lmp LiNKTEMP User manual

L NKTEMP
TM
Non-contact
Infrared Thermometer
User Manual
LMP001

Product Description
Intended Use
Product Features
Safety Precautions
Using the Thermometer
Taking a Measurement
Temperature Mode
Temperature Unit
Viewing Data Memory
Deleting Data Memory
Setting Alarm Temperature
Audio Setting
High Temperature Alarm
Low Battery Detection
Changing the Battery
Technical Specications
Product Maintenance
Cleaning and Disinfecting Instructions
Accessories Included
Guidelines and Compliance
Classication
Troubleshooting
Symbol References
TABLE OF CONTENTS
THE MANUFACTURER RESERVES THE RIGHT TO
ALTER THE SPECIFICATIONS OF THE PRODUCT
WITHOUT PRIOR NOTIFICATION.
1
1
1
1
3
4
5
6
6
6
7
7
7
8
8
9
10
10
12
12
12
12
16

Product Description
LinkTemp™ is a professional grade thermometer that measures body
(forehead), surface and room/ambient temperature instantly and
accurately. The LinkTemp Non-contact Infrared Thermometer can be
used as a non-invasive screening tool in a multi-patient environment.
Intended Use
LinkTemp is a non-contact, infrared thermometer intended to
measure forehead temperature in a professional healthcare
environment.
Product Features
• Infrared sensor accurately reads body temperature 1.2” to 2”
(3 to 5 cm) from forehead
• Large, 2-color LCD
• Stores the last 32 temperature readings
• Audible voice function (English & Spanish)
• Visual/audible alarm for high temperature readings
• Temperature displays in either Fahrenheit or Celsius
• Performs up to 100,000 readings
Safety Precautions
Follow the maintenance advice specied in this instruction manual.
• This device must only be used for the purposes described in this
instruction manual.
1

• This device must only be used in an ambient temperature range of
between 50˚F - 104˚F (10˚C - 40˚C).
• Always keep device in a clean, dry area.
• Do not expose the thermometer to electric shocks.
• Do not expose the thermometer to extreme temperature
conditions.
• Do not use this device in relative humidity higher than 85%.
• Do not touch the infrared sensor glass with your ngers.
• Clean the infrared sensor glass with a cotton swab lightly moistened
with 90% isopropyl alcohol.
• Follow the manufacturer’s recommended cleaning and disinfecting
instructions.
• Do not expose the thermometer to sunlight or to water.
• Do not drop the thermometer.
• Do not attempt to repair this device yourself.
• The device is not suitable for use in the presence of ammable
anesthetic mixtures with oxygen or nitrous oxide.
• The device does not require calibration.
• The device is not repairable and contains no user serviceable parts.
• Modication of the thermometer is prohibited and will invalidate
the warranty.
• Remove batteries if thermometer will not be used for a long period
of time.
• Check that the thermometer functions safely and is in proper
working condition before each use.
• Contact the manufacturer for assistance or reporting.
• Keep the thermometer out of reach of children and pets.
2

The LinkTemp Non-contact Infrared Thermometer is pre-set
at the factory.
It is not necessary to calibrate the device when starting it up.
However, in order to obtain reliable results, you are advised to allow
the thermometer to acclimate to room/ambient temperature for 15
to 20 minutes before using each time there is a signicant change in
environmental temperature. It is also important to allow a 3 to 5
second intervals between measurements.
Using the Thermometer
Audio On/O
(English/Spanish)
2-color LCD
(Blue & Red)
Memory
MODE
ON/SCAN
Lanyard
3

Fig.1
Taking a Measurement
1. Press ON/SCAN button to power on the thermometer. When the
devise is ready, the LCD screen will turn blue.
2. Position the thermometer between 1.2” to 2” (3 to 5 cm) from the
patient’s forehead. (Fig.1)
NOTE: Move hair out of the way before taking a measurement.
NOTE: If the forehead area is or has been covered with sweat,
makeup or a hat, rst clean and dry the forehead area
before taking a measurement.
Taking a Measurement
Taking a Measurement
Surface Temp Mode
Body Temp Mode
Room Temp Mode
Celsius
Fahrenheit
Memory
Low Battery
Indicator
Audio ON
(English or Spanish)
Temperature
Display Value
Measurement distance
from forehead is 1.2” - 2“
Target area for
measurement
4

