  9. Micromeritics ASAP 2420 1-mmHg Transducer Option User manual

Micromeritics ASAP 2420 1-mmHg Transducer Option User manual

Feb 2010 1
Installing the ASAP 2420 1-mmHg Transducer Option
(Units with Serial Number 199 and below)
P/N 242-42805-00
This document contains instructions for installing the 1-mmHg Transducer option only or for
installing both the 1-mmHg transducer and micropore options on an ASAP 2420 analyzer.
Follow the set of instructions that applies to the customer’s system.
Tools Required
The following tools are required for installation of the transducer option:
• 1/2-in. wrench
• 9/16-in. wrench
• 3/4-in. wrench
• Phillips screwdriver
• Small flat-blade screwdriver
Installing the 1-mmHg Transducer Option Only
This set of instructions describes how to install the 1-mmHg Transducer option on an ASAP
2420 system equipped with the Micropore option.
1. Backfill the manifold to 760 mmHg.
2. Close the ASAP 2420 software.
3. Turn off power to the vacuum pumps and to the analyzer.
4. Remove the four retaining screws from the top panel of the ASAP 2420.
Installing the 1-mmHg Transducer Option Only ASAP 2420 1-mmHg Transducer Option
2 Feb 2010
5. Lift the panel up and off the analyzer.
6. The 10-mmHg transducer must be removed from the top of the old transducer block and
installed on the front of the new block, and the 1-mmHg transducer must be installed on
top of the new block. For reference, the following photograph shows the location of the
transducers after installation.
Location of 1-mmHg transducer mounted on top of transducer block
Location of 10-mmHg transducer mounted on front of transducer block
Analyzer Top Panel
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer Option Only
Feb 2010 3
7. To remove the 10-mmHg transducer, first disconnect the 10-mmHg transducer ribbon
cable from the J8 connector on the transducer board (P/N 202-17706-001).
8. Use a wrench to loosen the nut that connects the 10-mmHg transducer to the transducer
block. Retain the nut, you will use it to reinstall the transducer.
9. Remove the transducer from the block.
J8
10-mmHg
transducer
Connector nut
Transducer
block
Installing the 1-mmHg Transducer Option Only ASAP 2420 1-mmHg Transducer Option
4 Feb 2010
10. Use a wrench to loosen the nuts that connect the transducer block to the manifold tee
connectors and to the gas inlet line, then remove the block.
11. Remove the block mounting screws, connector nuts, and the spacer and ferrules from each
tee connector.
Remove connector nuts
Connector nut
Rear ferrule
Front ferrule
Spacer
Front and rear
ferrules (under
nut)
Screw
Lock washer
Old Block
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer Option Only
Feb 2010 5
12. Assemble the new transducer block, spacers, ferrules and connector nuts.
13. Insert the mounting screws and lock washers into the block and tighten.
14. Use a wrench to tighten the nuts that connect the transducer block to the manifold tee
connectors and to the gas inlet line.
Connector nut
Front ferrule
Spacer
Front and rear
ferrules (under
nut)
Rear ferrule
Screw
Lock washer
New Block
Use new spacers, ferrules, screws and lock washers from the upgrade kit. Do not
replace the ones you removed, they may be damaged.
Tighten connector nuts
Installing the 1-mmHg Transducer Option Only ASAP 2420 1-mmHg Transducer Option
6 Feb 2010
15. Place the connector nut, and front and rear ferrules on the 1-mmHg transducer.
16. Install the transducer on top of the transducer block; tighten securely with a wrench.
17. Connect the ribbon cable to the 1-mmHg transducer and to J7 on the transducer board
(202-17706-001).
Nut
Front ferrule
Rear ferrule
Transducer block
Connector nut (between
transducer and block)
J7
1-mmHg transducer
Transducer board
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer Option Only
Feb 2010 7
18. Install the 10-mmHg transducer, with new front and rear ferrules, in the front of the
transducer block. Use the connector nut you removed earlier; tighten the nut securely with
a wrench.
19. Make sure the zero adjustment on the 10-mmHg transducer is accessible. You may have to
