6. Integrated Pneumatic Backup Test:
a. Verifypre-useassemblyowmeter
setto10L/min
b. TurnINOmaxDSIRbackupswitchON
c. Verify measured values
NO 14-26ppm
NO2≤<≤ 1.0ppm
d. TurnbackupswitchOFF
7. Performance Test:
a.VerifyO2owmeterissetto10L/min
b. SetINOMAX
doseto
40ppm
c. Verify
monitored
values
d. SetINOMAXdoseto0ppm
• “SetDoseisZero,PleaseCloseCylinderValve”
reminderwillappear-DONOTclosecylindervalveat
thistime
e. TurnoxygenowmeterOFF
8. INOblender Test:
a. Removeinjectormodulefrompre-useassembly
b. RemoveO2tubingfromowmeterandattachto
INOblenderoutlet
c. SetINOblenderowto10L/min,INOMAXdoseto
40ppm
d. ConrmacceptableNOvalue
NO 32-48ppm
e. SetINOblenderdoseandowto0
Pre-Use Checkout
WARNING:ThisreferencecardisnotasubstitutefortheINOmaxDSIRandINOblenderOperation
andMaintenancemanuals.Refertothesemanualsforallapplicablecautionsandwarnings.
AcceptableNOValue 35-45ppm
AcceptableNO2Value <1.5ppm
FiO295% ± 3%
1. TurndeviceON,lowcalibrationwillbeginandcomplete
2. Initial Connections:
 Conrmattachmentofthefollowing:
a. Waterseparatorcartridge,watertrapbottle,and
patientgassamplelineinplace
b. Injectormodulecableandtubingareconnected
c. PluginpowercordandverifyACpowerlightisON
d. Regulatorto800ppmINOMAXcylinder
e. RegulatorhosetoINOmaxDSIRinlet
f. INOblenderhoseconnectedandwhitelockinplace
g. Oxygensource(50psig)tobackofINOblender
h. IRcableinplace
3.Assemblepre-useset-upconnectors(diagramonback
ofcard).DonotturnO2owmeterONyet.
4. High Pressure Leak Test:
 Open/closeINOMAX800ppmcylindervalve
a. Verify,atleast500psigcylinderpressure
b. Verify,nodecreaseincylinderpressurefor
30seconds
5. Manual Purge/Alarm Verication:
a. PressCANCELtoexitpre-usewizard.
b. VerifytheINOMAXcylindervalveisclosed.
c. SettheO2owmeterto10L/min
d. PurgeINOmaxDSIR.
•SettheINOMAXdoseto40ppm.
•“CylinderValveClosed”alarmwilloccur.
•Cylindergaugepressureshoulddropto0psig.
•MeasuredNO2willincreaseandthendecreaseas
NO2ispurgedfromthesystem.
•“LowCylinderPressure”alarmwilloccur.
e. OpentheINOMAXcylindervalve.
f. TurntheINOMAXdosetozero.The“SetDoseis
Zero,CloseCylinderValve”indicatorwillappear.
Thisindicatorwilldisplayanytimethesetdose
isreturnedtozero;however,duringthispre-use
procedure,leavethecylinderopenandtouchthe
screentoresettheindicator.
End of Procedure
If device is not used within 10 minutes, depressurize cylinder regulator
(Thefollowinginstructionsareprovidedforwhentheon-screenpre-usewizardisnotused.)
2e 2f
2g
2h
2a
8d
8e
8c
8c
8a
8a
8e
8b
6b
6d
Series 3 software
20753Rev01
2014-08
MallinckrodtManufacturingLLC,6603FemriteDrive,Madison,WI53718-6801USA
ForassistancecontactTechnicalSupport877-566-9466.
“88” Cylinder Duration
FLOW
5 L/min 10 L/min 20 L/min 40 L/min
539Days 19.5Days 9.8Days 4.9Days
10 19.4Days 9.7Days 4.8Days 2.4Days
20 9.6Days 4.8Days 2.4Days 1.2Days
40 4.7Days 2.3Days 1.2Days 14Hours
80 2.2Days 1.1Days 13.3Hours 6.6Hours
INOMAX Dose (ppm)
FiO2Dilution
Set FiO2
.21 .40 .60 .80 1.00
10 0.21 0.40 0.59 0.79 0.99
20 0.20 0.39 0.59 0.78 0.98
40 0.20 0.38 0.57 0.76 0.95
80 0.19 0.36 0.54 0.72 0.90
Actual FiO2
INOMAX Dose (ppm)
!
1. O2Flowmeter
2. O2Tubing
3. 15Mx4.5mmAdapter
4. 22M/15Fx22M/15FAdapter
5. InjectorModule
6. 300mmof22mmhose
7. PatientGasSampleLinewithNaon
8. GasSampleTee
9. InjectorModuleElectricalCable
10.NO/N2InjectorTube
3
1
10
9
8
25
4
7
6
10 LPM
Off
Pre-Use Assembly
*Allcalculationsinthetableabovearebasedonafull
cylinder,138bar(2000psig),1963liter“88”cylinder,with
acylinderchangeat14bar(200psig).Theguresare
calculatedbasedonatotalcontinuousbreathingcircuit
gasowandacylinderconversionfactorof14.2litersper
bar/0.98litersperpsig.
•INOMAXow=[Desireddose×totalventilatorow]÷
[Cylinderconcentration-desireddose]
•Cylindervolume=Cylinderconversionfactor×cylinder
pressure(bar/psig)
•Cylinder duration (hours) =(Cylindervolume ÷INOMAX
owrate)÷60
!
!

