Mastech MS8910A User manual

MS8910A
Smart SMD Tester
USER'S MANUAL
1. Introduction
Thank you for purchasing a MS8910A smart SMD test er.
The MS8910A smart SMD tester is a professional
instrument for measuring inductance, cap acita nce and
resistance. It has many features, such as automatic
identification, automatic measuremen t range , high
measurement accuracy and speed, wide measu ring
range and so on.
An ordinary multimeter only provides DC mode f or
resistance measurement, while the MS8910 A provid es
both AC and DC measurement modes. A variety of test
frequencies up to 100Khz can be provide d for ind uctan ce,
capacitance and resistance in AC mode to m eet the
actual needs better.
Correct usage can ensure that the instrument w ill wor k
precisely for a long time. Please read the instr uctio ns
carefully before using and operate the instr ument s trict ly
in accordance with the instructions .
01 02 0 3 0 4 05
1
2
3
4
5
6
7
Test Pin (Gold-plated)
Test Pin (Gold-plated)
Pin Holder
“HOLD” Push Button
“Hz/RMS” Push Button
“POWER”Push Button
“function” Push Button
8LCD
9Battery cover
1.2 Panel description 1.2 Inspection
When you get a new LCR meter, please check the
instrument and its accessories. If something is dam aged
or missed, please contact the store you bought the
instrument from for adjustment or replacement.
2. Safety instructions:
Operating environment and conditi on:
• Elevation <2000 m
• Relative humidity (RH) ≦ 80%RH
• Operating temperature 0 – 40°C
Note: DO NOT input voltage at the measurement port.
When measuring capacitance, please discharge first
then measure, otherwise, the meter will be damaged.
Storage and maintenance: Do not use alcohol or other
solvents to clean the meter. If it will not be used for
long time, please remove battery and put the meter in
a dry and clean environment.
3. Description:
3.1 Definition Description
APO: Auto power off
H: Display hold state
AUTO: This c harac ter sho wing on the LCD means that
the tester works in LCR automatic range mode
PAL: par allel c onnec tion me asure ment mo de
SER: seriesl connection measurement mode
C: Capacitance series/parallel co nnect ion mea surem ent
mode
L: Inductance series/parallel connecti on meas ureme nt
mode
R: Resistance series /parallel connectio n measu remen t
mode
DCR: Resistance DC measurement mode
D: Wa stage f actor
Q: Quality factor
12100KHZ: frequency
0.15VRMS: VRMS
3.1 Impedance parameter description (see Figure Ⅰ)
Z= R +jX =|Z|∠θ R =|Z|Cosθ X = |Z|Sinθ
s s s- s
X /R =Tanθ= Q=1/D
s s
If θ > 0, it means that the measured object is resistant,
if θ < 0, it means that that the measured object is
capacitive.
3.2 Series-parallel connection mode description
This meter has series and parallel measureme nt mode s.
When the capacitance value of the measured obj ect is
large or inductance value is small, use the seri es mode
for more accurate results. When the capacitance val ue of
the measured object is small or inductance val ue is lar ge,
use the parallel mode for more accurate result s. Th is
meter can select the measurement mode automa tical ly
according to the measured object. When the mea sured
impedance in 10KΩbelow is set to SER(Seri es mode ),
when the measured impedance in 10 K Ωabove se t to PAL
(parallel mode).
4. Functional c haracteristics descrip tion:
1. 6000 count
2. LC automatic range measurement
3. R analytical measurement
4. DCR mode of resistance measurement
5. D/Q display:The automatic selection for the
C+D/L+Q/R+Q
6. Measuring voltage : 0.5VRMS/0.1VRMS
7. The te st freq uency AC m ode :10 0/120 /1k/1 0k/10 0k
8. Battery electric power display : Long time no t to oper ate
the power off au tomat icall y
9. See table 1-3 for measurement accuracy and sc ope
Remarks: This accuracy is the measurement
standard. In DUT jack, the meter-specific probe
should be used if required. Measurement with probe
may be influenced by external environment. To
avoid inaccurate measurements, please keep away
from strong magnetic sources.
12100kHz 0.15VRMS
LCR
DCR
SER PAL
kMΩ
μmH
pnμF
kMΩ
DQ
R
θ
AUTO
12100kHz 0.15VRMS
LCR
DCR
SER PAL
kMΩ
μmH
pnμF
kMΩ
DQ
R
θ
AUTO
1346 8
9
7
5
2
12100kHz 0.15VRMS
LCR
DCR
SER PAL
kMΩ
μmH
pnμF
kMΩ
DQ
R
θ
AUTO
06
Measuring voltage Range Accuracy
0.5VRMS
0.5VRMS
0.5VRMS
60.00Ω