cut away some of the foam if it interferes with the transducer.
10-mmHg transducer Connector nut
Use new ferrules from the upgrade kit. Do not replace the ones you removed, they may
be damaged.
Zero
adjustment
Installing the 1-mmHg Transducer Option Only ASAP 2420 1-mmHg Transducer Option
8 Feb 2010
20. Attach the ribbon cable from the 10-mmHg transducer to the J8 connector on the
transducer board.
21. Replace the top panel.
22. Replace and tighten the top panel retaining screws.
23. Turn on power to the analyzer and to the vacuum pumps.
24. Install the updated software (refer to the ASAP 2420 Operator’s Manual for instructions).
25. Open the ASAP 2420 software.
J8
10-mmHg
transducer
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer Option Only
Feb 2010 9
26. Select Unit [n] > Show Instrument Schematic > Enable Manual Control.
27. Open valves 7 and 8. Make sure the 1-mmHg and 10-mmHg transducer readings are
displayed on the schematic.
28. Perform a leak test (refer to Appendix D of the ASAP 2420 Operator’s Manual).
29. Calibrate the 1-mmHg, 10-mmHg and 1000-mmHg transducers (refer to Product
Calibration Procedure for the ASAP 2420 (P/N 242-34000-76 Rev C).
30. Perform the following procedures from the Product Calibration Procedure for the ASAP
2420, P/N 242-34000-76 Rev C:
• 5.8 Analysis Servo and Transducer Accuracy Test
• 5.10 System Volume Calibration
• 5.11 Blank Tube Test (Nitrogen)
• 5.12 Reference Chamber Analysis
• 5.13 Nitrogen Standards Test (N2 Surface Area)
• 5.14 Performing a Krypton Blank Tube Test
• 5.15 Krypton Standards Test
1-mmHg transducer 10-mmHg transducer
Installing the 1-mmHg Transducer and Micropore Options ASAP 2420 1-mmHg Transducer Option
10 Feb 2010
Installing the 1-mmHg Transducer and Micropore Options
This set of instructions describes how to install the 1-mmHg Transducer option and the
Micropore option on ASAP 2420 Krypton systems.
1. Backfill the manifold to 760 mmHg.
2. Close the ASAP 2420 software.
3. Turn off power to the vacuum pumps and to the analyzer.
4. Remove the four retaining screws from the top panel of the ASAP 2420.
5. Lift the panel up and off the analyzer.
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer and Micropore Options
Feb 2010 11
6. The 10-mmHg transducer must be removed from the top of the old transducer block and
installed on the front of the new block, and the 1-mmHg transducer must be installed on
top of the new block. For reference, the following photograph shows the location of the
transducers after installation
7. To remove the 10-mmHg transducer, first disconnect the 10-mmHg transducer ribbon
cable from the J8 connector on the transducer board (P/N 202-17706-001).
Location of 1-mmHg transducer mounted on top transducer block
Location of 10-mmHg transducer mounted on front of transducer block
Analyzer Top Panel
J8
Installing the 1-mmHg Transducer and Micropore Options ASAP 2420 1-mmHg Transducer Option
12 Feb 2010
8. Use a wrench to loosen the nut that connects the 10-mmHg transducer to the transducer
block. Retain the nut, you will use it to reinstall the transducer
9. Remove the transducer from the block.
10. Use a wrench to loosen the nuts that connect the transducer block to the manifold tee
connectors and to the gas inlet line, then remove the block.
10-mmHg
transducer
Connector nut
Transducer
block
Remove connector nuts
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer and Micropore Options
Feb 2010 13
11. Remove the block mounting screws, connector nuts, and the spacer and ferrules from each
tee connector.
12. Assemble the new transducer block, spacers, ferrules and connector nuts.
13. Insert the mounting screws and lock washers into the block and tighten.
Connector nut
Rear ferrule
Front ferrule
Spacer
Front and rear
ferrules (under
nut)
Screw
Lock washer
Old Block
Connector nut
Front ferrule
Spacer
Front and rear
ferrules (under
nut)
Rear ferrule
Screw
Lock washer
New Block
Use the new spacers, ferrules, screws and lock washers in the upgrade kit. Do not
replace the ones you removed, they may be damaged.
