MedViron Rise II Manual

medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
Rise II Bassinets
Operations & Maintenance Manual
WARNING:
To avoid tipping hazards of this device, make
sure that the basket or bed component sits firmly in the
base and that all doors, drawers and casters are secure.

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
2
Symbols used in this manual:
Warnin
g
.
The warning symbol is used to identify critical hazards.
Pinch Point.
This symbol is used to identify a location where an injury may occur.
Important
N
ote
.
This symbol is used to identify important information to remember.
Contact
Corporate Headquarters: Email: customerservice@medviron.com
Medviron LLC Toll free: 1-844-633-8476
9380 Pentatech Drive
Zeeland, Michigan 49464 USA Website: medviron.com
Mailing address:
Medviron LLC
P.O. Box 286
Zeeland, Michigan 49464 USA

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
3
Table of Contents:
Symbols used in this manual ______________________________________________________________ 2
Manufacturer Contact Information_________________________________________________________ 2
Guardian Bassinet Description ____________________________________________________________ 4
Purpose __________________________________________________________________________________ 4
Features __________________________________________________________________________________ 5
Specifications _____________________________________________________________________________ 5
Contraindications _________________________________________________________________________ 6
Operating Instructions:
Moving the Guardian Bassinet ______________________________________________________6
Trendelenberg positioning _________________________________________________________ 7
Serial Number & Cleaning Instructions Location ___________________________________________ 8
Safety and Risks ___________________________________________________________________________ 9
Cleaning Instructions _____________________________________________________________________ 11
Preventative Maintenance List ____________________________________________________________ 12
Service Parts _____________________________________________________________________________ 13
Warranty _________________________________________________________________________________ 14

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
4
Description
MedViron Rise II Bassinets have two distinct parts, a plastic tub with mattress which sits on a
mobile storage cabinet with drawer and easily adjustable surface height. The bassinet has a
pull-out work surface which is accessible from either side. The bassinet’s modern design
conveys the warmth of home, while also implying competent, current healthcare.
Purpose
The functional design of the bassinet accommodates the care and transport of an infant
patient, the capability of Trendelenberg positioning of the bassinet tub, the storage of care
items for the infant and height adjustability.
The bassinet is intended to be sold to and for use by professionally licensed staff and facilities
caring for the labor and delivery of children.
The bassinet has been designed and manufactured under and to all pertaining ISO 13485
standards.
The bassinet is intended for indoor use only.
THE IMPROPER USE OF THE BASSINET CAN LEAD TO INJURY TO THE
INFANT OR TO THE HEALTHCARE PROFESSIONAL OPERATING THE
BASSINET. USE THE BASSINET ONLY AS INTENDED, AS DEFINED IN THIS
MANUAL.
CHANGES MADE TO THE BASSINET OR ITS PARTS CAN LEAD TO INJURY
TO THE INFANT OR THE HEALTHCARE PROFESSIONAL OPERATING THE
BASSINET. MAKING ANY CHANGES TO THE BASSINET ALSO VOIDS ITS
WARRANTY.

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
5
Features:
Two design goals for the bassinet were lower cost of ownership for the hospital and
improving the experience of mother, caregiver, and family.
1. Four height-adjustment handles, one in each corner, can be used independently to
raise or lower the bassinet’s height.
2. The bassinet’s stainless-steel push handles and all surface areas are easily accessible
and easy to clean. The tub and handle design allow a clear view of the infant.
3. Trendelenberg and Reverse-Trendelenberg positioning.
4. The bassinet has a pass-thru pull-out work surface accessible from either side.
5. The base panel has been designed to reduce the chances of caregivers hitting their
shins while the bassinet is in motion. The base panel’s concave curves also allow
caregivers to stand closer while caring for the infant.
6. KYDEX® thermofilm has been used for impact resistance and durability.
Specifications:
Width 37.5in (95 cm)
Dept
h
21.5in
(55 cm)
Surface low height
30
.5
in
(7
7
cm)
Surface high height
40.25
(
1.02 m
)
Drawer internal
dimensions
22.5
in W x
13.25
in L x 6in D

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
6
Contraindications:
1. The bassinet tub is not intended to be used without the mobile storage cabinet
beneath.
2. The bassinet’s plastic drawer is not intended to be used for anything other than as a
drawer.
3. The bassinet is not intended to be used for infants that can sit, stand or roll over.
4. The bassinet tub or drawer boxes are not intended to be used for bathing.
Operation: Moving the Guardian Bassinet
Stand at either the left or right side of the bassinet.
Make certain the bassinet tub is secure and that drawer and pull-out work surface are secure.
Place both hands on the bassinet push handle and push forward gently.
Make certain that the bassinet is always in front of you. Do not pull the bassinet behind you.
Make certain that the floor is free of obstruction and beware of transitions in the floor
NEVER PUSH THE BASSINET FROM THE FRONT OR REAR.
NEVER PULL THE BASSINET BEHIND YOU.
TRANSPORT OF AN INFANT MUST OCCUR IN THE LOWEST HEIGHT
POSITION ONLY.
Right
side
Left
side
Push
only
in this direction

