6. Integrated Pneumatic Backup Test:
a. Verifypre-useassemblyowmeter
setto10L/min
b. TurnINOmaxDSIRbackupswitchON
c. Verify measured values
NO 14-26ppm
NO2≤<≤ 1.0ppm
d. TurnbackupswitchOFF
7. Performance Test:
a.VerifyO2owmeterissetto10L/min
b. SetINOMAX
doseto
40ppm
c. Verify
monitored
values
d. SetINOMAXdoseto0ppm
• “SetDoseisZero,PleaseCloseCylinderValve”
reminderwillappear-DONOTclosecylindervalveat
thistime
e. TurnoxygenowmeterOFF
8. INOblender Test:
a. Removeinjectormodulefrompre-useassembly
b. RemoveO2tubingfromowmeterandattachto
INOblenderoutlet
c. SetINOblenderowto10L/min,INOMAXdoseto
40ppm
d. ConrmacceptableNOvalue
NO 32-48ppm
e. SetINOblenderdoseandowto0
Pre-Use Checkout
WARNING:ThisreferencecardisnotasubstitutefortheINOmaxDSIRandINOblenderOperation
andMaintenancemanuals.Refertothesemanualsforallapplicablecautionsandwarnings.
AcceptableNOValue 35-45ppm
AcceptableNO2Value <1.5ppm
FiO295% ± 3%
1. TurndeviceON,lowcalibrationwillbeginandcomplete
2. Initial Connections:
 Conrmattachmentofthefollowing:
a. Waterseparatorcartridge,watertrapbottle,and
patientgassamplelineinplace
b. Injectormodulecableandtubingareconnected
c. PluginpowercordandverifyACpowerlightisON
d. Regulatorto800ppmINOMAXcylinder
e. RegulatorhosetoINOmaxDSIRinlet
f. INOblenderhoseconnectedandwhitelockinplace
g. Oxygensource(50psig)tobackofINOblender
h. IRcableinplace
3.Assemblepre-useset-upconnectors(diagramonback
ofcard).DonotturnO2owmeterONyet.
4. High Pressure Leak Test:
 Open/closeINOMAX800ppmcylindervalve
a. Verify,atleast500psigcylinderpressure
b. Verify,nodecreaseincylinderpressurefor
30seconds
5. Manual Purge/Alarm Verication:
a. PressCANCELtoexitpre-usewizard.
b. VerifytheINOMAXcylindervalveisclosed.
c. SettheO2owmeterto10L/min
d. PurgeINOmaxDSIR.
•SettheINOMAXdoseto40ppm.
•“CylinderValveClosed”alarmwilloccur.
•Cylindergaugepressureshoulddropto0psig.
•MeasuredNO2willincreaseandthendecreaseas
NO2ispurgedfromthesystem.
•“LowCylinderPressure”alarmwilloccur.
e. OpentheINOMAXcylindervalve.
f. TurntheINOMAXdosetozero.The“SetDoseis
Zero,CloseCylinderValve”indicatorwillappear.
Thisindicatorwilldisplayanytimethesetdose
isreturnedtozero;however,duringthispre-use
procedure,leavethecylinderopenandtouchthe
screentoresettheindicator.
End of Procedure
If device is not used within 10 minutes, depressurize cylinder regulator
(Thefollowinginstructionsareprovidedforwhentheon-screenpre-usewizardisnotused.)
2e 2f
2g
2h
2a
8d
8e
8c
8c
8a
8a
8e
8b
6b
6d
Series 3 software
20753Rev01
2014-08
MallinckrodtManufacturingLLC,6603FemriteDrive,Madison,WI53718-6801USA
ForassistancecontactTechnicalSupport877-566-9466.
“88” Cylinder Duration
FLOW
5 L/min 10 L/min 20 L/min 40 L/min
539Days 19.5Days 9.8Days 4.9Days
10 19.4Days 9.7Days 4.8Days 2.4Days
20 9.6Days 4.8Days 2.4Days 1.2Days
40 4.7Days 2.3Days 1.2Days 14Hours
80 2.2Days 1.1Days 13.3Hours 6.6Hours
INOMAX Dose (ppm)
FiO2Dilution
Set FiO2
.21 .40 .60 .80 1.00
10 0.21 0.40 0.59 0.79 0.99
20 0.20 0.39 0.59 0.78 0.98
40 0.20 0.38 0.57 0.76 0.95
80 0.19 0.36 0.54 0.72 0.90
Actual FiO2
INOMAX Dose (ppm)
!
1. O2Flowmeter
2. O2Tubing
3. 15Mx4.5mmAdapter
4. 22M/15Fx22M/15FAdapter
5. InjectorModule
6. 300mmof22mmhose
7. PatientGasSampleLinewithNaon
8. GasSampleTee
9. InjectorModuleElectricalCable
10.NO/N2InjectorTube
3
1
10
9
8
25
4
7
6
10 LPM
Off
Pre-Use Assembly
*Allcalculationsinthetableabovearebasedonafull
cylinder,138bar(2000psig),1963liter“88”cylinder,with
acylinderchangeat14bar(200psig).Theguresare
calculatedbasedonatotalcontinuousbreathingcircuit
gasowandacylinderconversionfactorof14.2litersper
bar/0.98litersperpsig.
•INOMAXow=[Desireddose×totalventilatorow]÷
[Cylinderconcentration-desireddose]
•Cylindervolume=Cylinderconversionfactor×cylinder
pressure(bar/psig)
•Cylinder duration (hours) =(Cylindervolume ÷INOMAX
owrate)÷60
!
!

