SPECIFICATIONS
Range 0.00 to 5.00 ppm
Resolution 0.01 ppm
Accuracy ±0.04 ppm ±2% of reading
@ 25 °C
Typical EMC Dev. ±0.01 ppm
Light Source Light Emitting Diode
@525nm
Light Detector Silicon Photocell
Method
Adaptation of the EPA
Phenantroline method
315B, for natural and treated
waters. The reaction between
iron and reagent causes an
orange tint in the sample.
Environment 0 to 50°C (32 to 122 °F)
max. 95% RH non-condensing
Battery Type 1 x 1.5V AAA
Auto-Shut O After 3 minutes of non-use
and 2 minutes after reading
Dimensions 87.5 x 63.7 x 35.2 mm
(3.4 x 2.5 x 1.4”)
Weight 65 g (2.29 oz.)
OPTIONAL ACCESSORIES
Reagent Sets
MI528-25 Iron powder reagent, 25 pcs.
Other Accessories
3000100 Glass cuvettes with caps (2 pcs.)
3000200 Tissue fow wiping cuvettes
(2pcs.)
3000300 1.5V AAA batteries (4 pcs.)
CERTIFICATION
Milwaukee Instruments conform to
the CE European Directives.
Disposal of Electrical & Electronic
Equipment. Do not treat this product
as household waste. Hand it over
to the appropriate collection point
for the recycling of electrical and
electronic equipment.
Disposal of waste batteries. This product
contains batteries. Do not dispose of them with
other household waste. Hand them over to the
appropriate collection point for recycling.
Please note: proper product and battery disposal
prevents potential negative consequences
for human health and the environment. For
detailed information, contact your local
household waste disposal service or go to
www.milwaukeeinstruments.com (USA & CAN)
or www.milwaukeeinst.com.
RECOMMENDATION
Before using this product, make sure it is entirely
suitable for your specific application and for the
environment in which it is used. Any modification
introduced by the user to the supplied equipment
may compromise the tester’s performance. For
your and the tester’s safety do not use or store
the tester in hazardous environment. To avoid
damage or burn, do not perform any measurement
in microwave ovens.
WARRANTY
This instrument is warranted against defects
in materials and manufacturing for a period of
2years from the date of purchase. This warranty
is limited to repair or free of charge replacement if
the instrument cannot be repaired. Damage due to
accidents, misuse, tampering or lack of prescribed
maintenance is not covered by warranty. If
service is required, contact your local Milwaukee
Instruments Technical Service. If the repair is not
covered by the warranty, you will be notified of the
charges incurred. When shipping any instrument,
make sure it is properly packaged for complete
protection.
ISTMW14 11/20
USER MANUAL
SATISFACTION
GUARANTEED
milwaukeeinstruments.com (USA & CAN)
milwaukeeinst.com
MW14 Digital
Iron Tester
PRELIMINARY EXAMINATION
Please examine this product carefully. Make
sure that the instrument is not damaged.
If any damage occured during shipment,
please notify your Dealer.
Each MW14 tester is supplied complete with:
yTwo Sample Cuvettes and Caps
ySix powder reagents for Iron High Range
y1 x 1.5V AAA Battery
yInstruction Manual
FUNCTIONAL DESCRIPTION
1. Dust cover
2. Cuvette with cap
3. Cuvette holder
4. Liquid Crystal Display
5. Button
OPERATION
yTurn the tester on by pressing the button.
After all the segments are displayed, “C.1”,
“Add” appears with “Press” blinking, the
tester is ready.
yFill the cuvette with 10 mL of unreacted
sample and replace the cap. Place the
cuvette into the tester and close the
tester’s cap.
yPress the button. When the display shows
“Add”, “C.2” with “Press” blinking the tester
is zeroed.
yRemove the cuvette, open it and add the
content of one packet of MI528-0 reagent.
Replace the cap and shake gently for
2 minutes until the powder is completely
dissolved. Replace the cuvette into the tester.
yWait for 2 minutes and 30 seconds and
then press the button or press and hold
the button until the timer is displayed on
the LCD.
yThe instrument directly displays the
concentration of iron in ppm. The tester
automatically turns o after 2minutes.
BATTERY REPLACEMENT
To save the battery, the instrument shuts down
after 3 minutes of non-use and 2 minutes
after reading. One fresh battery lasts for a
minimum of 5000 measurements, depending
on the light level. When the battery capacity
is under 10 % “bAt” appears on the LCD at
start-up.
If the battery is empty and accurate
measurements can’t be taken any more, the
instrument shows “bAd” then ”bAt” each
for 1 second and turns o. To restart the
instrument, the battery must be replaced with
a fresh one.
Iron
ppm
MW14
To replace the instrument’s battery, follow the
steps:
yTurn the instrument o by holding the
button until the tester shuts o.
yTurn the instrument upside down
and remove the battery cover with a
screwdriver.
yRemove the battery from its location and
replace it with a fresh one.
yInsert the battery cover and replace the
screw with a screwdriver.