This manual suits for next models

1

Other Mallinckrodt Medical Equipment manuals

Mallinckrodt INOmax DS User manual

Mallinckrodt

Mallinckrodt INOmax DS User manual

Mallinckrodt INOmax DSIR User manual

Mallinckrodt

Mallinckrodt INOmax DSIR User manual

Mallinckrodt INOblender 10026 User manual

Mallinckrodt

Mallinckrodt INOblender 10026 User manual

Mallinckrodt INOmax DSIR User manual

Mallinckrodt

Mallinckrodt INOmax DSIR User manual

Mallinckrodt Nellcor N-20PA User manual

Mallinckrodt

Mallinckrodt Nellcor N-20PA User manual

Mallinckrodt Nellcor N-20 User manual

Mallinckrodt

Mallinckrodt Nellcor N-20 User manual

Mallinckrodt INOmax DSIR User manual

Mallinckrodt

Mallinckrodt INOmax DSIR User manual

Mallinckrodt NELLCOR NPB-195 User manual

Mallinckrodt

Mallinckrodt NELLCOR NPB-195 User manual

Mallinckrodt INOMAX DSir Plus User manual

Mallinckrodt

Mallinckrodt INOMAX DSir Plus User manual

Popular Medical Equipment manuals by other brands

Wohler VE Series operating manual

Wohler

Wohler VE Series operating manual

Arjo Maxi Air Quick reference guide

Arjo

Arjo Maxi Air Quick reference guide

Mobiclinic Mar instruction manual

Mobiclinic

Mobiclinic Mar instruction manual

BD EnCor Enspire Quick reference guide

BD

BD EnCor Enspire Quick reference guide

Arjo MAA8000 Instructions for use

Arjo

Arjo MAA8000 Instructions for use

BodyGuardian Heart Patient Instruction Manual

BodyGuardian

BodyGuardian Heart Patient Instruction Manual

Ossur PROPRIO FOOT Instructions for use

Ossur

Ossur PROPRIO FOOT Instructions for use

Stryker Endoscopy Arthroscopy Pump Standard manual

Stryker Endoscopy

Stryker Endoscopy Arthroscopy Pump Standard manual

Microlife NEB 10 manual

Microlife

Microlife NEB 10 manual

Pulsed Harmonix True Pulse A2000 quick start guide

Pulsed Harmonix

Pulsed Harmonix True Pulse A2000 quick start guide

Primus drive P1752 installation instructions

Primus

Primus drive P1752 installation instructions

Flaem RF6 Basic 2 operating instructions

Flaem

Flaem RF6 Basic 2 operating instructions

Helena Laboratories Actalyke XL Operator's manual

Helena Laboratories

Helena Laboratories Actalyke XL Operator's manual

Halyard ON-Q PAIN RELIEF SYSTEM PATIENT GUIDELINES

Halyard

Halyard ON-Q PAIN RELIEF SYSTEM PATIENT GUIDELINES

Mettler Electronics Sonicator Plus 992 Maintenance manual

Mettler Electronics

Mettler Electronics Sonicator Plus 992 Maintenance manual

SonoSite C1.9 PLUS Service manual

SonoSite

SonoSite C1.9 PLUS Service manual

Kodak M35 X-OMAT installation instructions

Kodak

Kodak M35 X-OMAT installation instructions

Pari Vortex Instructions for use

Pari

Pari Vortex Instructions for use

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.