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
7
Operation: Trendelenberg Positioning
To put the bassinet tub into the Trendelenberg or Reverse Trendelenberg position, simply
raise either end of the tub and then lower it so that the hook on the side of the tub engages
the metal rail that bisects the push handle.
DO NOT LET GO OF THE TUB UNTIL YOU ARE CERTAIN THE TUB HOOK IS
FIRMLY SEATED ON THE RAIL.
NEVER PLACE ANY FOREIGN OBJECTS INSIDE OR ON TOP OF THE
BASSINET TUB.
Lift this side
and engage the
hook
for the
Trendelenberg
Position.
Lift this side and
engage the
hook
for the Reverse
Trendelenberg
Position.

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
8
Serial Number & Cleaning Instructions Location
The serial number and
cleaning instructions are
on a label on the side
panel of the bassinet.
Stand in front of the
bassinet and push the
lower drawer inward, away
from you, to reveal the
label.
Contact Information
Corporate Headquarters: Email: customerservice@medviron.com
MedViron LLC Toll free: 1-844-633-8476
9380 Pentatech Drive
Zeeland, Michigan 49464 USA Website: medviron.com
Mailing address:
MedViron LLC
P.O. Box 286
Zeeland, Michigan 49464 USA

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
9
Safety and Risks:
TO AVOID TIPPING HAZARDS OF THIS DEVICE, MAKE
SURE
THAT THE
BASKET
OR BED COMPONENT FITS FIRMLY IN THE BASE, AND THAT THE DRAWER
AND CASTERS ARE SECURE.
TRANSPORT OF AN INFANT SHOULD OCCUR IN THE LOWEST HEIGHT
POSITION ONLY.
WHEN PUTTING THE BASSINET TUB INTO THE TRE
ND
ELENBURG OR REVERSE
TRENDELENBERG POSITION, ALWAYS VERIFY THAT THE TUB HOOK IS
SECURE AND SEATED FIRMLY ON THE RAIL.
OBJECTS SHOULD NOT BE LEFT IN OR ON TOP OF THE BASSINET INFANT TUB
AT ANY TIME.
THE MAXIMUM WEIGHT LOAD OF THE BASSINET DRAWER IS 15 POUNDS (7
KG).
THE MAXIMUM WEIGHT LOAD OF THE
PULL
-
OUT
WORK SURFACE IS 10LBS
(4.5 KG). THIS IS NECESSARY TO AVOID TIPPING THE BASSINET.
NEVER PLACE AN INFANT ON THE
PULL
-
OUT
WORK SURFACE.
WHEN MOVING THE BASSINET, THE OPERATOR SHOULD PUSH ONLY FROM
THE EITHER END USING THE BASSINET HANDLES.

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
10
Safety and Risks:
THE PULL
-
OUT WORK SURFACE SHOULD BE FULLY CLOSED AND IN THE
LOCKED POSITION BEFORE MOVING THE BASSINET.
PINCH POINTS EXIST AT THE LOCATIONS INDICATED IN THE DRAWINGS
BELOW. PLEASE BE CAREFUL.

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
11
Product Care and Disinfection
Clean with a damp cloth or mild detergent solution. MedViron furniture can withstand a
wide range of hospital cleaners and disinfectants. When using any of these products, read
and follow the manufactures directions carefully. Do not use abrasive or solvent cleaner. A
CDC recommended chlorinated bleach solution of 5.25% -6.15% sodium hypochlorite (i.e.,
household bleach) diluted in water at 1:100 also can be used.
Infant tub: Always use a soft cloth or sponge on the bassinet’s infant tub for washing and
rinsing. Liquid dish detergent is an acceptable cleaner. Never use abrasives or solvent
cleaners. Always follow up the application of any cleaning solution with a warm water
rinse.
Clean the tub and mattress after each patient. Do not immerse the mattress in cleaning
solution. Make certain the mattress and tub are completely dry after each cleaning.
After cleaning, it is important that the bassinet is rinsed with warm water
using a clean damp cloth and then wiped with a clean dry cloth to remove any
moisture.

