manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Milwaukee
  6. •
  7. Measuring Instrument
  8. •
  9. Milwaukee MW802 User manual

Milwaukee MW802 User manual

This manual suits for next models

1

Other Milwaukee Measuring Instrument manuals

Milwaukee MR32ATC User manual

Milwaukee

Milwaukee MR32ATC User manual

Milwaukee PRO MI415 User manual

Milwaukee

Milwaukee PRO MI415 User manual

Milwaukee MW 14 User manual

Milwaukee

Milwaukee MW 14 User manual

Milwaukee MI405 User manual

Milwaukee

Milwaukee MI405 User manual

Milwaukee MA887 User manual

Milwaukee

Milwaukee MA887 User manual

Milwaukee M12 3PLA User manual

Milwaukee

Milwaukee M12 3PLA User manual

Milwaukee L4 CLL User manual

Milwaukee

Milwaukee L4 CLL User manual

Milwaukee MW190 User manual

Milwaukee

Milwaukee MW190 User manual

Milwaukee 3421 User manual

Milwaukee

Milwaukee 3421 User manual

Milwaukee M12 C4PLA User manual

Milwaukee

Milwaukee M12 C4PLA User manual

Milwaukee 2270-20 User manual

Milwaukee

Milwaukee 2270-20 User manual

Milwaukee PH55 User manual

Milwaukee

Milwaukee PH55 User manual

Milwaukee MW10 User manual

Milwaukee

Milwaukee MW10 User manual

Milwaukee Smart Monitoring System SMS110 User manual

Milwaukee

Milwaukee Smart Monitoring System SMS110 User manual

Milwaukee CALIBRATION 2238-20 Operating instructions

Milwaukee

Milwaukee CALIBRATION 2238-20 Operating instructions

Milwaukee LDM 80 User manual

Milwaukee

Milwaukee LDM 80 User manual

Milwaukee SM700 User manual

Milwaukee

Milwaukee SM700 User manual

Milwaukee LDM 45 User manual

Milwaukee

Milwaukee LDM 45 User manual

Milwaukee REDLITHIUM 3521-20 User manual

Milwaukee

Milwaukee REDLITHIUM 3521-20 User manual

Milwaukee Mi453 User manual

Milwaukee

Milwaukee Mi453 User manual

Milwaukee 2235-40 User manual

Milwaukee

Milwaukee 2235-40 User manual

Milwaukee MW101 User manual

Milwaukee

Milwaukee MW101 User manual

Milwaukee L4 CLLP User manual

Milwaukee

Milwaukee L4 CLLP User manual

Milwaukee 2235-20 User manual

Milwaukee

Milwaukee 2235-20 User manual

Popular Measuring Instrument manuals by other brands

Suaoki D40 user manual

Suaoki

Suaoki D40 user manual

Elster BK-G1.6 operating instructions

Elster

Elster BK-G1.6 operating instructions

Endress+Hauser Deltabar FMD71 technical information

Endress+Hauser

Endress+Hauser Deltabar FMD71 technical information

Duratool D03128 manual

Duratool

Duratool D03128 manual

Vega VEGAPULS 65 Quick setup guide

Vega

Vega VEGAPULS 65 Quick setup guide

Tonghui TH2618B Operation manual

Tonghui

Tonghui TH2618B Operation manual

Tenmars TM-801 user manual

Tenmars

Tenmars TM-801 user manual

MSR 145 user manual

MSR

MSR 145 user manual

TPS WP-82 manual

TPS

TPS WP-82 manual

PCB Piezotronics 003C30 Installation and operating manual

PCB Piezotronics

PCB Piezotronics 003C30 Installation and operating manual

Tenma 72-10465 manual

Tenma

Tenma 72-10465 manual

Endress+Hauser Proline Promag 50 Brief operating instructions

Endress+Hauser

Endress+Hauser Proline Promag 50 Brief operating instructions

Siemens SITRANS F Coriolis FCT030 Function manual

Siemens

Siemens SITRANS F Coriolis FCT030 Function manual

KLINGER CMF V Series instruction manual

KLINGER

KLINGER CMF V Series instruction manual

Keysight M8290A Getting started guide

Keysight

Keysight M8290A Getting started guide

Acterna MS1400 Operation manual

Acterna

Acterna MS1400 Operation manual

R&S ZPH manual

R&S

R&S ZPH manual

Datacolor ColorReader DC10-2 user guide

Datacolor

Datacolor ColorReader DC10-2 user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