This manual suits for next models

1

Other Mallinckrodt Medical Equipment manuals

Mallinckrodt INOMAX DSir Plus User manual

Mallinckrodt

Mallinckrodt INOMAX DSir Plus User manual

Mallinckrodt INOmax DS User manual

Mallinckrodt

Mallinckrodt INOmax DS User manual

Mallinckrodt NELLCOR NPB-195 User manual

Mallinckrodt

Mallinckrodt NELLCOR NPB-195 User manual

Mallinckrodt Nellcor N-20 User manual

Mallinckrodt

Mallinckrodt Nellcor N-20 User manual

Mallinckrodt INOmax DSIR User manual

Mallinckrodt

Mallinckrodt INOmax DSIR User manual

Mallinckrodt INOmax DSIR User manual

Mallinckrodt

Mallinckrodt INOmax DSIR User manual

Mallinckrodt Nellcor N-20PA User manual

Mallinckrodt

Mallinckrodt Nellcor N-20PA User manual

Mallinckrodt INOmax DSIR User manual

Mallinckrodt

Mallinckrodt INOmax DSIR User manual

Mallinckrodt INOblender 10026 User manual

Mallinckrodt

Mallinckrodt INOblender 10026 User manual

Popular Medical Equipment manuals by other brands

Itamar Medical WatchPAT Operation manual

Itamar Medical

Itamar Medical WatchPAT Operation manual

PROTEOR 1P650 KNEE Assembling and adjustment instructions

PROTEOR

PROTEOR 1P650 KNEE Assembling and adjustment instructions

Liberate Medical VentFree user manual

Liberate Medical

Liberate Medical VentFree user manual

Dicarre DA20 user guide

Dicarre

Dicarre DA20 user guide

Stryker Medical M Series Operation manual

Stryker Medical

Stryker Medical M Series Operation manual

Invacare ScanBed 750 user manual

Invacare

Invacare ScanBed 750 user manual

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Getinge

Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics Sonicator 730 Maintenance manual

Mettler Electronics

Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care R1100 Mounting instruction

Pressalit Care

Pressalit Care R1100 Mounting instruction

Denas MS DENAS-T operating manual

Denas MS

Denas MS DENAS-T operating manual

bort medical ActiveColor quick guide

bort medical

bort medical ActiveColor quick guide

AccuVein AV400 user manual

AccuVein

AccuVein AV400 user manual

agiliti NP Trio user manual

agiliti

agiliti NP Trio user manual

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Rocket Medical

Rocket Medical Rocket Digital Oocyte Aspiration Pump Service manual

Monteris Medical Neuroblate Instructions for use

Monteris Medical

Monteris Medical Neuroblate Instructions for use

Otto Block 8E33 7 manual

Otto Block

Otto Block 8E33 7 manual

ADVANCED HOME CARE Helios Guide

ADVANCED HOME CARE

ADVANCED HOME CARE Helios Guide

NeuLog NUL-208 Guide

NeuLog

NeuLog NUL-208 Guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.