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
12
Preventative Maintenance List:
WARNING: To avoid tipping hazards of this device, make sure that
the basket or bed component sits firmly in the base and that all
doors, drawers and casters are secure.
_____ Casters are secure. (Check that hex nuts are tight: 4 per caster)
_____ Casters: Casters should turn smoothly and should be free of any debris. Caster brakes
should hold firmly, the wheel should not turn when brake is applied.
_____ Plastic drawer: Should be clean and free of damage.
_____ Drawer slides: Drawers should open smoothly and quietly.
_____ Bassinet tub: Trendelenburg hooks must be intact and firmly attached to the tub.
Bassinet tub should be free of damage.
_____ Warning labels must be intact and legible.
_____ Check mattress for tears, punctures, or wear.
_____ Verify that the pull-out work surface can be secured.
Serial number
Date
Completed
Completed by

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
13
Service Parts
Part Name
Part Number
Bassinet Tub
523
-
0150
-
0000
Mattress
516
-
0000
-
0000
Inset Drawer Pull
557
-
0105
-
0000
Stainless Steel Drawer Pull
557
-
0036
-
0028
Plastic Drawer Liner
558
-
0063
-
0000
Caster Replacement Kit (4
casters w/hardware
)
556
-
0305
-
0000
Drawer Front
530
-
M009
-
Kxxx
Please contact your dealer or sales representative to order service parts.

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
14
Limited Lifetime Warranty
MedViron® warrants the products it manufactures against defects in material and
workmanship under normal use and service for the expected useful life of the product
from the date of shipment. This limited warranty is valid only to the original purchaser
using the product for its intended purpose. This warranty does not apply to expected
changes resulting from normal wear and tear, use and aging of the product. Specifically
excluded from this limited warranty are defects and damage resulting from abuse,
neglect, misuse, or excessive use. Product alteration, substitution of components or
materials, improper cleaning or repair, and installation and handling damages are also
not covered by this warranty. All other express or implied warranties made by agents,
dealers, sales representatives or any other representatives, including any warranty of
merchantability or fitness for a particular purpose, are hereby excluded.
There are no representations as to the capacity or performance of the products sold
hereunder except as set forth in the price list or quotation specifications, if any, and
such representations are expressly conditioned upon the correctness of the data
furnished by purchaser and upon the products being properly installed and maintained.
MedViron’s sole obligation shall be to repair or replace any defective product or product
component covered under this warranty, including shipping and labor, during the warranty
period. In no event shall MedViron be liable for any consequential or incidental damages,
including lost profits or revenues incurred by the purchaser with respect to any goods
furnished by MedViron. Expenses incurred by purchaser in repairing or replacing MedViron
product must be pre-approved in writing by MedViron
In the event a product has to be returned to the factory for verification of a defect and/or
repair, it must be pre-approved in writing by MedViron freight pre-paid, and include a
copy of the original invoice with the date and name of original purchaser.
MedViron hereby assigns to the original owner any and all warranties offered by its
component suppliers. To the extent the terms of these warranties differ from the terms of
the Limited Warranty, the terms of the suppliers’ component warranties shall apply.
However, MedViron does not guarantee that the component supplier will comply with any
terms of the warranty of such supplier.
Any MedViron products with the serial number removed or altered are excluded from this
warranty.
MEDVIRON WILL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES
INCLUDING ANY LOSS, EXPENSE, OR DAMAGE OTHER THAN TO THE PRODUCT ITSELF
THAT MAY RESULT FROM A DEFECT IN THE PRODUCT. SOME STATES DO NOT ALLOW FOR
THE EXCLUSION OR LIMITATION OF CONSEQUENTIAL AND INCIDENTAL DAMAGES, SO THE
ABOVE LIMITATION OR EXCLUSION MAY NOT APPLY.
MEDVIRON MAKES NO IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. THESE AND ALL OTHER IMPLIED WARRANTIES ARE SPECIFICALLY
DISCLAIMED. SOME STATES DO NOT ALLOW AN EXCLUSION OR LIMITATION ON HOW
LONG AN IMPLIED WARRANTY LASTS, SO THE ABOVE LIMITATIONS OR EXCLUSIONS MAY
NOT APPLY.
THIS WARRANTY GIVES YOU, THE ORIGINAL PURCHASER, SPECIFIC LEGAL RIGHTS AND YOU
MAY ALSO HAVE OTHER RIGHTS THAT VARY FROM STATE TO STATE. EXCEPT FOR SUCH
OTHER RIGHTS, THE REMEDIES PROVIDED UNDER THIS WARRANTY STATE THE LIMIT OF
MEDVIRON’S RESPONSIBILITIES.

Rise II Bassinets
medviron.com | 1-844-633-8476 | 984-0061-0000 (Rev 2)
15
Table of contents
Other MedViron Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Getinge
Getinge Arjohuntleigh Nimbus 3 Professional Instructions for use

Mettler Electronics
Mettler Electronics Sonicator 730 Maintenance manual

Pressalit Care
Pressalit Care R1100 Mounting instruction

Denas MS
Denas MS DENAS-T operating manual

bort medical
bort medical ActiveColor quick guide

AccuVein
AccuVein AV400 user manual