Other Milwaukee Test Equipment manuals

Milwaukee pH55 PRO User manual

Milwaukee

Milwaukee pH55 PRO User manual

Milwaukee MW803 User manual

Milwaukee

Milwaukee MW803 User manual

Milwaukee MW14 User manual

Milwaukee

Milwaukee MW14 User manual

Milwaukee MW12 User manual

Milwaukee

Milwaukee MW12 User manual

Milwaukee CD97 User manual

Milwaukee

Milwaukee CD97 User manual

Milwaukee ORP57 User manual

Milwaukee

Milwaukee ORP57 User manual

Milwaukee pH54 User manual

Milwaukee

Milwaukee pH54 User manual

Milwaukee pH600 User manual

Milwaukee

Milwaukee pH600 User manual

Milwaukee MW10 User manual

Milwaukee

Milwaukee MW10 User manual

Milwaukee 2237-40 User manual

Milwaukee

Milwaukee 2237-40 User manual

Milwaukee MW804 User manual

Milwaukee

Milwaukee MW804 User manual

Milwaukee MW11 User manual

Milwaukee

Milwaukee MW11 User manual

Milwaukee pH51 User manual

Milwaukee

Milwaukee pH51 User manual

Milwaukee 2210-20 User manual

Milwaukee

Milwaukee 2210-20 User manual

Milwaukee pH58 User manual

Milwaukee

Milwaukee pH58 User manual

Milwaukee ORP57 User manual

Milwaukee

Milwaukee ORP57 User manual

Milwaukee HEAVY DUTY 2216-40 User manual

Milwaukee

Milwaukee HEAVY DUTY 2216-40 User manual

Milwaukee EC59 User manual

Milwaukee

Milwaukee EC59 User manual

Popular Test Equipment manuals by other brands

Unit UT501A operating manual

Unit

Unit UT501A operating manual

VOLTCRAFT VC-55 LCD operating instructions

VOLTCRAFT

VOLTCRAFT VC-55 LCD operating instructions

SCHNIER EASY Checker 80 operating manual

SCHNIER

SCHNIER EASY Checker 80 operating manual

Amprobe AM-42 user manual

Amprobe

Amprobe AM-42 user manual

Devine CABT100 user manual

Devine

Devine CABT100 user manual

thomann STARVILLE LF-6 user manual

thomann

thomann STARVILLE LF-6 user manual

Bosch MOT 240 operating instructions

Bosch

Bosch MOT 240 operating instructions

YOKOGAWA 30032A user manual

YOKOGAWA

YOKOGAWA 30032A user manual

MARTINDALE PD240 instruction manual

MARTINDALE

MARTINDALE PD240 instruction manual

GW Instek GPT-9500 Series user manual

GW Instek

GW Instek GPT-9500 Series user manual

NIEUWKOOP EC2620 user manual

NIEUWKOOP

NIEUWKOOP EC2620 user manual

OfiTE 120-57 instruction manual

OfiTE

OfiTE 120-57 instruction manual

Motomaster 011-3003-0 instruction manual

Motomaster

Motomaster 011-3003-0 instruction manual

PeakTech 1245 Operation manual

PeakTech

PeakTech 1245 Operation manual

VOLTCRAFT DS-01 operating instructions

VOLTCRAFT

VOLTCRAFT DS-01 operating instructions

Keysight Technologies Infiniium 90000A Series User's quick start guide

Keysight Technologies

Keysight Technologies Infiniium 90000A Series User's quick start guide

Huazheng Electric Manufacturing HZJQ-X1 manual

Huazheng Electric Manufacturing

Huazheng Electric Manufacturing HZJQ-X1 manual

shumatt CRT067 instruction manual

shumatt

shumatt CRT067 instruction manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.