume and will give an
error message if the subject cannot reach 1 liter of breath.
Read the result.
After taking a test, the unit will display the results in large numbers on the display.
The result is stored in memory and available for viewing and printing at a later date.
Momentarily press the Power button to return to the test mode.
Conducting a Passive Test
The unit monitors pressure at the sample port and
automatically takes a passive test when it senses the
subject’s breath flow.
Passive testing can be done either with or without a cup,
which makes passive testing on multiple subjects easier.
Turn the unit on.
From the Blowing Man icon, press the right arrow key to enter Passive Test mode.
Verify the display shows “Passive Test” and there is no Blowing Man icon.
Hold the unit sample port on top of the unit about between 1-6 inches from the subject’s
mouth.
Have the subject blow toward the port.
The unit will display a graph of the subject’s breath flow and when high enough, take the
sample.
Read the result. The unit will only display “POS” or “NEG”. It will not display the actual
numerical result.
1”– 6”