manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Megger
  6. •
  7. Measuring Instrument
  8. •
  9. Megger DCM320 User manual

Megger DCM320 User manual

Other Megger Measuring Instrument manuals

Megger MJ15 User manual

Megger

Megger MJ15 User manual

Megger DLRO100EB User manual

Megger

Megger DLRO100EB User manual

Megger DCM330 User manual

Megger

Megger DCM330 User manual

Megger SVERKER 900 Installation and operating instructions

Megger

Megger SVERKER 900 Installation and operating instructions

Megger HVB 10 User manual

Megger

Megger HVB 10 User manual

Megger Ducter DLRO2X User manual

Megger

Megger Ducter DLRO2X User manual

Megger MPQ1000 User manual

Megger

Megger MPQ1000 User manual

Megger DCM2000P User manual

Megger

Megger DCM2000P User manual

Megger DLRO100E User manual

Megger

Megger DLRO100E User manual

Megger S1-1054/2 User manual

Megger

Megger S1-1054/2 User manual

Megger DLRO User manual

Megger

Megger DLRO User manual

Megger DCM1500S User manual

Megger

Megger DCM1500S User manual

Megger HT1000/2 User manual

Megger

Megger HT1000/2 User manual

Megger DET10C User manual

Megger

Megger DET10C User manual

Megger HV DAC-200 User manual

Megger

Megger HV DAC-200 User manual

Megger ICLAMP User manual

Megger

Megger ICLAMP User manual

Megger SmartFuse 250 User manual

Megger

Megger SmartFuse 250 User manual

Megger TDR900 User manual

Megger

Megger TDR900 User manual

Megger PSI410 User manual

Megger

Megger PSI410 User manual

Megger DET14C User manual

Megger

Megger DET14C User manual

Megger MJOLNER 200 User manual

Megger

Megger MJOLNER 200 User manual

Megger PVM210 User manual

Megger

Megger PVM210 User manual

Megger MLDM50 User manual

Megger

Megger MLDM50 User manual

Megger KF-LAB Mk II User manual

Megger

Megger KF-LAB Mk II User manual

Popular Measuring Instrument manuals by other brands

Powerfix Profi 278296 Operation and safety notes

Powerfix Profi

Powerfix Profi 278296 Operation and safety notes

Test Equipment Depot GVT-427B user manual

Test Equipment Depot

Test Equipment Depot GVT-427B user manual

Fieldpiece ACH Operator's manual

Fieldpiece

Fieldpiece ACH Operator's manual

FLYSURFER VIRON3 user manual

FLYSURFER

FLYSURFER VIRON3 user manual

GMW TG uni 1 operating manual

GMW

GMW TG uni 1 operating manual

Downeaster Wind & Weather Medallion Series instruction manual

Downeaster

Downeaster Wind & Weather Medallion Series instruction manual

Hanna Instruments HI96725C instruction manual

Hanna Instruments

Hanna Instruments HI96725C instruction manual

Nokeval KMR260 quick guide

Nokeval

Nokeval KMR260 quick guide

HOKUYO AUTOMATIC UBG-05LN instruction manual

HOKUYO AUTOMATIC

HOKUYO AUTOMATIC UBG-05LN instruction manual

Fluke 96000 Series Operator's manual

Fluke

Fluke 96000 Series Operator's manual

Test Products International SP565 user manual

Test Products International

Test Products International SP565 user manual

General Sleep Zmachine Insight+ DT-200 Service manual

General Sleep

General Sleep Zmachine Insight+ DT-200 Service manual

Sensa Core Lacto Spark user manual

Sensa Core

Sensa Core Lacto Spark user manual

VOLTCRAFT UTS-1980 operating instructions

VOLTCRAFT

VOLTCRAFT UTS-1980 operating instructions

EBCHQ 94915 Operation manual

EBCHQ

EBCHQ 94915 Operation manual

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Balluff

Balluff BIP LD2-T017-01-EP-S4 Series user guide

Pulsar IMP Lite Series instruction manual

Pulsar

Pulsar IMP Lite Series instruction manual

ETCR 6800 user manual

ETCR

ETCR 6800 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

Power On / Off AC V / DC V
Auto Power Off Resistance
The meter will automatically shut itself
off after approximately 10 minutes after
power on.
Continuity
Data Hold
User Guide
DCM320 Clampmeter
Warning
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
Safety sheet
. . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . .
. . . . . . . . . Read First
Safety Information
To ensure safe operation and service
of the Meter, follow these instructions.
Failure to observe warnings can result
in severe injury or death.
-Avoid working alone so assistance can
be rendered.
- To enhance safety, test leads should be
disconnected from instrument when not
in use.
- Do not use test leads or the Clamp
Meter if they look damaged.
- Do not use the Meter if the Clamp
Meter is not operating properly or if it
is wet.
- Use the Clamp Meter only as specified
in the Instruction card or the protection
provided by the Clamp Meter might be
impaired.
- Special precautions are necessary
when operating in situations where
