IMPORTANT: This guide is provided as a convenience and for General Information only. Never
use this product without clearly and thoroughly understanding the most recent revision of the
INOmax DS, INOmax DSIR and INOvent Operation Manuals. The Operation Manual is the source
for specic, updated information regarding warnings, cautions, checklists, diagrams, and/or
instructions contained in this guide.
See set up diagram on the reverse side of this document
For assistance contact Technical Support 877-566-9466
WARNING:
1. Avoid recirculation of gases.
Undesired recirculation of gases
will occur if fresh gas ows are less
than the patient minute volume and
may result in:
• Higher NO2levels due to the
limited ability of the carbon
dioxide absorbent to remove
NO2.
• Higher concentrations
than those set due to NO
recirculated through the
absorber.
• Reduction in O2concentrations
because nitrogen is the
balance gas for nitric oxide
and will be present in the re-
circulated gases.
2. If the Injector Module was used
in the wet/humidied part of the
breathing circuit, it should be
sterilized between each patient use.
Caution:
1. Note airow direction on the
Injector Module: the ow out of the
absorber must pass through the
Injector Module in the direction of
the arrow on the module.
2. Nitrous oxide (N2O) will affect
the Set NO versus the measured
NO value. For a 50% N2O, 50%
O2composition, the measured
NO value will be approximately
7% indicated less than the
same Set NO value of 100%
O2. For example, at a Set NO
of 20 ppm, measured NO will be
approximately 18 ppm.
3. Similarly, the effect of 2%
v/v isourane will result in a
high measured NO value of
approximately 3% indicated for the
same Set NO value at 100% O2.
4. Sudden changes in anesthetic
agent concentration may cause
brief transient changes in the
measured NO and NO2values.
Connection to a Circle Anesthesia Ventilator System
IMPORTANT: This guide is provided as a convenience and for General Information only. Never
use this product without clearly and thoroughly understanding the most recent revision of the
INOmax DS, INOmax DSIR and INOvent Operation Manuals. The Operation Manual is the source
for specic, updated information regarding warnings, cautions, checklists, diagrams, and/or
instructions contained in this guide.
1. Patient wye
2. Gas sample line
3. Sample line inlet
4. INOmax DS / INOmax DSIR /
INOvent delivery system
5. Bellows assembly
6. Ventilator drive gas line
7. Ventilator (rear view)
8. Absorber
9. Absorber expiratory port
10. Absorber inspiratory port
11. Injector Module
A. input
B. output
12. Injector module cable
13. Injector module tubing
14. Inspiratory hose
15. 22M / 15F X 22M / 15F Adapter
16. Sample Tee
See applicable warnings and cautions on page 1 of this document.
Note:
1. With a circle anesthesia breathing circuit, the
INOmax DS / INOmax DSIR / INOvent delivery
system will perform as specied in the technical
specications with
fresh gas ow rates equal to or
greater than the patient minute volume.
2. The breathing circuit between the sample tee and
the patient Y should be between 6 and 12 inches
(150 -300mm) long: greater than 6 inches to
minimize the sampling of mixed inspired/expired
concentrations and less than 12 inches to help
ensure correct NO2measurement.
3. For OR ventilation systems with the inspiratory
ow measurements at the inspiratory port of the
absorber, place the Injector Module upstream of
the inspiratory ow sensor.
Connection to a Circle Anesthesia Ventilator System
Mallinckrodt Manufacturing LLC, 6603 Femrite Drive, Madison, WI 53718-6801 USA
For assistance contact Technical Support 877-566-9466.
1
16 15 14 11b 8
6
11 11a
457
9
2 3
10
13 12
INOmax DS
INOmax DSIR
INOvent
Part No. 20731 Rev-01
2014-08

This manual suits for next models

2

Other Mallinckrodt Medical Equipment manuals

Mallinckrodt INOmax DSIR User manual

Mallinckrodt

Mallinckrodt INOmax DSIR User manual

Mallinckrodt NELLCOR NPB-195 User manual

Mallinckrodt

Mallinckrodt NELLCOR NPB-195 User manual

Mallinckrodt Nellcor N-20PA User manual

Mallinckrodt

Mallinckrodt Nellcor N-20PA User manual

Mallinckrodt INOmax DS IR User manual

Mallinckrodt

Mallinckrodt INOmax DS IR User manual

Mallinckrodt INOblender 10026 User manual

Mallinckrodt

Mallinckrodt INOblender 10026 User manual

Mallinckrodt INOmax DSIR User manual

Mallinckrodt

Mallinckrodt INOmax DSIR User manual

Mallinckrodt Nellcor N-20 User manual

Mallinckrodt

Mallinckrodt Nellcor N-20 User manual

Mallinckrodt INOmax DSIR User manual

Mallinckrodt

Mallinckrodt INOmax DSIR User manual

Mallinckrodt INOMAX DSir Plus User manual

Mallinckrodt

Mallinckrodt INOMAX DSir Plus User manual

Popular Medical Equipment manuals by other brands

STIHLER ELECTRONIC ASTODIA DIA100 Instructions for use

STIHLER ELECTRONIC

STIHLER ELECTRONIC ASTODIA DIA100 Instructions for use

Canon Leisegang OptiK 2 Setup instructions

Canon

Canon Leisegang OptiK 2 Setup instructions

Carevature Dreal 3S45FU Setup guide

Carevature

Carevature Dreal 3S45FU Setup guide

Arjo Maxi Sky 600 ECS Instructions for use

Arjo

Arjo Maxi Sky 600 ECS Instructions for use

Kimberly-Clark B BRAUN Easypump C-block RA instructions

Kimberly-Clark

Kimberly-Clark B BRAUN Easypump C-block RA instructions

Blatchford LINX22L1S user guide

Blatchford

Blatchford LINX22L1S user guide

Ossur VARI-FLEX Instructions for use

Ossur

Ossur VARI-FLEX Instructions for use

GSK VoltActive Instructions for use

GSK

GSK VoltActive Instructions for use

Air Liquide Presence PR Instructions for use

Air Liquide

Air Liquide Presence PR Instructions for use

DT Research 554GR Basic operation guide

DT Research

DT Research 554GR Basic operation guide

Atos Medical Provox Vega XtraSeal manual

Atos Medical

Atos Medical Provox Vega XtraSeal manual

Stryker IsoAir 2941 Operation manual

Stryker

Stryker IsoAir 2941 Operation manual

Bort activemed elbow support Instructions for use

Bort

Bort activemed elbow support Instructions for use

Medline MDS80600 Quick Release

Medline

Medline MDS80600 Quick Release

bort medical StabiloGen Sport Instructions for use

bort medical

bort medical StabiloGen Sport Instructions for use

Stryker ENDOSCOPY Standard Service manual

Stryker

Stryker ENDOSCOPY Standard Service manual

DRI Sleeper excel Instructions for use and care

DRI

DRI Sleeper excel Instructions for use and care

Buffalo filter ViroVac VV120 Service manual

Buffalo filter

Buffalo filter ViroVac VV120 Service manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.