3. Press ON/SCAN button to take measurement.
4. A measurement will be displayed on the LCD screen.
• To repeat the process, return to step 2.
• Allow 3 to 5 seconds between measurements.
5. The thermometer automatically turns o after 30 seconds of
inactivity.
Temperature Mode
The
LinkTemp Non-contact Infrared Thermometer can take body,
surface or room/ambient temperature.
1. While the thermometer is turned on, press the MODE button to
toggle through the three temperature mode options - BODY,
SURFACE or ROOM.
NOTE: When rst using the thermometer, the default is set to
BODY mode.
NOTE: Measurement ranges for each temperature mode are as follows:
BODY mode: 89.6˚F to 109˚F (32˚C to 42.9˚C)
SURFACE mode: 32˚F to 140˚F (0˚C to 60˚C)
ROOM mode: 32˚F to 104˚F (0˚C to 40˚C)
NOTE: Once the thermometer is powered on, the last temperature
mode measured will default as the current temperature mode.
NOTE: Always use the BODY mode to measure core body temperature.
5

Temperature Unit
While the thermometer is turned on, press the MODE and the
buttons at the same time for 3 seconds, the screen will display F-1.
Select the button for Celsius and MODE button for Fahrenheit.
Conrm by pressing MEM button. Then to exit the setting, press MEM
button quickly two times.
NOTE: The LinkTemp Non-contact Infrared Thermometer default
setting is Fahrenheit.
Viewing Data Memory
1. While the thermometer is turned on, press MEM button, which will
display the last temperature reading.
2. Press MEM button again to see the next reading stored of the last
32 measurements taken.
3. To exit data memory mode, press ON/SCAN button to go to
measurement mode or wait 30 seconds for thermometer to
automatically turn o.
Deleting Data Memory
1. While the thermometer is turned on, press MEM button for 3 seconds
to delete all data in memory.
2. When data memory is deleted you will see CLR on the screen.
3. To exit deleting data memory mode, press ON/SCAN button to go to
measurement mode or wait 30 seconds for thermometer to
automatically turn o.
2. Press MEM button again to see the next reading stored of the last
6

Setting Alarm Temperature
1. While the thermometer is turned on, press the MODE and the
button at the same time for 3 seconds. The screen will display F-1.
2. Press the MEM button one additional time to get F-2. Select MODE
button to add 0.1˚F or 0.1˚C. Select the button to minus 0.1˚F
or 0.1˚C.
3. Conrm by pressing MEM button.
4. Exit the setting by pressing MEM button again.
NOTE: For temperature unit Fahrenheit, the alarm range is 96.8˚F to
109.2˚F. If the temperature unit is Celsius, the alarm range is
36˚C to 42.9˚C.
NOTE: High temperature alarm default setting is 100.4˚F (38˚C).
Audio Setting
Press the button to choose audio language type or to shut
audio o.
1. While the thermometer is turned on, press the button to toggle
between English (EN), Spanish (SP) and audio o (OFF).
2. To exit audio setting, press ON/SCAN button to go to measurement
mode or wait 30 seconds for thermometer to automatically turn o.
High Temperature Alarm
BODY mode
1. If the temperature taken is less than the setting alarm value, the
buzzer sends out 1 sound and reports the current temperature in a
blue backlit screen. When the temperature is more than or equal to
the setting alarm value, the buzzer sends out 5 sounds and reports
7

the current temperature in a red backlit screen.
NOTE: High temperature alarm default setting is 100.4˚F (38˚C).
NOTE: When operating the thermometer in a room/ambient
temperature less than 50˚F (10˚C) or more than 104˚F (40˚C),
the BODY and SURFACE temp modes will display an ERR1
message and send out 2 sounds in a blue backlit screen.
Low Battery Detection
If the battery voltage is not more than 2.7V, the LinkTemp Non-contact
Infrared Thermometer will display the ashed symbol to remind
you to replace both batteries.
Changing the Battery
The LinkTemp Non-contact Infrared Thermometer comes with 2 AA
alkaline batteries. When the LCD screen displays , both batteries
should be replaced. Open the lid on the underside of the thermometer
and replace the batteries, taking great care with the correct battery
positioning. (Fig.2)
Temp Non-contact Infrared Thermometer comes with 2 AA
alkaline batteries. When the LCD screen displays , both batteries
should be replaced. Open the lid on the underside of the thermometer
and replace the batteries, taking great care with the correct battery
Temp Non-contact Infrared Thermometer comes with 2 AA
Fig.2
8

