
7 VG9F 70 en 17
To conclude press the ?key when the desired
value is shown on the display.
NOTE: If nothing else is defined, please select type 1 in any
class.
Select parameters as follows:
4.4.5 Actuator size, ACTS
This parameter defines the actuator size.
Look at e.g. the type code on the machine plate in
the Neles B1-series actuators to check the size. If 3rd
party actuator is used, please check the actuator
stroke volume.
After selecting ACTS on the display, press the ?key
to enter the edit state and ACTS starts to blink.
Select between the following values:
S 1 = Neles B1J8 actuator
(or stroke volume <1 dm3/ <61 in3)
S 3 = B1J10 (1-3 dm3/ 61-183 in3)
S10 = B1J12-16 (3-10 dm3/ 183-610 in3)
S30 = B1J20-25 (10-30 dm3/ 610-1831 in3)
L30 = B1J32- (>30 dm3/ >1831 in3)
Use the +and -keys to change the value.
To conclude press the ?key when the desired
value is shown on the display.
4.4.6 Spool type, STYP
This parameter defines the spool type and size in VG9000F.
Look at the machine plate in the device to check the
typecode.
After selecting STYP on the display, press the ?key
to enter the edit state and STYP starts to blink.
Select between the following values:
15 = VG9_12 or VG9_15
35 = VG9235
37 = VG9237
Use the +and -keys to change the value.
To conclude press the ?key when the desired
value is shown on the display.
4.4.7 Language selection, LANG
Select between three languages EnG, GEr or FrE
(English, German or French) using the +and -keys.
To conclude press the ?key when the desired
value is shown on the display.
4.5 Valve travel calibration
Select CAL from the menu by using +or -keys and press
the ?key.
NOTE: When device usage is hybrid the valve needs to be in
the normal position (not fail safe) for calibration to be
started.
4.5.1 AUTO CAL calibration function
During calibration process a blinking text "CAL run" will be
show on the display. If calibration ends successfully, a text
"CALIBRATION SUCCESSFUL" will be shown. Calibration can be
cancelled with the =key, which will show a text "CALIBRATION
CANCELLED". If calibration fails, the reason will be shown, eg.
"CALIBRATION START FAILED", " POSITION SENSOR RANGE
TOO SMALL", "CALIBRATION TIMEOUT" or "CALIBRATION
FAILED". After calibration the device will return to the main
menu (measurement monitoring).
4.6 Testing, TESt
Select between two tests Partial Stroke Test (PARt
tst) or Pneumatics test (PNEU tst).
To conclude press the ?key when the desired
value is shown on the display.
4.6.1 Partial Stroke Test, PARt tst
Table 5 Selecting extra pneumatics instrumentation
parameters
Instrumentation type Parameter (EXTI)
Volume booster (VB) Booster type 1 (bo1)
Quick exhaust (QEV) Quick exhaust 1 (qE1)
Combination of VB and QEV Combinationtype1(co1)
NOTE:
When QEV or volume booster is used, VG with standard
spool valve (VG_15_) is required.
CAUTION:
It is important to select the correct actuator size
because this parameter is used in device control.
Erroneous value may cause instability.
CAUTION:
It is important to select the correct spool type because
this parameter is used in device control.
Erroneous value may cause instability.
WARNING:
Automatic calibration drives the valve against the
mechanical open and closed travel limits of the valve-
actuator assembly and a tuning procedure is per-
formed. Make sure that these procedures can be safely
executed.
NOTE:
Make sure to run the pneumatics test before the auto-
matic travel calibration.
Fig. 15 Calibration selection
NOTE:
Valve position needs to be in the normal operating posi-
tion, supply pressure needs to be in valid range, no supply
pressure drop is allowed and any test cannot be active
when calibration is started.
Pneumatics test needs to be performed succesfully before
the calibration.
WARNING:
Partial Stroke Test moves the valve according to the
stroke size. Make sure that this procedure can be
safely executed.