USER MANUAL
PORTABLE pH/EC/TDS METER
MODEL: MW801, MW802
Smart Combined Meter
WARRANTY:
This instrument is warranted from all defects in materials
and manufacturing for a period of two years from the
date of purchase.
The probe is warranted for a period of six months.
If during this period the repair or replacement of parts
is required, where the damage is not due to negligence
or erroneous operation by the user, please return the
parts to either distributor or our office and the repair will
be effected free of change.
Note: We reserve the right to modify the design,
construction and appearance of our products
without advance notice.
PRELIMINARY EXAMINATION
Remove the instrument from the packing material
and examine it carefully to make sure that no
damage has occurred during shipment. If noticeable
damage is found, notify your Dealer.
Each meter is supplied complete with:
•MA850DIN preamplified non refillable double
junction combination pH electrode with built
in EC/TDS probe and temperature sensor;
1m (3.3') cable.
• pH7.01, 1413 µS/cm & 1382 ppm (for MW801)
or 1500 ppm (for MW802) solutions, 20 mL
each.
• Instruction Manual.
• 1 x 9V battery.
OPERATION
Slide the battery compartment cover located on
the back of the meter off, and install the battery
into the battery clip connector while observing
polarity.
Connect the probe to the meter
securely by aligning the pins
with the socket and pushing
the plug in.
Always remove the electrode
protective cap before use.
Make sure that the meter has
been calibrated before taking
any measurements (see
calibration procedure).
Turn the instrument on by pressing the ON/OFF key.
pH MEASUREMENTS
If the electrode has been left dry, soak the tip in
a pH7 (M10007) buffer solution for a few minutes
to reactivate it.
Submerge the tip of the electrode into the sample
to be measured, select the pH mode and allow
the reading to stabilize before taking measurements.
EC/TDS MEASUREMENTS
Immerse the tip of the electrode into the sample
to be tested.
Select the measurement range (EC or TDS)
and wait for the temperature sensor to reach
thermal equilibrium with the samples before
taking measurements.
AFTER MEASUREMENTS the instrument should
be switched off and the probe cleaned and stored
with a few drops of storage solution in the
protective cap.
CALIBRATION PROCEDURE:
Remove the protective cap from the electrode.
Switch the meter on.
Select the range pressing the appropriate key for
pH, EC or TDS.
pH Calibration
A) PREPARATION
Single point manual calibration. Choose buffer.
If you are going to measure samples near pH 7
use pH 7.01 buffer solution (M10007B) for calibration;
use pH 4.01 buffer solution
(M10004B) for acidic
measurements, or pH 10.01
(M10010B) for alkaline
measurements.
B) PROCEDURE
Rinse the tip of the electrode and immerse it in
the calibration solution. Wait a couple of minutes
for the reading to stabilize.
Measure the temperature of the
buffer solution with a thermometer.
Adjust the calibration knob until the
LCD shows the pH value of the
buffer at the measured temperature
(see pH versus temperature chart).
The calibration is now complete and the meter is
ready for use.
Milwauke
e
Milwauke
e
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
EC/TDS calibration:
A) PREPARATION
Single point EC calibration.
Choose appropriate standard:
Model MW801: use 1382 ppm
(M10032B) for TDS or 1.413 μS/cm (M10031B)
for EC calibration.
Model MW802: use 1500 ppm (M10442B) for
TDS or 1.413 mS/cm (M10031B) for EC calibration.
B) PROCEDURE
Immerse the tip of the electrode in the standard
solution.
Wait for the reading to stabilize, and thermal
equilibrium be reached.
Turn the EC/TDS calibration
knob until the display shows
the EC or TDS reading at
25°C.
The calibration is now complete
and the meter is ready for use.
The instrument should be re-calibrated at least
once a month, or when the probe or battery is
replaced.
Note:The conversion between EC and TDS is
made by a built-in circuit. If calibration
is made in EC range, TDS range is
automatically calibrated (or vice versa).
BATTERY REPLACEMENT:
The meters are powered by a 9V battery that is
in a compartment located in the rear of the
instrument.
When the battery becomes weak, the instrument
automatically switches off.
Replacement must only take place in a non-hazardous
area using an alkaline 9V battery. Slide off the
battery compartment cover and replace the old
battery with a new one. Make sure that the battery
contacts are fully engaged in the connector. Seat
the battery in is compartment and replace the
cover.
pH VERSUS TEMPERATURE CHART
OPTIONAL ACCESSORIES
M10007B pH7.01 buffer solution,
25x20 mL sachet
M10004B pH4.01 buffer solution,
25x20 mL sachet
M10010B pH10.01 buffer solution,
25x20 mL sachet
M10031B 1413 μS/cm calibration
solution, 25x20 mL sachet
M10032B 1382 ppm calibration
solution, 25x20 mL sachet
M10442B 1500 ppm calibration
solution, 25x20 mL sachet
MA9015 Electrode storage
solution, 220 mL bottle
MA850DIN pH/EC/TDS spare probe
with 1m cable
PMETSEULAVHp
C°F°4009AM7009AM0109AM
02310.431.723.01
51400.401.742.01
010500.470.781.01
519500.440.721.01
028600.430.760.01
527710.410.710.01
036820.400.769.9
535930.499.629.9
0440140.489.658.9
5431150.489.658.9
0522160.489.628.9
5513170.489.697.9
0604190.489.677.9
5694111.499.667.9
0785121.499.657.9
SPECIFICATIONS:
MW801 MW802
RANGE 0.0 to 14.0 pH 0.00 to 14.00 pH
0 to 1990 ppm 0 to 4000 ppm
0 to 1990 μS/cm 0.00 to 6.00 mS/cm
RESOLUTION 0.1 pH 0.10 pH
10 ppm 10 ppm
10 μS/cm 0.01 mS/cm
ACCURACY ± 0.2 pH ±0.20 pH
(@ 20ºC) ±2% Full scale ±2% Full scale
CALIBRATION M10007B (pH 7.01)
SOLUTIONS M10031B (1.413 mS/cm)
M10032B (1382ppm) M10442B (1500ppm)
CONVERSION 0.5 0.68
FACTOR
CALIBRATION Manual, at 1 point
TEMPERATURE COMPENSATION
Automatic from 0 to 50ºC
PROBE MA850DIN combination pH/EC/TDS probe
ENVIRONMENT 0 to 50ºC, 95%RH max
BATTERY LIFE /TYPE 1x9V alkaline
approx. 150 hours of use
DIMENSIONS 143 x 80 x 32 mm
WEIGHT 220 g (with battery)