SPECIFICATIONS
Range 0.0 to 14.0 pH
Resolution 0.1 pH
Accuracy (@25 °C) ±0.1 pH
Calibration Manual, at 2 points
through trimmers
pH electrode MA73600
Environment 0 to 50 °C; 100% RH
Battery type 3 x 1.5V, alkaline
Battery life more than 1500 hours
of continuous use
Dimensions 165 x 30 x 30 mm
Weight 80 g
ACCESSORIES
MA73600 Replaceable pH electrode
M10004B pH 4.01 buer, 20 ml sachet,
25pcs.
M10007B pH 7.01 buer, 20 ml sachet,
25pcs.
M10010B pH 10.01 buer, 20 ml sachet,
25pcs.
MA9015 Electrode storage solution,
230ml
MA9016 Electrode cleaning solution,
230ml
M10000B Electrode rinse solution, 20 ml
sachet, 25 pcs.
CERTIFICATION
Milwaukee Instruments conform to the CE European
Directives.
Disposal of Electrical & Electronic Equipment. Do not
treat this product as household waste. Hand it over to the
appropriate collection point for the recycling of electrical
and electronic equipment.
Disposal of waste batteries. This product contains
batteries. Do not dispose of them with other household
waste. Hand them over to the appropriate collection point
for recycling.
Please note: proper product and battery disposal
prevents potential negative consequences for human
health and the environment. For detailed information,
contact your local household waste disposal service or
go to www.milwaukeeinstruments.com (USA & CAN)
or www.milwaukeeinst.com.
RECOMMENDATION
Before using this product, make sure it is entirely suitable for
your specific application and for the environment in which it is
used. Any modification introduced by the user to the supplied
equipment may compromise the meter’s performance. For
your and the meter’s safety do not use or store the meter
in hazardous environment. To avoid damage or burn, do not
perform any measurement in microwave ovens.
WARRANTY
This instrument is warranted against defects in materials
and manufacturing for a period of 2years from the date of
purchase. Electrode is warranted for 1 year. This warranty
is limited to repair or free of charge replacement if the
instrument cannot be repaired. Damage due to accidents,
misuse, tampering or lack of prescribed maintenance is
not covered by warranty. If service is required, contact
your local Milwaukee Instruments Technical Service. If the
repair is not covered by the warranty, you will be notified
of the charges incurred. When shipping any meter, make
sure it is properly packaged for complete protection.
Milwaukee Instruments reserves the right to make
improvements in design, construction and appearance of
its products without advance notice.
ISTPH51 09/20
USER MANUAL
SATISFACTION
GUARANTEED
milwaukeeinstruments.com (USA & CAN)
milwaukeeinst.com
pH51
PRO Waterproof
pH Tester
CONDITIONING
Remove the cap and immerse the pH51
in MA9015 storage solution for 2 hours to
activate the electrode.
CALIBRATION
Unscrew the battery compartment cap on
the top of the meter and perform one of the
following two-point calibration procedure,
according to the required application:
Acid calibration for measurements below
pH7:
yImmerse the electrode in pH7.01 (M10007)
buer solution.
yAllow the reading to stabilize and with a
small screwdriver turn the “7” calibration
trimmer inside the battery compartment
until the display shows “7.0”.
yRinse the electrode with tap water and
immerse it in pH4.01 (M10004) buer
solution.
yAllow the reading to stabilize and with a
small screwdriver turn the “4” calibration
trimmer until the display shows “4.0”.
Alkaline calibration for measurements
above pH7:
yImmerse the electrode in pH7.01 (M10007)
buer solution.
yAllow the reading to stabilize and with a
small screwdriver turn the “7” calibration
trimmer until the display shows “7.0”.
yRinse the electrode with tap water and
immerse it in pH10.01 (M10010) buer
solution.
yAllow the reading to stabilize and with a
small screwdriver turn the “4” calibration
trimmer until the display shows “10.0”.
Replace the battery compartment cap.
OPERATION
yRemove the protective cap.
yTurn the pH51 on by pressing the ON key.
yImmerse the electrode in the solution to be
tested.
yStir gently and wait for the reading to
stabilize.
yAfter use, rinse the electrode with water to
minimize contamination.
yStore the electrode with a few drops
of storage (MA9015) or pH7 (M10007)
solution in the protective cap.
yAlways replace the protective cap after use.
Never use distilled or deionized water for
storage purposes.
ELECTRODE REPLACEMENT
The electrode can be easily replaced in the
following way:
yRemove the protective cap.
yUnscrew the plastic ring on the top of the
electrode.
yPull out the MA73600 pH electrode and
replace it with a new one (aligning the
colored pin towards the LCD side).
yMake sure the gaskets are in place before
screwing back the ring.
BATTERY REPLACEMENT
To replace the batteries unscrew the battery
compartment cap and replace all three 1.5V
batteries while paying attention to their
polarity. Make sure the gasket is in place
before screwing back the cap.
Batteries should only be replaced in a
nonhazardous area using the battery type
specified in this instruction manual.

