lmp LiNKTEMP User manual

L NKTEMPTM
Device Type
Measurement Distance
Measuring Range
Displayed Temperature Range
Display Type
Display Resolution
Accuracy
Operating Temperature
Operating Humidity
Storage Temperature
Storage Humidity
Pressure Altitude
Consumption
Memory
Automatic Power Off
Batteries
Unit Weight
Unit Size
Warranty
Non-Contact Infrared Thermometer
Non-contact infrared thermometer
1.2” to 2” (3 to 5 cm) from forehead
Body temperature: 89.6°F to 109°F (32°C to 42.9°C)
Surface temperature: 32°F to 140°F (0°C to 60°C)
Ambient temperature: 32°F to 104°F (0°C to 40°C)
89.6°F to 109.2°F (32°C to 42.9°C)
Large, 2-color LCD
0.1°F (0.1°C)
89.6°F to 94.8°F=±0.6°F (32°C to 34.9°C=±0.3°C)
95°F to 107.6°F=±0.4°F (35°C to 42°C=±0.2°C)
107.8°F to 109.2°F=±0.6°F (42.1°C to 42.9°C=±0.3°C)
50°F to 104°F (10°C to 40°C)
Up to 85% non condensing
-4°F to 131°F (-20°C to 55°C)
Up to 95% non condensing
700 hPa - 1060 hPa
≤450 mW
Stores up to 32 readings
30 seconds after last measurement
2 AAbatteries
3.48 oz without batteries
6.81” x 2.05” x 1.77” (173 x 52 x 45 mm) in LxWxH
18 month limited warranty
For more information on the LinkTemp™ Non-Contact Infrared Thermometer, please contact
your local distributor or Links Medical Products customer service at (888) 425-1149.
9247 Research Drive
Irvine, CA 92618 USA
www.linksmed.com

L NKTEMPTM
Non-Contact Infrared Thermometer
Quick Start Guide
Surface Temp Mode
Body Temp Mode
Room Temp Mode
Celsius
Fahrenheit
Memory
Low Battery
Indicator
Audio ON
(English & Spanish)
Temperature
Display Value
MEM
2-color LCD
MODE
logo
Audio On/O
(English/Spanish)
ON/SCAN
CHANGING TEMPERATURE MODE
CHANGING TEMPERATURE UNIT
While the thermometer is turned on, press the MODE
button to toggle through the three temperature mode
options - BODY, SURFACE or ROOM.
NOTE: When rst using the thermometer, the default is
set to BODY mode.
NOTE: Measurement ranges for each temperature
mode are as follows:
BODY mode: 89.6˚F to 109˚F (32˚C to 42.9˚C)
SURFACE mode: 32˚F to 140˚F (0˚C to 60˚C)
ROOM mode: 32˚F to 104˚F (0˚C to 40˚C)
NOTE: Once the thermometer is powered on, the last
temperature mode measured will default as the
current temperature mode.
NOTE: Always use the BODY mode to measure internal
human temperature.
While the thermometer is turned on, press the MODE
andthe buttons at the same time for 3 seconds, the
screen will display F-1. Select the button for Celsius
and MODE button for Fahrenheit. Conrm by pressing
MEM button. Then to exit the setting, press MEM
button quickly two times.
B
1. Press ON/SCAN button to power on the thermometer.
When the devise is ready, the LCD screen will turn blue.
2. Position the thermometer between 1.2” to 2” (3 to 5 cm)
from the patient’s forehead.
3. Press ON/SCAN button to take measurement.
4. A measurement will be displayed on the LCD screen.
• To repeat the process, return to step 2.
• Allow 3 to 5 seconds between measurements.
5. The thermometer automatically turns o after
30 seconds of inactivity.
Note: High temperature
alarm default setting is
100.4˚F (38˚C).
TAKING A MEASUREMENT
1.2” - 2”
Temperature reading is hampered
by hair, perspiration or makeup.
Temperature hampered by an air
draft or dramatic change in ambient
temperature.
Temperature readings are too close
together and the thermometer did
not have time to reboot.
Measuring too far from the forehead.
The ambient temp < 30˚F (0˚C).
Measuring after cold compress,
taking medicine or bathing.
Make sure there is no obstruction
or dampness prior to taking
temperature.
Make sure there is no air blowing in
the area of use, this could aect the
infrared reading.
Pause for 3-5 seconds minimum
between two readings; a 5 second
pause is recommended.
Take measurements at the
recommended distance of 1.2” to 2”
(3 to 5 cm) from forehead.
Make sure measuring environment
temperature is higher than 32˚F (0˚C).
If any of these apply, make sure to
wait 30 minutes and test again.
Reasons for Lo message display Advice Reasons for Hi message display
When body temperature is above
109.2˚F (42.9˚C).
F2 value is set higher than desired.
The room/ambient temperature is
above 104˚F (40˚C).
The surface temperature exceeds
140˚F (60˚C).
Advice
Temperature reading exceeds
measurement range of 89.6˚F to
109.2˚F (32˚C to 42.9˚C). Make sure
there is no hot air blowing in the area
of use. If necessary, immediately seek
medical attention.
Reset high alarm temperature setting
closer to or at default value of
100.4˚F (38˚C).
Make sure to operate thermometer
between 50˚F to 104˚F (10˚C to 40˚C).
Temperature reading exceeds
measurement range of 32˚F to 140F
(0˚C to 60˚C).
USING THE THERMOMETER
TROUBLESHOOTING
LMP001
9247 Research Drive
Irvine, CA 92618 USA
888.425.1149
www.linksmed.com
If the battery voltage is not more than 2.7V, the
LinkTemp Non-contact Infrared Thermometer will
display the ashed symbol to remind you to
replace both batteries.
LOW BATTERY DETECTION
CHANGING TEMPERATURE MODE
TAKING A MEASUREMENT
USING THE THERMOMETER
CHANGING THE BATTERIES
The LinkTemp Non-contact Infrared Thermometer
comes with 2 AA alkaline batteries. When the LCD screen
displays , both batteries should be replaced. Open the
lid on the underside of the thermometer and replace the
batteries, taking great care with the correct battery
positioning.