Technical Specications
NOTE: The LinkTemp Non-contact Infrared Thermometer can take
temperature readings below 89.6˚F or above 109.2˚F (32˚C or
above 42.9˚C) but precision is not guaranteed outside of
this range.
Device Type
Measurement Distance
Measuring Range
Display Type
Display Resolution
Accuracy
Operating Temperature
Operating Humidity
Storage Temperature
Storage Humidity
Pressure Altitude
Consumption
Memory
Automatic Power Off
Batteries
Unit Weight
Unit Size
Warranty
Non-contact infrared thermometer
1.2 to 2 inches (3 to 5 cm) from forehead
Body temperature: 89.6°F to 109°F (32°C to 42.9°C)
Surface temperature: 32°F to 140°F (0°C to 60°C)
Ambient temperature: 32°F to 104°F (0°C to 40°C)
Large, 2-color LCD
0.1°F (0.1°C)
89.6°F to 94.8°F=±0.6°F (32°C to 34.9°C=±0.3°C)
95°F to 107.6°F=±0.4°F (35°C to 42°C=±0.2°C)
107.8°F to 109.2°F=±0.6°F (42.1°C to 42.9°C=±0.3°C)
50°F to 104°F (10°C to 40°C)
Up to 85% non condensing
-4°F to 131°F (-20°C to 55°C)
Up to 95% non condensing
700 hPa - 1060 hPa
≤450 mW
Stores last 32 readings
30 seconds after last measurement
2 AA batteries
3.48 oz without batteries
6.81” x 2.05” x 1.77” (173 x 52 x 45 mm) in L x W x H
18 months
9

NOTE: The LinkTemp Non-contact Infrared Thermometer was
conceived for professional use and should provide
approximately 100,000 readings.
Product Maintenance
• The infrared sensor is the most important and fragile part of the
thermometer. Take care when using, storing or cleaning.
• Clean and/or disinfect the thermometer according to your facility’s
infection control protocol.
• Clean infrared sensor glass with cotton swab dampened with 90%
isopropyl alcohol solution.
• Please dispose of batteries properly.
• Remove the battery when thermometer is not used for an extended
period of time.
• Do not expose the thermometer to water or excessive sunlight.
• Any impact to the thermometer may damage the product.
Cleaning and Disinfecting Instructions
CAUTION: Never submerge the thermometer in water or any
other liquid.
CAUTION: Never use abrasive cleaning agents, thinners or benzene
for cleaning and never immerse the instrument in water
or other cleaning liquids.
CAUTION: Never insert a sharp object into the scanner area or any
other open surface on the thermometer.
10

CAUTION: Do not use unapproved cleaning or disinfection agents.
Use of these agents may cause damage to components.
CAUTION: Do not use chemicals other than isopropyl or ethyl
alcohol on the lens.
Cleaning the Sensor Window
Periodically, slightly moisten a cotton swab or cloth with isopropyl
(90% minimum) alcohol and gently wipe the surface of the lens using a
side-to-side (not circular) motion. Avoid touching the lens except when
cleaning is required.
Cleaning & Disinfecting the Thermometer
Following installation, it is recommended to disinfect the LinkTemp
thermometer.
The LinkTemp thermometer is compatible with most common
disinfection chemicals products. It can be subject to the following
exposures with no degradation in safety or ecacy.
• 1/100 1% Bleach
• Alcohol prep pads
Disinfection should be carried out as per standard clinic procedures.
No special precautions or procedures are required for the LinkTemp
thermometer.
If you have any questions, please contact our customer service
department at (888) 425-1149.
11