Installing the 1-mmHg Transducer and Micropore Options ASAP 2420 1-mmHg Transducer Option
14 Feb 2010
14. Use a wrench to tighten the nuts that connect the transducer block to the manifold tee
connectors and to the gas inlet line.
15. Place the connector nut, and front and rear ferrules on the 1-mmHg transducer.
Tighten connector nuts
Nut
Front ferrule
Rear ferrule
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer and Micropore Options
Feb 2010 15
16. Install the transducer on top of the transducer block; tighten securely with a wrench.
17. Connect the ribbon cable to the 1-mmHg transducer and to J7 on the transducer board
(202-17706-001).
18. Install the 10-mmHg transducer, with new front and rear ferrules, in the front of the
transducer block. Use the connector nut you removed earlier; tighten the nut securely with
a wrench.
Transducer block
Connector nut (between
transducer and block)
J7
1-mmHg transducer
Transducer board
10-mmHg transducer Connector nut
Use new ferrules from the upgrade kit. Do not replace the ones you removed, they may
be damaged.
Installing the 1-mmHg Transducer and Micropore Options ASAP 2420 1-mmHg Transducer Option
16 Feb 2010
Make sure the zero adjustment on the 10-mmHg is accessible. You may have to cut away
some of the foam if it interferes with the transducer.
19. Attach the ribbon cable from the 10-mmHg transducer to the J8 connector on the
transducer board.
Zero
adjustment
J8
10-mmHg
transducer
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer and Micropore Options
Feb 2010 17
20. Plug the Micropore key into the J10 connector on the transducer board. Use a screwdriver
to tighten the retaining screws.
21. Replace the top panel.
22. Replace and tighten the top panel retaining screws.
23. Cover each of the Dewars.
24. Replace the port fitting in each of the six analysis ports as described in steps
25 through 29. The port fittings must be replaced to accommodate the heating mantles
used for degassing on analysis ports.
25. Using a wrench, remove the sample tube fitting from the analyzer.
Micropore key (J10)
Place a Dewar cover on each Dewar before replacing port fittings. This prevents parts
from falling into the Dewars and possibly causing breakage.
Sample tube
fitting
Installing the 1-mmHg Transducer and Micropore Options ASAP 2420 1-mmHg Transducer Option
18 Feb 2010
26. Remove the filter and O-ring.
27. The filter and O-ring can be used with the new port fittings unless either is damaged. A
damaged filter or O-ring should be replaced with a new one.
28. Place the filter and the O-ring on the new port fitting.
29. Install the port fitting; securely tighten the fitting with a wrench to prevent leaks.
30. Install the updated software (refer to the ASAP 2420 Operator’s Manual for instructions).
31. Turn on power to the analyzer and to the vacuum pumps.
32. Open the ASAP 2420 software.
O-ring
Filter
To avoid analysis problems, the filter and O-ring must be clean. Do not touch
them with bare hands.
ASAP 2420 1-mmHg Transducer Option Installing the 1-mmHg Transducer and Micropore Options
Feb 2010 19
33. Select Unit [n] > Show Instrument Schematic > Enable Manual Control.
34. Open valves 7 and 8. Make sure the 1-mmHg and 10-mmHg transducer readings are
displayed on the schematic.
35. Perform a leak test (refer to Appendix D of the ASAP 2420 Operator’s Manual).
36. Calibrate the 1-mmHg, 10-mmHg and 1000-mmHg transducers (refer to Product
Calibration Procedure for the ASAP 2420 (P/N 242-34000-76 Rev C).
37. Perform the following procedures from the Product Calibration Procedure for the ASAP
2420, P/N 242-34000-76 Rev C:
• 5.8 Analysis Servo and Transducer Accuracy Test
• 5.10 System Volume Calibration
• 5.11 Blank Tube Test (Nitrogen)
• 5.12 Reference Chamber Analysis
• 5.13 Nitrogen Standards Test (N2 Surface Area)
• 5.14 Performing a Krypton Blank Tube Test
• 5.15 Krypton Standards Test
• 5.19 Micropore Standards Test
1-mmHg transducer 10-mmHg transducer
Installing the 1-mmHg Transducer and Micropore Options ASAP 2420 1-mmHg Transducer Option
20 Feb 2010