600.0Ω
6.000KΩ
2.0%+10d
1.0%+5d
1.0%+5d
0.5VRMS
0.5VRMS
0.5VRMS
60.00KΩ
600.0KΩ
6.000MΩ
1.0%+5d
1.0%+5d
2.0%+5d
0.5VRMS
0.1VRMS
0.1VRMS
20.00MΩ
60.00Ω
600.0Ω
2.0%+5d
2.0%+10d
1.5%+5d
0.1VRMS
0.1VRMS
0.5VRMS
6.000KΩ
60.00KΩ
600.0KΩ
1.5%+5d
1.5%+5d
15%+5d
0.1VRMS
0.1VRMS
6.000MΩ
20.00MΩ
3.0%+5d
3.0%+5d
07 08 0 9 1 0 00-05-3854
11
Table 1 Re sista nce mea surem ent sco pe
Note: The accuracy of Q< 0.1 assessment criteria
Measurement Frequencies:
(100HZ/120HZ/1KHZ/10KHZ:60.00Ω~6M Ω posit ion)
(100HZ/120HZ/1KHZ/10KHZ:20MΩ position)
minimum resolution:0.1Ω
minimum test range:>0.1Ω
Table 2 Ca pacit ance me asure ment sc ope
Measuring voltage Range Accur acy
0.5VRMS
0.5VRMS
0.5VRMS
600.0pF
6.000nF
60.00nF
2.0%+8d
1.0%+5d
1.0%+5d
0.5VRMS
0.5VRMS
0.5VRMS
600.0nF
6.000uF
60.00uF
1.0%+5d
2.0%+5d
2.0%+5d
0.5VRMS
0.5VRMS
0.1VRMS
600.0uF
6.000mF
600.0pF
2.0%+10d
2.0%+5d
2.0%+5d
0.1VRMS
0.1VRMS
0.1VRMS
6.000nF
60.00nF
600.0nF
1.5%+5d
1.5%+5d
1.5%+5d
0.1VRMS
0.1VRMS
6.000uF
60.00uF
1.5%+5d
3.0%+5d
0.1VRMS
0.1VRMS
600.0uF
6.000mF
3.0%+5d
3.0%+5d
when D <0.1, if D> 0.1, it should be multiplied by the
extraction of a root of 1.5Measurement Frequencies:
(100HZ/120HZ:60.00nF~600.0nF~6.0 00uF~ 60.00 uF~
600.0uF~6.000mF position)
(1KHZ:60.00nF~600.0nF~6.000uF~60 .00uF ~600. 0uF
position,In 600uF position maximum test ra nge<3 00uF)
(10KHZ:600.0pF~6.000nF~60.00nF~6 00.0n F~6.0 00uF
position)
minimum resolution:0.1pF
minimum test range:>0.8pF
Note: This accuracy is the measurement standard
Table 3. I nduct ance me asure ment sc ope
Measuring voltage Range Accur acy
0.5VRMS
0.5VRMS
0.5VRMS
600.0uH
6000uH
60.00mH
2.0%+5d
2.0%+5d
1.0%+5d
0.5VRMS
0.5VRMS
0.5VRMS
600.0mH
6.000H
60.00H
1.0%+5d
1.0%+5d
1.0%+5d
0.5VRMS
0.1VRMS
200.0H
600.0uH
1.0%+5d
3.0%+5d
0.1VRMS
0.1VRMS
0.1VRMS
6000uH
60.00mH
600.0mH
3.0%+5d
1.5%+5d
1.5%+5d
0.1VRMS
0.1VRMS
6.000H
60.00H
1.5%+5d
1.0%+5d
0.1VRMS 200.0H 1.0%+5d
Note: This accuracy is the measurement standard
when D <0.1, if D> 0.1, it should be multiplied by the
extraction of a root of 1.5
Measurement Frequencies:
(100HZ/120HZ:60.00mH~600.0mH~6.0 00H~6 0.00H ~
200.0H position)
(1KHZ:6000uH~60.00mH~600.0mH~6.0 00H~6 0.00H
position).
(10KHZ:600.0uH~6000uH~60.00 mH~60 0.0mH ~6.00 0H
position)
minimum resolution:0.1uH
minimum test range:>1.6uH
Table 4: DCR res istan ce meas ureme nt scop e
Measuring voltage Range Accur acy
0.5VRMS
0.5VRMS
0.5VRMS
60.00Ω
600.0Ω
6.000KΩ
2.0%+10d
1.0%+5d
1.0%+5d
0.5VRMS
0.5VRMS
0.5VRMS
60.00KΩ
600.0KΩ
6.000MΩ
1.0%+5d
1.0%+5d
2.0%+5d
0.5VRMS 20.00MΩ2.0%+5d
minimum resolution:0.01Ω
4 Measurement operation description:
4.2 Parameter selection function
In AC mode, each press the "Hz / RMS" key for a short
time, you will switch the test frequency to 1KHz-> 1KHz -
> 120Hz-> 100Hz-> 1KHz, press the "Hz / RMS" key for
one second, you will switch the measure voltag e to
0.5VRMS-> 0.1VRMS -> 0.5VRMS .
4.3 Measuring function
When the meter after startup, the default is LC autom atic
measurement range, short press the "functi on" ke y, th en
enter the R single measurement mode, then pre ss the"
function" key, then L C autom atic ra nge me asure ment,
cycle. Long press the"function" key,th en ente r the DC R
mode resistance measurement, short press t he "fun ction "
key, th en LC autom atic ra nge me asure ment .
4.1 Power on/off function:
Short press the “POWER” key to activate the meter and
long press for one second to shut it down.
4.4 Data Hold function:
press “H” button to stop reading measurement val ue and
show the current measurement value on the main d ispla y
continuously.
5. Ad ditio nal fun ction s
5.1 Auto pow er off fu nctio n: To prolong battery life,
when the external power supply is not used, APO wil l
display on the LCD, which means that auto power off
is available. The ins trume nt will a utoma tical ly powe r
off wit hout an y opera tion fo r 5 minut es.
5.2 Battery power detection function
The meter has a battery power detection function. Bat tery
power includes four levels and displa ys on the L CD
screen. When displays for battery, please r eplac e
battery. Th is mete r uses 3V *1 CR2032 button battery.
Batteries with the same model should be used for b atter y
replacement. Please remember that you can’ t use the
meter until the rear cover is tightened.

