PROGRAMMABLE ITEMS
The following items can be monitored and programmed. For detailed explanations, please refer to the Operation Manual for
Model PU-2x (EM-9255), Section A.
[GROUP 01]
ITEM MDFY. DATA INPUT DISPLAY CONTENTS
01 S MAINTENANCE SWITCH (lock command)
0 MTSW: MON.MODE MONITOR mode: data monitoring only
1 MTSW: PRG.MODE PROGRAM mode: “P” marked data modifiable
02 P Alphabets & No. TG: XXXXXXXXXX Tag name entry (10 characters max.)
03
05
06
P
D
D
Number (%)
No input
No input
OUTPER XXX.XX
INPPER XXX.XX
INPVAL XXX.XX
Output status monitor (%) & simulation output
Input status monitor (%)
Input status monitor in actual value, same unit as selected in
ITEM 10.
10 TYPE OF TRANSMITTER INPUT
D 11 mV-1: 5 – 100mV
D 12 mV-2: 0.05 – 1V
P 13 mV-3: 0.5 – 10V
P 14 mV-6: 1 – 5V
D 15 mV-A: 4 – 20mA
D 16 mV-H: 10 – 50mA
11 P COARSE INPUT RANGE SELECTION
mV-1
(± mV)
mV-2
(± V)
mV-3
(± V)
mV-6/A/H
0 INPRNG: XXXX 5 0.05 0.5
No
Adjustment
Needed
1 INPRNG: XXXX 10 0.10 1.0
2 INPRNG: XXXX 20 0.20 2.0
3 INPRNG: XXXX 50 0.50 5.0
4 INPRNG: XXXX 100 1.00 10.0
14
15
P
P
Actual value
Actual value
SCLLOW XXXXXX
SCLHIG XXXXXX
Input range scaling 0% value
Input range scaling 100% value
Use the same engineering unit as the coarse range selected in
ITEM 11.
19 P Percentage FINZER Fine zero adjustment
Initial status shows actual bias (%). When data is entered, out-
put (%) is shown.
20 P Percentage FINSPN Fine span adjustment
Initial status shows actual gain (%). When data is entered, out-
put (%) is shown.
Modication Code
D: No modification (writing) possible. Used only for monitoring (reading).
S: Modifiable at any time.
P: Modifiable only when the MAINTENANCE SWITCH is in the “PRG” mode.
MAINTENANCE
Regular calibration procedure is explained below:
■CALIBRATION
Warm up the unit for at least 10 minutes. Apply 0%, 25%,
50%, 75% and 100% input signal. Check that the output
signal for the respective input signal remains within accu-
racy described in the data sheet. When the output is out of
tolerance, recalibrate the unit according to the “ADJUST-
MENT PROCEDURE” explained earlier.
LIGHTNING SURGE PROTECTION
M-System offers a series of lightning surge protector for
protection against induced lightning surges. Please contact
M-System to choose appropriate models.