This manual suits for next models

1

Other Micromeritics Measuring Instrument manuals

Micromeritics SmartPrep 065 User manual

Micromeritics

Micromeritics SmartPrep 065 User manual

Micromeritics ACCUPYC II 1345 Series User manual

Micromeritics

Micromeritics ACCUPYC II 1345 Series User manual

Micromeritics AccuPyc 1330 User manual

Micromeritics

Micromeritics AccuPyc 1330 User manual

Micromeritics GEMINI VII 2390 User manual

Micromeritics

Micromeritics GEMINI VII 2390 User manual

Micromeritics GEMINI VII KEYPAD User manual

Micromeritics

Micromeritics GEMINI VII KEYPAD User manual

Micromeritics SediGraph III 5120 User manual

Micromeritics

Micromeritics SediGraph III 5120 User manual

Micromeritics TriStar II 3020 User manual

Micromeritics

Micromeritics TriStar II 3020 User manual

Micromeritics ACCUPYC II 1345 Series User manual

Micromeritics

Micromeritics ACCUPYC II 1345 Series User manual

Micromeritics ASAP 2020 Confirm User manual

Micromeritics

Micromeritics ASAP 2020 Confirm User manual

Micromeritics Flowsorb II 2300 User manual

Micromeritics

Micromeritics Flowsorb II 2300 User manual

Micromeritics AquaPrep 055 User manual

Micromeritics

Micromeritics AquaPrep 055 User manual

Micromeritics MIC SAS II 5800 User manual

Micromeritics

Micromeritics MIC SAS II 5800 User manual

Micromeritics TriStar II 3020 User manual

Micromeritics

Micromeritics TriStar II 3020 User manual

Micromeritics AutoPore V Series User manual

Micromeritics

Micromeritics AutoPore V Series User manual

Micromeritics ASAP 2460 User manual

Micromeritics

Micromeritics ASAP 2460 User manual

Micromeritics TriStar II 3020 User manual

Micromeritics

Micromeritics TriStar II 3020 User manual

Micromeritics GEMINI VII 2390 User manual

Micromeritics

Micromeritics GEMINI VII 2390 User manual

Micromeritics DVVA II 4000 User manual

Micromeritics

Micromeritics DVVA II 4000 User manual

Micromeritics 3Flex User manual

Micromeritics

Micromeritics 3Flex User manual

Micromeritics GEMINI VII 2390 User manual

Micromeritics

Micromeritics GEMINI VII 2390 User manual

Micromeritics TriStar II 3020 Technical specifications

Micromeritics

Micromeritics TriStar II 3020 Technical specifications

Micromeritics ACCUPYC III User manual

Micromeritics

Micromeritics ACCUPYC III User manual

Micromeritics AccuPyc II 1340 User manual

Micromeritics

Micromeritics AccuPyc II 1340 User manual

Micromeritics Gemini VII 2390a User manual

Micromeritics

Micromeritics Gemini VII 2390a User manual

Popular Measuring Instrument manuals by other brands

Bender RCM420 Series manual

Bender

Bender RCM420 Series manual

Lumel ND03 user manual

Lumel

Lumel ND03 user manual

TLV EF200R-C technical information

TLV

TLV EF200R-C technical information

MG DLS-A1 instruction manual

MG

MG DLS-A1 instruction manual

Hanna Instruments Checker HI772 instruction manual

Hanna Instruments

Hanna Instruments Checker HI772 instruction manual

Wilcom F6121A operating instructions

Wilcom

Wilcom F6121A operating instructions

Bosch FAA-420-RI installation guide

Bosch

Bosch FAA-420-RI installation guide

AEMC instruments Simple Logger AmpFlex AL24-2500 user manual

AEMC instruments

AEMC instruments Simple Logger AmpFlex AL24-2500 user manual

Polhemus PATRIOT DB installation guide

Polhemus

Polhemus PATRIOT DB installation guide

Omtech 3020 user manual

Omtech

Omtech 3020 user manual

EBCHQ 95608 quick start guide

EBCHQ

EBCHQ 95608 quick start guide

Megger MCCV6000-18 Guide

Megger

Megger MCCV6000-18 Guide

baxter ML-132392 Operator's & installation manual

baxter

baxter ML-132392 Operator's & installation manual

Vega VEGAPULS 61 Quick setup guide

Vega

Vega VEGAPULS 61 Quick setup guide

techem HT2 installation instructions

techem

techem HT2 installation instructions

Envent 331S user manual

Envent

Envent 331S user manual

Lovibond MD200 instruction manual

Lovibond

Lovibond MD200 instruction manual

Siemens QFM81 Series Technical instructions

Siemens

Siemens QFM81 Series Technical instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.