Other Milwaukee Test Equipment manuals

Milwaukee 2237-40 User manual

Milwaukee

Milwaukee 2237-40 User manual

Milwaukee 2210-20 User manual

Milwaukee

Milwaukee 2210-20 User manual

Milwaukee MW804 User manual

Milwaukee

Milwaukee MW804 User manual

Milwaukee MW10 User manual

Milwaukee

Milwaukee MW10 User manual

Milwaukee CD97 User manual

Milwaukee

Milwaukee CD97 User manual

Milwaukee MW12 User manual

Milwaukee

Milwaukee MW12 User manual

Milwaukee pH54 User manual

Milwaukee

Milwaukee pH54 User manual

Milwaukee pH55 PRO User manual

Milwaukee

Milwaukee pH55 PRO User manual

Milwaukee MW11 User manual

Milwaukee

Milwaukee MW11 User manual

Milwaukee pH58 User manual

Milwaukee

Milwaukee pH58 User manual

Milwaukee HEAVY DUTY 2216-40 User manual

Milwaukee

Milwaukee HEAVY DUTY 2216-40 User manual

Milwaukee MW14 User manual

Milwaukee

Milwaukee MW14 User manual

Milwaukee pH600 User manual

Milwaukee

Milwaukee pH600 User manual

Milwaukee ORP57 User manual

Milwaukee

Milwaukee ORP57 User manual

Milwaukee MW803 User manual

Milwaukee

Milwaukee MW803 User manual

Milwaukee EC59 User manual

Milwaukee

Milwaukee EC59 User manual

Milwaukee MW14 User manual

Milwaukee

Milwaukee MW14 User manual

Milwaukee ORP57 User manual

Milwaukee

Milwaukee ORP57 User manual

Popular Test Equipment manuals by other brands

Uson SPRINT LC owner's guide

Uson

Uson SPRINT LC owner's guide

Tektronix TDS 500B Technical reference

Tektronix

Tektronix TDS 500B Technical reference

Agilent Technologies e6601a Getting started guide

Agilent Technologies

Agilent Technologies e6601a Getting started guide

HT FULLTEST3 user manual

HT

HT FULLTEST3 user manual

Wavecom WCM-RCD manual

Wavecom

Wavecom WCM-RCD manual

Tietzsch MultiSafe DSP-HW 2 User instructions

Tietzsch

Tietzsch MultiSafe DSP-HW 2 User instructions

JDS Uniphase SmartClass E1 reference guide

JDS Uniphase

JDS Uniphase SmartClass E1 reference guide

Perel EEM800 user manual

Perel

Perel EEM800 user manual

Hioki 3550 instruction manual

Hioki

Hioki 3550 instruction manual

agratronix MT-PRO+ Operator's manual

agratronix

agratronix MT-PRO+ Operator's manual

Crane Electronics WM10K Operator's manual

Crane Electronics

Crane Electronics WM10K Operator's manual

LeCroy HDO6000 Getting started guide

LeCroy

LeCroy HDO6000 Getting started guide

GE DMS Go+ operating manual

GE

GE DMS Go+ operating manual

HT COMBI521 user manual

HT

HT COMBI521 user manual

Applent Instruments AT851 user guide

Applent Instruments

Applent Instruments AT851 user guide

TESTO 830-T4 instruction manual

TESTO

TESTO 830-T4 instruction manual

Camille Bauer GOSSEN METRAWATT GEOHM C operating instructions

Camille Bauer

Camille Bauer GOSSEN METRAWATT GEOHM C operating instructions

Rigol DS-5000 Series user manual

Rigol

Rigol DS-5000 Series user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.