Other Mastech Test Equipment manuals

Mastech MS2301 User manual

Mastech

Mastech MS2301 User manual

Mastech MS8910 User manual

Mastech

Mastech MS8910 User manual

Mastech MS5215 User manual

Mastech

Mastech MS5215 User manual

Mastech MS6810 User manual

Mastech

Mastech MS6810 User manual

Mastech MS8922H User manual

Mastech

Mastech MS8922H User manual

Mastech MS5910 User manual

Mastech

Mastech MS5910 User manual

Mastech MS8910 User manual

Mastech

Mastech MS8910 User manual

Mastech MS7222 User manual

Mastech

Mastech MS7222 User manual

Mastech MS2302 User manual

Mastech

Mastech MS2302 User manual

Mastech KPS-TL300 User manual

Mastech

Mastech KPS-TL300 User manual

Mastech MS5902RTB User manual

Mastech

Mastech MS5902RTB User manual

Mastech MS2601 User manual

Mastech

Mastech MS2601 User manual

Mastech MS2309 User manual

Mastech

Mastech MS2309 User manual

Mastech Analog Earth Resistance Tester User manual

Mastech

Mastech Analog Earth Resistance Tester User manual

Mastech MS6813 User manual

Mastech

Mastech MS6813 User manual

Mastech ms5908 User manual

Mastech

Mastech ms5908 User manual

Mastech MS7220 User manual

Mastech

Mastech MS7220 User manual

Popular Test Equipment manuals by other brands

Virtins technology DSO-2810E manual

Virtins technology

Virtins technology DSO-2810E manual

Televes H30 Series 593101 instruction manual

Televes

Televes H30 Series 593101 instruction manual

AWS AWS-QCM user guide

AWS

AWS AWS-QCM user guide

Novatech 1450B manual

Novatech

Novatech 1450B manual

Rigol MSO7000 Series user guide

Rigol

Rigol MSO7000 Series user guide

Hantek DSO8072E user manual

Hantek

Hantek DSO8072E user manual

Midtronics PBT-50 instruction manual

Midtronics

Midtronics PBT-50 instruction manual

Berkeley Varitronics Systems Mantis user manual

Berkeley Varitronics Systems

Berkeley Varitronics Systems Mantis user manual

Endress+Hauser RID 261 operating instructions

Endress+Hauser

Endress+Hauser RID 261 operating instructions

RS PRO RS MT-6600 instruction manual

RS PRO

RS PRO RS MT-6600 instruction manual

Major tech MT700 instruction manual

Major tech

Major tech MT700 instruction manual

Megger BT 500-IS-1 operating manual

Megger

Megger BT 500-IS-1 operating manual

pico Technology PicoScope 2203 user guide

pico Technology

pico Technology PicoScope 2203 user guide

WIKA CPH6000 operating instructions

WIKA

WIKA CPH6000 operating instructions

VOLTCRAFT DSO-6202E user manual

VOLTCRAFT

VOLTCRAFT DSO-6202E user manual

VOP LC-CAST 60A user manual

VOP

VOP LC-CAST 60A user manual

VeEX Fiber Scope Inspector Quick reference guide

VeEX

VeEX Fiber Scope Inspector Quick reference guide

Fluke 718 30G manual

Fluke

Fluke 718 30G manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.