Accessories Included
• User manual: 1 pc
• AA alkaline batteries: 2 pcs
• Lanyard: 1 pc
• Quick reference guide: 1 pc
Guidelines and Compliance
This device complies with the EU Directive 93/42/EEC concerning
medical products, the ISO80601-2-56 and the European Standard
EN60601-1-2 and is subject to particular precautions with regard to
electromagnetic compatibility.
Classication
1. Internally powered equipment
2. Type BF applied part
3. IP22
4. Continuous operation
Troubleshooting
If you have problems while using the LinkTemp Non-contact Infrared
Thermometer, please refer to the following guide to help resolve the
problem. If the problem persists, please contact our customer service
department at (888) 425-1149.
1. The LinkTemp Non-contact Infrared Thermometer can display
temperature in either Fahrenheit or in Celsius.
12

2. Under BODY mode:
• When the measured value is < 89.6˚F (32˚C), the thermometer
beeps once and reads Lo.
• When the measured value is ≥ 89.6˚F (32˚C) and < the setting alarm
temperature (factory default setting is 100.4˚F (38˚C), the
thermometer beeps and reads the temperature.
• When the measured value is ≥ setting alarm temperature (factory
default setting is 100.4˚F (38˚C) and ≤ 109.2˚F (42.9˚C), the
thermometer beeps 5 times and reads the temperature.
• When the measured value is > 109.2˚F (42.9˚C), after the
thermometer beeps 5 sounds and reads Hi.
3. Under the SURFACE TEMP mode:
• When measured value is < 32˚F (0˚C),the thermometer beeps once
and reads Lo.
• When the measured value is ≥ 32˚F (0˚C) and ≤ 140˚ (60˚C), the
thermometer beeps once and reads the temperature.
• When the measured value is > 140˚F (60˚C), the thermometer
beeps once and reads Hi.
4. Under the ROOM mode:
• When measured value < 32˚F (0˚C),the thermometer beeps once
and reads Lo.
• When the measured value ≥ 32˚ (0˚) and ≤ 104˚F (40˚C), the
thermometer beeps once and reads the temperature
• When the measured value > 104˚F (40˚C), the thermometer beeps
once and reads Hi.
13

5. Under the SURFACE TEMP mode:
• When measured value is < 32˚F (0˚C),the thermometer beeps once
and reads Lo.
• When the measured value is ≥ 32˚F (0˚C) and ≤ 140˚ (60˚C), the
thermometer beeps once and reads the temperature.
• When the measured value is > 140˚F (60˚C), the thermometer
beeps once and reads Hi.
6. Under the ROOM mode:
• When measured value < 32˚F (0˚C),the thermometer beeps once
and reads Lo.
• When the measured value ≥ 32˚ (0˚) and ≤ 104˚F (40˚C), the
thermometer beeps once and reads the temperature.
• When the measured value > 104˚F (40˚C), the thermometer beeps
once and reads Hi.
14

Temperature reading is hampered
by hair or perspiration.
Temperature hampered by an air
draft or dramatic change in ambient
temperature.
Temperature readings are too close
together and the thermometer did
not have time to reboot.
Measuring too far from the forehead.
The ambient temp < 30˚F (0˚C).
Measuring after cold compress,
taking medicine or bathing.
Make sure there is no obstruction
or dampness prior to taking
temperature.
Make sure there is no air blowing in
the area of use, this could aect the
infrared reading.
Pause for 3-5 seconds minimum
between two readings; a 5 second
pause is recommended.
Take measurements at the
recommended distance of 1.2” to 2”
(3 to 5 cm) from forehead.
Make sure measuring environment
temperature is higher than 32˚F (0˚C).
If any of these apply, make sure to
wait 30 minutes and test again.
Reasons for Lo message display Advice
Reasons for Hi message display
When body temperature is above
109.2˚F (42.9˚C).
F2 value is set higher than desired.
The room/ambient temperature is
above 104˚F (40˚C).
The surface temperature exceeds
140˚F (60˚C).
Advice
Temperature reading exceeds
measurement range of 89.6˚F to
109.2˚F (32˚C to 42.9˚C). Make sure
there is no hot air blowing in the area
of use. If necessary, immediately seek
medical attention.
Reset high alarm temperature setting
closer to or at default value of
100.4˚F (38˚C).
Make sure to operate thermometer
between 50˚F to 104˚F (10˚C to 40˚C).
Temperature reading exceeds
measurement range of 32˚F to 140F
(0˚C to 60˚C).
15
Other manuals for LiNKTEMP
1
Table of contents
Other lmp Thermometer manuals