exposed live parts at dangerous
voltages may be encountered. Personal
protective equipment (not supplied with
this instrument) should be used.
Symbols as marked on the
Tester and Instruction card
Risk of electric shock
See instruction card
Equipment protected by double or
reinforced insulation
Battery
Earth
Conforms to EU directives
FApplication around and removal
from hazardous live conductors is
permitted
Do not discard this product or
throw away
Maintenance
Do not attempt to repair this Clamp Meter.
It contains no user-serviceable parts.
Repair or serving should only be
performed by qualified personal.
Cleaning
Periodically wipe the case with a dry
cloth and detergent do not use
abrasives or solvents.
ACA
1-3 Electrical Specifications
Accuracy is ±(% reading + number
of digits) at 23°C ±5°C < 80%RH.
Temperature coefficient :
Add 0.2 x (Specified accuracy) / °C,
< 18°C, > 28°C .
DC / AC Volts
Range DC Accuracy AC Accuracy
±(1.0% + 2dgt) ±(1.5% + 5dgt)
50Hz ~ 500Hz
200.0V
600V
Over voltage protection :
DC / AC 600V
Input Impedance :
10MΩ// less than 100pF.
CMRR / NMRR :
(Common Mode Rejection Ratio)
(Normal Mode Rejection Ratio)
VAC : CMRR > 60dB at DC,
50Hz / 60Hz
VDC : CMRR > 100dB at DC,
50Hz / 60Hz
NMRR > 50dB at DC,
50Hz / 60Hz
AC Conversion Type :
Average sensing rms indication.
Range Accuracy
200.0 Ω*3 ±(1.0% + 5 dgt) *2
2.000 KΩ*3
20.00 KΩ*3
200.0 KΩ*3
2.000MΩ*3 ±(1.0% + 2 dgt)
20.00 MΩ ±(1.9% + 5 dgt) *1
±(0.7% + 2 dgt)
Overload protection : 600V rms
Open circuit Voltage :
-1.3V approx.
* 1 < 100 dgt rolling.
* 2 < 10 dgt rolling.
* 3 The maximum LCD reading is
1400 counts.
Continuity :
Internal beeper activates if the
resistance of the circuit under test
is less than 20Ω. It will then turn off
if the resistance is increased
beyond 50Ω. Response time is
approximately 50m sec.
Overload protection : 600V rms.
Function Range Accuracy
0.0 ~ 40.0A ±(1.9% + 10 dgt)
40.0 ~ 200.0A ±(1.9% + 5 dgt)
201 ~ 400A ±(1.9% + 5 dgt)
A
(50~60Hz)
Overload protection : 600A rms
AC Conversion Type :
Average sensing rms indication.
Position Error : ±1.0% of reading.
AC Current Limited Warranty
This Meter is warranted to the original
purchaser against defects in material
and workmanship for 1 year from the
date of purchase. During this warranty
period, manufacturer will, at its option,
replace or repair the defective unit,
subject to verification of the defect or
malfunction.
This warranty does not cover ,
disposable batteries, or damage from
abuse, neglect, accident, unauthoriz-
ed repair, alteration, contamination,
or abnormal conditions of operation
or handling.
Any implied warranties arising out of
the sale of this product, including but
not limited to implied warranties of
merchantability and fitness for a
particular purpose, are limited to the
above. The manufacturer shall not be
liable for loss of use of the instrument
or other incidental or consequential
damages, expenses, or economic
loss, or for any claim or claims for
such damage, expense or economic
loss. Some states or countries laws
vary, so the above limitations or
exclusions may not apply to you.
Operating altitude : 2000m (6562 ft)
Conductor Size : 27mm diameter.
Pollution degree : 2
EMC : EN 61326-1
Shock vibration :
Sinusoidal vibration per MIL-T-
28800E (5 ~ 55 Hz, 3g maximum).
Drop Protection :
1.2m drop to hardwood on concrete
floor.
Dimensions (W x H x D) :
56mm x 188mm x 28mm
Weight : 225g including battery.
Accessories : Battery (installed),
Carrying case, Test lead and User
manual.
ⅠThe circuits are not connected to
mains.
ⅡThe circuits are directly connected
to Low-voltage installation.
ⅢThe building installation.
ⅣThe source of the Low-voltage
installation.
Resistance
I
I
I
I
a
I + (-I) = 0
Specifications
1-1 General Specifications
LCD display digits :
3 1/2 digit large scale LCD readout.
Display count : 2000 counts.
Measuring rate : 1.5 times / sec.
Overrange display :
“OL” is displayed for “Ω” functions,
shows the real value for “A” and “V”
function.
Automatic power off time :
Approximately 10 minutes after
power on.
Low battery indicator :
=is displayed.
Power requirement :
1.5V AAA x 2 battery.
Rechargeable batteries are not
suitable for use with this instrument.
1-2 Environmental Conditions
Indoor Use.
Calibration :
One year calibration cycle.
Operating temperature :
0°C ~ 30°C (≦80% RH)
30°C ~ 40°C (≦75% RH)
40°C ~ 50°C (≦45%RH)
Storage temperature :
-20 to +60°C, 0 to 80% RH (batteries
not fitted).
Overvoltage category :
IEC 61010-1 600V CAT.Ⅲ.
CAT Application field
Battery Replacement
CAT.Ⅲ.600V
with respect to
earth for the jaw.
Tactile barrier for
Hand Guard.
Do not hold the
meter in front of
the Tactile Barrier.
rr